Manual Omega 27

48 pages 0.65 mb
Download

Go to site of 48

Summary
  • Omega 27 - page 1

    Need for Control? Panels & Assemblies http://www .omega.com e-mail: info@omega.com ® U ser’ s Guide C L 23A, 24, 25, 26, 27 Digital Calibrator/Thermometer ...

  • Omega 27 - page 2

    Servicing North America: USA: One Omega Drive, Box 4047 ISO 9001 Certified Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: canada@omega.com For immediate technical or application assistance: USA and Canada: Sales Service: ...

  • Omega 27 - page 3

    T ABLE OF CONTENTS General Information ......................................................................... 2 Common Specifications ................................................................... 2 Unique Specifications ...................................................................... 5 Features........................................ ...

  • Omega 27 - page 4

    GENERAL INFORMA TION This manual provides operating instructions and maintenance information for five temperature instruments. These instruments are high performance cali- brator-thermometers capable of simulating and measuring a wide-variety of sensors. In addition, features such as data storage and ramping further enhance their versatility. It is ...

  • Omega 27 - page 5

    3 RTD; ± 0.2°F (rdg ≥ – 50°F) RTD; ± 0.5°F (rdg < – 50°F) RTD; ± 0.2°F ± 0.04% rdg (meter mode) THERMISTOR; ± 0.5F OHMS; ± 0.02% range * Exclusive of sensor errors, and lead resistance induced errors. RESOLUTION: TEMPERATURE; 0.1°/1° F/C RESOLUTION: OHMS; 0.01 Ω /0.1 Ω /1 Ω REPEATABILITY (1 week at constant ambient tempe ...

  • Omega 27 - page 6

    4 KEYPAD: 12 momentary switches with tactile feedback select; • On/Off • Change/Enter • Clear/–0 • Sensor Select/1 • °F/°C/2 • Resolution/3 • Operate/4 • Store/5 • Recall/6 • Calibrator/Meter/Up Ramp/7 • Calibrator Step/Down Ramp/8 • Calibrator Ramp/Meter Hold/9 DISPLAY: 5 digit LCD, 0.4" height, and decimal point. ...

  • Omega 27 - page 7

    5 BATTERY LIFE, CONTINUOUS: 10 hrs. typical, alkaline; 3 hrs typical, Ni-Cd (rechargeable). BATTERY INDICATOR: Display indicates BAT when less than 10% of life remains. DIMENSIONS, WEIGHT: 7.0" × 2.9 × 1.1". Net weight 12 oz. UNIQUE SPECIFICA TIONS NOTE: The specifications below are specific to the model(s) as identified by model number ...

  • Omega 27 - page 8

    ACCURACY: (18°C to 28°C ambient, 2 years): ± 0.5 °F (rdg ≥ –50°F). ± 0.04% rdg (rdg > 1250°F) ± 1.0°F (rdg < –50°F) OHMS; ± 0.04% range TEMPERATURE COEFFICIENT: 18°C to 28°C; included in accuracy specifications. From 0°C to 18°C, and 28°C to 50°C. K, J, T, E; ± (0.006% rdg + 0.03°C)/°C RTD, THERMISTOR, OHMS; ± (0.00 ...

  • Omega 27 - page 9

    UNP ACKING YOUR UNIT Remove the Packing List and verify that you have received all equipment. If you have any questions about the shipment, please call the OMEGA Customer Service Department at 1-800-622-2378 or (203) 359-1660. When you receive the shipment, inspect the container and equipment for any signs of damage. Note any evidence of rough hand ...

  • Omega 27 - page 10

    WARNING Do not touch a temperature probe sheath when measuring excessively high or low temperatures, or toxic substances. WARNING Do not attempt to measure temperatures beyond the range of the probe being used. Probe damage or personal injury could result from exceeding a probe's maximum temperature rating. WARNING The American National Safety ...

  • Omega 27 - page 11

    CAUTION Keep inputs electrically isolated from outputs. Do not con- nect an external voltage source between the two connectors, and avoid any external electrical pathways between input and output (T2 & T1). Failure to do so will disturb meter readings and calibrator output levels. Damage to the instrument is also possible. Where the meter and c ...

  • Omega 27 - page 12

    2. Remove the old battery by lifting the battery extractor loop. 3. Place the new battery in the battery compartment. Be sure to observe proper polarity. 4. Re-install the battery cover before resuming use of the instrument. NOTES: • Less than 10% of battery life remains when the BAT annuncia- tor turns on. • When the battery is excessively dis ...

  • Omega 27 - page 13

    OPERA TING INSTRUCTIONS 1. DIAGNOSTICS AND ERROR MESSAGES Unit self-diagnostics provide fault condition readouts which are described below: PROBLEM: LIKELY CAUSE: Blank display, unit does not (1) Improper battery installation. power-up. (1) Check battery polarity. (2) Dead battery. BAT Annunciator. (1) Low battery voltage, install a new battery. If ...

  • Omega 27 - page 14

    If the unit is turning on for the first time after a new battery is installed, it will automatically default to the METER mode (ie. temperature measurement operation). It also defaults to K-type thermocouples, and °F readings with 0.1° resolution. The corresponding LCD annunciators are enabled: 12 Figure 3. Display Test Figure 2. Keypad ...

  • Omega 27 - page 15

    METER: (unit functioning as a thermometer) K: (K-type thermocouple) °F: (Fahrenheit readings) OPR: (Instrument is operational) If no input thermocouple is connected to the unit at this time, the numeric segments will indicate "OPEN". Plugging a thermocouple into the meter jack (connector T2, on the top right side) will give actual thermo ...

  • Omega 27 - page 16

    14 Figure 4. 2-Wire Resistance Measurement Figure 5. 3-Wire Resistance Measurement 6. Select display resolution with 0.1°/1° key. NOTE: Set-up is retained during power-off. Hold: Meter readings can be put on hold at any time by depressing the HOLD key. At this time, the HOLD annunciator turns on, the reading is frozen, and all keys (except ON/OFF ...

  • Omega 27 - page 17

    15 Lead resistance errors are compensated for in 3-wire configurations. However full compensation requires equal resistance in each lead. This configuration is common with 100-ohm RTDs. In 4-wire configurations, accuracy is unaffected by lead resistance, and resistance differences between leads. NOTE: These instruments measure input voltages on all ...

  • Omega 27 - page 18

    To Save Readings: 1. While in the METER OPR mode, depress the STORE key. Note that the OPR annunciator is replaced by STO. 2. Next depress one of the memory location numbers (keys 0, 1, . . ., 9). The corresponding memory annunciator turns on and the meter returns momentarily to the OPR mode, displaying thermocouple temperature. eg. STORE 7 NOTES: ...

  • Omega 27 - page 19

    eg. RECALL 3 eg. RECALL 5 eg. RECALL 2 eg. etc. 4. To return the meter to the operate mode, depress OPR after recalling the last reading. All active location annunciators will turn on again, together with the OPR annunciator. NOTE: Recalling data from unused (ie, empty) locations will give a zero reading. To Erase Data Locations (Meter Mode): 1. In ...

  • Omega 27 - page 20

    18 6. Select the resolution desired (0.1°/1° key). NOTE: With 1° resolution, one less key stroke will be required to enter calibrator output temperatures via the keypad. 7. The unit accepts numeric inputs by first changing the keypad to its num- ber-entry mode (NUM) by depressing the CHANGE/ENTER key once. The numeric display will flash, the OPR ...

  • Omega 27 - page 21

    Figure 10. 4-Wire Resistance Simulation 19 Figure 8. 2-Wire Resistance Simulation Figure 9. 3-Wire Resistance Simulation GRN RED BLK WHT NOTE: Colors indicated apply to Model 84937 Calibration Cable. mA (current) source Calibrator (out) Thermometer (in) I out V out ...

  • Omega 27 - page 22

    8. STORAGE AND RECALL OF CALIBRATOR SETTINGS While operating as a calibrator, the unit can store and recall up to 10 output settings (each with independent parameters, eg. °F/°C, sensor type, and res- olution) in random access memory. To Store Outputs: 1. While in the CALIB OPR mode, depress the STORE key. Note that the OPR annunciator is replace ...

  • Omega 27 - page 23

    To Recall Outputs: 1. While in the CALIB OPR mode, set-ups saved earlier in memory can be recalled to program the calibrator output. First depress the RECALL key to recall a set-up. 1. Note that the RCL annunciator turns on, OPR turns off, and all memory locations annunciators are blanked. 2. Next hit the key with the number corresponding to the de ...

  • Omega 27 - page 24

    9. STEP FUNCTION When operating in the CALIBRATOR mode, the STEP key will sequentially recall data stored in memory (up to 10 steps). At each step, the calibrator will display and output these parameters. The STEP function is non-functional in METER mode. 10. RAMP FUNCTION: CL25 and CL27 This feature, available only on Models CL25 and CL27, allows ...

  • Omega 27 - page 25

    SER VICE INFORMA TION WARNING All service information is intended for qualified electronic mainte- nance personnel only. 1. CALIBRATION PROCEDURES This is a two-part procedure. Products with thermocouple functions only, fol- low the Thermocouple Calibration procedure below. Products with resistance functions (RTD, thermistor, ohms) follow first the ...

  • Omega 27 - page 26

    24 NOTE: For calibrators with the older, single-jumper configura- tion, refer to calibration procedure beginning on page 39. 1 4. Install calibration cover in place of bottom-cover. 1 5. Re-install battery. 1 6. Hook up test-equipment, calibration cables, and U.U.T. per Figure 10. 1 7. Turn on DMM and TC simulator for warmup. Allow at least 30 minu ...

  • Omega 27 - page 27

    Calibration Adjustments: NOTE: Do not deviate from the calibration adjustment sequence that follows. This will ensure that adjustments to be stored in EEPROM go to the correct locations. 1. Set meter-mode zero and gain adjustments per Table 1. 2. Hit OPR key (RTD annunciator turns off). 3. Set U.U.T. to CALIB, OPR, K, °F. 4. Turn U.U.T. OFF, then ...

  • Omega 27 - page 28

    Sensor U.U.T. DMM Step Select Output Adjust Reading* 1 K 32.0°F P1 0.000 mV ± 1µV 2 K 2500.0°F P41 54.856 mV ± 1µV 3 Repeat Steps 1 & 2 As Required 4 J 32.0°F P1 0.000 mV ± 1µV 5 T 32.0°F P1 0.000 mV ± 1µV 6 E 32.0°F P1 0.000 mV ± 1µV 7 N 32.0°F P1 0.000 mV ± 1µV 8 B 1000.0°F P1 1.439 mV ± 1µV 9 R 32.0°F P1 0.000 mV ± 1µ ...

  • Omega 27 - page 29

    27 Figure 13. Calibration Setup, Calibrator Mode. Figure 11. Calibration Setup, Copper Mode. Figure 12. Calibration Setup, Meter Mode. ...

  • Omega 27 - page 30

    B. RESISTANCE CALIBRATION NOTE: Thermocouple and resistance calibrations are indepen- dent. However, potentiometer P1 is used in both thermocouple and resistance calibrations to set EEPROM constants. Follow cal- ibration procedures carefully to ensure that thermocouple and resistance calibrations do not disturb one another. Test Equipment Required: ...

  • Omega 27 - page 31

    29 Figure 14. Resistance Calibration Setup ...

  • Omega 27 - page 32

    Sensor Unit Unit Step Select Input Adjust Verify Reading* 1 RTD-100 390.26 Ω P1** 1562.0°F±0.5°F 2 RTD-100 100.00 Ω √ 32.0±0.1°F 3 RTD-100 18.49 Ω √ –328.0±0.2°F 4 RTD-1000 3,902.6 Ω P1** 1562.0°F±0.5°F 5 RTD-1000 1000.0 Ω √ 32.0+0.1°F 6 1,000 Ω 900.00 Ω P1 900.00 Ω 7 1,000 Ω 1.00 Ω √ 1.00± 0.10 Ω 8 10,00 ...

  • Omega 27 - page 33

    31 Current Sensor Unit Source DMM Step Select Output Input Adjust Verify Reading* 1 RTD-100 1562.0°F 1mA P38 0.39026v 2 RTD-100 32.0°F 1mA √ 0.10000 ± 0.00003v 3 RTD-100 –328.0°F 1mA √ 0.01849 ± 0.00006v 4 RTD-1000 1562.0°F 100µA P51 0.39026v 5 RTD-1000 32.0°F 100µ A √ 0.10000 ± 0.00003v 6 1,000 Ω 900.00 Ω 1mA √ 0.90000 ± 0 ...

  • Omega 27 - page 34

    Current Sensor Unit Source DMM Step Select Output Input Adjust Verify Reading* 1 RTD-100 1562.0°F 1mA P38 0.39026v 2 RTD-100 32.0°F 1mA √ 0.10000 ± 0.00006v 3 RTD-100 –328.0°F 1mA √ 0.01849 ± 0.00012v 4 1,000 Ω 900.00 Ω 1mA √ 0.90000 ± 0.00020v 5 1,000 Ω 1.00 Ω 1mA √ 0.00100 ± 0.00020v 6 100,000 Ω 90,000 Ω 10µA P51 0.9 ...

  • Omega 27 - page 35

    33 Model Software Revision CL23A 6.7 CL24 7.7 CL25 3.7 CL26 5.6 CL27 4.16 (Software revision is marked on a label attached to integrated-circuit U1 on the mother-board). "Copper-mode" is a non-compensated mode of operation during which cold-junction compensation of inputs and outputs is disabled. This permits the use of copper-wire interc ...

  • Omega 27 - page 36

    4. Select type K thermocouple (“K” on display) by pressing the SENSOR SELECT function button until type K is reached. 5. Select the tenth degree display (with a digit appearing after the decimal point on the display) by pressing the 0.0.1 function button. 6. Select °F (“°F” on display) by pressing the °F °C function button. 7. Press the ...

  • Omega 27 - page 37

    19. Press the SENSOR SELECT function button. The sensor type will change to type T. The “J” will disappear from the display and a “T” will appear on the display. The voltmeter should read 0.000 mV. 20. Press the SENSOR SELECT function button. The sensor type will change to type E. The “T” will disappear from the display and an “E” w ...

  • Omega 27 - page 38

    32. Press the SENSOR SELECT function button. The sensor type will change to type C. A “C” will appear momentarily in the center of the display. “600.0” will then appear in the center of the display and both the “T” and “R” will appear along the top row of the display. 33. Press the CHANGE function button. The “600.0 °F” display ...

  • Omega 27 - page 39

    3. Select the calibrator mode (“CALIB” on display) by pressing the CALIB/METER function button. 4. Select type K thermocouple (“K” on display) by pressing the SENSOR SELECT function button until type K is reached. 5. Select the tenth degree display with a digit appearing after the decimal point on the display) by pressing the 0.0.1 function ...

  • Omega 27 - page 40

    4. MECHANICAL PARTS DIAGRAM 38 ...

  • Omega 27 - page 41

    39 CALIBRA TION PROCEDURES Test Equipment Required: 1. Thermocouple simulator (OMEGA CL25, or equivalent). 2. Microvolt null-meter (Keithley 155, or equivalent) 3. Calibration cover (OMEGA 820-307-4). 4. Calibration cables, T.C. types K, J, T & E (as required) per Figure1 below. 5. Low-thermal (eg. copper) shorting-link for null-meter. NOTE: Th ...

  • Omega 27 - page 42

    40 NOTE: • With J1 removed, the digital-filters in the U.U.T. are dis- abled. This is to speed up U.U.T. response to calibra- tion adjustments. In this mode ± 0.1°F of reading noise is normal. NOTE: • Do not deviate from the calibration adjustment sequence that follows. This will ensure that zero adjust- ments for each TC type will be stored ...

  • Omega 27 - page 43

    41 23. Select Type-E thermocouple. 24. Turn off U.U.T. 25. Replace T calibration-cable with E calibration-cable. 26. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization. 27. Apply 32°F (Type E). Adjust P1 for a reading of 32.0°F (exclusive of noise). 28. Hit OPR key to store E-meter zero-adjustment. 29. Select Type-K thermocouple. 30. Sele ...

  • Omega 27 - page 44

    42 47. Set both the U.U.T. and TC-simulator to output 32.0°F (Type T). Adjust P1 for a null-meter reading of 0± 3µV. 48. Hit OPR key to store T-calib zero adjustment. NOTE: Steps 49 thru 54 apply to Model CL24 only. For Model CL23A go next to step 55. 49. Select Type-E thermocouple. 50. Turn off U.U.T. 51. Replace T calibration-cable with E cali ...

  • Omega 27 - page 45

    43 NOTES ...

  • Omega 27 - page 46

    44 NOTES ...

  • Omega 27 - page 47

    WARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 37 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal three (3) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers rec ...

  • Omega 27 - page 48

    M1554/ 1105 Where Do I Find Ev erything I Need for Process M easurement and Control? OM EGA…Of Course! TEMPERATURE M U Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies M U Wire: Thermocouple, RTD & Thermistor M U Calibrators & Ice Point References M U Recorders, Controllers & Process Monitors M U Infrared ...

Manufacturer Omega Category Thermometer

Documents that we receive from a manufacturer of a Omega 27 can be divided into several groups. They are, among others:
- Omega technical drawings
- 27 manuals
- Omega product data sheets
- information booklets
- or energy labels Omega 27
All of them are important, but the most important information from the point of view of use of the device are in the user manual Omega 27.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Omega 27, service manual, brief instructions and user manuals Omega 27. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Omega 27.

Similar manuals

A complete manual for the device Omega 27, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Omega 27 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Omega 27.

A complete Omega manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Omega 27 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Omega 27, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Omega 27, that we can find in the current document
3. Tips how to use the basic functions of the device Omega 27 - which should help us in our first steps of using Omega 27
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Omega 27
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Omega 27 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Omega 27?

Use the form below

If you did not solve your problem by using a manual Omega 27, ask a question using the form below. If a user had a similar problem with Omega 27 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)