Manual Fluke 192

116 pages 2.29 mb
Download

Go to site of 116

Summary
  • Fluke 192 - page 1

    4822 872 00983 Oct ober 2000, Rev. 2, 2/ 01 © 2000 Fluke Corpor ation. All rights r eserv ed. Printed in the Netherlands. All product names ar e trademark s of their res pectiv e companies. Fluke 192/196/199 ScopeM eter Us ers M anual ...

  • Fluke 192 - page 2

    ...

  • Fluke 192 - page 3

    LI M I TED WARRANTY & LI M IT ATI O N O F LI ABI LI T Y Each Fluke pr oduct is w arranted to be free fr om defects in material and workmanship under normal use and ser vice. T he warranty period is three year s and begins on the date of shipment. Parts , product r epairs and s ervic es are war ranted for 90 day s. This warranty extends only t o ...

  • Fluke 192 - page 4

    SERVICE CENTERS To locat e an author iz ed s er v ic e c enter, visit us on t he W orld W ide W eb: http://www.f luke .com or call F luke using any of the phone num ber s lis t ed below: +1- 888- 993- 5853 in U. S. A . and Canada +31- 40- 2675200 in E ur ope +1- 425- 356- 5500 f rom ot her count ries ...

  • Fluke 192 - page 5

    i Table of Contents Cha pte r Title Page Un packi ng th e T est T ool Ki t ...................................................................................... 2 Safety I nf or mat ion: Read First .................................................................................. 4 1 Us ing Th e Sc op e ............................................ ...

  • Fluke 192 - page 6

    Fl u ke 192/196/ 199 Users Manual ii 2 Us ing Th e Multime ter.............................................................................................. 21 Making Meter Connect ions ........................................................................................ 21 Maki ng Mu lti meter Me asu rements..................................... ...

  • Fluke 192 - page 7

    Conte nts (continued) iii 6 Us ing Mem ory , PC and Prin ter .............................................................................. 57 Saving and Recalling ................................................................................................ 57 Documenting Screens .................................................................. ...

  • Fluke 192 - page 8

    Fl u ke 192/196/ 199 Users Manual iv 9 Sp ecifica tion s ......................................................................................................... 87 Int roduction ............................................................................................................... 87 Dual Input O scilloscope .............................. ...

  • Fluke 192 - page 9

    Unpacking the Test Tool Kit 1 Declar ation of Conform ity for Fluke 192/196/199 ScopeMeter ® tes t tool s Manuf acturer Fluke Industr ial B.V. Lelyweg 1 7602 EA Almelo The Netherlands Statem ent of Conf or m i t y Based on test r esults using appropr iat e st andar ds, the product is in conformit y with Electr omagnet ic Com pat ibility Dir ect iv ...

  • Fluke 192 - page 10

    Fluke 192/ 196/ 199 Users Manual 2 Unpacking the Test Tool Kit The following i tems ar e included in your test t ool kit: Not e When new, t he r echar geable NiMH batter y is not fully charged. See Chapter 8. F L U K E S C O P E M E T E R P R O D U C T R E G I S T R A T I O N C A R D 4 3 7 1 2 9 11 12 10 13 5 6 (2x) (2x) (2x) (2x) (2x) (2x) (2x) 8 ...

  • Fluke 192 - page 11

    Unpacking the Test Tool Kit 3 # Descript ion 1 ScopeMeter Test Tool 2 Batter y Charger ( count ry dependent ) 3 10:1 Voltage Probe Set ( r ed) a) 10:1 Volt age Probe ( r ed) b) Hook Clip for Probe Tip (r ed) c) Gr ound Lead with Hook Clip (red) d) Gr ound Lead with M ini Alligator Clip (black) e) 4-m m Test Probe for Pr obe Tip (r ed) f) Ground Spr ...

  • Fluke 192 - page 12

    Fluke 192/ 196/ 199 Users Manual 4 Safety Infor mation: Read Firs t Carefully read t he f ollowing safety infor mation bef or e using the test t ool. Specific warning and caution statem ent s, wher e t hey apply, appear t hr oughout t he m anual. A “Warni ng” i denti f i es condit i ons and acti ons t hat pose hazard( s) to the user . A “Caut ...

  • Fluke 192 - page 13

    Safety I nf or mat ion: Read First 5 War nin g To avoid el ect r i cal shock or f i re i f a t est t ool input i s connect ed t o more t han 42 V peak ( 3 0 Vrms) or on ci r cui ts of m ore t han 4800 VA: • Use only insul ated vol t age pr obes and t est leads ( and adapt er s) suppl ied w i t h the test tool , or equi val ents as speci f i ed in ...

  • Fluke 192 - page 14

    Fluke 192/ 196/ 199 Users Manual 6 The term s ‘ Is olated ’ or ‘ Electrically floating ’ are used in this manual to indicate a m easur em ent in which the t est tool input BNC or banana jack is connected to a voltage differ ent f rom ear th gr ound. The isolated input connector s have no exposed m et al and are fully insulated to pr ot ect ...

  • Fluke 192 - page 15

    7 Chapter 1 Using The Scope About this C h apter This chapter pr ovides a step- by- s t ep intr oduct ion to the scope funct ions of t he t est t ool. The intr oduct ion does not cover all of t he c apabilit ies of t he scope f unct ions but gives basic examples to show how to use the menus and perfor m basic oper at ions. Powe ring the Test Tool F ...

  • Fluke 192 - page 16

    Fluke 192/ 196/ 199 Users Manual 8 Resetting the Test Tool If you wan t to res et the te st too l to th e fac tory se ttin gs, d o the f ollowing: 1 Tu rn th e tes t too l o ff. 2 Press and hold. 3 Press and release. The test t ool tur ns on, and you should hear a double beep, indicating the reset was successf ul. 4 Release. Now l ook at t he displ ...

  • Fluke 192 - page 17

    Using The Scope Navigating a Menu 1 9 Navigating a Menu The following example shows how to use the t est tool’ s menus to select a f unct ion. Subsequently f ollow steps 1 through 4 t o open the scope m enu and t o choose an ite m. 1 Press the SCO PE key to display the labels that def ine the present use for t he four blue funct ion keys at t he ...

  • Fluke 192 - page 18

    Fluke 192/ 196/ 199 Users Manual 10 Hiding K ey Labels and Menus You can hide a menu or key label at any time: Hide any key label or menu. To display menus or key labels, press one of t he yellow menu keys, e. g. t he SC OPE key. Input Connecti ons Look at t he t op of t he t est t ool. The t est tool has four signal i nputs: t wo saf et y BNC jack ...

  • Fluke 192 - page 19

    Using The Scope Making Scope Connections 1 11 Making Sc ope Connections To make dual input scope measur em ent s, connect the red voltage pr obe t o input A, and t he gray voltage pr obe to input B. Connect t he shor t ground leads of each voltage probe t o its ow n ref er ence pot ent ial. ( See Figure 6.) Not e To maximally benefit f r om having ...

  • Fluke 192 - page 20

    Fluke 192/ 196/ 199 Users Manual 12 Displayi ng an Unknown S ignal with Connect-and-V iew™ The Connect-and- View featur e let s t he t est t ool display complex, unknown signals automat ically. This funct ion optimizes the posit ion, r ange, t ime base, and tr iggering and assures a stable display of virtually any waveform . I f the signal change ...

  • Fluke 192 - page 21

    Using The Scope Making Automat ic Scope Measur em ent s 1 13 Making Automa tic Scope Measur ements The test t ool off ers a wide range of aut omat ic scope measurem ent s. You can display two numeric readings: R EADING 1 and R EADING 2 . These readings are select able independently, and t he m easur em ent s can be done on the input A or input B wa ...

  • Fluke 192 - page 22

    Fluke 192/ 196/ 199 Users Manual 14 Figure 8 shows an example of the scr een. Note that t he Peak-Peak reading for input B appear s next t o t he input A frequency r eading at t he t op of the screen. Figure 8. Hz and V peak-peak as Scope Readings Freezing the S creen You can freeze t he scr een ( all readings and waveform s) at any t ime. 1 Freeze ...

  • Fluke 192 - page 23

    Using The Scope Using Average, Persist ence and G litch Captur e 1 15 Using Average, Persistence and Glitch Captur e Usin g Average f o r Smo o th i ng W avefo rms To smooth t he waveform , do t he f ollowi ng: 1 Display the SCOPE key label s. 2 Open t he Wavef orm O pt i ons menu. 3 Jump to Wavef orm: 4 Select Average... to open t he Average menu ...

  • Fluke 192 - page 24

    Fluke 192/ 196/ 199 Users Manual 16 Usin g Persi st en ce to Di spl ay Wavefo rms When persistence is selected, t he t est t ool displays the upper and lower boundaries of dynam ic wavefor m s. 1 Display the SCOPE key label s. 2 Open t he Wavef orm O pt i ons menu. 3 Jump to Wavef orm: and Select Persist ence , t hen select Infinite . 4 Start m oni ...

  • Fluke 192 - page 25

    Using The Scope Using Average, Persist ence and G litch Captur e 1 17 Disp l ayin g G li t ches To capture glitches on a wavef or m , do t he f ollowi ng: 1 Display the SCOPE key label s. 2 Open t he Wavef orm O pt i ons menu. 3 Select Display G l i t ches: Yes 4 Exit the m enu. You can use this function t o display events (glitches or other asynch ...

  • Fluke 192 - page 26

    Fluke 192/ 196/ 199 Users Manual 18 Acquir ing Waveform s Sele cting AC -Coupling After a r eset , t he test t ool is dc-coupled so that ac and dc voltages appear on t he scr een. Use ac-coupling when you wi sh t o observe a small ac signal that rides on a dc signal. To select ac-coupling, do the f ollowing: 1 Display the INPUT A key label s. 2 Hig ...

  • Fluke 192 - page 27

    Using The Scope Acquiring Waveform s 1 19 Wo rkin g wi th No isy W avef o rms To suppress high frequency noise on wavefor m s, you can limit the working bandwidth to 10 kHz or 20 M Hz. This funct ion smoot hes t he displayed waveform . For t he same reason, it impr oves t r igger ing on the wavef or m . To choose HF reject, do t he f ollowing: 1 Di ...

  • Fluke 192 - page 28

    Fluke 192/ 196/ 199 Users Manual 20 3 Jump to Wavef orm: and Select Math ema tics... to open the Mathem atics m enu. 4 Select Function: A+B , A- B , A*B or A vs B . 5 Se le ct a sca le fa ctor to fi t the mathem at ical result waveform ont o the display, and ret ur n. The sensitivity range of t he m at hem atical result is equal to the sensit ivity ...

  • Fluke 192 - page 29

    21 Chapter 2 Using The Multimeter About this C h apter This chapter pr ovides a step- by- s t ep intr oduct ion to the multimet er f unct ions of the t est t ool (hereaf t er called “meter”). T he in tr oduct ion gives basic examples to show how to use the m enus and per f or m basic oper at ions. Making Meter C onnections Use the two 4- m m sa ...

  • Fluke 192 - page 30

    Fluke 192/ 196/ 199 Users Manual 22 Making Multimeter Measurements The screen displays the numer ic readings of t he measurem ent s on t he met er input. Measu ri ng Resi stan ce V al ues To measure a r esist ance, do t he f ollowing: 1 Connect the r ed and black test leads f r om t he 4-mm banana jack inputs t o t he r esistor . 2 Display the METE ...

  • Fluke 192 - page 31

    Using The Mul t im et er Making Multimet er M easur em ents 2 23 Makin g a Cu rrent M easu rement You can measure cur r ent in bot h Scope mode and M et er mode. Scope m ode has t he advantage of t wo wavefor m s being displayed w hile you perform m easur ements. Meter m ode has t he advantage of high measur em ent resolution. The next example expl ...

  • Fluke 192 - page 32

    Fluke 192/ 196/ 199 Users Manual 24 6 Observe t he sensitivity of t he current probe. Highlight the corresponding sensitivity in the menu, e. g. 10 mV/ A . 7 Accept the cur r ent m easur em ent . Now, you will s ee a sc r een lik e in Figur e 14. Figure 14. Ampere M easur em ent Readi ngs ...

  • Fluke 192 - page 33

    Using The Mul t im et er Freezing the Readings 2 25 Freezing the R eadings You can freeze t he displayed readings at any t ime. 1 Freeze the scr een. HOLD appear s at t he t op r ight of t he r eading area. 2 Resume your measur em ent . You can use this funct ion to hold accurat e readings for later exam ination. Not e For saving screens into m em ...

  • Fluke 192 - page 34

    Fluke 192/ 196/ 199 Users Manual 26 Making Relative Measur ements A relative measurem ent displays the pr esent measurem ent r esult relative t o a def ined ref er ence value. The following example show s how to perf or m a r elat ive voltage measur em ent . Fir st obt ain a refer ence value: 1 Display the METER key labels. 2 Measure a volt age t o ...

  • Fluke 192 - page 35

    27 Chapter 3 Using The Recorder Functions About this C h apter This chapter pr ovides a step- by- s t ep intr oduct ion to the recorder f unct ions of the t est t ool. The introduct ion gives examples to show how to use t he m enus and per f or m basic operations. Opening the Recor der Main Menu First choose a m easur em ent in scope or m et er m o ...

  • Fluke 192 - page 36

    Fluke 192/ 196/ 199 Users Manual 28 Plotting Measur ements Over Tim e (TrendP lot™) Use the TrendPlot f unct ion to plot a gr aph of Scope or Meter m easur em ent s as f unction of t ime. Not e Because the navigations for t he dual input TrendPlot (Scope) and t he single input TrendPlot (Met er) ar e ident ical, only TrendPlot ( Scope) is explain ...

  • Fluke 192 - page 37

    Using The Recorder Funct ions Plotting Measur em ent s O v er Time ( Tr endPlot™) 3 29 Figure 17. TrendPl ot Readi ng When the Scope is in autom at ic m ode, aut om atic vert ical scaling is used to fit the Tr endPlot gr aph on t he screen. 5 Set RECORDER to STOP to freez e the recor der f unction. 6 Set RECORDER to RUN to continue. Disp l ayin g ...

  • Fluke 192 - page 38

    Fluke 192/ 196/ 199 Users Manual 30 Cha nging the Re cor d er Options At the r ight bot tom of t he display you can choose to display the time elapsed fr om st ar t and t he actual time of the day. To change the t ime r ef er ence, pr oceed f rom st ep 6 as fo llows: 7 Open t he Recorder O pt i ons menu. 8 Select Time of Day and jum p t o the next ...

  • Fluke 192 - page 39

    Using The Recorder Funct ions Recording Scope Waveform s I n Deep Mem or y ( Scope Record ™) 3 31 Recor ding Scope Wavefor ms In Deep Me mory (Sc o p e R ec ord ™) The SCOPE RECORD function is a roll mode t hat logs one or two long waveforms. This function can be used t o m onitor waveforms like mot ion cont r ol signals or the power- on event ...

  • Fluke 192 - page 40

    Fluke 192/ 196/ 199 Users Manual 32 Disp l ayin g Reco rded Dat a In Normal view, the sam ples t hat r oll off t he scr een ar e stored in deep m em or y. W hen t he m em or y is f ull, recording cont inues by shift ing the dat a in m em ory and deleting the fir st sam ples out of m em or y. In View All mode, t he com plete memor y cont ent s ar e ...

  • Fluke 192 - page 41

    Using The Recorder Funct ions Recording Scope Waveform s I n Deep Mem or y ( Scope Record ™) 3 33 Us ing Sc ope R ec ord in Trigger ed Single S weep Mode To record, f or example, t he power - on event of an Uninterrupt able Power Supply (UPS) it might be usef ul t o start recording on an ext er nal tr igger signal. The next example explains a typ ...

  • Fluke 192 - page 42

    Fluke 192/ 196/ 199 Users Manual 34 The test t ool start s r ecor ding af t er r eceiving the st ar t signal. All samples are cont inuously saved in deep memor y unt il the memor y is f ull. The last nine r ecor ded divisions are displ ayed on the screen. Use View All to display the full memory cont ent s. Not e To learn mor e about t he Single Sho ...

  • Fluke 192 - page 43

    35 Chapter 4 Using Replay, Zoom and Cursors About this C h apter This chapt er cov er s t he c apabilit ies of t he analysis functions Curso r , Zoom , and Re pla y . These f unct ions can be used with one or mor e of t he pr im ar y f unct ions Scope, TrendPlot or Scope Record. It is possible to combine two or t hr ee analysis funct ions. A typica ...

  • Fluke 192 - page 44

    Fluke 192/ 196/ 199 Users Manual 36 Repl ayi ng Step -b y- S t ep To step t hrough t he last scope screens, do t he f ollowing: 1 From scope m ode, open the REPLA Y menu. Observ e t hat t he t race is fr ozen and that R EPLAY appears at the top of t he scr een ( see Figur e 21) . 2 Step thr ough t he pr evious screens. 3 Step thr ough t he next scr ...

  • Fluke 192 - page 45

    Using Replay, Zoom and Cursor s Replaying the 100 Most Recent Scope Screens 4 37 Re play ing Continuously You can also replay the stored scr eens cont inuously, like playing a video tape. To replay continuously, do t he f ollowing: 1 From Scope m ode, open the REPLA Y menu. Observ e t hat t he t race is fr ozen and REPL AY appears at t he top of th ...

  • Fluke 192 - page 46

    Fluke 192/ 196/ 199 Users Manual 38 Zooming in on a W aveform To obtain a mor e det ailed view of a waveform , you can zoom in on a waveform using the ZOOM function. To zoom in on a waveform , do t he following: 1 Display the ZOOM key label s. Observ e t hat t he t race is fr ozen, HOLD appears at the t op of t he screen, and t he waveform is magni ...

  • Fluke 192 - page 47

    Using Replay, Zoom and Cursor s Zooming in on a Wavefor m 4 39 Dis playing the Zoomed W aveform The VIEW A LL featur e is useful when you quickly need to see the complet e wavefor m and t hen r eturn t o the zoomed part . 4 Display the complete waveform . Press repeatedly t o t oggle bet ween the zoom ed part of the wavefor m and t he com plet e wa ...

  • Fluke 192 - page 48

    Fluke 192/ 196/ 199 Users Manual 40 Making Cur sor Measur ements Cursors allow you to make precise digital measur em ent s on waveforms. This can be done on live w aveform s, recorded wavefor m s, and on saved wavef or m s. Usin g Hori zo n tal Cu rsors o n a W avefo rm To use the curs or s f or a voltage m easur em ent, do t he fo llowing: 1 From ...

  • Fluke 192 - page 49

    Using Replay, Zoom and Cursor s Making Cursor M easur em ent s 4 41 Usin g Verti cal Cu rsors o n a Wavef o rm To use the curs or s f or a t ime m easur em ent, do t he fo llowing: 1 From scope m ode, display the cursor key labels. 2 Press to highlight . Obse rve that t wo vert ical cursors ar e displayed. Marker s ( — ) identif y the point where ...

  • Fluke 192 - page 50

    Fluke 192/ 196/ 199 Users Manual 42 Usin g Curso rs on a A+B, A-B or A*B W avefo rm Cursor m easur em ent s on a A*B wavefor m give a reading in Watts if input A measur es ( m illi)Volts and input B me as u r e s ( milli) Amp e r e s . For other cur s or m easur em ent s on a A+B, A- B or A*B wavefo r m no r eading will be available if t he input A ...

  • Fluke 192 - page 51

    43 Chapter 5 Triggering on Wavefor ms About this C h apter This chapter pr ovides an intr oduct ion t o t he t r igger funct ions of t he t est t ool. Tr iggering t ells the t est t ool when to begin displayi ng the wavefor m . You can use fully autom at ic t r iggering, t ake cont r ol of one or m or e main tr igger f unct ions (semi-aut om at ic ...

  • Fluke 192 - page 52

    Fluke 192/ 196/ 199 Users Manual 44 Setting Tr igger Level and Slope The Connect-and- View ™ function enables hands-of f triggering t o display complex unknown signal s. When your t est t ool is in manual range, do t he f ollowi ng: Perfor m an aut o set . AUTO appears at t he t op r ight of the scr een. Automat ic tr iggering assures a st able d ...

  • Fluke 192 - page 53

    Trigger i ng on Wavefor m s Using Trigger Delay or Pre-t r igger 5 45 Using Tr igger Del ay or P re-trigger You can begin to display the waveform som e t ime bef ore or aft er t he t r igger point has been det ect ed. Initially, you have 2 divisions of pre-tr igger view (negative delay). To set t he t r igger delay, do t he f ollowing: 5 Hold down ...

  • Fluke 192 - page 54

    Fluke 192/ 196/ 199 Users Manual 46 Autom atic Tr igger Options In t he t r igger m enu, sett ings for aut omatic t r igger ing can be changed as follows. (See also Chapter 1: “ Displayi ng an Unknown Si gnal with Connect-and-View” ) 1 Displ ay the TRIGGE R key l abels. Not e The TRIGGE R key l abels can differ depending on the latest t r igger ...

  • Fluke 192 - page 55

    Trigger i ng on Wavefor m s Triggering on Edges 5 47 Tri ggeri ng on Edges If t he signal is instable or has a very low frequency, use edge trigger ing to obt ain full manual trigger cont rol. To trigger on r ising edges of t he input A wavefor m , do the fo llowing: 1 Displ ay the TRIGGE R key l abels. 2 Open t he Tr i gger O pti ons m enu. 3 Open ...

  • Fluke 192 - page 56

    Fluke 192/ 196/ 199 Users Manual 48 Tri g geri ng on No isy W avefo rms To reduce jitter on t he scr een when t r iggering on noisy waveforms, you can use a noise rejection filter. Cont inue from st ep 3 of t he pr evious example as follows: 4 Select On Tri gger , jump to Noi se reje ct F ilter . 5 Set Noise rej ect Fil t er to On . Observ e t hat ...

  • Fluke 192 - page 57

    Trigger i ng on Wavefor m s Triggering on Edges 5 49 The test tool will now have a s cr een like Figure 28. 6 Arm t he t est t ool f or a new single shot. Tip The test t ool stores all single shots in the replay memor y. Use the Replay function t o look at all the st or ed single shots. Figure 28. Maki ng a Si ngl e Shot M easur e me nt ...

  • Fluke 192 - page 58

    Fluke 192/ 196/ 199 Users Manual 50 Trigger ing on External W aveform s Use external tr igger ing when you want to display waveform s on inputs A and B while triggering on a third signal. You can choose external tr iggering with aut om at ic trigger ing or with edge t r igger ing. 1 Supply a signal to the r ed and black 4-m m banana jack inputs. Se ...

  • Fluke 192 - page 59

    Trigger i ng on Wavefor m s Triggering on Video Signals 5 51 Tri ggeri ng on Video S ignals To trigger on a video signal, first select t he st andar d of the video signal you are going to measur e: 1 Apply a video si gnal to the r ed input A. 2 Displ ay the TRIGGE R key l abels. 3 Open t he Tr i gger O pti ons m enu. 4 Select Video on A to open t h ...

  • Fluke 192 - page 60

    Fluke 192/ 196/ 199 Users Manual 52 Trigge ring on Video Fr ame s Use FIELD 1 or FI ELD 2 t o t rigger either on the f irst half of t he f r am e ( odd) or on t he second half of t he f r ame (even). To trigger on t he second half of t he f r ame, do t he fo llowing: 7 Choose FIELD 2 . The signal part of t he even f ield is di splayed on the screen ...

  • Fluke 192 - page 61

    Trigger i ng on Wavefor m s Triggering on Pulses 5 53 Tri ggering on P ulses Use pulse wi dth t r iggering to isolate and display specific pulses that you can qualify by tim e, such as glitches, missing pulses, bursts or signal dropouts. Detecti ng Narro w P u lses To set t he t est t ool t o t r igger on nar r ow positive pulses shorter t han 5 ms ...

  • Fluke 192 - page 62

    Fluke 192/ 196/ 199 Users Manual 54 To set t he pulse width to 5 m s, do the f ollowing: 7 Enable the arrow keys to adjust the pulse width. 8 Select 5 m s . All narrow positive pulses shorter t han 5 m s ar e now displayed on the screen. ( See Figure 31. ) Tip The test t ool stores all trigger ed scr eens in the replay memor y . For exam ple, if yo ...

  • Fluke 192 - page 63

    Trigger i ng on Wavefor m s Triggering on Pulses 5 55 Finding Miss ing Pulses The next example cover s f inding missing pulses in a train of positive pulses. I n t his exam ple it is assumed t hat t he pulses have a 100 ms distance between t he r ising edges. If t he t ime accident ly increases to 200 m s, a pulse is missing. To set t he t est t oo ...

  • Fluke 192 - page 64

    Fluke 192/ 196/ 199 Users Manual 56 To set t he pulse width to 150 m s, cont inue as follows: 7 Enable the arrow keys to adjust the pulse width. 8 Select 150 ms. Figure 32. Tri gger i ng on M i ssing Pul ses ...

  • Fluke 192 - page 65

    57 Chapter 6 Using Memory, PC and Printer About this C h apter This chapter pr ovides a step- by- s t ep intr oduct ion to the general funct ions of t he test t ool that can be used in the three m ain m odes: Scope, M et er , or Recorder . You will find inform at ion on pr inter and com put er comm unicat ion at t he end of t his chapt er . Saving ...

  • Fluke 192 - page 66

    Fluke 192/ 196/ 199 Users Manual 58 Savin g S creen s wi th Asso ciated Setu ps To save a screen in memor y location 10, do t he following: 1 Display the SAVE / PRIN T key labels. From t his point t he scr een is f r ozen unt il you hide the SAVE / PRI NT key labels again. 2 Open t he Save menu. Observ e t hat f r ee memor y locations are indicated ...

  • Fluke 192 - page 67

    Using M em or y, PC and Pr i nt er Saving and Recal ling 6 59 Deleti ng Screen s w it h Associ at ed Setu ps To delete all screens and associated setups, cont inue from st ep 2 of t he pr evious example as follows: 3 Delete all saved screens and setups. To delete only one screen and setup, cont inue from st ep 2 of t he pr evious example as follows ...

  • Fluke 192 - page 68

    Fluke 192/ 196/ 199 Users Manual 60 Re ca lling a Set up Configur ation To recall the setup conf igur at ion f r om m em or y 2, do t he fo llowing: 1 Display the SAVE / PRIN T key labels. 2 Open t he Recall menu. 3 Highli ght SCREEN + SET UP location 2. 4 Use RECALL SETUP to rec al l th e saved setup. Observ e t hat RUN appear s at t he top right ...

  • Fluke 192 - page 69

    Using M em or y, PC and Pr i nt er Documenting Screens 6 61 Docum enting S creens With t he FlukeView ® software you can upload wavefor m data and screen bit m aps t o your PC or not ebook computer f or fur t her processing. Print ing can also be done by connecting the t est t ool direct ly t o a pr inter . Conne ct ing to a C omputer To connect t ...

  • Fluke 192 - page 70

    Fluke 192/ 196/ 199 Users Manual 62 Conne ct ing to a P rinte r To print a scr een direct ly t o a pr inter , use one of the following adapters: The Opt ically Isolated RS-232 Adapt er / Cable (PM9080) t o connect a ser ial printer to t he O PTI CAL PORT of t he t est t ool. ( See Figure 34. ) The Print Adapter Cable (PAC91, opt ional) t o connect ...

  • Fluke 192 - page 71

    Using M em or y, PC and Pr i nt er Documenting Screens 6 63 Set ting up the Printing C onfigura tion This example demonstr at es how t o set up t he t est t ool for printing on a postscr ipt pr inter with a 9600 baud r at e: 1 Display the USER O PTIO NS key labels. 2 Open t he User O pt i ons menu. 3 Open t he Pr i nte r Se t up subm enu. 4 Select ...

  • Fluke 192 - page 72

    Fluke 192/ 196/ 199 Users Manual 64 ...

  • Fluke 192 - page 73

    65 Chapter 7 Tips About this C h apter This chapter gives you inform ation and tips on how you can make the best use of the t est t ool. Using the S tandard Accessor ies The following illust rat ions show t he use of t he standar d accessories such as voltage probes, t est leads, and t he various clips. ...

  • Fluke 192 - page 74

    Fluke 192/ 196/ 199 Users Manual 66 Figure 36. Probing Usi ng 2- m m Heavy Duty Test Probes War nin g To avoid el ect r i cal shock or fi re, do not connect t he ground spr i ng to vol t ages hi gher than 30 Vrm s f rom ear t h gr ound. Figur e 37. HF Vol t age Pr obe Connect i on Usi ng Gr ound Spr i ng Figure 38. Heavy Duty Fixed Connecti ons for ...

  • Fluke 192 - page 75

    Tips Using the Standard Accessories 7 67 Figure 40. Elect r oni c Connections f or Scope Measurem ent s Using Hook Cli ps and A lligato r Clip Gr ounding Figure 41. Fixed El ect roni c Connecti ons for Scope Measurem ent s wi t h Fi xed Ext e rnal Tri gger i ng Figure 42. Manual Pr obi ng f or M eter M easur em ent s using 2- m m Test Pr obes Figur ...

  • Fluke 192 - page 76

    Fluke 192/ 196/ 199 Users Manual 68 Using the I ndependently Floating Isolated Inputs You can use the independently float ing isolated inputs t o measure signals that ar e independent ly f loating f r om each other. Independent ly float ing isolated inputs of f er addit ional safet y and measur em ent capabilit ies compar ed to input s with common ...

  • Fluke 192 - page 77

    Tips Using the Independent ly Floating I solat ed I nput s 7 69 Figure 44. Parasit i c capacit ance bet w een pr obes, inst r ument and envir onm ent M MOT IO N/MOT OR C ONTR OLLE R DC BUS DC BUS DIGITA L GROUND + - Figure 45. Parasit i c capacit ance bet w een analog and digi t al ref er ence ANALOG I NPUT DIGIT AL CONTROLLER ANALOG GROUND DIG ITA ...

  • Fluke 192 - page 78

    Fluke 192/ 196/ 199 Users Manual 70 Using the T ilt Stand The test t ool is equipped with a tilt stand, allowing view ing from an angle whil e placed on a table. From t his position you can access the O PTICAL PORT at t he side of t he test t ool. The t ypical position is show n in Figure 48. Figur e 48. Usi ng t he Ti lt St and Resetting the Test ...

  • Fluke 192 - page 79

    Tips Changing the Inf or m at ion Language 7 71 Changing the Infor mation Language During operation of t he t est t ool, m essages m ay appear at t he bot t om of t he screen. You can select t he language in whi ch these m essages ar e displayed. The com bination of selectable languages (one or m or e) depends on t he version order ed. In t his exa ...

  • Fluke 192 - page 80

    Fluke 192/ 196/ 199 Users Manual 72 Changing D ate and Tim e The test t ool has a date and t im e clock. For exam ple, t o change the dat e t o 19 April, 1999, do the f ollowing: 1 Display the USER key labels. 2 Open t he User O pt i ons menu. 3 Open Date Adj ust menu. 4 Choose 1999, jump to M onth . 5 Choose 04, jump to Da y . 6 Choose 19, jump to ...

  • Fluke 192 - page 81

    Tips Sa vi ng Ba ttery L ife 7 73 Saving B attery Life When operat ed on t he bat ter y ( no bat t er y charger connected), the t est t ool conserves power by shut t ing itself down. I f you have not pr essed a key f or at least 30 minutes, t he t est t ool turns itself of f aut om at ically. Not e If t he power adapt er is connected, t here is no ...

  • Fluke 192 - page 82

    Fluke 192/ 196/ 199 Users Manual 74 Changing the A uto Set O ptions With the next procedure you can choose how aut o set behaves when you press the AUTO (auto set ) key. 1 Display the USER key labels. 2 Open t he User O pt i ons menu. 3 Open t he Auto Se t Ad just menu. If t he f r equency r ange is set t o > 15 Hz, t he Connect - and- View func ...

  • Fluke 192 - page 83

    75 Chapter 8 Maintaining the Test Tool About this C h apter This chapter cover s basic m aintenance pr ocedur es t hat can be perfor m ed by t he user. For com plete ser vice, disassembly, repair, and calibration inform at ion, see t he Service M anual. You will f ind t he par t number of t he Service Manual in the section “ Par t s and Accessori ...

  • Fluke 192 - page 84

    Fluke 192/ 196/ 199 Users Manual 76 Charging the B atteries At delivery, t he NiMH batt er ies may be em pt y and m ust be charged for 4 hour s ( with t he t est t ool tur ned of f ) t o reach full charge. W hen f ully charged, t he bat ter ies provide 4 hours of use. When batt er y power is used, t he bat t ery indicator at the top of t he scr een ...

  • Fluke 192 - page 85

    Mai nt ai ning t he Test Tool Extending Batt er y O perat ion Time 8 77 Extendi ng Batter y Operation Tim e Typicall y, NiMH batt eries always meet t he specif ied operating t ime. However, if t he bat t er ies have been extrem ely discharged ( f or exam ple, when empt y bat t eries were stor ed f or a long period) it is possible that t he batt er ...

  • Fluke 192 - page 86

    Fluke 192/ 196/ 199 Users Manual 78 Replacing the NiMH Batter y Pack BP 190 Usually i t should not be necessar y t o replace the bat t er y pack. However, if r eplacem ent is needed, t his should be done by qualified personnel only. Contact your near est Fluke center f or m or e inf or m at ion. Calibr ating the Voltage P robes To meet f ull user s ...

  • Fluke 192 - page 87

    Mai nt ai ning t he Test Tool Calibrating the Voltage Probes 8 79 Repeat steps 2 and 3 and pr oceed as f ollows: 5 Select Probe Cal with t he ar r ow keys, then accept . A message appears asking you whether t o st art t he 10:1 probe calibration. 6 Start t he pr obe calibration. A message appears t elling you how to connect the pr obe. Connect the ...

  • Fluke 192 - page 88

    Fluke 192/ 196/ 199 Users Manual 80 Displayi ng Cali brati on Inform ation You can display version number and calibration date: 1 Display the USER key labels. 2 Open t he Ver si on & Cal i brat i on menu. The screen gives you inform at ion about t he m odel number with sof t ware version, t he calibration number with latest calibration dat e, a ...

  • Fluke 192 - page 89

    Mai nt ai ning t he Test Tool Parts and Accessories 8 81 Table 1. St andard Accessori es Item Orderi ng Code Batter y Char ger , available models: Universal Europe 230 V, 50 and 60 Hz North Amer ica 120 V, 50 and 60 Hz United Kingdom 240 V, 50 and 60 Hz Japan 100 V, 50 and 60 Hz Australia 240 V, 50 and 60 Hz Universal 115 V/230 V, 50 and 60 Hz * * ...

  • Fluke 192 - page 90

    Fluke 192/ 196/ 199 Users Manual 82 Item Orderi ng Code Voltage Probe Set (G ray), designed for use with t he Fluke ScopeMet er 190 series test t ool. The set includes the following items (not available separately): • 10:1 Voltage Probe ( gr ay) • 4-mm Test Probe for Pr obe Tip (gr ay) • Hook Cli p for Pr obe Tip (gr ay) • Ground Lead with ...

  • Fluke 192 - page 91

    Mai nt ai ning t he Test Tool Parts and Accessories 8 83 Item Orderi ng Code Replacement Set f or Voltage Pr obe The set includes the following items (not available separately): • 2x 4-mm Test Probe for Pr obe Tip (r ed and gr ay) • 3x Hook Cli p for Pr obe Tip (2 r ed, 1 gr ay) • 2x Ground Lead with Hook Clip (red and gray) • 2x Ground Lea ...

  • Fluke 192 - page 92

    Fluke 192/ 196/ 199 Users Manual 84 Op t io n al Accesso ries Item Orderi ng Code Software & Cable Carrying Case Kit Set contains t he f ollowing parts: Opt ically Isolated RS-232 Adapter / Cable Har d Ca r ry in g Cas e FlukeView ® S copeMet er ® Soft ware for Windows 95 ® , 98 ® , Me ® , 2000 ® and NT4 ® SCC190 PM9080 C190 SW90W Opt ic ...

  • Fluke 192 - page 93

    Mai nt ai ning t he Test Tool Troubleshooting 8 85 Troubl eshooting The Te st Tool Doe s N ot Star t Up • The batter ies m ay be complet ely empt y. I n this case the t est t ool will not start up, even if it is powered by the bat t er y char ger . Charge the bat t er ies f irs t : power the t est t ool with the bat ter y char ger without t ur ni ...

  • Fluke 192 - page 94

    Fluke 192/ 196/ 199 Users Manual 86 Fluke View Doe s N ot Rec ognize The Te st Tool • Make sure t hat the t est t ool is tur ned on. • Make sure t hat the inter f ace cable is proper ly connected bet ween the t est tool and the PC. • Make sure t hat t he correct CO M por t has been selected in FlukeView . If not , change t he COM por t sett i ...

  • Fluke 192 - page 95

    87 Chapter 9 Specifications Intr oduction Perform ance Charact er i st ics FLUKE guarantees the pr oper t ies expressed in num er ical values with the stat ed t olerance. Specified non-t olerance numerical values indicate those that could be nominally expected f r om t he m ean of a r ange of identical ScopeMeter t est t ools. Environm ent al Data ...

  • Fluke 192 - page 96

    Fluke 192/ 196/ 199 Users Manual 88 Dual Input Oscilloscope Isola ted Inputs A and B (V er tic al) Bandwidth, DC Coupled FL UK E 19 9 ....................................... 200 MHz (-3 dB) FL UK E 19 6 ....................................... 100 MHz (-3 dB) FL UK E 19 2 ......................................... 60 MH z (-3 dB ) Lower Frequency Lim ...

  • Fluke 192 - page 97

    Specifi cati ons Dual Input Oscilloscope 9 89 Record Length Scope Record Mode .......... 27500 points on each input Scope Normal Mode ............ 1000 point s on each input Scope Glitch Capture M ode . . . 500 points on each input Glit ch De t ec t ion 5 µ s to 120 s/div ....... disp lay s gl itch es a s fast a s 50 ns Wav efo rm Di spl ay ...... ...

  • Fluke 192 - page 98

    Fluke 192/ 196/ 199 Users Manual 90 Pulse Width Trigger Screen Update ........................... On Tri gger , Single Shot Trigger Conditions .......... <T , >T , ≈ T (±10 %) , ≠ T(±10 %) So urc e .........................................................................A Po la rity ...................................Po si tive o r nega ...

  • Fluke 192 - page 99

    Specifi cati ons Automat ic Scope Measur em ent s 9 91 AC Vol tag e ( V AC) Maxi mum Vol tage wi th 10:1 probe ................................................ 1000 V di rect (1:1 )......................................................... 300 V Maxi mum Re sol ution wi th 10:1 probe ................................................... 1 mV di rect ( ...

  • Fluke 192 - page 100

    Fluke 192/ 196/ 199 Users Manual 92 Ampere s (AMP) With O pt ional Current Pr obe or Curr ent Shunt Ranges ........................... same a s VD C, V AC, V AC+ DC Probe Sensitivity ................ 100 µV/A, 1 m V/ A, 10 m V/ A, 100 mV/A, 1 V/ A, 10 V/ A, and 100 V/ A Ac cura cy ........................ same as V DC , VA C, VA C+D C (add curr en ...

  • Fluke 192 - page 101

    Specifi cati ons Automat ic Scope Measur em ent s 9 93 Power Po we r Facto r ..................... ratio betw een Watts and VA Range ...................................................... 0.0 0 to 1.0 0 Watt ................................. R MS r eadi ng of mult iplication corresponding samples of input A ( volts ) and Input B (am per es) Full Sc ...

  • Fluke 192 - page 102

    Fluke 192/ 196/ 199 Users Manual 94 Meter Met er Input Input Coupling .......................................................... DC Frequency Response ..................... DC to 10 kH z (-3 dB ) Input I m pedance .............. 1 M Ω (±1 %)/ / 10 pF ( ±1. 5 pF) Max. I nput Voltage ............................ 1000 V CAT II 600 V CAT III (For d ...

  • Fluke 192 - page 103

    Specifi cati ons DMM Measur ements on M eter I nputs 9 95 Diode Maximum Voltage Reading .................................... 2.8 V Open Circuit Voltage .............................................. <4 V Ac cura cy ............................................. ± ( 2 % +5 count s) Mea sure ment C urren t ............................ 0.5 mA , ± ...

  • Fluke 192 - page 104

    Fluke 192/ 196/ 199 Users Manual 96 Ampere s (AMP) With O pt ional Current Pr obe or Curr ent Shunt Ranges ........................... same a s VD C, V AC, V AC+ DC Probe Sensitivity ................ 100 µV/A, 1 m V/ A, 10 m V/ A, 100 mV/A, 1 V/ A, 10 V/ A, and 100 V/ A Ac cura cy ........................ same as V DC , VA C, VA C+D C (add curr en ...

  • Fluke 192 - page 105

    Specifi cati ons Zoom, Replay and Cursor s 9 97 Zoom, Replay and C urs ors Zoom Horizontal Magnification Scope Record ........................................... up to 100x TrendPlot ..................................................... up to 50x Sc op e ............................................................ up to 8x Repl ay Displays a maximu ...

  • Fluke 192 - page 106

    Fluke 192/ 196/ 199 Users Manual 98 Probe Calibr at ion Manual pulse adjustment and aut om at ic DC adjustment with probe check. Generat or O ut pu t.................................... 3 Vpp / 500 Hz square wave Memory Number of Scope M em or ies ....................................... 10 Each memory can cont ain two wavefor m s plus corresponding ...

  • Fluke 192 - page 107

    Specifi cati ons Environmental 9 99 Envir onmental En vi ronmen tal ....................... MIL-P RF- 28800F, Class 2 Temperat ure Operat ing: ba ttery o nly......................... 0 to 5 0 ° C (32 to 122 ° F) power adapter .................... 0 to 40 ° C (32 to 104 ° F) Sto rag e .......................... -20 to + 60 ° C (-4 t o +140 ° F ...

  • Fluke 192 - page 108

    Fluke 192/ 196/ 199 Users Manual 100 Safety Designed for measur em ent s on 1000 V Categor y I I Installations, 600 V Categor y III Instal lati ons, P oll utio n Degree 2, per : • ANSI/ISA S82.01- 1994 • EN61010-1 (1993) ( I EC1010-1) • CAN/CSA-C22.2 No.1010.1-92 • UL3111-1 Max. I nput Vol tages Input A and B directly ...................... ...

  • Fluke 192 - page 109

    Specifi cati ons 10:1 Probe 9 101 Figure 52. Safe Handl i ng: M ax. I nput Vol tage Betw een Scope References, and Betw een Scope References and M et er Ref er ence 10:1 Pr obe Safety Max. I nput Vol tage .......................... 1000 V CAT II 600 V CAT III Max. Fl oat i ng Volt age from any term inal to ground ................... 1000 V CAT II 6 ...

  • Fluke 192 - page 110

    Fluke 192/ 196/ 199 Users Manual 102 0.01 1 2 5 10 20 50 10 0 20 0 50 0 100 0 0 .02 0 .05 0. 1 0.2 0.5 1 2 5 10 20 50 1 00 2 00 FREQ UEN C Y ( M H z) MAX. I N P U T VO L T AGE ( Vr m s ) CA T I I I CA T I I Figur e 53. M ax. Volt age Fr om Pr obe Ti p t o Gr ound and From Pr obe Ti p to Pr obe Ref er ence 0. 01 1 2 5 10 20 50 100 200 500 10 00 0. 0 ...

  • Fluke 192 - page 111

    Specifi cati ons Electromagnet ic I m m unit y 9 103 Ele ct ro mag n et ic I mmun it y The Fluke 190 series, including standard accessories, conf or ms with the EEC directive 89/ 336 f or EMC imm unity , as defined by EN-61326-1, with t he addition of t he f ollowi ng tables. Scope Mode ( 10 ms / di v) : Trace di stur bance w i t h VPS200 volt age ...

  • Fluke 192 - page 112

    Fluke 192/ 196/ 199 Users Manual 104 ...

  • Fluke 192 - page 113

    105 Index —Numerics— 10-1 Voltage Probe, 81 2-mm Test Pr obes, 3, 82 4-mm Test Pr obes, 3, 81 —A— Accessories, 65, 80 AC-Coupling, 18 Acquiring the Wavefor m, 18 Altitude, 99, 101 Ampere M easur em ent , 23 Amperes, 92, 96 Analysis Functions, 35, 97 AS200 Accessory Set, 3, 82 Auto Ranges, 25 Au to S et, 90 Automat ic Connect-and- View Trigg ...

  • Fluke 192 - page 114

    Fl u ke 192/196/ 199 Users Manual 106 Charger, 81 Charging, 76 Cleaning, 75 Clear Menu, 10, 70 Common Leads, 3 Connect-and- View, 43, 90 Connecting a Computer , 61 Connecting A Printer, 62 Connections, 10, 21 Continuity, 94 Contrast , 71 CS20MA Current Shunt , 84 Current M easur em ent , 23 Current Probe, 23 Current Shunt , 84 Cursors Measur em ent ...

  • Fluke 192 - page 115

    In dex ( continued ) 107 Max. Float ing Voltage, 100, 101 Max. I nput Voltage, 100 Measurem ent Connections, 10, 21 Measurem ent s, 13, 22 Measurem ent s on M et er Input s, 94 Mechanical, 98 Memory , 98 Meter Connect ions, 67 Meter M easur em ent s, 22 Multi Met er M easur em ent s, 22 —N— Navigating a Menu, 9 Ni MH Ba ttery, 75 , 76 Noisy Wav ...

  • Fluke 192 - page 116

    Fl u ke 192/196/ 199 Users Manual 108 Scope Record, 31, 96 Screen Contrast , 71 Screen Without M enus, 10, 70 Seria l Pr in t er , 6 2 Service Manual, 84 Shock, 99 Single S hot, 48 Single S weep Mode, 32 Slope, 44, 89 Slow V ariations, 28 Smooth, 15 So ft Ca se, 84 Softwar e, 3, 84 Softwar e Version, 80 Specifications, 87 Spike Capture, 17 Stand, 7 ...

Manufacturer Fluke Category Marine Radio

Documents that we receive from a manufacturer of a Fluke 192 can be divided into several groups. They are, among others:
- Fluke technical drawings
- 192 manuals
- Fluke product data sheets
- information booklets
- or energy labels Fluke 192
All of them are important, but the most important information from the point of view of use of the device are in the user manual Fluke 192.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Fluke 192, service manual, brief instructions and user manuals Fluke 192. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Fluke 192.

Similar manuals

A complete manual for the device Fluke 192, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Fluke 192 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Fluke 192.

A complete Fluke manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Fluke 192 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Fluke 192, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Fluke 192, that we can find in the current document
3. Tips how to use the basic functions of the device Fluke 192 - which should help us in our first steps of using Fluke 192
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Fluke 192
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Fluke 192 in a specific country, if it was not possible to solve the problem on our own.