    OXYGEN SA TURA TION MONITOR PULSOX -300i ® Instruction Manual This instrument must be used according to the instructions of the doctor . ...

  • Konica Minolta PULSOX-300i - page 2

    Safety Symbols W arnings and precautions noted in this manual and the instrument are indicated by the following markings, designed to prevent accidents caused by erroneous handling of the equipment. This indicates text consisting of a warning or precaution relating to safety . Please read the text carefully and use the equipment safely . This indic ...

  • Konica Minolta PULSOX-300i - page 3

    1 Foreword Safety Precautions T o ens ure correct use of th is instrument, read th e following points carefully and adhere to them. After you have read this manual, keep it in a safe place where it can be referred to anytime a question arises. W ARNING (Failure to adhere to the following points may result in death or serious injury .) The instrumen ...

  • Konica Minolta PULSOX-300i - page 4

    2 Foreword CAUTION (Failure to adhere to the following points may result in injury or damage to the in - strument or other property .) Do not use batteries other than those specified by KONICA MINOL T A SENSING, INC. When installing batteries in the instrument, make sure that they are correctly oriented according to the mark. If alkali uid from ...

  • Konica Minolta PULSOX-300i - page 5

    3 Foreword Contents Safety Symbols Safety Precautions .............................................................1 Foreword ............................................................................4 Safety Notes ........................................................................... 4 Package Contents....................................... ...

  • Konica Minolta PULSOX-300i - page 6

    4 Foreword Foreword Safety Notes CAUTION: U.S.Federal law restricts this device to sale by or on the order of a physician. The instrument is designed for measurement of the oxygen saturation (SpO 2 ) of arterial blood and the pulse rate. Do not use it for any other purposes, such as warning of sleep apnea and breathing abnormalities. The Oxygen Sat ...

  • Konica Minolta PULSOX-300i - page 7

    5 Foreword Package Contents Before using the instrument, make sure that the following items are present. 1) PULSOX-300i main body × 1 2) Wristband WB-300 × 1 3) AAA-size alkaline battery × 1 4) Instruction manuals • PULSOX-300i instruction manual × 1 • Separate manual (Using manual for Probes) × 1 ✻ The probe is optional. About Probe The ...

  • Konica Minolta PULSOX-300i - page 8

    6 Foreword System Diagrams and Accessories The instrument consists of the following items. 05,3/ 8I Optional accessories Standard accessories Wristband WB-300 AAA-size alkaline battery Finger Clip Probe SR-5C Spot Check Probe SP-5C Monitor Probe LM-5C Personal Probe SD-5C Extension cable EC-300 USB adapter UA-300 Standard accessories Op ...

  • Konica Minolta PULSOX-300i - page 9

    7 Foreword Names of Parts PULSOX-300i Power button LCD Display Backlight button Probe connector Wristband Battery cover External output terminal Used to start measurement (Power-ON) or end measurement (Power-OFF). Measurement results and messages are displayed. Connect the probe or extension cable EC-300 to this connector . Press it to turn ON the ...

  • Konica Minolta PULSOX-300i - page 10

    8 Foreword Notes on Use • This instrument should be used under the following operat - ing conditions. • T emperature of 0 to 40°C (32 to 104°F), relative humidity of 30 to 85%, with no condensation. • Atmospheric pressure of 700 to 1060 hPa (altitude of – 400 to 3000m). • This is a precision instrument. T o avoid the possibility of it b ...

  • Konica Minolta PULSOX-300i - page 11

    9 Foreword Differenc e f ro m t he SaO 2 value to displayed value <T able 1> <T able 2> SaO 2 HbCO SaO 2 Methemoglobin 1% 5% 10% 1% 5% 10% 50 50 70 70 90 90 100 100 Measured value is always lower than th e a ctual value. Displayed value is sometimes higher than th e a ctual value if SaO 2 is around 50%. -0.1 -0.7 -1.5 +0.2 +1.3 +3.2 -0. ...

  • Konica Minolta PULSOX-300i - page 12

    10 Foreword Notes on Storage • This instrument should be stored under the following storage conditions. • T emperature of –10 to 60°C (14 to 140°F), relative humid- ity of 10 to 95%, with no condensation. • Atmospheric pressure of 700 to 1060 hPa (altitude of –400 to 3000m). • When storing the instrument: • Do not store the instrume ...

  • Konica Minolta PULSOX-300i - page 13

    11 Foreword Disposal Method • When disposing of the used battery , insulate the polarity ter - minals with insulating tape etc. If the terminals of the battery come into contact with metal objects, heat generation, explo- sion or re may result. • Make sure that the battery is either disposed of or recycled correctly in accordance with local ...

  • Konica Minolta PULSOX-300i - page 14

    12 Preparations Preparations Before starting measurements using this instrument, follow the procedure given below to ensure that measurements are stable. 1) Attaching the W rist Band Attach the Wrist Band WB-300 (supplied as a standard accessory) to the instrument. Since the WB-300 is made of elastic material, it allows for exibility and comfort ...

  • Konica Minolta PULSOX-300i - page 15

    13 Preparations 4 Pass the other end of the band outward through band xture B. ◆  T ake care not to twist the band. 5 Fold back and secure the band with the hook- and-loop tape section. ◆  When attaching the instrument to an arm, adjust the band length and fitting at this hook- and-loop tape section. 3 Fold up the inserted hook-and-loo ...

  • Konica Minolta PULSOX-300i - page 16

    14 Preparations 2) Connecting the probe Connect the probe’s connector plug into the Probe Connector on the instrument. ◆ Make sure that probe’s connector plug is oriented correctly . ◆ Do not connect it in with excessive force. Doing so may damage the connector . ◆  Do not touch the connector terminals by hand. Contact failure may o ...

  • Konica Minolta PULSOX-300i - page 17

    15 Preparations 3) Installing the battery 1 T urn the battery cover clockwise approximately 90 degrees to open the cover . ◆  When turning the cover , keep the flat part of your finger on the cover ’ s catch. This facilitates opening the cover with minimum force. 2 Insert the battery into the battery compartment according to the polarity mar ...

  • Konica Minolta PULSOX-300i - page 18

    16 Preparations 4) Attaching the PULSOX to the Wrist 1 Remove the hook-and- loop section of the band from the side with no probe connector . 2 Attach PULSOX-300i to the arm by inserting your hand through the Wrist Band. 3 Adjust the fastening position of the hook-and- loop tape to adjust the tension of the Band. ...

  • Konica Minolta PULSOX-300i - page 19

    17 Measurement Measurement 1) Starting the Measurement 1 Connect the probe to the PULSOX-300i correctly . 2 Press the Power button for a long time, the Power is ON, the backlight is ON, and all the elements on the display is displayed. ◆  After 30 seconds since the Power is ON, the Backlight is OFF automatically . 3 As shown the right, the set ...

  • Konica Minolta PULSOX-300i - page 20

    18 Measurement 4 Next, it will cause the file number of the next measurement to appear . ◆  The measurement data obtained from power-ON (start of measurement) up to power-OFF (end of measurement) will be stored in one le. 5 And, if the time for Auto- Power-Off was set in the instrument, the setting time is displayed. ◆  The time for Au ...

  • Konica Minolta PULSOX-300i - page 21

    19 Measurement 2) Reading the display When the instrument is turned ON, all the elements of the LCD will light up. The pulse level, pulse rate and SpO 2 value are displayed during measurement. Pulse rate (P .R.) If an error occurs, an error message mainly about P .R. will be displayed. Battery mark Indicates the remain ing batter y capacity in thre ...

  • Konica Minolta PULSOX-300i - page 22

    20 Measurement 3) Ending the Measurement About Auto-Power-Off function By connecting the instrument to a personal computer and using the optional software, the time to be elapsed from power-on (start of measurement) to power-of f (end of measurement) can be set as the auto-power-off timer . If an auto-power-off timer is set, the measurement is n ...

  • Konica Minolta PULSOX-300i - page 23

    21 Ext.Connect External Connection The PULSOX-300i has an external output terminal for outputting measurement data stored in the instrument's memory to a computer . In addition, when the Data Analyzing Software is installed on the computer and the PULSOX-300i is connected to the computer using UA-300, the date and/or Auto- power-off timer (whi ...

  • Konica Minolta PULSOX-300i - page 24

    22 Ext.Connect <Regarding the display in communication mode> ● When the PULSOX-300i is connected to a computer , the PULSOX-300i will enter UA communication mode automatically . In UA communication mode, the display will appear as shown below: <Notes on communication mode> ● It is impossible to take measurements while the inst ...

  • Konica Minolta PULSOX-300i - page 25

    23 Check Points T rouble Shootings Error Messages The instrument display error messages when the error occurs. Check and correspond to them correctly . %RROR-ESSAGES #AUSE 3OLUTION s  #ONNECTTHEPROBETOTHE 05,3/8IPROPERLY7HENUSING THEOPTIONAL%XTENSION#ABLE %#CONNECTTHEC ...

  • Konica Minolta PULSOX-300i - page 26

    24 Check Points s  #HECKTHATTHEPROBEISPROPERLY ATTACHED TOTHEPATIENT )FTHEPROBE ISPROPERLYATTACHEDBUTTHIS ME SSAG ES TIL LA PPEA RS TH EC IRCU LAT ION  INT HEARE ABEI NGMEA SURED MAYNO T BE GOOD -O VET HEPR OBE TOA NARE A ...

  • Konica Minolta PULSOX-300i - page 27

    25 Check Points %RROR-ESSAGES #AUSE 3OLUTION s  !TTACHTHEPROBETOTHEBODY CORRECTLY)FTHISMESSAGEIS DISPLAYEDWHENTHEUNITISATTACHED TOTHEPATIENTCORRECTLYTHE MEASUREMENTCONDITIONSARE REGARDEDASBLOODCIRCULATIONOR OTHERCIRCULATORYPROBLEMS? ...

  • Konica Minolta PULSOX-300i - page 28

    26 Check Points %RROR-ESSAGES #AUSE 3OLUTION 4HISINDICATESTHATTHEMEASUREMENT VALUEISOUTOFTHEGUARANTEED ACCURACYRANGE ・  4HEOXYGENSATURATION 3 P/  ISALUEBELOWTHE LOWERLIMITOFTHE MEASUREMENTRANGE4HE MEASUREDVALUE ...

  • Konica Minolta PULSOX-300i - page 29

    27 Check Points %RROR-ESSAGES #AUSE 3OLUTION s  4U R N T H E  0 OW E R  / FF   A N D T U R N  /.  AGAIN s  )FTHESEMESSAGEREAPPEAREVEN THOUGHTHEABOVEPOINTSHAVEBEEN CHECKEDTHEINSTRUMENTMAYBE MALFUNCTIONING#ONTACTTOTHE NEARESTDISTRIBUTEROR ...

  • Konica Minolta PULSOX-300i - page 30

    28 Check Points Maintenance and Inspection • Before using the instrument, make sure that there is no dam - age to the instrument, no damage or wire-breakage in the probe cable and the instrument operates correctly and safely . Cleaning • Dampen a soft cloth with neutral detergent or water , wipe the instrument with it, then wipe off carefully w ...

  • Konica Minolta PULSOX-300i - page 31

    29 Specications Specication ¢ Dual-wavelength pulse type (665nm/880nm) arterial blood oxygen satura- tion measuring instrument • Name: Oxygen Saturation Monitor PULSOX-300i ¢ Functions: • Measuring range: SpO 2 : 0 to 100% SpO 2 Pulse rate: 30 to 230 bpm • Accuracy: SpO 2 : ±2% SpO 2 (70 to 100% range, 1S.D.) Pulse rate: ±2 bp ...

  • Konica Minolta PULSOX-300i - page 32

    30 Specications ¢ Weight 56g (including battery/excluding wrist band and probes) ¢ Usable life 6 years [Veried by KONICA MINOL T A SENSING, INC. (based on own data)] Probes (SR-5C, SP-5C, LM-5C, SD-5C) ¢ LED-SPD probe ¢ T ype • Finger Clip Probe SR-5C • Spot Check Probe SP-5C • Monitor Probe LM-5C • Personal Probe SD-5 ...

  • Konica Minolta PULSOX-300i - page 33

    31 Appendix Appendix Calculation Method for Displayed V alues <Pulse level meter> The pulse level meter displays the transmitted light amount ratio calculated by the following equation. The pulse level is indicated in eight levels (0 to 15%, full scale if over 10%). <Oxygen saturation (SpO 2 )> Updates the moving average of the oxygen s ...

  • Konica Minolta PULSOX-300i - page 34

    32 Appendix EMC Guidance Guidance and manufactures declaration - electromagnetic emissions Guidance and manufactures declaration - electromagnetic immunity The Oxygen Saturation Monitor Model PULSO X-300i is intended for use in the electromagnetic environment specified below. The customer or the user of the Model PULSO X-300i should assure t ...

  • Konica Minolta PULSOX-300i - page 35

    33 Appendix Guidance and manufactures declaration - electromagnetic immunity The Oxygen Saturation Monitor Model PULSO X-300i is intended for use in the electromagnetic environment specified below. The customer or the user of the Model PULSO X-300i should assure that it is used in such anenvironment. Immunity test IEC 60601 T est level Complian ...

  • Konica Minolta PULSOX-300i - page 36

    34 Appendix ¢ Spectral Absorption of Hb and HbO 2 SpO 2 =  100 (% SpO 2 ) C(HbO 2 ) C(HbO 2 ) +C(Hb) Absorption coefficient Wa velength Red light Infrared light Measurement Principle This instrument measures the oxygen saturation (SpO 2 ) in arte - rial blood and pulse rate continuously in a non-invasive method by applying op ...

  • Konica Minolta PULSOX-300i - page 37

    35 Appendix Relation between Oxygen Saturation and Partial Pressure The relation between oxygen saturation (SpO 2 ) and oxygen partial pressure (PaO 2 ) is shown in the graph below. SpO 2 is the oxygen saturation as measured by pulse oximeters. ¢ Oxygen Saturation vs. Oxygen Partial Pressure The curve of the above graph may shift to the right o ...

  • Konica Minolta PULSOX-300i - page 38

    36 Appendix MEMO ...

  • Konica Minolta PULSOX-300i - page 39

    ...

  • Konica Minolta PULSOX-300i - page 40

    AJF AKK Printed in Japan 9222-1732-36 © 2007-2009 KONICA MINOL T A SENSING, INC. Manufacturer 3-91, Daisen-nishimachi, Sakai-ku, Sakai-shi, Osaka 590-8551, Japan ...

Manufacturer Konica Minolta Category Oxygen Equipment

Documents that we receive from a manufacturer of a Konica Minolta PULSOX-300i can be divided into several groups. They are, among others:
- Konica Minolta technical drawings
- PULSOX-300i manuals
- Konica Minolta product data sheets
- information booklets
- or energy labels Konica Minolta PULSOX-300i
All of them are important, but the most important information from the point of view of use of the device are in the user manual Konica Minolta PULSOX-300i.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Konica Minolta PULSOX-300i, service manual, brief instructions and user manuals Konica Minolta PULSOX-300i. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Konica Minolta PULSOX-300i.

A complete manual for the device Konica Minolta PULSOX-300i, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Konica Minolta PULSOX-300i by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Konica Minolta PULSOX-300i.

A complete Konica Minolta manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Konica Minolta PULSOX-300i - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Konica Minolta PULSOX-300i, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Konica Minolta PULSOX-300i, that we can find in the current document
3. Tips how to use the basic functions of the device Konica Minolta PULSOX-300i - which should help us in our first steps of using Konica Minolta PULSOX-300i
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Konica Minolta PULSOX-300i
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Konica Minolta PULSOX-300i in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Konica Minolta PULSOX-300i?

Use the form below

If you did not solve your problem by using a manual Konica Minolta PULSOX-300i, ask a question using the form below. If a user had a similar problem with Konica Minolta PULSOX-300i it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)