    User's Manual Y oko gawa Ele c tr i c Corp or ation AXF A11G Magnetic Flowmeter Remote Converter [Hardware Edition/Software Edition] IM 01E20C01-01E IM 01E20C01-01E 5th Edition ...

  • APC AXFA11G - page 2

    i CONTENTS IM 01E20C01-01E 5th Edition: Oct. 2006(KP) All Rights Reserved, Copyright © 2003, Yokogawa Electric Corporation Contents 1. INTRODUCTION ................................................................................................... 1-1 1.1 Using the Magnetic Flowmeter Safely ......................................................... ...

  • APC AXFA11G - page 3

    ii CONTENTS IM 01E20C01-01E 6. PARAMETER DESCRIPTION ............................................................................ 6-1 6.1 Parameters ....................................................................................................... 6-1 6.2 Parameter Lists ......................................................................... ...

  • APC AXFA11G - page 4

    iii CONTENTS IM 01E20C01-01E 9. ACTUAL OPERATION ......................................................................................... 9-1 9.1 Pre-operation Zero Adjustment ....................................................................... 9-1 9.1.1 Zero Adjustment Using Display Unit Switches ....................................... 9-2 9.1 ...

  • APC AXFA11G - page 5

    IM 01E20C01-01E 1-1 1. INTRODUCTION 1. INTR ODUCTION This instrument has been adjusted at the factory before shipment. To ensure correct use of the instrument, please read this manual thoroughly and fully understand how to operate the instrument before operating it.  Regarding This User's Manual • This manual should be provided to the end ...

  • APC AXFA11G - page 6

    IM 01E20C01-01E 1-2 1. INTRODUCTION 1.1 Using the Magnetic Flo wmeter Safel y WARNING (1) Installation •I nstallation of the magnetic flowmeter must be performed by expert engineer or skilled person- nel. No operator shall be permitted to perform procedures relating to installation. • The magnetic flowmeter is a heavy instrument. Be careful tha ...

  • APC AXFA11G - page 7

    IM 01E20C01-01E 1-3 1. INTRODUCTION 1.3 Combination Remote Flo wtubes IMPORTANT •T he AXFA11 Magnetic Flowmeter Converter should be used in combination with the follow- ing remote flowtubes: AXFA11G ⇔ AXF002  -N to AXF26L  -N Contact Yokogawa before using it in combina- tion with flowtubes other than those listed above. • The model AXF ...

  • APC AXFA11G - page 8

    IM 01E20C01-01E 2-1 2. HANDLING PRECAUTIONS 2. HANDLING PRECA UTIONS This instrument has been inspected carefully at the factory before shipment. When the instrument is delivered, visually check that no damage has occurred during transportation. Read this section carefully as it contains important information on handling this instrument. Refer to t ...

  • APC AXFA11G - page 9

    IM 01E20C01-01E 3-1 3. INSTALLATION 3. INST ALLA TION WARNING Installation of the magnetic flowmeter must be performed by expert engineer or skilled personnel. No operator shall be permitted to perform procedures relating to installation. 3.1 Installation Location IMPORTANT Install the instrument in a location where it is not exposed to direct sunl ...

  • APC AXFA11G - page 10

    IM 01E20C01-01E 3-2 3. INSTALLATION  2-inch Pipe Mounting Pa ss the four clamp screws through the mounting fi xture, position it on the 2-inch pipe, and then fasten the AXF A11 in place. F0302.EPS 2-inch pipe Mounting fixture Wa s h e r Clamp screw Figure 3.2.2 2-inch Pipe Mounting  Panel Mounting 172 (6.8) R3MAX 203 (8.0) F0303.EPS Fit the A ...

  • APC AXFA11G - page 11

    IM 01E20C01-01E 4-1 4. WIRING 4. WIRING This section describes the wiring on the converter side only. For information relating to wiring on the flowtube side, refer to the user’s manual of the AXF Remote Flowtube (IM 01E20D01-01E). WARNING The wiring of the magnetic flowmeter must be performed by expert engineer or skilled person- nel. No operato ...

  • APC AXFA11G - page 12

    IM 01E20C01-01E 4-2 4. WIRING 20 (0.8)  10.5 (0.4) AXFC L (Specified Dimensions) 150  5 8 (0.3) max. 150  5 70 (2.76) 60 (2.36) 90 (3.54) (5.9) (5.9) 55 (2.17) 90 (3.54) 50 (1.97) 25 (0.98) SA SB A C B A C B 8(0.3) max. White Black Red White Black Red F0402.EPS Unit : mm (approx. inch) On the conv er ter side On the flowtube side Figure 4. ...

  • APC AXFA11G - page 13

    IM 01E20C01-01E 4-3 4. WIRING (1) When there are no particular optional specifications The wiring port is sealed with a cap (not water-proof) that must be removed before wiring. At this time, handle the wiring port in accordance with the JIS C0920-1982 mentioned above. (2) Wiring using waterproof glands IMPORTANT To prevent water or condensation fr ...

  • APC AXFA11G - page 14

    IM 01E20C01-01E 4-4 4. WIRING 4.4 Wiring Connections 4.4.1 Removing Cover While supporting the front of the cover with your hand, flip the connecting screw protec- tive cover over, and remove the four connecting screws. F0409.EPS Figure 4.4.1 Removing the Front Cover 4.4.2 Terminal Configuration When the cover is removed, the connection terminals w ...

  • APC AXFA11G - page 15

    IM 01E20C01-01E 4-5 4. WIRING IMPORTANT Do not wire the terminal without terminal sym- bols in terminal layout labels. 4.4.3 Precautions for Wiring of Power Supply Cables When connecting to the power supply, observe the points below. Failure to comply with these warnings may result in an electric shock or damage to the instrument. WARNING • Ensur ...

  • APC AXFA11G - page 16

    IM 01E20C01-01E 4-6 4. WIRING 4.4.5 Grounding CAUTION Be sure to connect the protective grounding of the AXFA11 with a cable of 2mm 2 or larger cross section in oder to avoid electrical shock to the operators and maintenance engineers and to prevent the influence of external noise. Connect the grounding wire to the mark (100  or less). IMPORTANT ...

  • APC AXFA11G - page 17

    IM 01E20C01-01E 4-7 4. WIRING In case of explosion proof type for CENELEC ATEX, FM, and CSA certification, connect wiring as shown in the figure below. In case of the explosion proof type, the protective grounding of remote flowtube must be connected to a suitable IS grounding system. In that case, (functional grounding terminal) need not be connec ...

  • APC AXFA11G - page 18

    IM 01E20C01-01E 4-8 4. WIRING  Pulse Output IMPORTANT • As this is a transistor contact (insulated type), give attention to proper voltage and polarity when wiring. • Do not apply a voltage larger than 30V DC or a current larger than 0.2A in order to prevent damage to the instrument. • When input filter constant of the electronic counter i ...

  • APC AXFA11G - page 19

    IM 01E20C01-01E 4-9 4. WIRING  Status Output / Alarm Output IMPORTANT Since this is an isolated transistor output, be careful of voltage and polarity when wiring. Do not apply a voltage larger than 30V DC or a current larger than 0.2A in order to prevent damage to the instrument. This output cannot switch an AC load. To switch an AC load, an int ...

  • APC AXFA11G - page 20

    IM 01E20C01-01E 5-1 5. BASIC OPERATING PROCEDURES 5. BASIC OPERA TING PR OCEDURES (USING THE DISPLA Y UNIT) The modification of data settings from the display unit can be carried out using the three setting switches (infrared switches) - namely, the , , and switches. This chapter will provide a description of basic data configuration and the method ...

  • APC AXFA11G - page 21

    IM 01E20C01-01E 5-2 5. BASIC OPERATING PROCEDURES (3) Display items Display ed items and rev ersed-character indication T0501.EPS Displays the instantaneous flow rate for the span as a percentage. Displays the actual reading for instantaneous flow rate. Displays the instantaneous flow rate for the span as a current output v alue. Displays the insta ...

  • APC AXFA11G - page 22

    IM 01E20C01-01E 5-3 5. BASIC OPERATING PROCEDURES T ouch SHIFT + key FR m 3 /h : : : 100 3274m 3 FTL 0 F0503-1.EPS Setting switches Star t: Display Mode Entry Mode The Entry Mode will be omitted for a period of 1 minute when returning from the Setting Mode to the Display Mode . 5.2 Displa y Unit Setting Methods NOTE Before changing any settings, be ...

  • APC AXFA11G - page 23

    IM 01E20C01-01E 5-4 5. BASIC OPERATING PROCEDURES Setting Mode Y es Setting Mode P:Protect B:Easy Setup C:Basic Setup Setting Mode No Ye s Setting Mode No Y es F0503-2.EPS The rev ersed-character (i.e . the cursor position) indicates the item that is currently selected. Setting Mode To P arameter Search Mode A screen is displayed to confirm whether ...

  • APC AXFA11G - page 24

    IM 01E20C01-01E 5-5 5. BASIC OPERATING PROCEDURES 5.2.2 Setting Mode When the Setting Mode has been activated using the procedure from Section 5.2.1, param- eters can be selected for setting. NOTE If no operations are carried out for a period of 10 minutes in Setting Mode, the system will automatically return to the Display Mode. Format for Paramet ...

  • APC AXFA11G - page 25

    IM 01E20C01-01E 5-6 5. BASIC OPERATING PROCEDURES Sub-item Parameter Search Mode has been accessed in this screen. Touch the switch to move the cursor to B21: Base Flow Unit . The cursor has been moved to B21: Base Flow Unit in this screen. (Sub-item selection screen (A)) In this screen, Parameter Replacement Mode has been called up using the switc ...

  • APC AXFA11G - page 26

    IM 01E20C01-01E 5-7 5. BASIC OPERATING PROCEDURES 5.3.2 Setting Example for Numeric-Type Data: Flow rate span This example describes the setting of the flow rate span for the numeric-type parameter B23: Flow Span from 100 l/min to 120 l/min. Setting Mode Condition Touch the switch to access B: Easy Setup . NOTE The  and  symbols to the left o ...

  • APC AXFA11G - page 27

    IM 01E20C01-01E 5-8 5. BASIC OPERATING PROCEDURES When the switch is touched, the entire display flashes on and off. Confirm that the setting has been correctly changed to “120”, and then fix this value by touching the switch once again. The system returns automatically to sub-item selection screen (B). NOTE If the input value is outside the va ...

  • APC AXFA11G - page 28

    IM 01E20C01-01E 5-9 5. BASIC OPERATING PROCEDURES Upon selection of C: Basic Setup , the cursor will be positioned at C10: Tag No . (Sub-item selection screen (C)) Touch the switch to access Parameter Replacement Mode. The cursor will flash on and off on the left of the tag number. As “FI-1201” is to be setup in this example, touch the switch t ...

  • APC AXFA11G - page 29

    IM 01E20C01-01E 6-1 6. PARAMETER DESCRIPTION 6. PARAMETER DESCRIPTION 6.1 P arameters With the exception of parameters that were specified by the customer upon ordering, all of the AXFA11’s internal parameters will initially be set to default values. Actions such as the modification of display details can then be carried out whenever necessary. I ...

  • APC AXFA11G - page 30

    IM 01E20C01-01E 6-2 6. PARAMETER DESCRIPTION 6.3 P arameter List Over vie w (1) Item A (Menu A): Display items Menu A contains the instantaneous flow rate, totalization values, and other items relevant to display. Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Desc ...

  • APC AXFA11G - page 31

    IM 01E20C01-01E 6-3 6. PARAMETER DESCRIPTION Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Description B22 Base Time Unit (TIME UNIT) W/ d /h /min /s /s (*) Selects time units for the flow rate span. Linked with C41. B24 Flow Decimal Pnt (FLOW DECIMAL) W Auto 0 1 ...

  • APC AXFA11G - page 32

    IM 01E20C01-01E 6-4 6. PARAMETER DESCRIPTION (3) Item C (Menu C): Basic Setting items Menu C principally contains the basic setting items for the flowtube. Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Description T0604-1.EPS C00 C10 Basic Setup (BASIC SETUP) Tag ...

  • APC AXFA11G - page 33

    IM 01E20C01-01E 6-5 6. PARAMETER DESCRIPTION Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Description T0604-2.EPS C60 — (SELF CHECK) See “6.5 Alarm Functions”. Good Error R m/s 3 C44 Velocity Check (VELOCITY CHK) Display of the span setting using flow veloc ...

  • APC AXFA11G - page 34

    IM 01E20C01-01E 6-6 6. PARAMETER DESCRIPTION (5) Item E (Menu E): Pulse Setting items Menu E contains items relevant to pulse output. This is used to set parameters such as the pulse scale and width. T0606.EPS Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Descript ...

  • APC AXFA11G - page 35

    IM 01E20C01-01E 6-7 6. PARAMETER DESCRIPTION T0607-2.EPS Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Description F14 SO1/2 Active Mode (SO ACT MODE) W Closed(On) Act Open(Off) Act Closed(On) Act Selects whether SO1/SO2 output will be set to “On Active” or ? ...

  • APC AXFA11G - page 36

    IM 01E20C01-01E 6-8 6. PARAMETER DESCRIPTION T0608-2.EPS Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Description G14 H/L Alarm Hys (H/L ALM HYS) W% 0 0 to 10 5 % Sets hysteresis width for high-low flow rate limit alarm G20 Alm Out Act Mode ( ALM OUT ACT) W Close ...

  • APC AXFA11G - page 37

    IM 01E20C01-01E 6-9 6. PARAMETER DESCRIPTION (8) Item H (Menu H): Display Setting items Menu H contains setting items that are relevant to display on the display unit. T0609.EPS Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Description H10 Display Select1 (DISP SE ...

  • APC AXFA11G - page 38

    IM 01E20C01-01E 6-10 6. PARAMETER DESCRIPTION T0610-2.EPS Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Description J30 Power Synch (POWER SYNCH) WN o Yes Yes Selects whether or not the internal frequency is to be synchronized with the power supply frequency. J40 ...

  • APC AXFA11G - page 39

    IM 01E20C01-01E 6-11 6. PARAMETER DESCRIPTION (12) Item N (Menu N): Loop Test Setting items Menu N contains items that are relevant to the execution of loop testing. T0613.EPS Item Name Display unit (BRAIN) Data range Display unit /BRAIN Default value (*): Indicated item Units R/W Position of decimal point Description N10 Test Mode (TEST MODE) N00 ...

  • APC AXFA11G - page 40

    IM 01E20C01-01E 6-12 6. PARAMETER DESCRIPTION 6.4 P arameter Description (1) Menu B: Easy Setup items Those parameters with a high frequency of use have been grouped together in Easy Setup. All basic functions can be controlled using only the parameters from this block. Parameters from Menu B share identical names with those from other menus; howev ...

  • APC AXFA11G - page 41

    IM 01E20C01-01E 6-13 6. PARAMETER DESCRIPTION number of 2 or greater for the next digit to the right, regardless of the position of the decimal point. Example: A value of 333.33 is represented by the character string 33333, and since this exceeds 32000, it cannot be set. In such a case, the value 333.3 should be set instead. [B24: Flow Decimal Pnt] ...

  • APC AXFA11G - page 42

    IM 01E20C01-01E 6-14 6. PARAMETER DESCRIPTION Example 1: To count in 1 Ml (mega-liter) steps with flow rate span = 100 m 3 /h Since 1 Ml = 10 3 x m 3 , k Unit/P is set for B30/D10 , and 1 is set for B31/D11 . “x10 3 m 3 ” is indicated for the totalized units in the Display Mode. Example 2: To count in 10 l (liter) steps with flow rate span = 10 ...

  • APC AXFA11G - page 43

    IM 01E20C01-01E 6-15 6. PARAMETER DESCRIPTION B41/H11: Display Select2 and B42/H12: Display Select3 as described below. (For more details, refer to Chapter 5: Basic Operating Procedures.) CAUTION It is not possible to set Display Select1 to “Off”. [B41: Display Select2] Setting of the second line for display unit → This setting is linked with ...

  • APC AXFA11G - page 44

    IM 01E20C01-01E 6-16 6. PARAMETER DESCRIPTION selected, a setting alarm will be displayed. The setting should be returned to “Standard DF” in such a case. •When dual frequency excitation mode is changed, perform zero adjustment. For details on zero adjustment, refer to chapter 9. [C21: Low MF] Setting of the low-frequency meter factor This pa ...

  • APC AXFA11G - page 45

    IM 01E20C01-01E 6-17 6. PARAMETER DESCRIPTION [C45: Density Unit] Setting of the density units for mass flow rate This parameter selects the units for density as required when making settings using C46: Mass Density . [C46: Mass Density] Setting of the density for mass flow rate This parameter is necessary in situations where t, kg, g, klb or lb ha ...

  • APC AXFA11G - page 46

    IM 01E20C01-01E 6-18 6. PARAMETER DESCRIPTION [D20: Total Execution] Operation setting for the totalization function This parameter sets “Start” and “Stop” of the totaliza- tion function, in addition to performing the preset function for the forward totalized value and the reverse totalized value. *: The preset function starts the count for ...

  • APC AXFA11G - page 47

    IM 01E20C01-01E 6-19 6. PARAMETER DESCRIPTION and “dl” is set for D31: Ttl User Unit . “dl” is indicated for the totalized units in the Display Mode and is counted in 1 dl steps. (4) Menu E: Pulse Setting items Menu E contains items relevant to pulse output. [E10: Pulse Unit] Setting of the pulse units → Refer to the description of parame ...

  • APC AXFA11G - page 48

    IM 01E20C01-01E 6-20 6. PARAMETER DESCRIPTION (5) Menu F: Status Functions Setting items Menu F contains setting items relevant to status Input/Output functions. [F10: SO1 Function] Setting of the function for the SO1 status output terminal This parameter sets the function for the SO1 (status output 1) terminal. Setting No Function Stop output (i.e ...

  • APC AXFA11G - page 49

    IM 01E20C01-01E 6-21 6. PARAMETER DESCRIPTION [F12: SI1 Function] Setting of the funtion for the SI1 status input terminal This parameter sets the function for the SI1 (status input 1) terminal. Setting No Function No input function 0% Signal Lock 0% signal lock via external status input. Ext Auto Zero Automatic zero adjustment via ex ternal status ...

  • APC AXFA11G - page 50

    IM 01E20C01-01E 6-22 6. PARAMETER DESCRIPTION [F15: SI1/2 Active Mode] Setting of the active mode for status input This parameter sets the active mode for the terminals SI1 and SI2. Active modes cannot be set individually for these two terminals. Setting Function Short Activ e Open Activ e When the status input is set to "Short", occurren ...

  • APC AXFA11G - page 51

    IM 01E20C01-01E 6-23 6. PARAMETER DESCRIPTION Parameter setting sequence (for automatic multiple ranges switching) F10: SO1 Function Select a function F11: SO2 Function Select a function F14: SO1/2 Active Mode Select whether SO1 and SO2 output is to be “Closed (On) Act” or “Open (Off) Act” Function Selection No Function: Output is stopped ( ...

  • APC AXFA11G - page 52

    IM 01E20C01-01E 6-24 6. PARAMETER DESCRIPTION Multiple Ranges Switching via External Status Input Operations are performed in accordance with the following table when the active mode has been set to “Closed (On) Act” using F14: SO1/2 Active Mode and when the active mode has been set to “Short Active” using F15: SI1/2 Active Mode . Operating ...

  • APC AXFA11G - page 53

    IM 01E20C01-01E 6-25 6. PARAMETER DESCRIPTION Parameter setting sequence (for multiple ranges switching via external status input) F10: SO1 Function Select a function F11: SO2 Function Select a function F14: SO1/2 Active Mode Select whether SO1 and SO2 output is to be “Closed (on) Act” or “Open (off) Act. ” Function Selection No Function: O ...

  • APC AXFA11G - page 54

    IM 01E20C01-01E 6-26 6. PARAMETER DESCRIPTION [G13: High High Alarm] High-high alarm setting This parameter sets the high-high limit (HH) alarm value, and this is done using a % value of the maxi- mum span. 䊉 A setting value of 110% indicates that the alarm is disabled. NOTE Setting of -110% or 110% results in the corre- sponding function being d ...

  • APC AXFA11G - page 55

    IM 01E20C01-01E 6-27 6. PARAMETER DESCRIPTION Output Example 3 The high alarm (H) is set to 80% or more of the flow rate span; the high-high alarm (HH), to 90% or more. Settings are: G10: Low Alarm = -110% G11: High Alarm = 80% G12: Low Low Alarm = -110% G13: High High Alarm = 90% 0T i me (t) H HH SO1 SO2 HH AL Instantaneous flow rate 50 80 90 100 ...

  • APC AXFA11G - page 56

    IM 01E20C01-01E 6-28 6. PARAMETER DESCRIPTION Example Alarm occurs Closed (On) Active Alarm occurs Open (Off) Active F0614.EPS Closed(On) Open(Off) [G21: 4-20mA Alarm Out] Setting of the current output during an alarm occurrence This parameter can be used to set up the current output during alarm occurrence in advance. Fixed at 2.4 mA or less Fixed ...

  • APC AXFA11G - page 57

    IM 01E20C01-01E 6-29 6. PARAMETER DESCRIPTION [G40: Operation Time] Display of operation time This parameter is used to display the operation time. The operation time is the total time that is counted while the device works actually. When the power supply is off, the operation time is not counted. For example, “1D23:45” indicates an operation t ...

  • APC AXFA11G - page 58

    IM 01E20C01-01E 6-30 6. PARAMETER DESCRIPTION This parameter selects the display content of the second line for display unit. [H12: Display Select3] Setting of the third line for display unit → Refer to the description for parameter B42 This parameter selects the display content of the third line for display unit. [H20: Display Cycle] Setting of ...

  • APC AXFA11G - page 59

    IM 01E20C01-01E 6-31 6. PARAMETER DESCRIPTION [J20: Flow Direction] Setting of the flow direction Upon shipment from the manufacturing plant, the system is setup such that flow in the same direction, as shown by the direction of the arrow mark on the flowtube, will be measured as forward flow; however, this parameter can be used to set “Reverse? ...

  • APC AXFA11G - page 60

    IM 01E20C01-01E 6-32 6. PARAMETER DESCRIPTION [J23: Pulsing Flow] Selection of pulsing flow support In a situation where pulsating flow causes error in the average flow value, due to the application of a plunger pump, this parameter provides functionality whereby calculation is controlled and variations in flow rate are followed. Setting Function N ...

  • APC AXFA11G - page 61

    IM 01E20C01-01E 6-33 6. PARAMETER DESCRIPTION Setting Function Halt the adhesion diagnostic function No Ye s T0642.EPS Carry out the adhesion diagnostic function NOTE Adhesion Diagnostic Function 䊏 This function diagnose adhesion using elec- trode resistance values. 䊏 When “Adhesion check” has been set for B41/ H11: Display Select 2 or B42/ ...

  • APC AXFA11G - page 62

    IM 01E20C01-01E 6-34 6. PARAMETER DESCRIPTION (3) “Normal” will be restored when the power is turned off or when 30 minutes have elapsed since entry to Test Mode. (4) In Test Mode, the warning 83: Fix Cur Wng will be displayed as a warning message. (For more details, refer to Section 6.5 Alarm Functions.) [N11: Test Output Value] Setting for te ...

  • APC AXFA11G - page 63

    IM 01E20C01-01E 6-35 6. PARAMETER DESCRIPTION [P21: Enable Wrt Passwd] Setting of password to release the write protection function When the correct password is input, write protection will be released for a period of 10 minutes; further- more, this period will be extended by a further 10 minutes each time a parameter is overwritten. P21:Enable Wrt ...

  • APC AXFA11G - page 64

    IM 01E20C01-01E 6-36 6. PARAMETER DESCRIPTION 6.5.2 Alarm Selection The display and output differs depending on the alarm levels. Certain types of alarm may or may not be recognized as alarms, according to the settings of certain parameters. The parameters that are relevant to this function as follows. [G20: Alm Out Act Mode] Setting of the active ...

  • APC AXFA11G - page 65

    IM 01E20C01-01E 6-37 6. PARAMETER DESCRIPTION (3) Display and output condition for setting alarm occurrences 50 Span > 10m/s 51 Span < 0.1m/s 52 TTL>10000p/s 53 TTL<0.0001p/s 54 4-20 Lmt Err 55 Multi Rng Err 56 H/L HH/LL Set 57 Dens Set Err 60 PLS >10000p/s 61 PLS > 5000p/s 62 PLS > 1000p/s 63 PLS > 500p/s 64 PLS > 25p/s ...

  • APC AXFA11G - page 66

    IM 01E20C01-01E 6-38 6. PARAMETER DESCRIPTION 6.5.3 Alarms & Warning Messages System Alarms (Device breakdown or inability to obtain correct measurements .) Display unit/BRAIN ( 䊐 60) content Alarm counter measure message on display unit 10:uP Fault Contact nearest office or service center Microprocessor (CPU) failure 11:EEPR OM Fault Contact ...

  • APC AXFA11G - page 67

    IM 01E20C01-01E 6-39 6. PARAMETER DESCRIPTION Setting Alarms (Device is normal but errors hav e been made in the setting of parameters .) Display unit/BRAIN ( 䊐 60) content Alarm counter measure message on display unit 50:Span > 10m/s Check parameter C40, C41, and C42 Span flow v elocity setting is 11 m/s or more 51:Span < 0.1m/s Check para ...

  • APC AXFA11G - page 68

    IM 01E20C01-01E 7-1 7. OPERATION VIA BRAIN TERMINAL (BT200) 7. OPERA TION VIA BRAIN TERMINAL (BT200) 7.1 BT200 Basic Operations 7.1.1 Key Layout and Display F0701.EPS LCD (21 characters ⴛ 8 lines) Po w e r ON/OFF ke y ENT ke y • Enters selected items • Sets up input data on the connected instrument. • Runs BT200 functions. Alphanumeric ke y ...

  • APC AXFA11G - page 69

    IM 01E20C01-01E 7-2 7. OPERATION VIA BRAIN TERMINAL (BT200) Use the function key [F2] CAPS to select between uppercase and lowercase (for letters only). The case toggles between uppercase and lowercase each time you press [F2] CAPS. Entry Boy Ke y-in Sequence F0807.EPS CODE CAPS CLR ESC Entering uppercase CODE caps CLR ESC Entering lowercase ( B ) ...

  • APC AXFA11G - page 70

    IM 01E20C01-01E 7-3 7. OPERATION VIA BRAIN TERMINAL (BT200) 7.2 AXF A11 Operation Using a BT200 This section describes procedures for setting parameters using a BRAIN Terminal (BT200). For more details regarding AXFA11 functions, refer to Chapter 6: Parameter Description; and for more details regarding BT200 operation methods, refer to the BT200 Us ...

  • APC AXFA11G - page 71

    IM 01E20C01-01E 7-4 7. OPERATION VIA BRAIN TERMINAL (BT200) 7.2.3 BT200 Screens & Flow Rate Data Display Use the following procedure to display flow rate data on the BT200. • The display of flow rate data is updated every 5 seconds. The Flow Rate Screen will be display ed. A maximum of three items of data can be display ed. The or ke y can be ...

  • APC AXFA11G - page 72

    IM 01E20C01-01E 7-5 7. OPERATION VIA BRAIN TERMINAL (BT200) 7.3.1 BT200 Setting of Selection-Type Data: Flow rate units In this example, the flow rate units specified by the selection-type parameter B21: Flow Unit are changed from m 3 to l (Liter). With the cursor aligned with “l (Liter)”, press the key. X 2 X 2 MENU A:DISPLAY B:EASY SETUP C:BA ...

  • APC AXFA11G - page 73

    IM 01E20C01-01E 7-6 7. OPERATION VIA BRAIN TERMINAL (BT200) 7.3.2 BT200 Setting of Numeric-Type Data: Flow rate span In this example, the flow rate span specified by the numeric-type parameter B23: Flow Span is changed from 100.000 l/min. to 120.000 l/m. or In this mode, ne w values can be entered into the highlighted field using the alphanumeric k ...

  • APC AXFA11G - page 74

    IM 01E20C01-01E 7-7 7. OPERATION VIA BRAIN TERMINAL (BT200) 7.3.3 BT200 Setting of Alphanumeric-Type Data: Tag number In this example, the tag number specified by the alphanumeric-type parameter C10: TAG NO is changed from “FI-1101” to “FI-1201”. The highlighted data will begin to flash on and off: Press the ke y once again to confirm your ...

  • APC AXFA11G - page 75

    IM 01E20C01-01E 8-1 8. OPERATION VIA HART COMMUNICATOR 8. OPERA TION VIA HAR T COMMUNICA T OR CAUTION Matching of communicator DD and instrument DD Before using the Model 275 HART Communicator, check that the DD (Device Description) installed in the communicator matches to that of t he instruments to be set up. To check the DD in the instrument or ...

  • APC AXFA11G - page 76

    IM 01E20C01-01E 8-2 8. OPERATION VIA HART COMMUNICATOR 8.1 Conditions of Comm unication Line 8.1.1 Interconnectwion between AXFA11 and HART Communicator The HART Communicator can interface with the AXFA11 from the control room, the AXFA11 site, or any other wiring termination point in the loop, provided there is a minimum load resistance of 230  ...

  • APC AXFA11G - page 77

    IM 01E20C01-01E 8-3 8. OPERATION VIA HART COMMUNICATOR 8.2 Basic Operation of the HAR T Comm unicator (Model 275) 8.2.1 Keys and Functions LCD (Liquid crystal display) (21 characters × 8 lines) Communication Cable Function K eys Functions of the ke ys are indicated on the display . Hot ke y Open the Hot ke y Menu as f ollows: 1. PV Span 2. Wr t Pr ...

  • APC AXFA11G - page 78

    IM 01E20C01-01E 8-4 8. OPERATION VIA HART COMMUNICATOR 8.2.2 Display The HART Communicator automatically searches for the AXFA11 on the 4 to 20 mA loop when it is turned on. When the HART Communicator is connected to the AXFA11, it displays the “ Online ” menu as shown below. (If AXFA11 is not found, the communicator displays the message “No ...

  • APC AXFA11G - page 79

    IM 01E20C01-01E 8-5 8. OPERATION VIA HART COMMUNICATOR 8.2.4 Entering, Setting and Sending Data The data which are input with the keys are set in the HART Communicator by pressing ENTER ( F4 ). Then, by pressing SEND ( F2 ), the data are sent to the AXFA11. Note that the data are not set in the AXFA11 if SEND ( F2 ) is not pressed. All the data set ...

  • APC AXFA11G - page 80

    IM 01E20C01-01E 8-6 8. OPERATION VIA HART COMMUNICATOR NOTE Data changed with the HART Communicator is sent to the AXFA11 by pressing SEND (F2) of the HART Communicator. 8.3.3 Checking for Problems The self-diagnostic function of the AXFA11 is explained in Section 6.5 “Alarm Functions” By using the HART Communicator, it is also possible to carr ...

  • APC AXFA11G - page 81

    IM 01E20C01-01E 8-7 8. OPERATION VIA HART COMMUNICATOR 8.3.4.2 Example of Setting Numeric-Type Data: PV Span This example describes the setting of the flow rate span for the numeric-type parameter “PV Span” from 100 l/min to 120 l/min. There are two ways to open the “PV Span” setting display. Following the menu tree configration, “PV Span ...

  • APC AXFA11G - page 82

    IM 01E20C01-01E 8-8 8. OPERATION VIA HART COMMUNICATOR 8.3.4.3 Example of Setting Alphanumeric-Type Data: Tag Number This example describes the setting of the tag number for the alphanumeric-type parameter ‘ Tag ’ from “FI-1101” to “FI-1201”. 1 Display Operation 2 Select “ Tag ”. The display for the Tag setting appears. (The default ...

  • APC AXFA11G - page 83

    IM 01E20C01-01E 8-9 8. OPERATION VIA HART COMMUNICATOR 8.3.4.4 Example 1 of Other Settings: Meter Factor The “ Meter Factor ” is engraved on the data plate of the combined flowtube. The meter factor is required to compute the correct electromotive force proportional to the fluid velocity, and is determined by actual flow-test calibration at the ...

  • APC AXFA11G - page 84

    IM 01E20C01-01E 8-10 8. OPERATION VIA HART COMMUNICATOR 8.3.4.5 Example 2 of Other Settings: Power Frequency (For DC version only) IMPORTANT In situation where a DC power supply is used for converters, set the local com- mercial power frequency in area where the converter is lnstalled. The flowmeter is set to 50.00 Hz at the factory. Set “No” f ...

  • APC AXFA11G - page 85

    IM 01E20C01-01E 8-11 8. OPERATION VIA HART COMMUNICATOR 8.3.4.6 Example 3 of Other Settings: Trim Analog Output Fine output adjustment is carried out with “ D/A trim ” or “ Scaled D/A trim ”. • D/A trim “ D/A trim ” is to be carried out if the calibration digital ammeter does not read exactly 4.000 mA and 20.000 mA with output signals ...

  • APC AXFA11G - page 86

    IM 01E20C01-01E 8-12 8. OPERATION VIA HART COMMUNICATOR F4 (ENTER) AXFA11 : FI-1201 Enter meter value 20.000 HELP DEL ABORT ENTER 8 9 AXFA11 : FI-1201 Fld dev output 20.000 mA equal to reference meter? 1 Yes 2 No HELP DEL ABORT ENTER F4 (ENTER) F4 (OK) F3 (HOME) AXFA11 : FI-1201 NOTE-Loop may be returned to automatic control DEL DEL ABORT OK 10 11 ...

  • APC AXFA11G - page 87

    IM 01E20C01-01E 8-13 8. OPERATION VIA HART COMMUNICATOR Example 2: For adjustments using a voltmeter F0816-1.EPS AXFA11 : FI-1201 Adjustment 1 Auto Zero Exe 2 Magflow Zero 3 D/A trim 4 Scaled D/A trim HELP SAVE HOME ENTER 1 2 Select the “ Scaled D/A trim ”. Press OK (F4) . AXFA11 : FI-1201 WARN-Loop should be removed from automatic control DEL ...

  • APC AXFA11G - page 88

    IM 01E20C01-01E 8-14 8. OPERATION VIA HART COMMUNICATOR 9 AXFA11 : FI-1201 Enter meter value 1.000000 1.000000 DEL DEL ABORT ENTER F4 (ENTER) 10 AXFA11 : FI-1201 Scaled output: 1.000 equal readout device? 1 Yes 2 No DEL DEL ABORT ENTER F4 (ENTER) Voltmeter reading: 1.010 Enter the voltmeter’s reading of 1.010, and press ENTER (F4) . (The output o ...

  • APC AXFA11G - page 89

    IM 01E20C01-01E 8-15 8. OPERATION VIA HART COMMUNICATOR 8.3.4.7 Example 4 of Other Settings: Burst Mode The AXFA11 continuously sends its stored data when the “ Burst mode ” is set to “On”. Any one of instantaneous flow rate, output in %, totalization values or current output can be selected and sent. The data is sent intermittently as a di ...

  • APC AXFA11G - page 90

    IM 01E20C01-01E 8-16 8. OPERATION VIA HART COMMUNICATOR HART Communicator Polling Never Poll Never Poll Ask Before Polling Always Poll Digital Poll HELP SAVE ESC ENTER 6 Select “ Digital Poll ” and press ENTER (F4) . F4 (ENTER) × 3 5. HART Output Online Menu Main Menu 4. Utility 1. Configure Communication 1. Polling AXFA11 : FI-1201 HART Outpu ...

  • APC AXFA11G - page 91

    IM 01E20C01-01E 8-17 8. OPERATION VIA HART COMMUNICATOR Example: Communication when set in the multidrop mode. F0820-1.EPS HART Communicator Online 1 1: FIC-1A 2 2: FIC-2A 3 3: PIC-1A DEL DEL ESC ENTER 1 AXFA11 : FI-1201 Online 1 Device Setup 2 PV 3 PV AO 4 PV Span HELP SAVE HOME ENTER 2 The HART Communicator searches for field devices that are set ...

  • APC AXFA11G - page 92

    IM 01E20C01-01E 8-18 8. OPERATION VIA HART COMMUNICATOR Setting Password Example: Set the password to “ 1 2 3 4 ” F0821-1.EPS ABORT ENTER 3. New Password AXFA11 : FI-1201 Hot key 1 PV Span 1 m/s 2 Wrt Protect Menu HELP HOME ENTER 1 2 AXFA11 : FI-1201 Enter new password to change state of write protect HELP DEL ABORT ENTER 3 Open “ Wrt Protect ...

  • APC AXFA11G - page 93

    IM 01E20C01-01E 8-19 8. OPERATION VIA HART COMMUNICATOR Changing Password Example: Change the password from “ 1 2 3 4 ” to “ 6 7 8 9 A ”. F0822-1.EPS ABORT ABORT OK AXFA11 : FI-1201 Enter current password to enable to write for 10 minutes: Hot key 2. Wrt Protect Menu 2. Enable Wrt 10min HELP DEL ABORT ENTER 1 2 AXFA11 : FI-1201 If you wish ...

  • APC AXFA11G - page 94

    IM 01E20C01-01E 8-20 8. OPERATION VIA HART COMMUNICATOR 9 F0822-3.EPS ABORT SAVE AXFA11 : FI-1201 Wrt Protect Menu 1 Write protect Yes 2 Enable Wrt 10min 3 New Password 4 Software seal Keep Press twice, and return to the “ Online ” menu. × 2 NOTE 1. “ Enable Wrt 10min ” releases write protection status for 10 minutes. While write protectio ...

  • APC AXFA11G - page 95

    IM 01E20C01-01E 8-21 8. OPERATION VIA HART COMMUNICATOR 8.3.5 Menu Tree 1 Device setup 1 Process V ariables PV % rnge PV PV A0 To t l Rev erse T otl Dif T otl 2 Diag/Service 1 T est/Status 1 Status Status group 1(System alarms) Status group 3(Process alarms) Status group 5(Setting alarms) Status group 6(Setting alarms) Status group 7(Setting alarms ...

  • APC AXFA11G - page 96

    IM 01E20C01-01E 8-22 8. OPERATION VIA HART COMMUNICATOR Pa ra meter of BRAIN protocol Read/Write 4 Detailed Setup 1 Basic Setup 1 T ag 2 PV Damping 3 MF Set Standard DF Low MF High MF Enhanced DF Lo w MF (EDF) High MF (EDF) 4 Select Flow T ube 5 Nominal Size unit 6 Nominal Size 7 Base Flow Unit 8 Base Time Unit 9 PV Span Flow Decimal Pnt V elocity ...

  • APC AXFA11G - page 97

    IM 01E20C01-01E 8-23 8. OPERATION VIA HART COMMUNICATOR Pa ra meter of BRAIN protocol Read/Write 5 Aux 1 4-20mA Low Cut 2 4-20mA Low Lmt 3 4-20mA High Lmt 4 Flow Direction 5 Rate Limit 6 Dead Time 7 Pulsing Flow 8 T/P Damp Select 9 P ow er Synch P ow er F requency Memo 1 Memo 2 Memo 3 Software Re v No J10 J11 J12 J20 J21 J22 J23 J24 J30 J31 J40 J41 ...

  • APC AXFA11G - page 98

    IM 01E20C01-01E 8-24 8. OPERATION VIA HART COMMUNICATOR Pa ra meter of BRAIN protocol Read/Write P20 P21 P22 P23 R W W R B23/C42 W 4 Detailed setup 5 HAR T output 1 P oll addr 2 Num req preams 3 Burst mode 4 Burst option 6 Device inf o 1 Field de vice info 1 Manuf acturer 2 T ag 3 Descriptor 4 Message 5 Date 6 Device id 7 Write protect 8 Revision # ...

  • APC AXFA11G - page 99

    IM 01E20C01-01E 9-1 9. ACTUAL OPERATION 9. ACTU AL OPERA TION After you have installed the flowtube into the process piping, wired the input/output terminals, set up the required parameters, and performed a pre-operation zero adjustment, the magnetic flowmeter should output an accurate flow signal from its terminals as soon as flow of the fluid to ...

  • APC AXFA11G - page 100

    IM 01E20C01-01E 9-2 9. ACTUAL OPERATION 9.1.1 Zero Adjustment Using Display Unit Switches This section describes the procedure for zero adjustment using the display unit switches. (For more details regarding setting methods using these switches, refer to Chapter 5: Basic Operating Procedures.) The parameters for zero adjustment are B50/M10: Auto Ze ...

  • APC AXFA11G - page 101

    IM 01E20C01-01E 9-3 9. ACTUAL OPERATION NOTE The results of M10: Auto Zero Exe can be displayed using M11: Magflow Zero . Alterna- tively, if the results of the automatic zero adjustment exceed the rated value, the warn- ing 82: Auto Zero Wng will be displayed. 9.1.2 Zero Adjustment via External Status Input This section describes the procedure for ...

  • APC AXFA11G - page 102

    IM 01E20C01-01E 9-4 9. ACTUAL OPERATION Touch the switch to select “Ext Auto Zero (Zero adjustment via external status input)”. In order to request confirmation, the entire display flashes on and off. Touch the switch once again at this time to fix selection of the automatic zero adjustment function. The system returns automatically to sub-item ...

  • APC AXFA11G - page 103

    IM 01E20C01-01E 10-1 10. MAINTENANCE 10. MAINTENANCE Fuse holder F1001.EPS Figure 10.1.1 Fuse Assembly 10.1.2 Display Unit Replacement 10.1.2.1 Removing the Display Unit (1) Turn off the power. (2) Remove the cover. (3) Hold the display unit with your hand and remove the two mounting screws. Remove the connector by lifting it upward, taking care no ...

  • APC AXFA11G - page 104

    IM 01E20C01-01E 10-2 10. MAINTENANCE 10.1.2.2 Assembling the Display Unit (1) Align the display unit with the protrusion of the connector facing forward and then make the required connection. (2) Secure the unit using its two mounting screws. (3) Replace the cover. Connector Display unit mounting screws (x 2) F1002.EPS Figure 10.1.2 Display Unit As ...

  • APC AXFA11G - page 105

    IM 01E20C01-01E 10-3 10. MAINTENANCE F1004.EPS Low High Enable Protect Switch 1 2 Burnout setting switch Switch 2 1 Write protect setting switch Figure 10.1.4 Switch Configuration 10.2.2 Setting of Write Protect Switch By setting the write protect function to “Protect” it is possible to prevent the overwriting of parameters. Write protection ca ...

  • APC AXFA11G - page 106

    IM 01E20C01-01E 10-4 10. MAINTENANCE 10.3 T r oubleshooting Although magnetic flowmeters rarely require maintenance, failures may occur when the instrument is not operated correctly. This section describes troubleshooting procedures where the cause of the breakdown is identified through receiver indication. 10.3.1 No Indication Is an error being di ...

  • APC AXFA11G - page 107

    IM 01E20C01-01E 10-5 10. MAINTENANCE 10.3.2 Unstable Zero START Check for the presence of valve leakage. • Confirm the flowmeter grounding. • Confirm whether or not the grounding wire is disconnected from the earth flange. Recheck the fluid to be measured. Particular care must be taken in the case of horizontal mounting. In order to ensure comp ...

  • APC AXFA11G - page 108

    IM 01E20C01-01E 10-6 10. MAINTENANCE 10.3.3 Disagreement Between Indication and Actual Flow NO NO NO NO YES NO NO YES YES YES NO YES YES NO NO YES Set the parameters correctly. Examine the condition of the fluid in the flowtube, of bubbles, and of grounding. Execute zero adjustment when the flowtube is filled completely with fluid and when the flui ...

  • APC AXFA11G - page 109

    IM 01E20C01-01E 11-1 11. OUTLINE Cable Length for Specific Applications: Use the following formula to determine the cable length for specific applications: L  65  10 6 (R  C)  (Cf  10,000) C where: L= length in meters or feet R= resistance in Ω (including barrier resistance) C= cable capacitance in pF/m or pF/ft Cf = maximum shunt ...

  • APC AXFA11G - page 110

    IM 01E20C01-01E 11-2 11. OUTLINE Functions How to Set Parameters: The indicator’s LCD and three infra-red switches enable users to set parameters without opening the case cover. Parameters can also be set by means of the HHT(Hand- held terminal). Displayed Languages: Users can choose a language from among English, Japanese, German, French, Italia ...

  • APC AXFA11G - page 111

    IM 01E20C01-01E 11-3 11. OUTLINE Polyurethane Rubber/Natural Soft Rubber/EPDM Rubber Lining Size mm (in.) V  0.3 (1.0)  1.0 mm/s 0.3  V  10 (1.0) (33)  0.35% of Rate 25 (1.0) to 400 (16) Flow V elocity V m/s (ft/s) Standard Accurac y (Calibration code B) V  0.3 (1.0)  1.75 mm/s 0.3  V  1 (1.0) (3.3) 1  V  10 (3.3) ( ...

  • APC AXFA11G - page 112

    IM 01E20C01-01E 11-4 11. OUTLINE Size 2200 mm (88 in.) to 2600 mm (104 in.) 2.4 2.0 1.6 1.2 0.85 0.8 0.4 0 012 6 8 1 0 V[m/s] F16-3.EPS Pulse Output Accuracy % of Rate High grade Accuracy High grade Accuracy Standard Accuracy Current Output: Pulse output accuracy plus ⫾ 0.05% of Span Repeatability: ⫾ 0.1% of Rate (V ⭌ 1 m/s (3.3 ft/s)) ⫾ 0. ...

  • APC AXFA11G - page 113

    IM 01E20C01-01E 11-5 11. OUTLINE ■ ACCESSORIES Fuse (Time lag fuse): 1 pc. • Power supply code 1; T2.5 A, 250 V • Power supply code 2; T3.15A, 250 V Mounting bracket: 1 set ■ MODEL AND SUFFIX CODE AXFA11 Magnetic Flowmeter Remote Converter: Model Suffix Code Description AXF A11 Use Output Signal and Communication Po w er Supply Electrical C ...

  • APC AXFA11G - page 114

    IM 01E20C01-01E 11-6 11. OUTLINE ■ OPTIONAL SPECIFICATIONS FOR AXFA11 REMOTE CONVERTER Item Specification Code Burn Out Down Lightning Protector DC Noise Cut Circuit A lightning protector is built into the po wer terminals. In case of power supply code 2 (24 V AC/DC), this optional code is mandatory . The output lev el is set to 0 mA during a CPU ...

  • APC AXFA11G - page 115

    IM 01E20C01-01E 11-7 11. OUTLINE ■ EXTERNAL DIMENSIONS ● Remote Converter AXFA11G Pipe mounting 120(4.72) 180(7.09) 4-ø6 Hole (Surface mounting hole) 194(7.64) 210(8.27) 65 (67) (2.64) 194(7.64) 182(7.17) Pa nel mounting 100(3.94) 1 to 10(0.04 to 0.39) (P anel thickness) Mounting Screw (M6  100(3.94)) 172(6.77) R3 MAX 203(7.99) P anel cutou ...

  • APC AXFA11G - page 116

    IM 01E20C01-01E 11-8 11. OUTLINE ● Remote Converter AXFA11G /RK (AM11 Replacement Bracket) Pipe mounting 133.5(5.26) 180(7.09) 194(7.64) 210(8.27) (80.5) (3.17) 49 (1.93) 29 (1.14) 24 (0.94) 65(2.56) (80.5) (3.17) 31 (1.22) 31 (1.22) 31 (1.22) 31 (1.22) FF0AXF A11E.EPS Brack et for replacement (For pipe mounting) 4-ø6(0.24) (Surface mounting hol ...

  • APC AXFA11G - page 117

    IM 01E20C01-01E 11-9 11. OUTLINE ● Dedicated Signal Cable AXFC-4-L  20(0.79) CONVERTER FLOW TUBE ø10.5(0.41) AXFC L(SPECIFIED LENGTH) 150  5 (5.91) 8(0.31)Max. 150  5 (5.91) D G 90(3.54) 55(2.17) 90(3.54) F E SA SB A C B A C B 8(0.31)Max. (WHITE) (BLACK) (RED) (WHITE) (BLACK) (RED) Model AXFC-4 D 70(2.76) E 25(0.98) F 50(1.97) G 6 ...

  • APC AXFA11G - page 118

    IM 01E20C01-01E REVISION RECORD Title: AXFA11G Magnetic Flowmeter Remote Converter [Hardware Edition/Software Edition] Manual No.: IM 01E20C01-01E Edition Date P age Revised item 5th Oct. 2006 2-1 4-1 4-5 4-8 5-1 5-3 6-3 to 6-7 6-30 6-32 7-3 8-1 11-1 11-4 11-5 11-6 2.2 Added the fuse of “P ower supply code 2 (24 V AC/DC)”. 4.1 Added the caution ...

Manufacturer APC Category TV Converter Box

Documents that we receive from a manufacturer of a APC AXFA11G can be divided into several groups. They are, among others:
- APC technical drawings
- AXFA11G manuals
- APC product data sheets
- information booklets
- or energy labels APC AXFA11G
All of them are important, but the most important information from the point of view of use of the device are in the user manual APC AXFA11G.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals APC AXFA11G, service manual, brief instructions and user manuals APC AXFA11G. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product APC AXFA11G.

Similar manuals

A complete manual for the device APC AXFA11G, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use APC AXFA11G by users. Manuals are usually written by a technical writer, but in a language understandable to all users of APC AXFA11G.

A complete APC manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual APC AXFA11G - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the APC AXFA11G, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the APC AXFA11G, that we can find in the current document
3. Tips how to use the basic functions of the device APC AXFA11G - which should help us in our first steps of using APC AXFA11G
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with APC AXFA11G
5. FAQ - Frequently Asked Questions