  • Fluke 3125 - page 1

    3125 Detachable Hot Plate Users’ Manual Rev. 622101 Hart Scientific http://www.rongqe.cn/ ...

  • Fluke 3125 - page 2

    Limited W arranty & Limitation of Liability Each product from Fluke's Hart Scientific Di vision ("Hart") is warranted to be free from defects in mate - rial and workmanship under normal use and service. The warranty period is one year for the Surface Cal - ibrator . The warranty period be gins on the date of the shipment. Parts, ...

  • Fluke 3125 - page 3

    T able of Contents 1 Bef ore Y ou Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 W arnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.2 Cautio ...

  • Fluke 3125 - page 4

    ...

  • Fluke 3125 - page 5

    1 Bef ore Y ou Start 1.1 Symbols Used T able 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manu al. Symbol Description AC ( Alternating Current) AC- D C Batter y CE Complies with European Uni on Directiv es DC Double Insulated Electric Shock Fuse PE Ground Hot Surf ace (Burn Haza ...

  • Fluke 3125 - page 6

    Symbol Description Canadian Standards Associati on O VER VOL T AGE ( Installation) CA TEGORY II, Polluti on Degree 2 per IEC1010- 1 re - f ers to the le v el of Impul se Withstand V oltage protection provided. Equipment of O VER VOL T AGE CA TEGOR Y II is energy-consumi ng equipment to be supplied f rom the fix ed install ation. Examples include ho ...

  • Fluke 3125 - page 7

    cautions. Before working inside the equipment, turn the power off and disconnect the power cord. • Always replace the fuse with one of the same rating, voltage, and type. • This instrument is intended for indoor use only. • Overhead clearance is required. DO NO T place this instrument under a cabinet or other structure. • DO NO T use this u ...

  • Fluke 3125 - page 8

    1.2.2 Cautions • DO NO T plug the unit into 230V if the heater switches and fuse holder read 115V . This action will cause the fuses to blow and may damage the instrument. • Components and heater lifetime can be shortened by continuous high temperature operation. • DO NO T change the values of the calibration constants from the factory set va ...

  • Fluke 3125 - page 9

    NETHERLANDS Phone: +31-402-675300 T elefax: +31-402-675321 E-mail: ServiceDesk@fluke.nl Fluke Int'l Corporation Service Center - Instrimpex Room 2301 Sciteck T ower 22 Jianguomenwai Dajie Chao Y ang District Beijing 100004, PRC CHIN A Phone: +86-10-6-512-3436 T elefax: +86-10-6-512-3437 E-mail: xingye.han@fluke.com.cn Fluke South East Asia Pte ...

  • Fluke 3125 - page 10

    ...

  • Fluke 3125 - page 11

    2 Intr oduction Hart Scientif ic’ s Model 3125 Detachable Hot Plate is controlled by Hart Scien - tif ic’ s 2200 controller and uses a precision pl atinum R TD as a sensor with a heater to control the t emperature. The 3125 and 2200 are a matched set . The controller display sho ws the temperat ure and also the set-point tempera - ture. The tem ...

  • Fluke 3125 - page 12

    ...

  • Fluke 3125 - page 13

    3 Specifications and En vir onmental Conditions 3.1 Specifications Temperature Range 35°C to 400°C (95°F to 752°F) Display Accuracy ±0.5°C to 200°C ±1.0°C to 400°C Stability ±0.2°C to 300°C ±0.3°C to 400°C Resolution 0.01° Uniformity ±0.3°C at 100°C ±0.6°C at 200°C ±0.9°C at 300°C ±1.4°C at 400°C Heating Time 25 ° C to ...

  • Fluke 3125 - page 14

    • ambient relati ve humidity: maximum 80% for temperature <31°C, de - creasing linearly to 50% at 40°C • pressure: 75kPa-106kPa • mains voltage within ±10% of nominal • vibrations in the calibration en vironment should be minimized • altitude does not effect the performance or safety of the unit 10 ...

  • Fluke 3125 - page 15

    4 T r oubleshooting 4.1 T roub leshooting Pr ob lems, P ossib le Causes, and Solutions In the ev ent that the instrument a ppears to function abnormally , this section may help to fi nd and solve the problem. Se v eral possible problem condi tions are described along with likely causes and solutions. If a problem arises, please read this section ca ...

  • Fluke 3125 - page 16

    Problem Possible Causes and Solutions The display shows any of the following: err 1 , err 2 , err 3 , err 4 , Err 5 , Err 6 ,o r Err 7 Controller probl em. The error messages signi fy the following pr oblems with the controlle r. Err 1 - a RAM error Err 2 - a NVRAM error Err 3 - a Structure error Err 4 - an ADC setup e rror Err 5 - an ADC ready err ...

Manufacturer Fluke Category Food Warmer

Documents that we receive from a manufacturer of a Fluke 3125 can be divided into several groups. They are, among others:
- Fluke technical drawings
- 3125 manuals
- Fluke product data sheets
- information booklets
- or energy labels Fluke 3125
All of them are important, but the most important information from the point of view of use of the device are in the user manual Fluke 3125.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Fluke 3125, service manual, brief instructions and user manuals Fluke 3125. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Fluke 3125.

Similar manuals

A complete manual for the device Fluke 3125, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Fluke 3125 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Fluke 3125.

A complete Fluke manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Fluke 3125 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Fluke 3125, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Fluke 3125, that we can find in the current document
3. Tips how to use the basic functions of the device Fluke 3125 - which should help us in our first steps of using Fluke 3125
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Fluke 3125
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Fluke 3125 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Fluke 3125?

Use the form below

If you did not solve your problem by using a manual Fluke 3125, ask a question using the form below. If a user had a similar problem with Fluke 3125 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)