    Agilent Acqiris Broadband High-Speed Digitizers Using 89600 V ector Signal Analyzer Software Application Note V ector signal analysis using Agilent Acqiris broadband digitizers. ...

  • Agilent Technologies 89600 - page 2

    2 1 Overview 3 1.1 89600 VSA software 4 1.2 Required software revisions 4 2 Install 89600 VSA Software 5 3 Hardware Confi guration 6 3.1 I/O between computer and digitizer 6 3.2 Anti-alias fi lters and alias exposure-free mode 7 4 Operation 8 4.1 Basic operation 8 4.2 Sampling modes: full rate and user rate 8 4.3 Differences from 89600 VSA operat ...

  • Agilent Technologies 89600 - page 3

    3 3 This product note describes the characteristics, setup, and operation of a broadband vector signal analyzer (VSA) comprised of an Agilent Acqiris U1066A (DC440, DC438) or U1065A (DC282, DC252, DC222) high-speed digitizer and the 89600 vector signal analyzer software. By taking advantage of the layered architecture and extensive digital signal p ...

  • Agilent Technologies 89600 - page 4

    4 1.1 89600 VSA software The 89600 vector signal analyzer software is the heart of the broadband VSA. This software provides fl exible tools for analyzing and demodulating even the most advanced digital modu- lations, including those not defi ned by an established standard. The VSA makes measurements on signals in the time and frequency domains u ...

  • Agilent Technologies 89600 - page 5

    5 5 2 Install 89600 VSA Software The 89600 VSA software must be installed on a computer . Installation for standard operation Insert the Agilent T echnologies 89600 Series Software CD-ROM in your standard PC and follow the instructions in the installation wizard. When the Installation Manager window opens, click Install 89600 Series Software . In t ...

  • Agilent Technologies 89600 - page 6

    6 3 Hardware Confi guration Connect the digitizer to the computer as shown in Figure 1a or 1b. There are two options for the connection, by remote computer or by embedded computer . 3.1 I/O between computer and digitizer Operation of the 89600 Series software on an embedded Compact PCI computer is similar to operation on a standard PC. The instruc ...

  • Agilent Technologies 89600 - page 7

    3.2 Anti-alias fi lters and alias exposure-free mode The broadband VSA can be confi gured to be essentially alias-free by using the full rate sampling mode and installing an external lowpass fi lter on the digitizer input. Use BNC-f to SMA -m and SMA -f to BNC-m adapters to connect the fi lter’ s SMA connectors to the digitizer in case the di ...

  • Agilent Technologies 89600 - page 8

    4 Operation 4.1 Basic operation Operation of the broadband VSA is substantially the same as that of the 89600 VSA. See the following sections for some differences or special confi gurations. There is an online T utorial and Getting Started Guide under the Help/Roadmap toolbar in the 89600 software. 4.2 Sampling modes: full rate and user rate Full ...

  • Agilent Technologies 89600 - page 9

    User rate mode The user rate setting allows you to directly control the digitizer’ s sample rate. In this mode you select the digitizer sample rate to be used. The actual sample rate selected will be the highest rate that is less than or equal to the “user sample rate” setting. By controlling the digitizer’ s sample rate you can directly in ...

  • Agilent Technologies 89600 - page 10

    This table summarizes the difference in approach to alias protection between the various sample modes: Another concern in user rate mode is that you may see unwanted amplitude variability in two channel measurements. This can occur if any part of the mea- surement frequency span falls within the regions defi ned by: (2*n+1) x F S /2 ± 0.1 x F S w ...

  • Agilent Technologies 89600 - page 11

    11 The operating region curves for the broadband VSAs are different from those of standard VSAs because the Main Time Length may be limited for some sample rate, center frequency , and span combinations. The differences are due to three hardware characteristics of the ultra-wideband ADCs in the Agilent Acqiris digitizers: • A much higher sampling ...

  • Agilent Technologies 89600 - page 12

    12 Figure 3. The user-settable parameter , frequency points, begins to determine the boundaries of the time record length and frequency span operating region 5.2 Operating region curves for Agilent Acqiris digitizers Several interrelated characteristics of both the Acqiris digitizers and the 89600 VSA software combine to form a region within which ...

  • Agilent Technologies 89600 - page 13

    13 In addition to the time record length limits, there are limits to the allowable range of spans. These are represented as vertical lines on the operating region graphs (Figure 4). The upper span limit is set by the digitizer’ s sample rate. For the highest sample rate supported by the digitizer this boundary may be further limited by the analog ...

  • Agilent Technologies 89600 - page 14

    14 Figure 5. The digitizer’ s available memory also constrains the operating region Figure 4. The allowable range of spans, which may differ by model, further bound the operating region of the digitizer 1.E-08 1.E-07 1.E-06 1.E-05 1.E-04 1.E-03 1.E-02 1.E-01 1.E+00 1.E+01 1.E+02 1.E+03 1.E+04 1.E+05 1.E+06 1.E+07 1.E+08 1.E+09 1.E+10 1.E+11 Span ...

  • Agilent Technologies 89600 - page 15

    15 In full rate sample mode with the use of appropriate lowpass fi ltering (T able 3) signal aliasing is completely avoided. In this case the “Operating Region” is renamed “ Alias Exposure Free Zone” to indicate this special condition. NOTE: Does not apply to HF operation. Although the operating regions for all digitizer models are similar ...

  • Agilent Technologies 89600 - page 16

    16 5.3 Recording limits for Acqiris digitizers The 89600 VSA software allows the acquisition of long records to be captured. Analysis is done as a post-processing operation. This feature is called Recording. Recordings can be much longer than time records in live measurements. Prior to taking a recording, live test measurements of Main Time length ...

  • Agilent Technologies 89600 - page 17

    17 As mentioned above, the user rate sampling modes increase the maximum Main Time length at the expense of possible aliasing of out-of-band signals into the frequency span of the measurement. The spectrum ranges subject to the alias conversion are called alias exposure zones. T o determine if any out-of-band signals exist in the exposure zones, ru ...

  • Agilent Technologies 89600 - page 18

    18 Figure 7. Alias Exposure Zone Checker results. The desired signal is at 150 MHz and an out-of- band signal is at –35 dBc and 642 MHz. The unwanted signal in the second alias zone is 5 dB above the –40 dBc threshold line. When no alias zones exist because the current sampling mode is alias free, no orange zone boxes will appear . In addition, ...

  • Agilent Technologies 89600 - page 19

    19 When the original center frequency/span is inadvertently tuned to an alias response of a signal whose frequency is not the center frequency , the checker will attempt to suggest the signal’ s frequency. The checker results at the desired center frequency location (box formed by white solid and dashed lines) will show no signal, and a large sig ...

  • Agilent Technologies 89600 - page 20

    20 Downconverting RF and Microwave Signals into the Range of the Agilent Acqiris Digitizers External components with wide bandwidths can be used to downconvert signals into the range of the broadband VSAs. The 89600 VSA software can account for the mixing equation of the external downconverter . While the downconverter cannot be controlled by the s ...

  • Agilent Technologies 89600 - page 21

    21 PC Requirements to Run 89600 Software The 89600 requires a PC connected via the PCI extender to the Acqiris digi- tizer or an embedded PC. The PC must meet or exceed the following minimum requirements: Appendix B Characteristic Microsoft® Windows® XP Professional CPU 600 MHz Pentium® or AMD-K6 > 600 MHz Pentium or AMD-K6 (> 2 GHz recomm ...

  • Agilent Technologies 89600 - page 22

    22 Span (Hz) Time record length (sec) Operating region Model: Acqiris U1065A (DC222, DC252, DC282) Memory: 256 Mpts and 1024 Mpts 1.E-01 1.E+00 1.E+01 1.E-02 1.E-04 1.E-03 1.E-05 5 Msa/s 10 Msa/s 20 Msa/s 50 Msa/s 100 Msa/s 250 Msa/s 200 Msa/s 500 Msa/s 1 Gsa/s 2 Gsa/s 4 Gsa/s 8 Gsa/s Minimum Time Rec Len 1.E-06 1.E-07 1.E-08 1.E +04 1.E +03 1.E +0 ...

  • Agilent Technologies 89600 - page 23

    23 For more product literature visit www .agilent.com/fi nd/89600 Related literature list 89600 Series V ector Signal Analyzers Installation and VXI Service Guide Go to www .agilent.com/fi nd/89600 then click on T echnical Support > Manuals. Understanding Time and Frequency Domain Interactions in the Agilent T echnologies 89400 Series V ector ...

  • Agilent Technologies 89600 - page 24

    Remove all doubt Our repair and calibration services will get your equipment back to you, performing like new , when prom- ised. Y ou will get full value out of your Agilent equipment through- out its lifetime. Y our equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnosti ...

Manufacturer Agilent Technologies Category Webcam

Documents that we receive from a manufacturer of a Agilent Technologies 89600 can be divided into several groups. They are, among others:
- Agilent Technologies technical drawings
- 89600 manuals
- Agilent Technologies product data sheets
- information booklets
- or energy labels Agilent Technologies 89600
All of them are important, but the most important information from the point of view of use of the device are in the user manual Agilent Technologies 89600.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Agilent Technologies 89600, service manual, brief instructions and user manuals Agilent Technologies 89600. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Agilent Technologies 89600.

A complete manual for the device Agilent Technologies 89600, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Agilent Technologies 89600 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Agilent Technologies 89600.

A complete Agilent Technologies manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Agilent Technologies 89600 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Agilent Technologies 89600, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Agilent Technologies 89600, that we can find in the current document
3. Tips how to use the basic functions of the device Agilent Technologies 89600 - which should help us in our first steps of using Agilent Technologies 89600
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Agilent Technologies 89600
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Agilent Technologies 89600 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Agilent Technologies 89600?

Use the form below

If you did not solve your problem by using a manual Agilent Technologies 89600, ask a question using the form below. If a user had a similar problem with Agilent Technologies 89600 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)