    PROGRAMMABLE PROCESS MONITORS USER ’ S MANUAL DP3600, DP3700 AND DP3800 SERIE S OMEGA ENGINEERING, INC., ONE OMEGA DRIVE, STAMFORD, CT 06907. MANUAL: M881/0399 ...

  • Omega Engineering DP3600 - page 2

    1 SPECIFICATIONS ............................................................................................................................................................................................... 3 DESCRIPTION ............................................................................................................................... ...

  • Omega Engineering DP3600 - page 3

    2 SCALING .............................................................................................................................................................................................................. 12 OFFSET ........................................................................................................................... ...

  • Omega Engineering DP3600 - page 4

    3 SPECIFICATIONS (typical @ 25 C and rated supply voltage unless otherwise specified). STANDARD INPUT: * 4 - 20 MA, 10 - 50 MA, 0 - 5 VDC, 0 - 10 VDC EXTERNAL HOLD: * An external low level signal applied to the HOLD pin on the rear connector will initiate hold function. ANALOG TO DIGITAL CONVERSION: * 41/2 Digit A/D conversion. * Dual slope integra ...

  • Omega Engineering DP3600 - page 5

    4 DESCRIPTION The DP3600 series of instrumen ts are micro - processor based intelligent monitors that are capable of performing several tasks in addition to displaying the process variable. The micro - processor technology used in the system design has provided these instruments a cost and performance edge that is simply not possible with analog sy ...

  • Omega Engineering DP3600 - page 6

    5 PROGRAM LIMITS 1. Process Limit1 2. Process Limit2 3. Setup limt 3 f or Rate/Process 4. Process Limit 3 5. Setup Limit4 for Process/Deviation 6. Process Limit 4 7. Timer 1 Limit 8. Timer 2 Limit 9. Deadband for limits 9. Limits Normally open/close (1 Thru 4) 10. Limits Latching/Non Latching 11. Audio Alarm On/Off PROGRAM SYSTEM 1. Display Options ...

  • Omega Engineering DP3600 - page 7

    6 Offset: Next configurable function is offset and is indicated by the message ‘OFFSET’ followed by current offset value. To change the v alue of flashing digit press ^v key. To move the flashing digit, push <NXT> key. After programming the desired value of offset, push the "SETUP" key to get to the next function. Tare: Tare mod ...

  • Omega Engineering DP3600 - page 8

    7 Selection of Limit 4 as Process or Deviation Limit: This step comes after programming Limit 3 and is indicated by the message ‘LT3 Pr’ (if Limit 3 is setup as Process limit ) or ‘LT3 dEN’ (if Limit 3 is setup as deviation limit). Alternate selection can be made by pushing ^v key. Push SETUP key after making the selection. Programming Limi ...

  • Omega Engineering DP3600 - page 9

    8 Audio Alarm On/Off: This unit supports an audio alarm which comes on when ever a limit value is exceeded. This alarm can be turned On or Off, as desired. The selection is made by pushing ^v key. If the alarm is ON, the display will read ‘ALR ON’ (Alarm On). Alternately, it will display ‘ALR OFF’ (for ‘Alarm Off’). Pus h SETUP key afte ...

  • Omega Engineering DP3600 - page 10

    9 Rate Time - Base: This step is for programming time - base for calculation of Rate. Units of time - base are seco nds. Again, the active digit will be flashing. To change the value, push ^v key. To activate another digit for change, use <NXT> key. On finishing, push SETUP key. Remote Tare/Hold: Remote Tare/Hold feature lets you select betwe ...

  • Omega Engineering DP3600 - page 11

    10 SET UP FLOW CHART: SETUP SCALING (SCALING) HIGH SCALE (HI SCLE) LOW SCALE (LO SCLE) OFFSET (OFFSET) TARE (TARE) PROGRAM PROCESS LIMIT 1 (PrG Lt1 -- 01.234) PROGRAM PROCESS LIMIT 2 (PrG Lt2 -- 01.234) LIMIT 3 RATE OR PROCESS (Lt3 rtE) or (Lt3 Pr) PROGRAM LIMIT 3 VALUE (PrG Lt3 -- 02.123) LIMIT4 PROCESS / DEVIATION (Lt3 dEN) or (Lt3 Pr) TIMER 1 LI ...

  • Omega Engineering DP3600 - page 12

    11 MAX. AND MIN. READIN GS (PEAKS) Displaying: To display the peak values of Process variable push Peak key once and let go. The display will first indicate ‘ HIGH Pt’ briefly and then the value of the highest Process variable value monitored by the system. A second push while high peak is being displayed will bring up low peak value on the dis ...

  • Omega Engineering DP3600 - page 13

    12 PROCESS CHANGE Displaying: The system will constantly monitor the total amount of change in process variable since start or last reset. This change is displayable in engineering units by pushing the ‘PR CNG’ k ey. This function is useful in tracking the change that has taken place in the process e.g. in a tank containing 300 gallons of water ...

  • Omega Engineering DP3600 - page 14

    13 TARE Tare function is used for taking out a preset amount of process variable from the total reading e.g. if we are measuring weight of a certain p roduct being put in a container but do not want the weight of the container to show on the display, then enter the weight of the container as tare value. The DP3600 unit will then automatically subtr ...

  • Omega Engineering DP3600 - page 15

    14 Timer Limits: Displaying: To display timer limits, first push the TIME key. While still in this mode go on to push LIMITS key. The display will briefly show TR1 LT (for timer 1 Limit) and then the previously programmed value. Pushing the LIMITS key one more time will display TR2 LT followed by pre - programmed timer 2 value. Programming of timer ...

  • Omega Engineering DP3600 - page 16

    15 Open Collector Option Whenever a limit is reached an open collector output provides a 5 Vdc signal at 50ma. on the output connector, e.g... Limit 1 provides its 5 volt output signal between pins 1 and 2. Figure 2. Open Collector Hookup Example HOLD FUNCTION The hold function allows to freeze the Process Variable reading in the display window ind ...

  • Omega Engineering DP3600 - page 17

    16 PIN NO DESCRIPTION 1 LIMIT 1 COMMON/OPEN COLLECTOR POSITIVE 2 LIMIT 1 NORMALLY OPEN/OPEN COLLECTOR NEGA TIVE 3 LIMIT 2 COMMON/OPEN COLLECTOR POSITIVE 4 LIMIT 2 NORMALLY OPEN/OPEN COLLECTOR NEGATIVE 5 LIMIT 3 COMMON/OPEN COLLECTOR POSITIVE 6 LIMIT 3 NORMALLY OPEN/OPEN COLLECTOR NEGATIVE 7 LIMIT 4 COMMON/OPEN COLLECTOR P OSITIVE 8 LIMIT 4 NORMALLY ...

  • Omega Engineering DP3600 - page 18

    17 TROUBLE SHOOTING If the meter locks on a particular display, unplug and plug in again the meter to unlock the display. A total reset defaults all scaling values to zero. To perform thi s reset, re - apply power with RESET key pressed in. Factory preset values for Current, Voltage and millivolt, respectively, can be programmed by simply re - appl ...

  • Omega Engineering DP3600 - page 19

    18 A 1 2 3 4 3 B 3.62" (91.9mm) 1.70" 43.2mm NOT DRAWN TO SCALE Back of the Meter Bracket Screw Driver Push Meter this way with the Screwdriver MOUNTING 1. Cut a hole in panel to the specifications shown in the figure 4 : Figure 4. Panel Cutout Dimensions 2. Insert pannel meter into the hole until it is flush with the pannel. 3. Refer to ...

Manufacturer Omega Engineering Category Computer Monitor

Documents that we receive from a manufacturer of a Omega Engineering DP3600 can be divided into several groups. They are, among others:
- Omega Engineering technical drawings
- DP3600 manuals
- Omega Engineering product data sheets
- information booklets
- or energy labels Omega Engineering DP3600
All of them are important, but the most important information from the point of view of use of the device are in the user manual Omega Engineering DP3600.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Omega Engineering DP3600, service manual, brief instructions and user manuals Omega Engineering DP3600. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Omega Engineering DP3600.

A complete manual for the device Omega Engineering DP3600, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Omega Engineering DP3600 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Omega Engineering DP3600.

A complete Omega Engineering manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Omega Engineering DP3600 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Omega Engineering DP3600, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Omega Engineering DP3600, that we can find in the current document
3. Tips how to use the basic functions of the device Omega Engineering DP3600 - which should help us in our first steps of using Omega Engineering DP3600
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Omega Engineering DP3600
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Omega Engineering DP3600 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Omega Engineering DP3600?

Use the form below

If you did not solve your problem by using a manual Omega Engineering DP3600, ask a question using the form below. If a user had a similar problem with Omega Engineering DP3600 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)