    Milwa uk ee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter ...

  • Milwaukee MW600 - page 2

    PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the meter. Install the battery into the battery clip connector while observing polarity. D.O. probes are shipped dry. Assemble the probe and prepare it for use, connect it to the meter and proceed as follows. 1. Remove the red and blac ...

  • Milwaukee MW600 - page 3

    PROBE POLARIZATION: • A dissolved oxygen probe must be polarized to function properly. • To polarize the probe, the fully assambled probe must be connected to the meter and the meter must be on. • During polarization (and during measurement), approximately 800 mV is applied to the cathode and anode inside the membrane and a chemical reaction ...

  • Milwaukee MW600 - page 4

    Zero Calibration: • Dip the probe into MA9070 zero oxygen solution and stir gently for 2-3 minutes. • Wait for the probe to reach it’s lowest stable reading. • Adjust the zero D.O. calibration trimmer (located on the front panel) until the display reads "0.0". Slope Calibration: It is suggested to perform the slope calibration in ...

  • Milwaukee MW600 - page 5

    TAKING MEASUREMENTS: Verify the probe is polarized and the probe and meter have been calibrated. Remove the protective cap from probe. Immerse the tip of the probe in the sample to be tested. For accurate dissolved oxygen measurements a minimum water movement of 0.3 m/sec is required. This is to ensure that the oxygen-depleted membrane surface is c ...

  • Milwaukee MW600 - page 6

    5 °C 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 0 m 14.6 13.8 13.1 12.4 11.8 11.3 10.8 10.3 9.9 9.5 9.1 8.7 8.4 8.1 7.8 7.5 7.3 7.1 6.8 6.6 6.4 300 m 14.1 13.3 12.7 12.0 11.4 10.9 10.4 9.9 9.7 9.2 8.8 8.4 8.1 7.8 7.5 7.2 7.1 6.9 6.6 6.4 6.2 600 m 13.6 12.9 12.2 11.6 11.0 10.5 10.1 9.6 9.2 8.7 8.5 8.1 7.8 7.5 7.3 7.0 6.8 6.6 6.3 6.2 ...

  • Milwaukee MW600 - page 7

    6 SALINITY COMPENSATION: The t able below illustrates the change in the solubility of oxygen in air saturated water as a result of chloride concentration or salinity . The t able can also be used to correct the displayed measurement. If you are making measurement s in salt water and know the chloride concentration (or salinity), you can multiply yo ...

  • Milwaukee MW600 - page 8

    7 PROBE & MEMBRANE MAINTENANCE: The D.O. probe body is made of reinforced plastic for maximum durability. A thermistor temperature sensor provides temperature measurements of the tested sample. It is always recommended to keep the protective cap on the probe when not in use, to protect the membrane against damage and dirt. To replace the membra ...

  • Milwaukee MW600 - page 9

    The Platinum cathode should always be bright and untarnished. If it is tarnished or stained, due to contact with certain gases or a damaged membrane cap, the cathode should be cleaned. You can use a clean lint-free cardboard or cloth. Rub the cathode very gently side to side 4-5 times. This will be enough to polish and remove any stains without dam ...

  • Milwaukee MW600 - page 10

    BATTERY REPLACEMENT: When the battery becomes weak the meter will display the low battery indicator " ". When this appears, only a few hours of battery life remain. A low battery will result in unreliable measurements. Prompt battery replacement is required. Battery replacement must take place in a non-hazardous area using an alkaline 9V ...

  • Milwaukee MW600 - page 11

    SPECIFICATIONS: MW600 RANGE 0.0 to 19.9 mg/L RESOLUTION 0.1 mg/L ACCURACY (@25°C) ±1.5% Full Scale TEMPERATURE Automatic from 0 to 30°C COMPENSATION CALIBRATION Manual on 2 points (zero and slope) LCD 3½ digits with symbols PROBE MA840 (included) ENVIRONMENT 0 to 50°C, 95% RH max. BATTERY TYPE 9V alkaline (included) BATTERY LIFE approximately ...

  • Milwaukee MW600 - page 12

    ISTMW600 01/10 Milwa uk ee Authorized Dealer: ...

Manufacturer Milwaukee Category Marine Instruments

Documents that we receive from a manufacturer of a Milwaukee MW600 can be divided into several groups. They are, among others:
- Milwaukee technical drawings
- MW600 manuals
- Milwaukee product data sheets
- information booklets
- or energy labels Milwaukee MW600
All of them are important, but the most important information from the point of view of use of the device are in the user manual Milwaukee MW600.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Milwaukee MW600, service manual, brief instructions and user manuals Milwaukee MW600. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Milwaukee MW600.

Similar manuals

A complete manual for the device Milwaukee MW600, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Milwaukee MW600 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Milwaukee MW600.

A complete Milwaukee manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Milwaukee MW600 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Milwaukee MW600, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Milwaukee MW600, that we can find in the current document
3. Tips how to use the basic functions of the device Milwaukee MW600 - which should help us in our first steps of using Milwaukee MW600
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Milwaukee MW600
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Milwaukee MW600 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Milwaukee MW600?

Use the form below

If you did not solve your problem by using a manual Milwaukee MW600, ask a question using the form below. If a user had a similar problem with Milwaukee MW600 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)