    User’s Guide 11970 SERIE S HARMONIC MIXERS (K, A, Q, U , V , and W Model s) Manufac t uring P art Number: 11970- 90031 Supersedes 11970- 90030 Printed in USA Octobe r 2003 © Copyright 1992, 1999, 2001 − 2003 Agi lent T ec hnol ogies , Inc. ...

  • Agilent Technologies 11970-90031 - page 2

    2 Notice The informati on contai ned in this document is subje ct to change without notic e. Agilent T ec hnologies makes no w arranty of any kind with regard to this material, including but not limite d to, the implied wa rrant ies of merchantab ility and fi tness for a particular purpose. Agilent T ec hnologies shall not be liable for err ors con ...

  • Agilent Technologies 11970-90031 - page 3

    3 LIMIT A TION OF W ARR ANTY The foregoin g w arranty shall not apply to def ect s resulting from improper or inad equat e maintenance by Buyer , Buyer -supplied softw are or interfacing, unauthorized modific atio n or misuse , operation out side of the environmental s pecificati ons for the product, or improper sit e pre paration or maintenance . ...

  • Agilent Technologies 11970-90031 - page 4

    4 ...

  • Agilent Technologies 11970-90031 - page 5

    Con tents 5 T able of Co n tent s 1. General Inf ormation Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Mixers Co vered by Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies 11970-90031 - page 6

    6 Con tents T abl e of Contents Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68 ...

  • Agilent Technologies 11970-90031 - page 7

    7 General I nformatio n 1 General Information ...

  • Agilent Technologies 11970-90031 - page 8

    8 Chapter 1 General Inf or mation Intro duction General In formatio n Introduction The Agilent Models 11970K, 11970A, 11970Q , 11970U , 11970V and 11970W are general-pur pose harmonic mixers with very flat frequenc y response charact eristics and low conversion los s . C ollec tively , they cover the frequency range of 18 to 110 GHz. The 11970K cov ...

  • Agilent Technologies 11970-90031 - page 9

    Chapter 1 9 General Inf or mation Intro duction General I nformatio n 11970 Serie s Harmonic Mixers F igure 1-1. Options Option 009 , show n in Figure 1-2. is a Mixer Connection Kit. It includ es th ree low- loss SMA cables (P art Number 5061-5458), one hex-he ad balld river (P art Number 8710-1 539) for tightenin g the wa veguide connect or screws ...

  • Agilent Technologies 11970-90031 - page 10

    10 Chapter 1 General Inf or mation Intro duction General In formatio n • PSA Series E4440A, E4446A, and E4448A spect rum analyzers requir e Option A YZ (LO OUT and IF IN connectors). The E4443A and E4445A do not supp ort external mixing. F igure 1-2. Mixer Conn ection Kit, Option 009 ...

  • Agilent Technologies 11970-90031 - page 11

    Chapter 1 11 General Inf or mation Specifications General I nformatio n Specifications Specific ations for the 11970 Series Mixers are listed in Ta b l e 1 - 1 . Thes e are the per form anc e standards agains t which the mixers are test ed (p erformance tests are provided in Cha pter 3 ,“Performance T ests ,” ). Typical or nominal op erating va ...

  • Agilent Technologies 11970-90031 - page 12

    12 Chapter 1 General Inf or mation Specifications General In formatio n T able 1- 1 1 1970 Series Specifications NOTE Unless otherwise st ated, all speci fications app ly for an IF of 321.4 MHz and for RF in put amplitude s of less than -20 dBm. GENERAL LO Amplitu de Range: 14 to 1 8 dBm 1 Bias Requirement s: None Calibration Acc uracy: Maximum CW ...

  • Agilent Technologies 11970-90031 - page 13

    Chapter 1 13 General Inf or mation Specifications General I nformatio n MODEL 1 197 0K MODEL 1 1970 A RF Frequency Range: 18 - 26 .5 GHz RF Frequency Range: 26.5 - 4 0 GHz LO Harmoni c Number: 6 LO Harmonic Number: 8 LO Input F requency Rang e: 2.95 - 4.3 6 GHz LO Input Fre quency Range : 3.27 - 4.96 GHz Maximum Conversio n Loss: 24 dB Maximum Conv ...

  • Agilent Technologies 11970-90031 - page 14

    14 Chapter 1 General Inf or mation Specifications General In formatio n MODEL 1 1970 V MODEL 1 1970W RF Frequency Range: 50 - 75 GHz RF Frequency Range: 75 - 1 10 GHz LO Harmoni c Number: 14 LO Harmonic Numb er: 18 LO Input F requency Rang e: 3.55 - 5.3 3 GHz LO Input Frequen cy Range: 4.15 - 6.09 GHz Maximum Conversio n Loss: 40 dB Maximum Conve r ...

  • Agilent Technologies 11970-90031 - page 15

    Chapter 1 15 General Inf or mation Specifications General I nformatio n T able 1-2 1 1970 Supplemental Characteristics NOTE Supplemen tal c haracteris tics are incl uded only as additional i nformation; they are not s pecifications . 3 dB IF Bandw id th: DC to 1.3 GHz Odd Order Mixing Produc t Suppressi on: 1 1970K/ A/Q/U: > 20 dB 1 1970V /W : & ...

  • Agilent Technologies 11970-90031 - page 16

    16 Chapter 1 General Inf or mation Specifications General In formatio n ...

  • Agilent Technologies 11970-90031 - page 17

    17 Operati on 2 Oper ation ...

  • Agilent Technologies 11970-90031 - page 18

    18 Chapter 2 Operation Intro duction Operati on Introductio n This section provid es inf ormation on how to make effective use of the 11970 mixers . Operating Precautions Refe r to th e se ct ion s be l o w fo r sp ecif ic par a m e te rs to fo llow pr ior to m i xe r oper a tion . WA R N I N G Do not ex ceed th e maxi mum rati ngs lis ted belo w o ...

  • Agilent Technologies 11970-90031 - page 19

    Chapter 2 19 Operation Getting Started Operati on Getting Started The 11970 series of millimet er wa ve mixers have no bias or back-shor t adjustments . 11970 Mixers requir e an LO power of 14 to 18 dBm at the LO input. If the spec tru m analyzer used with the 11970 does not ha ve sufficient LO power , an 11975A Amplifier or an equivalent is requir ...

  • Agilent Technologies 11970-90031 - page 20

    20 Chapter 2 Operation Getting Started Operati on If you ar e using an 119 70Q, 11970U, 11970V o r 11970W Mix er , and the shoulder o f its wa ve guide flange i s not p roperl y a ligned with th e flan ge of t he de vi ce unde r t est, amplit ude measure ment err ors ca n result . T o ensure pr oper alignment , tighten each of t he four flang e scr ...

  • Agilent Technologies 11970-90031 - page 21

    Chapter 2 21 Operation Using a Con version-L oss Data Disk with the ESA or PSA Series Analyzers Operati on Using a Conversion-Loss Data Disk with the ESA or PSA Series Analyzers The conver sio n-loss data f or your mixer ca n be quickly lo aded in the memory of an ESA or PSA series analyzer fr om a fl opp y disk. The following proc es s works with ...

  • Agilent Technologies 11970-90031 - page 22

    22 Chapter 2 Operation Using the Mixer s with the E4407B Spectrum Anal yzer (Option A YZ) Operati on Using the Mixers with the E4407B Spe ctrum Analyzer (Optio n A YZ) The Agilent T echnologies E4407B spectrum analyzer contains an e xtensive menu of functions that help with millime ter measurements . The following examples explai n how to connect e ...

  • Agilent Technologies 11970-90031 - page 23

    Chapter 2 23 Operation Using the Mixer s with the E4407B Spectrum Analyzer (Optio n A YZ) Operati on 2. On the analyzer , press Preset , Factory Pres et , if pr ese nt . 3. Select e xternal m ixing by pr essing Input/Output , Input Mixer , Input Mix er (Ext ). 4. The analyzer frequency band will be set to 26 .5 - 40 GHz (A ) . T o c hoose a differe ...

  • Agilent Technologies 11970-90031 - page 24

    24 Chapter 2 Operation Using the Mixer s with the E4407B Spectrum Anal yzer (Option A YZ) Operati on Signal Identif ication The IF output of a harmon ic mixer will con tain a sig nal at t he inte rmediate f requency of the analyz er wh enev er the harm onic freque ncy of the L O and the f requen cy o f the R F diff er by th e interme diate fr equen ...

  • Agilent Technologies 11970-90031 - page 25

    Chapter 2 25 Operation Using the Mixer s with the E4407B Spectrum Analyzer (Optio n A YZ) Operati on F igure 2-4. ...

  • Agilent Technologies 11970-90031 - page 26

    26 Chapter 2 Operation Using the Mix er s with the E4440A, E4446A, or E 4448A PSA S eries Spectrum Anal yzer ( Option AY Z ) Operati on Using th e Mixers with the E44 40A, E4446A, or E4 448A PSA Series Spectrum Analyzer (Option A YZ) Be sure to refer to “ Spectrum Analyze r Retrof it Require ments ” on page 9 fo r inst rument requiremen ts for ...

  • Agilent Technologies 11970-90031 - page 27

    Chapter 2 27 Operation Using the Mixer s with the E4440 A, E4446A, or E4448A PSA Series Spectrum Anal yzer (Option AY Z ) Operati on 2. On the analyzer , press Preset , Factory Pres et , if pr ese nt . 3. Select e xternal m ixing by pr essing Input/Output , Input Mixer , Input Mix er (Ext ). 4. The analyzer frequency band will be set to 26 .5 - 40 ...

  • Agilent Technologies 11970-90031 - page 28

    28 Chapter 2 Operation Using the Mix er s with the E4440A, E4446A, or E 4448A PSA S eries Spectrum Anal yzer ( Option AY Z ) Operati on Signal Identif ication The IF output of a harmon ic mixer will con tain a sig nal at t he inte rmediate f requency of the analyz er wh enev er the harm onic freque ncy of the L O and the f requen cy o f the R F dif ...

  • Agilent Technologies 11970-90031 - page 29

    Chapter 2 29 Operation Using the Mixer s with the E4440 A, E4446A, or E4448A PSA Series Spectrum Anal yzer (Option AY Z ) Operati on F igure 2-7. ...

  • Agilent Technologies 11970-90031 - page 30

    30 Chapter 2 Operation Using the Mixer s with the 856X Series Spectrum Analyzers Operati on Using the Mixers with the 856X Se ries Spectrum Analyzers External millimeter mixers can be used to extend the fre quency coverage of the 8560 E-Series and EC-Serie s spectrum analyzers. (The 8560E/ EC O ption 002 and Option 327 do not ha ve external mixing ...

  • Agilent Technologies 11970-90031 - page 31

    Chapter 2 31 Operation Using the Mixer s with the 856X Series Spectrum Analyzers Operati on Select the Frequency Band 2. Specify unpr ese lected external mi xi ng by press ing CONFIG , then EXT MXR PRE UNP R until UNPR is selecte d. 3. T o select a frequency above 18 GHz: a. Press A UX CTRL, EXTERNAL MIXE R to set the analyzer to external mixer m o ...

  • Agilent Technologies 11970-90031 - page 32

    32 Chapter 2 Operation Using the Mixer s with the 856X Series Spectrum Analyzers Operati on In this ex ample, we ’ ll look at U-band, which range s fr om 40 GHz to 60 G Hz, as shown in Figure 2 -9 . The LOCK HARM ONIC function “ lo cks ” the spectrum analyzer in that band, ensuri ng that the sp ectrum analyzer sweeps only the chosen band. LOC ...

  • Agilent Technologies 11970-90031 - page 33

    Chapter 2 33 Operation Using the Mixer s with the 856X Series Spectrum Analyzers Operati on F igure 2-10 Store and corre ct for conversion l oss. The second method fo r st oring conversion-lo ss information lets you sa ve individ ual conversion- loss data point s at s pec ific intervals acr oss the harmonic band, using CNV LOSS VS FR EQ . T o view ...

  • Agilent Technologies 11970-90031 - page 34

    34 Chapter 2 Operation Using the Mixer s with the 856X Series Spectrum Analyzers Operati on F igure 2-11 Signal Responses Produced by a 50 GHz Signal in U Band Identify signals with the fr equency-shift meth od 6. Signal-ident ification routin es that identify the signal and images are a vailable on inst ru m ents w i th fi r m wa re re v isi o ns ...

  • Agilent Technologies 11970-90031 - page 35

    Chapter 2 35 Operation Using the Mixer s with the 856X Series Spectrum Analyzers Operati on F igure 2-12 Response f or Invalid Signals F igure 2-13 Response fo r V alid Signals ...

  • Agilent Technologies 11970-90031 - page 36

    36 Chapter 2 Operation Using the Mixer s with the 856X Series Spectrum Analyzers Operati on Identify signals in wide freque ncy spans 7. SIG ID A T MKR ident ifies si gnals i n wide fr eque ncy spa ns, using har moni c search . SIG I D AT M K R automatically det ermines the proper frequency of a signal and di splays its value on the spectrum analyz ...

  • Agilent Technologies 11970-90031 - page 37

    Chapter 2 37 Operation Using the Mixer s with the 856X Series Spectrum Analyzers Operati on F igure 2-15 SIG ID A T MKR P erformed on a T rue Signal ...

  • Agilent Technologies 11970-90031 - page 38

    38 Chapter 2 Operation Using the Mixer s with MMS Analyzers Operati on Using the Mixers with MMS Analyzers Preliminary Operation NO TE This sec tion provides information f or the Agilent 70907A (or B), however the operation of the Agilent 70909A and Agi lent 70910A is similar . Minor differe nces may be noted , but the necessary de viations from th ...

  • Agilent Technologies 11970-90031 - page 39

    Chapter 2 39 Operation Using the Mixers wit h MMS Analyzers Operati on Operation Band Selection Use th e fol lowin g key se quen ce to enter t he ex tern al mi xing m ode an d to se lect t he de sired band of operation: MENU Select Input - choose exter nal mixer input, for example : “ IN 2 EM 7 0910A ” State ext mixer fulband KA QUVE Pres s the ...

  • Agilent Technologies 11970-90031 - page 40

    40 Chapter 2 Operation Using the Mixer s with MMS Analyzers Operati on Genera l Desc ripti ons of Agilen t 710 00 Seri es Spe ctru m Anal yzer External-Mixing Functi ons ext mixer Allows access to the f ollowing softkey func tions that contr ol the measurement range whe n an external mixer extend s the spectrum analyzer f requency range . The ext m ...

  • Agilent Technologies 11970-90031 - page 41

    Chapter 2 41 Operation Using the Mixers wit h MMS Analyzers Operati on CONV LOSS (conversi on loss) Offsets the ref erence level to compens ate for amplitude los se s at the acti ve input po rt. If ne cessary , use select input t o activat e the desi r ed inp u t po r t befo re sp ecify ing its co n v e rs ion- l os s offs et . T o clear the offs e ...

  • Agilent Technologies 11970-90031 - page 42

    42 Chapter 2 Operation Using the Mixer s with MMS Analyzers Operati on NO TE See th e tabl e below for a vailab le freq uency range s and r ela te d harm oni c num b ers. U se the IMA GE N ST ART and IMA GE N STOP so ftke ys for the image i dentif ication only . T able 2-3 Fr equency Ranges and Co rrespondi ng Harmonic Numbers Band/Range Harmonic N ...

  • Agilent Technologies 11970-90031 - page 43

    Chapter 2 43 Operation Using the Mixers wit h MMS Analyzers Operati on Conversion Loss V ersus Frequency Correction The Agilent 71000 Series Spectr um Analyzers with the Agilent 70907A (or B) External Mixer Inter face Module installed ha s the am plit ude-corre ction function (AM PCOR) availa ble by remote pro gramming. Use AMPCOR to compensate for ...

  • Agilent Technologies 11970-90031 - page 44

    44 Chapter 2 Operation Using the Mixer s with MMS Analyzers Operati on T o print out an existing AMPCOR table, use the progr am listed below: 10 dim A$ (1:20) [30] 20 OUTPUT 718; “ CONVLO SS 0DB; ” 30 OUTPUT 718; “ AMPCOR 50 GHZ, 46 .1DB, 52GHZ , 46.7DB, 5 3GHZ,47. 2DB; ” 40 ! 50 OUTPUT 718; “ AMPCOR ?; ” 60 ENTER 71 8 USING “ %, k ? ...

  • Agilent Technologies 11970-90031 - page 45

    Chapter 2 45 Operation Using the Mixer s with the 8566B Spectrum Analyzer Operati on Using th e Mixers with the 85 66B Spectrum Analyzer Set up the equipment 1. C o n n ect th e e xter n a l harm o n ic m i x e r to the sp e c tr um a n a ly zer , as sh o w n in Figure 2-17 . NO TE Good-qua lity shielded SMA-type cables should be used to co nnect t ...

  • Agilent Technologies 11970-90031 - page 46

    46 Chapter 2 Operation Using the Mixer s with the 8566B Spectrum Anal yzer Operati on “ Ha rmonic 6 L ” , where “ L ” stands for locked, or you coul d press Shift+ z to manually harmonic loc k . F igure 2-18 T YPICAL M ILLIMET ER SIGNAL D I SPLA Y A menu of millimeter measurement f uncti ons is accessed by pressi ng SHI FT , 1 , MHz . Even ...

  • Agilent Technologies 11970-90031 - page 47

    Chapter 2 47 Operation Using the Mixer s with the 8566B Spectrum Analyzer Operati on Signal Identification Figure 2 -18 shows a t ypical full-band display o f a single inpu t signal. Dir ect interpret ation of the displ a y is difficult becaus e of the large number of respon ses produce d by several local oscilla tor harmonics generated in the mixe ...

  • Agilent Technologies 11970-90031 - page 48

    48 Chapter 2 Operation Using the Mixer s with the 8566B Spectrum Anal yzer Operati on F igure 2-19 Sample Mixer Cal i b ration T able ...

  • Agilent Technologies 11970-90031 - page 49

    49 P erform ance T ests 3 Pe r f o r m a n c e Te s t s ...

  • Agilent Technologies 11970-90031 - page 50

    50 Chapter 3 P erformance T ests Intro duction P erformanc e T ests Introductio n This section contains instructions for testing the performanc e of the 11970 Series Mixers . P erformance te sts are used to c heck the mixers at incoming inspecti on and for periodic evaluation . The tests verify the sp ecification s listed for the mixers in T able 1 ...

  • Agilent Technologies 11970-90031 - page 51

    Chapter 3 51 P erformance T ests P erform ance T est Procedures P erform ance T ests P erformance T est Procedures Description Each perfo rmance test procedure is contained in a single paragraph. The fir st entry in each p aragr aph i s the sp ecific ation for th e para mete r bein g meas ured as desc ribe d in Ta b l e 1-1 This is followed by a ge ...

  • Agilent Technologies 11970-90031 - page 52

    52 Chapter 3 P erformance T ests P erform ance T est Procedures P erformanc e T ests T able 3-2. Recommended T est Equipment for 11 970A Instrument Critical Spe cifications Recommended Mode l Spectrum An alyzer LO and IF range s compatible w ith mixer 8566B Synthe sized Sw eeper Freq uency: 8 to 13.5 GHz Output Le vel: >0 dBm 8340A Amplifier Out ...

  • Agilent Technologies 11970-90031 - page 53

    Chapter 3 53 P erformance T ests P erform ance T est Procedures P erform ance T ests T able 3-3. Recommended T est Equipment for 11 970Q Instrument Critical Sp ecifications Recommended Model Spectrum An alyzer LO and IF ranges compat ible with mixer 8566B Synthesized S weeper Frequency: 11.0 t o 16.7 GHz Output Le vel: >0 dBm 8340A Amplifier Out ...

  • Agilent Technologies 11970-90031 - page 54

    54 Chapter 3 P erformance T ests P erform ance T est Procedures P erformanc e T ests T able 3-4. Recommended T est Equipment for 11 970U Instrument Critical Specifications Reco mmended M odel Spectrum Analyzer LO and IF ranges compatible with m ixer 8566B Synthe sized Sw eeper Frequency: 13. 3 to 20 GH z Out put L eve l: > 9 dBm 8340A Amplifier ...

  • Agilent Technologies 11970-90031 - page 55

    Chapter 3 55 P erformance T ests P erform ance T est Procedures P erform ance T ests T able 3-5. Recommended T est Equipment for 11 970V Instrument Critical Sp ecifications Recommended Model Spectrum Anal yzer LO and IF ranges comp atible wit h mixer 8566B Swept Source Compatible with V -Band Plug-In 8350B V -Band So urce Plug-In Frequency Rang e: ...

  • Agilent Technologies 11970-90031 - page 56

    56 Chapter 3 P erformance T ests P erform ance T est Procedures P erformanc e T ests T able 3-6. Recommended T est Equipment for 11 970W Instrument Critical S pecifications Recommended Mode l Spectrum Analyzer LO and IF ranges com patible with mixe r 8566B Swept S ource Compatible with W - Band Plug-In 8350B W -Band Source Plug-In F requency Range: ...

  • Agilent Technologies 11970-90031 - page 57

    Chapter 3 57 P erformance T ests Con version Lo ss and Frequency Response P erform ance T ests Conv ersion Lo ss and F requency Re sponse Specifications Conversi on Loss: F or a CW RF input power of less than -20 dBm 11970K: 24 d B maximum 11970A: 26 d B maximum 11970Q: 28 d B maximum 11970U: 28 dB maximum 11970V : 40 dB maximum 11970W : 47 dB m ax ...

  • Agilent Technologies 11970-90031 - page 58

    58 Chapter 3 P erformance T ests Con version Lo ss and Frequency Respo nse P erformanc e T ests Description The frequency r esponse and con version loss are checked at four LO powe r level s . A known input po wer is app lied to t he input o f the mixer . The I F output power is measured o n the 8566B Spect rum Analyzer . From thes e measurements , ...

  • Agilent Technologies 11970-90031 - page 59

    Chapter 3 59 P erformance T ests Con version Lo ss and Frequency Response P erform ance T ests 11970V : 50.0 GHz 11970W : 75.0 GHz ...

  • Agilent Technologies 11970-90031 - page 60

    60 Chapter 3 P erformance T ests Con version Lo ss and Frequency Respo nse P erformanc e T ests 9. Adjust the out put power of the si gna l gene rato r for a r eading of ap prox imately -10 dBm on the power meter for the 11970K, V or W and for approximately -3 dBm for the 11970A, Q or U . (Make sure that the unlevel ed light is not on for the 11970 ...

  • Agilent Technologies 11970-90031 - page 61

    Chapter 3 61 P erformance T ests Con version Lo ss and Frequency Response P erform ance T ests Then press CENTER FREQUENCY : 11970K: 1 , 8 , GHz 11970A: 2 , 6 , [ . ] 5 , GHz 11970Q: 3 , 3 , GHz 11970U: 4 , 0 , GHz 11970V : 5 , 0, GH z 11970W : 7 , 5 , GHz 14.Press MARKER , PEAK SEARCH . If n ecessa ry , pr ess MKR → CF and use DISPLA Y LINE ENTE ...

  • Agilent Technologies 11970-90031 - page 62

    62 Chapter 3 P erformance T ests Con version Lo ss and Frequency Respo nse P erformanc e T ests or Conversion Lo ss = (-10.03 dBm) - (-0.232) - (-39.78 dBm) - 8.93 dB = 21.05 dB 17.Record the conversion loss in Ta b l e 3 - 7 . ...

  • Agilent Technologies 11970-90031 - page 63

    Chapter 3 63 P erformance T ests Con version Lo ss and Frequency Response P erform ance T ests NO TE The convers ion los s indicated on the mixer calibr ation label includes the los s in the IF cable . If other than the spec ified cable is used, the n the loss in that cable must be compensat ed for when making amplitude measurements . 18.Increment ...

  • Agilent Technologies 11970-90031 - page 64

    64 Chapter 3 P erformance T ests Con version Lo ss and Frequency Respo nse P erformanc e T ests F igure 3-1 P erformance T est Setups (1 of 2) ...

  • Agilent Technologies 11970-90031 - page 65

    Chapter 3 65 P erformance T ests Con version Lo ss and Frequency Response P erform ance T ests F igure 3-2 P erformance T est Setups (2 of 2) ...

  • Agilent Technologies 11970-90031 - page 66

    66 Chapter 3 P erformance T ests Con version Lo ss and Frequency Respo nse P erformanc e T ests T able 3-7 Con ver sion Loss and Fr equency Response T est Record CONVERSIO N LOSS AND FRE QUENCY RESP ONSE for an LO POWER of_____ ________ dBm Model Number_ ________ ___ Date____ ________ _ Serial Number_ ________ _____T ested By _ _________ _______ Ma ...

  • Agilent Technologies 11970-90031 - page 67

    Chapter 3 67 P erformance T ests Con version Lo ss and Frequency Response P erform ance T ests Marke r Freq uency Marker Amp lit ude P ower Meter Readings Po w e r S e n s o r Cal F actor Directional Coupler Coupling F actor Conve rsi on Loss GHz dBm dBm % or dB dB dB Fr eq uency Respon se = _________ ______dB T able 3-7 Con ver sion Loss and Fr eq ...

  • Agilent Technologies 11970-90031 - page 68

    68 Chapter 3 P erformance T ests A VERAGE NOIS E LEVE L TEST P erformanc e T ests A VERAGE NOISE LEVEL TEST Specification 11970K: -110 dBm 11970A: -108 dBm 11970Q: -104 dBm 11970U: - 104 dBm 11970V : -92 dBm 11970W : -85 dBm Description The average displa yed noise level in a 1 kHz bandwidth, using external mixing with the 8566A/B Spectr um Analyze ...

  • Agilent Technologies 11970-90031 - page 69

    Chapter 3 69 P erformance T ests A VERAGE NOIS E LEVE L TEST P erform ance T ests 4. On the 1197 5A Amplif ier , set the rear panel ALC swit ch to ON. Then connect th e po wer sensor t o the fre e end of the c able conne cted to t he output o f the amplif ier . Set the po wer meter cal ibration fact or to the va lue sho wn on the po wer sensor cali ...

  • Agilent Technologies 11970-90031 - page 70

    70 Chapter 3 P erformance T ests A VERAGE NOIS E LEVE L TEST P erformanc e T ests CA UTION Make su re the 83 49A Amplifier , used in the si gnal generato r system f or 11970A, Q and U tests , is set for externa l leveling before you turn it on. F ailure to set this amplifier for external leveling may al low the amplifi er output t o rise about 20 d ...

  • Agilent Technologies 11970-90031 - page 71

    Chapter 3 71 P erformance T ests A VERAGE NOIS E LEVE L TEST P erform ance T ests F or 11970V : 5 , 0 , GHz F or 11970W : 7, 5 , GHz 10.Record the center frequency in Ta b l e 3 - 8 . 11.Set t he power mete r CAL F ACTOR to the value s hown on the sensor calibrat ion label for the center f requency i ndicated on the spectr um analyzer . 12.Set t he ...

  • Agilent Technologies 11970-90031 - page 72

    72 Chapter 3 P erformance T ests A VERAGE NOIS E LEVE L TEST P erformanc e T ests -104 dBm ( for the 11970Q) -104 dBm ( for the 11970U) -92 dBm (for the 11 970V) -85 dBm (for th e 11970W) 18.Enter t he following p ush button commands on the 85 66A/B: MARKER , OFF REFERENCE LEVEL , 0 , dBm 19.Repeat st eps 9 through 17 at the following center fre q ...

  • Agilent Technologies 11970-90031 - page 73

    73 Service 4 Ser vice ...

  • Agilent Technologies 11970-90031 - page 74

    74 Chapter 4 Ser vice Maintenance Service Mainte nance The only maintenance requi red for the 11970 Series Mixers is prevent ive maintenance. When you are not using your mixer , cover its waveguide input with its wa veguide cap . Also , though the 11970 Mixers can abso rb more punishment than is norma l for s u c h devices , you s hould avoid s ubj ...

  • Agilent Technologies 11970-90031 - page 75

    Chapter 4 75 Service Maintenance Service F igure 4-1 11970 Se rie s Mixer Schematic Diagram Ta b l e 4 - 1 Model LP F Fco (GHz) 11970K 4.4 11970A 5.0 11970Q 5.0 11970U 6.2 11970V 5.4 11970W 6.2 ...

  • Agilent Technologies 11970-90031 - page 76

    76 Chapter 4 Ser vice Maintenance Service T able 4-2 Accessories and Re placeable P arts Pa r t N u m b e r Description 5061-546 0: 5061-545 8 8710-051 0 8710-153 9 Mixer Con nector Kit ( Option 009 ), includes th e followin g three items: Cable, 1 meter long , SMA male connectors (3 required) Wrenc h, 5/16-i nch, open-end Bal l Driv er , 3/ 32-in ...

  • Agilent Technologies 11970-90031 - page 77

    Chapter 4 77 Service Maintenance Service T abl e 4-3 Agilen t T ec hnol ogies Sale s and Servic e Offices UNITED ST A TES Instrume nt Support Cent er Agil en t T echno lo gies (800) 403-080 1 EUROPEAN FIELD OPERA TIONS Headq uart ers Agil en t T echno lo gi es S . A. 150, Route du Nant-d ’ Avr i l 1217 Me yrin 2/ Geneva Switzerland (41 22) 78 0.8 ...

  • Agilent Technologies 11970-90031 - page 78

    78 Chapter 4 Ser vice Maintenance Service ...

Manufacturer Agilent Technologies Category Music Mixer

Documents that we receive from a manufacturer of a Agilent Technologies 11970-90031 can be divided into several groups. They are, among others:
- Agilent Technologies technical drawings
- 11970-90031 manuals
- Agilent Technologies product data sheets
- information booklets
- or energy labels Agilent Technologies 11970-90031
All of them are important, but the most important information from the point of view of use of the device are in the user manual Agilent Technologies 11970-90031.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Agilent Technologies 11970-90031, service manual, brief instructions and user manuals Agilent Technologies 11970-90031. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Agilent Technologies 11970-90031.

A complete manual for the device Agilent Technologies 11970-90031, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Agilent Technologies 11970-90031 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Agilent Technologies 11970-90031.

A complete Agilent Technologies manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Agilent Technologies 11970-90031 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Agilent Technologies 11970-90031, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Agilent Technologies 11970-90031, that we can find in the current document
3. Tips how to use the basic functions of the device Agilent Technologies 11970-90031 - which should help us in our first steps of using Agilent Technologies 11970-90031
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Agilent Technologies 11970-90031
5. FAQ - Frequently Asked Questions