    NI PXI/PCI-5124 Specifications 12-Bit 200 MS/s Digitizer This document lists the sp ecifications for the NI PX I/PCI-5124 (NI 5124) high-speed digitizer. Unless otherw ise noted, these specifications are valid for the following conditions: • All filter settings • All impedance selections • Sample clock set to 200 MS/s using onboard clock T yp ...

  • National Instruments NI PXI/PCI-5124 - page 2

    NI PXI/PCI-5124 Specifications 2 ni.com PFI 0 and PFI 1 (Programmabl e Function Interfac e, AUX Front Panel Connector) ................... .................... ................... .....20 TClk Specifications .............. .................... .................... ...................... .....21 Waveform Specifications ........... .................. ...

  • National Instruments NI PXI/PCI-5124 - page 3

    © National Instruments Corporat ion 3 NI PXI/PCI-5 124 Specifications Voltage Levels Full Scale (FS) Input Range and Programmable V ertical Offset 50 Ω 1M Ω — Range (V pk-pk ) V ertical Of fset Range (V) Range (V pk-pk ) V ertical Offs et Range (V) 0.2 ±0.1 0.2 ±0.1 0.4 ±0.2 0.4 ±0.2 1 ±0.5 1 ±0.5 2 ±1 2 ±1 4 ±2 4 ±2 10 0 10 ±5 20 0 ...

  • National Instruments NI PXI/PCI-5124 - page 4

    NI PXI/PCI-5124 Specifications 4 ni.com DC Drift Range (V pk-pk ) 50 Ω and 1 M Ω — 0.2, 0.4, 1, and 2 ±(0.057% of Input + 0.006% of FS +1 0 0 μ V) per °C 4, 10, and 20 (1 M Ω only) ±(0.057% of Input + 0.006% of FS + 900 µV) per °C A C Amplitude Accuracy 50 Ω 1M Ω W ithin ±5 °C of self-calibration temperature ±0.06 dB (±0.7%) at 50 ...

  • National Instruments NI PXI/PCI-5124 - page 5

    © National Instruments Corporat ion 5 NI PXI/PCI-5 124 Specifications Figure 1. NI 5124 Frequency Response (T ypical) A C Coupling Cutof f (–3 dB) 12 Hz A C coupling av ailable on 1M Ω only Passband Flatness Filter Settings Range (V pk-pk ) 50 Ω and 1 M Ω Referenced to 50 kHz Filters Off All ranges except 0.2 ±0.5 dB DC to 20 MHz ±1.0 dB 20 ...

  • National Instruments NI PXI/PCI-5124 - page 6

    NI PXI/PCI-5124 Specifications 6 ni.com Specif ication Va l u e Comments Spectral Characteristics Spurious-Free Dynamic Range with Harmonics (SFDR) (Physical), T ypical Range (V pk-pk ) 50 Ω 1M Ω Filters of f or antialias filter on 10 MHz, –1 dBFS input signal Includes the 2nd through the 5th harmonics Measured from DC to 100 MHz on NI PXI-5124 ...

  • National Instruments NI PXI/PCI-5124 - page 7

    © National Instruments Corporat ion 7 NI PXI/PCI-5 124 Specifications Intermodulation Distortion, T ypical 0.2 V pk-pk to 2.0 V pk-pk Ranges on 50 Ω Input Filters of f or antialias filter on T wo tones at 10.2 MHz and 11.2 MHz Each tone is –7 dBFS –75 dBc Signal-to-Noise Ratio (SNR), T ypical Range (V pk-pk ) 50 Ω 1M Ω Excludes harmonics 10 ...

  • National Instruments NI PXI/PCI-5124 - page 8

    NI PXI/PCI-5124 Specifications 8 ni.com Figure 2. NI 5124 Dynamic Performance, 50 Ω , 1 V pk-pk Range, 262,144 Point FF T (T ypical) –110 –100 –90 –80 –70 –60 –50 –40 –30 –20 –10 0 Amplitude (dBFS) 10 M 20 M 30 M 40 M 50 M 60 M 70 M 80 M 90 M 100 M 1 M F requency (Hz) 9.83 MHz, –1 dBFS Input Signal ...

  • National Instruments NI PXI/PCI-5124 - page 9

    © National Instruments Corporat ion 9 NI PXI/PCI-5 124 Specifications Specif ication Va l u e Comments RMS Noise Range (V pk-pk ) 50 Ω 1M Ω Noise filter on 50 Ω terminator connected to input 0.2 NI PXI-5124: 94 μ V rms (0.047% FS) NI PCI-5124: 106 μ V rms (0.053% FS) NI PXI-5124 : 104 μ V rms (0.052% FS) NI PCI-5124: 116 μ V rms (0.058% FS) ...

  • National Instruments NI PXI/PCI-5124 - page 10

    NI PXI/PCI-5124 Specifications 10 ni.com RMS Noise Range (V pk-pk ) 50 Ω 1M Ω Antiali as fil t e r on 50 Ω terminator connected to input 0.2 NI PXI-5124: 112 μ V rms (0.056% FS) NI PCI-5124: 126 μ V rms (0.063% FS) NI PXI-5124 : 130 μ V rms (0.065% FS) NI PCI-5124: 146 μ V rms (0.073% FS) 0.4 200 μ V rms (0.05% FS) 216 μ V rms (0.054% FS) ...

  • National Instruments NI PXI/PCI-5124 - page 11

    © National Instruments Corporat ion 11 NI PXI/PCI-5 124 Specifications RMS Noise Range (V pk-pk ) 50 Ω 1M Ω Filters off 50 Ω terminator connected to input 0.2 NI PXI-5124: 114 μ V rms (0.057% FS) NI PCI-5124: 128 μ V rms (0.064% FS) NI PXI-5124 : 164 μ V rms (0.082% FS) NI PCI-5124: 184 μ V rms (0.092% FS) 0.4 204 μ V rms (0.051% FS) 264 ? ...

  • National Instruments NI PXI/PCI-5124 - page 12

    NI PXI/PCI-5124 Specifications 12 ni.com Figure 3. Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Noise Filter Enabled, 1 M Ω Input Impedance Figure 4. Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Full Bandwidth, 50 Ω Input Impedance 60 n 50 n 40 n 30 n 20 n 10 n 1 n 1 k 10 k 100 k 1 M 10 M 100 M F requency ( ...

  • National Instruments NI PXI/PCI-5124 - page 13

    © National Instruments Corporat ion 13 NI PXI/PCI-5 124 Specifications Horizontal Sample Clock Specif ication Va l u e Comments Sources NI PXI-5124 NI PCI-5124 * Internal Sample Clock is locked to the Reference Clock or deriv ed from the onboard VCXO Internal, Onboard Clock (internal VCXO) * External, CLK IN (front panel SMB connector) External, P ...

  • National Instruments NI PXI/PCI-5124 - page 14

    NI PXI/PCI-5124 Specifications 14 ni.com Sample Clock Jitter , T ypical ≤ 1 ps rms (10 0 Hz to 100 kHz) ≤ 2p s r m s ( 1 0 0 H z t o 1M H z ) Includes the ef fects of the con verter aperture uncertainty and the clock circuitry jitter Excludes trigger jitter T imebase Frequency 200 MHz — T imebase Accuracy Not Phase-Locked to Reference Clock P ...

  • National Instruments NI PXI/PCI-5124 - page 15

    © National Instruments Corporat ion 15 NI PXI/PCI-5 124 Specifications External Sample Clock Sources NI PXI-5124 NI PCI-5124 — CLK IN (front panel SMB connector) PXI Star Trigger (backplane connector) CLK IN (front panel SMB connector) Frequenc y Range 50 MHz to 210 MHz (CLK IN) 50 MHz to 80 MHz (PXI Star T rigger) Di vide by n decimation av ail ...

  • National Instruments NI PXI/PCI-5124 - page 16

    NI PXI/PCI-5124 Specifications 16 ni.com Phase-Locked Loo p (PLL) Reference Clock Sample Clock Exporting Exported Sample Clock Destinations Destination Maximum Frequency * Decimated Sample Clock only CLK OUT (front panel SMB connector) 210 MHz PXI_T rig<0..6> (backplane connector) * 20 MHz PFI<0..1> (front panel 9-pin mini-circular DIN ...

  • National Instruments NI PXI/PCI-5124 - page 17

    © National Instruments Corporat ion 17 NI PXI/PCI-5 124 Specifications CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) CLK OUT (Sample Clock and Reference Clock Output, Front Panel Connector) T rigger Reference (Stop) T rigger Specification Va l u e Comments Input V oltage Range Sine wav e: 0.65 V pk-pk to 2.8 V pk-pk (0 dBm ...

  • National Instruments NI PXI/PCI-5124 - page 18

    NI PXI/PCI-5124 Specifications 18 ni.com Ti m e Resolution TDC Onboard Clock External Clock TDC = T ime to Digita l Con version Circuit On 50 ps N/A Of f 5n s External Clock Period Rearm T ime TDC Rearm T ime Ho ldoff s et to 0 On 10 μ s Of f 2 μ s Holdoff TDC Onboard Clock External Clock — On 10 μ s to 85.899 s N/A Of f 2 μ s to 85.8 99 s 20 ...

  • National Instruments NI PXI/PCI-5124 - page 19

    © National Instruments Corporat ion 19 NI PXI/PCI-5 124 Specifications TRIG (External T rigger , Front Panel Connector) T rigger Filters Lo w-Frequenc y (LF) Reject High-Frequency (HF) Reject — 50 kHz 50 kHz Digital Trigger (Digital Trig ger Type) Sources NI PXI-5124 NI PCI-5124 — PXI_T rig<0..6> (backplane connector) PFI<0..1> (fr ...

  • National Instruments NI PXI/PCI-5124 - page 20

    NI PXI/PCI-5124 Specifications 20 ni.com PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connector) Specif ication Va l u e Co mments Connector 9-pin mini-circu lar DIN — Direction Bi-directiona l — As an Input (Trigger) Destinations Start T rigger (Acquisition Arm) Reference (Stop) T rigger Arm Reference Trigger Advance T rig ...

  • National Instruments NI PXI/PCI-5124 - page 21

    © National Instruments Corporat ion 21 NI PXI/PCI-5 124 Specifications TClk Specifications National Instruments TClk synchronizati on method and the NI-TClk driver are used to align the sample clocks on any numb er of SMC-based modules in a chassis. For more inform ation about TClk synchronizati on, refer to the NI-TClk Synchronization Help , wh i ...

  • National Instruments NI PXI/PCI-5124 - page 22

    NI PXI/PCI-5124 Specifications 22 ni.com W aveform Specifications Specification Va l u e Comments Onboard Memory Size 8 MB per channel stan dard (4 megasamples per channel) * NI PXI-5124 only 32 MB per channel option (16 megasamples per channel) 256 MB per channel option (128 megasamples per channel) 512 MB per channel option (256 megasamples per c ...

  • National Instruments NI PXI/PCI-5124 - page 23

    © National Instruments Corporat ion 23 NI PXI/PCI-5 124 Specifications Calibration Power Specification Va l u e Comments Self-Calibrati on Self-calibration i s done on soft ware command. The calibration corrects for gain, offset, frequency response, triggering, and timing adjust ment errors for all input ranges. — External Calibrat ion (Factory ...

  • National Instruments NI PXI/PCI-5124 - page 24

    NI PXI/PCI-5124 Specifications 24 ni.com Software Specif ication Va l u e Comments Dri ver Software NI-SCOPE 2.7 or later NI-SCOPE is an IVI-comp liant driv er that allows you to configure, control, and calib rate the NI 5124. NI-SCOPE provides application program ming interf aces for many dev elopment environments. — Application Software NI-SCOP ...

  • National Instruments NI PXI/PCI-5124 - page 25

    © National Instruments Corporat ion 25 NI PXI/PCI-5 124 Specifications Environment NI PXI-5124 Note T o ensure that the NI PXI-5124 cools ef fectiv ely , follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5124 kit. The NI PXI-5124 is intended for indoor us e only . Specif ication Va l u e Comments Operati ...

  • National Instruments NI PXI/PCI-5124 - page 26

    NI PXI/PCI-5124 Specifications 26 ni.com NI PCI-5124 Note T o ensure that the NI PCI-5124 cools effecti v ely , make sure that the chassis in which it is used has activ e cooling that provides at least some airflow acros s the PCI card cage. T o maximize airflo w and extend the life of the de vice, leav e any adjacent PCI slots empty . Refer to the ...

  • National Instruments NI PXI/PCI-5124 - page 27

    © National Instruments Corporat ion 27 NI PXI/PCI-5 124 Specifications Safety , Electromagnetic Compatibility , and CE Compliance Safety This product is designed to meet the requirements of the follow ing standards of safety for electrical equi pment for measurement, control, and laboratory use: • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 6101 ...

  • National Instruments NI PXI/PCI-5124 - page 28

    NI PXI/PCI-5124 Specifications 28 ni.com Environmental Management National Instruments i s committed to designing and manu facturing products in an environmental ly responsible mann er. NI recognizes that eliminating certa in hazardous substances from our products is beneficial not only to t he environment but also to NI customers. For additional e ...

  • National Instruments NI PXI/PCI-5124 - page 29

    © National Instruments Corporat ion 29 NI PXI/PCI-5 124 Specifications Dimensions and W eight NI PXI-5124 Front Panel I ndicators Label Function Fo r m or e information, refer to the NI High-Speed Digitizers Help . A CCESS The A CCESS LED indicates the status of the PCI b us and the interface from the NI PXI-5124 to the co ntroller . AC T I V E Th ...

  • National Instruments NI PXI/PCI-5124 - page 30

    NI PXI/PCI-5124 Specifications 30 ni.com NI PCI-5124 Dimensions 35.5 × 2.0 × 11.3 cm (14.0 × 0.8 × 4.4 in.) We i g h t 455 g (16 oz) 2.0 cm (0.8 in.) 35.5 cm (14.0 in.) 11.3 cm (4.4 in.) ...

  • National Instruments NI PXI/PCI-5124 - page 31

    © National Instruments Corporat ion 31 NI PXI/PCI-5 124 Specifications Where to Go for Support The National Instrum ents W eb site is your complete resource for technical support. At ni.com/support you hav e access to e verything from troubleshooting and ap plication dev el opment self-help resources to email and phone assistance from NI Applicati ...

  • National Instruments NI PXI/PCI-5124 - page 32

    National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more inform ation about Natio nal Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s. For patents covering Natio ...

Manufacturer National Instruments Category Digital Photo Keychain

Documents that we receive from a manufacturer of a National Instruments NI PXI/PCI-5124 can be divided into several groups. They are, among others:
- National Instruments technical drawings
- NI PXI/PCI-5124 manuals
- National Instruments product data sheets
- information booklets
- or energy labels National Instruments NI PXI/PCI-5124
All of them are important, but the most important information from the point of view of use of the device are in the user manual National Instruments NI PXI/PCI-5124.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals National Instruments NI PXI/PCI-5124, service manual, brief instructions and user manuals National Instruments NI PXI/PCI-5124. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product National Instruments NI PXI/PCI-5124.

A complete manual for the device National Instruments NI PXI/PCI-5124, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use National Instruments NI PXI/PCI-5124 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of National Instruments NI PXI/PCI-5124.

A complete National Instruments manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual National Instruments NI PXI/PCI-5124 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the National Instruments NI PXI/PCI-5124, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the National Instruments NI PXI/PCI-5124, that we can find in the current document
3. Tips how to use the basic functions of the device National Instruments NI PXI/PCI-5124 - which should help us in our first steps of using National Instruments NI PXI/PCI-5124
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with National Instruments NI PXI/PCI-5124
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of National Instruments NI PXI/PCI-5124 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning National Instruments NI PXI/PCI-5124?

Use the form below

If you did not solve your problem by using a manual National Instruments NI PXI/PCI-5124, ask a question using the form below. If a user had a similar problem with National Instruments NI PXI/PCI-5124 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)