    HD 9021 INSTRUCTIONS MANUAL ...

  • Quantum HD 9021 - page 2

    HD 9021 1 2 5 4 6 3 7 9 8 11 10 13 19 21 20 22 23 24 26 25 27 12 14 15 16 17 18 30 29 28 31 32 33 34 35 ...

  • Quantum HD 9021 - page 3

    – 25 – HD 9021 ENGLISH QUANTUM - PHOTO - RADIOMETER ...

  • Quantum HD 9021 - page 4

    1 Input A socket, DIN 8-pole connector 2 Display 3 REL: the symbol indicates that the instrument is storing the Maximum, Minimum and Q energy values 4 MAX: the symbol indicates the maximum value of the chosen unit 5 MIN: the symbol indicates the minimum value of the chosen unit 6 A, B, A-B: the symbol indicates the relative value at the input A, B, ...

  • Quantum HD 9021 - page 5

    ENGLISH – 27 – Symbols lit besides the numbers Key Description All the symbols are lit for a few seconds after switching on. lx/fcd °C/°F RCD Key for switching the instrument on and off. The instrument switches off automatically about 8 minutes after this key has been pressed, as it has a built-in switch-off device. If any key other than the ...

  • Quantum HD 9021 - page 6

    shes. During the RCD phase the AUTO- POWER-OFF function is disabled, that is the instrument does not swit- ch off automatically. To switch off, press the ON/OFF key. If the stored data are not reset they remain in the memory until REC is pressed again. When the key is pressed for the first time it starts storing data in the memory, the second time ...

  • Quantum HD 9021 - page 7

    When the MAN key is pressed the user chooses what scale to operate on. There are 4 scales available, except for temperature measure- ment where there are 2. When this key is pressed the Q Energy integration time that has elapsed since the REC key was pressed appears on the display in seconds, minutes or hours. To quit the routine, press the key for ...

  • Quantum HD 9021 - page 8

    if a high value is to be set. This key has two functions; in the first, when Q is set, it increases the value; the second function enables the serial output. When the key is pressed only once, the figure shown on the display is sent to the serial output. If the key is pressed for more than three seconds the continuous serial output is enabled; this ...

  • Quantum HD 9021 - page 9

    ENGLISH – 31 – PROBE CONNECTION The HD 9021 may be fitted with either one or two probes; the instrument has two DIN 45326 8-pole inputs, input A and input B. If only one input is used, be careful when using the keys for input A or B. The instrument recognizes the type of probe connected, which is distingui- shed by a suitable code; it then gets ...

  • Quantum HD 9021 - page 10

    The light emitted by a lamp varies with the square of the current, so there may very well be variations in values during the test which the human eye does not detect. When measuring temperature choose the probe most suited for the case: immersion, penetration or surface probes. It is sufficient to immerse the probe in the liquid of which you want t ...

  • Quantum HD 9021 - page 11

    LOW BATTERY WARNING If the battery voltage falls below acceptable levels, a beep sounds every 30 seconds and the symbol appears. From that moment there remains about 1 hour autonomous operation, however the battery should be replaced as soon as possible. To change the battery turn the screw on the door of the battery compartment in an anti-clockwis ...

  • Quantum HD 9021 - page 12

    MAINTENANCE Storage conditions. * Temperature: -20 to +60°C * Humidity: less than 85% relative humidity. * Do not store the instrument in places where: 1 - There is a high degree of humidity. 2 - The instrument is exposed to direct sunlight. 3 - The instrument is exposed to a source of high temperature. 4 - There are strong vibrations. 5 - There i ...

  • Quantum HD 9021 - page 13

    ANALOG OUTPUT It is possible to link up with the SUB D male 9-pole connector of the instru- ment to obtain an analog output: ENGLISH – 35 – DTE DCE AD RS 232C - Data communication equipment - Data terminal equipment MODEM 14 15 16 17 18 19 20 21 22 23 24 25 123456789 1 0 1 1 1 2 1 3 SCR TXD RXD RTS CTS DSR GND DCD NC NC NC NC NC NC TC NC RC NC ...

  • Quantum HD 9021 - page 14

    ENGLISH – 36 – BLOCK DIAGRAM MICROPROCESSOR LCD LOW VOL T AGE DETECTOR RS232 KEYBOARD A/D CONVERTER INPUT A CONNECTOR INPUT B CONNECTOR Signal Signal Probe T ype Probe T ype (Range) (Range) 500 mV TR2 POWER SUPPL Y TR1 4V ...

  • Quantum HD 9021 - page 15

    ENGLISH – 37 – LAYOUT OF COMPONENTS J3 U5 J4 R19 R20 R12 R14 R1 R2 R45-51 R27 C13 C12 X1 R30 R28 C15 R32 R26 R41 C10 C1 1 C9 R16 R15 R37 R17 R18 R38 U2 U4 R23 C5 R33 R34 R9 TR2 B2 R33 R34 R23 J5 U3 R39 R6 R6 C4 R8 D1 R10 C14 R36 R35 R1 1 R29 C1 OUTB OUT A ICON1 U1 R5 R3 R25 R4 C3 C2 TR1 R40 R24 C7 R4C6 C8 C16 Q4 Q1 DZ1 G2 G3 TR2 TR1 ...

  • Quantum HD 9021 - page 16

    CALIBRATING THE INSTRUMENT HD 9021 1) Open the instrument. 2) Connect the prods of a precision meter to the following pins on the input A (or B) connector: the positive to pin 7 (V+) and the negative to pin 6 (GND). 3) With the instrument and the meter switched on, turn the trimmer TR1 until the display on the meter indicates 4Vdc. 4) Insert the ph ...

  • Quantum HD 9021 - page 17

    ENGLISH – 39 – CHARACTERISTICS - Microprocessor-controlled multifunction quantum-photo-radiometric indi- cator with LCD indication for measuring sources of light and temperature. Measures ILLUMINANCE (lux, foot-candle), IRRADIANCE (W/m 2 ) and RADIANCE (cd/m 2 ). - 2 inputs for probes and one for RS232C (with an optional adapter AD RS232C). - P ...

  • Quantum HD 9021 - page 18

    - Display: LCD 12 mm, function and photo-radiometric symbols. - Functions: Automatic and manual change of scale, integration time up to 1999 hours, auto power off. Storage and updating of the Maximum, Minimum and Q (Energy) values. Optoinsulated serial output RS232C with adapter AD RS232C. Analog output 1.9 mV/°C, 2.5 mV/lux, 250 mV/W. - Power sup ...

  • Quantum HD 9021 - page 19

    ORDER CODE HD 9021 K : The kit is composed of the HD 9021 instrument, diplomatic carrying case, without probe. HD 9021 : Photometric probe for measuring light, illuminance , photopic PHOT/C correction filter complying with CIE, diffuser for cosine correc- tion. Range 0÷ 200,000 lux. HD 9021 : Radiometric probe for measuring light energy, irradianc ...

  • Quantum HD 9021 - page 20

    PROBES The primary purpose of the probes is to detect and measure light in its various aspects; for this reason there are different probes available to suit the different requirements. Probes are required for measuring illuminance (lux, foot-candle), irradiance (W/m 2 ) and luminance (cd/m 2 ). Each sensor has its own work band completed by filters ...

  • Quantum HD 9021 - page 21

    HD 9021 RAD/C Radiometric probe for measuring the IRRADIANCE of light sources, irra- diance of the sun, etc. Radiometric filter and diffuser for angle correction according to the cosine law. Spectral measurement range 400÷ 900 nm, range from 1 microwatt/cm 2 to 200 milliwatt/cm 2 . The probe is used in greenhouses for measuring the power of artifi ...

  • Quantum HD 9021 - page 22

    The typical response curve of the probe is as follows: The study and field of application of ultraviolet rays has seen considerable growth in the last decades. The ultraviolet field is conventionally subdivi- ded into the following three "bands": - the band between the wavelengths from 400 to 315 nm. which includes radiations defined as t ...

  • Quantum HD 9021 - page 23

    UVC rays are useful thanks to their bactericide and ozonizing capacity, so they are used in the field of sterilization; another field of application that is being developed is in the polymerization of adhesives or resins (photoresi- st); as a drawback they cause more serious cases of erythema and conjunctivitis than UVB rays. Depending on the effec ...

  • Quantum HD 9021 - page 24

    HD 9021 UVB Radiometric probe for measuring radiation power, IRRADIANCE , in the UVB wave length 280÷ 315 nm, peak at 312 nm. Quartz diffuser for cosine correction. Measuring range from 10 nanowatt/cm 2 to 200 milliwatt/cm 2 . The typical response curve of the probe is as follows: HD 9021 UVC Radiometric probe for measuring radiation power, IRRADI ...

  • Quantum HD 9021 - page 25

    – 47 – HD 9021/Cd Probe for measuring LUMINANCE , measuring range from 0 to 1999x10 3 candles/cm 2 . Measuring angle 6° with opening surface 25 mm. CIE filter for correction according to the human eye. - The CANDLE (cd) is the light intensity in a specific direction of a source that emits monochrome radiation at a frequency of 540x10 12 Hz and ...

  • Quantum HD 9021 - page 26

    ENGLISH – 48 – CE CONFORMITY Safety EN61000-4-2, EN61010-1 level 3 Electrostatic discharge EN61000-4-2 level 3 Electric fast transients EN61000-4-4 level 3 Voltage variations EN61000-4-11 Electromagnetic interference IEC1000-4-3 sucseptibility Electromagnetic interference EN55020 class B emission GUARANTEE CONDITIONS All our appliances have bee ...

Manufacturer Quantum Category Marine Radio

Documents that we receive from a manufacturer of a Quantum HD 9021 can be divided into several groups. They are, among others:
- Quantum technical drawings
- HD 9021 manuals
- Quantum product data sheets
- information booklets
- or energy labels Quantum HD 9021
All of them are important, but the most important information from the point of view of use of the device are in the user manual Quantum HD 9021.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Quantum HD 9021, service manual, brief instructions and user manuals Quantum HD 9021. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Quantum HD 9021.

A complete manual for the device Quantum HD 9021, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Quantum HD 9021 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Quantum HD 9021.

A complete Quantum manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Quantum HD 9021 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Quantum HD 9021, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Quantum HD 9021, that we can find in the current document
3. Tips how to use the basic functions of the device Quantum HD 9021 - which should help us in our first steps of using Quantum HD 9021
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Quantum HD 9021
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Quantum HD 9021 in a specific country, if it was not possible to solve the problem on our own.