  • Agilent Technologies E1300B - page 1

    75000 S ER IES B Mainf r ames E1300B an d E 1301B Us er ’ s Man ual Copy rig ht © Ag il ent T echnol og i es, Inc. , 1989, 1990, 1991, 2006 M a nu a l P a r t N u m b e r : E 1300 - 90005 P r i n t e d : F e b r u a r y 2006 E d i t i o n 3 M i c r o f i c h e P a r t N u m b e r : E 1300 - 99005 P r i n t e d i n U . S . A . E 0206 ...

  • Agilent Technologies E1300B - page 2

    ...

  • Agilent Technologies E1300B - page 3

    Ce rt if ica ti on Agile nt T echnolo gies certifi es that th is p r oduct me t it s pu blished spe c ificatio ns a t th e time of ship ment from the f actory . Agil ent T ech- nolog ies f urth er cer tif ies that it s c alibrati on m e asure ment s a re t race able to the Un i ted St ates Na tion al Ins tit ute of St and ard s and Tec h- nolog y ( ...

  • Agilent Technologies E1300B - page 4

    F ra m e or cha ss is gro und t er m i nal — ty pi - cally c onn ects to the equ ipment’ s metal frame . Al ter n ati ng c urr en t ( A C). Dire ct cu r re nt (DC ) . Indi ca te s ha za r dous vol ta ge s . Ca lls at te nt i on t o a pr oce dur e, pr ac tic e , or condi t i on t hat co uld cause bodi l y i nj ury or deat h. Ca lls at te nt i on ...

  • Agilent Technologies E1300B - page 5

    Decl ar a ti on of Co nf ormit y according to I SO/IEC Guide 22 and EN 45014 Manu fa cturer’ s Na m e: Agile nt T e ch nol ogi es, Inc . Love la nd Ma nuf ac t uri ng Ce nt er Manu fa cturer’ s Ad dr ess: 815 14th S t re et S .W. Love la nd, Col ora do 80537 declares, t hat the product : Product Name: 75000 S er ie s B VX I M ainfr a me Model N ...

  • Agilent Technologies E1300B - page 6

    A gil e n t 75000 S erie s B D ocume n ta t i on M a nua l De scrip tio ns Instal lation and G etting Starte d G uide. Contai ns step -by -s tep ins tr ucti ons f or a ll a sp e c t s o f p lu g-in m o d u le a n d m a in fr a m e in st a lla t io n . T h is gu id e a ls o contai ns introd uctor y pr og rammi ng i nf ormati on and ex a mpl es . A g ...

  • Agilent Technologies E1300B - page 7

    1 Re late d Do cum e nts A gilent Ins trument BA S IC Us er’ s Handbook . Inc l ud es thre e books : A g ilen t Ins t r um e nt BA S IC P r og r a m m ing T e c hni q ue s , A g i le nt Ins tr um ent BA S IC I n ter f a cin g T ech n iqu es, and A g i le nt Ins tr um e nt B A S IC L ang uag e Ref e r e nc e . Us i n g A gi l ent I ns t r umen t B ...

  • Agilent Technologies E1300B - page 8

    A b o ut this Man ual M an ual Con t en t T h is m a n u a l s h o ws h o w t o u s e t h e A gi le n t E 1300/ E 1301 M a in fr a m e a n d h o w t o operate and pr og ram i ns trum en ts w i thi n the m ai nf r am e us i ng S CPI ( S tandar d C o m m a n d s f o r P ro g ra m m a b l e I n s t ru m e n t s ) c o m m a n ds a n d I EEE 488.2 Comm ...

  • Agilent Technologies E1300B - page 9

    T ab l e of C onten ts 1. G etting Starte d U sin g T h is C h a p t e r . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 M a in fr a m e D e scr ip t io n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 O p t io n a l M a in fr a m e M e m o r y . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E1300B - page 10

    U sin g S up p o r t e d T e r m in a ls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16 T h e Sup p o r t e d T e r m in a ls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16 U sin g t h e H P 700/ 22 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3- 1 7 U sin g t h e W Y SE Ø W Y -3 ...

  • Agilent Technologies E1300B - page 11

    T h e Se r v ic e R e q u e st B it . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5 C le a r in g t h e Se r v ic e R e q u e st E n a b le R e gist e r . . . . . . . . . . . . . . . . . . . . 6-5 Sta n d a r d E v e n t St a t u s R e gist e r . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6 U n m a skin g St a n d ...

  • Agilent Technologies E1300B - page 12

    *SR E < ma sk > . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-70 *SR E ? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-71 *ST B? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-71 *T R G . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies E1300B - page 13

    C . C onn e c tin g and C onf ig urin g a Dis pla y T er m inal U sin g t h is A p p e n d ix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1 O v e r v ie w . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C -1 Connecti ng a T er mi nal to the Ma i nf r ame . . . . . . . . . . . . . . ...

  • Agilent Technologies E1300B - page 14

    6 - Ta ble of C ontents ...

  • Agilent Technologies E1300B - page 15

    Ch a p t e r 1 G etting S tar ted Us i n g T h i s Ch ap t er T h is c h a p t e r d e s c r ib e s t h e A gi le n t E 1300B / E 1301B M a in fr a m e , d e fin e s t h e i ns trum en t c on c ept, and ex pl ai ns ho w pl ug -i n mo dul es are des i g nated as i ns trum en ts i n th e m ai nf r am e. Thi s c hapter al s o con tai ns i ntroduc to r ...

  • Agilent Technologies E1300B - page 16

    Tr i g O u t : Al l o w s an i n s tr um en t to o u tput a neg ati v e-g oi ng p ul s e to i ndi c ate the occ urr ence of som e ev ent such as cl os ing a channel on a S w itc hbox Ins trument. The s i g nal l ev el s are s tand ard TTL (0 V to 5 V). Th i s pul s e c an be u s ed to s y nchroni z e ex ter nal eq ui p men t to th e i ns tr um ent ...

  • Agilent Technologies E1300B - page 17

    1 I n strume nt De fini tion S CPI- c ompati b l e pl ug -i n mo dul es i ns tall ed i n th e m ai nf r am e are treated as i ndep end ent i ns t r uments each hav i ng a uni q ue secondar y G P IB add r ess . As s ho w n i n Fi g u re 1 -3 , eac h i n s tr um en t i s as s i g ned a ded i cated er ror q ueue, i n put an d o u t pu t b u f f ers , ...

  • Agilent Technologies E1300B - page 18

    Instr u m ent L ogical A d d r e sse s I n s t r u m e n t s a r e id e n t ifie d b y a lo gica l a d d r e ss whic h d ir e ct ly r e la t e s t o it s G P I B s ec on dary ad dr es s . I ns truments com e f rom the f actory w i th a pres et l og i c al ad d re s s . Y ou c an c hang e t he f actory setti ng d uring i ns tal l at i on (see the &q ...

  • Agilent Technologies E1300B - page 19

    Ex ample: Read ing the T ime Thi s prog r am reads and prin ts the ti me f r om th e S y s tem In s tr um ent’ s in ternal c lo c k . T h e c o m p u t e r u se d in t h e e x a m p l e is a n A gi le n t Se r i e s 200/ 300 c o m p u t e r w i th Ag i l ent B AS IC as the pr og ram l ang uag e. Th e c o m puter i n terf aces to the mai nf rame u ...

  • Agilent Technologies E1300B - page 20

    1-6 Gett ing St ar ted ...

  • Agilent Technologies E1300B - page 21

    Ch a p t e r 2 Us in g th e F r ont Panel Us i n g t h i s Ch a p t er T h is c h a p t e r sh o ws y o u h o w t o u s e t h e A gi le n t E 1301B M a in fr a m e ’ s fr o n t p a n e l k ey board and d i spl ay to oper ate i ns trum ents i n th e m ai nf r am e. It co n tai ns the f oll ow i ng sec ti ons: • F r o n t P a n e l F e a t u r e ...

  • Agilent Technologies E1300B - page 22

    1 Usi n g M enu s You can acc es s a S y s t em Ins trument menu and a v a r i e ty of other ins tr ument menus (d ep end i ng on i ns tal l ed ins tr uments) f rom the f ront p anel. The s e m enus i nc orp or ate t he mos t us ed f unc ti ons but do not p r ov i de a cc es s to al l of the i ns t r ument com m ands. If a par t i c ul ar f unc ti ...

  • Agilent Technologies E1300B - page 23

    • P r e ss Prev Menu to return to the prev i ou s menu w i thi n an i ns tr umen t menu or e s c ape f r om an i nput p r ompt. Press Sel ect I ns tr to retur n to the S elect an I n str um en t menu . Note that w hen y ou l eav e an i n s tr um ent an d retur n l ater , y ou retur n to th e s ame m en u l ocati on y ou w er e w hen y ou l ef t. ...

  • Agilent Technologies E1300B - page 24

    H o w to Disp la y Log ic a l A d d r e sse s o r I nstrum e nt I nfo rm a tio n H o w to Re se t the Syste m Note: The RE SET menu sel ec ti on i s equi v al ent to the D IA G : B O O T c omma nd w hi c h has the s ame ef fec t as c y c li ng pow er to the mainf r a me. Press ing Reset Ins t r from the S y s tem Ins t rument menu i s equiv al ent ...

  • Agilent Technologies E1300B - page 25

    Us i ng the Other I ns tru ment M e nu s The i ns tr ument menus al low y ou to acces s t he mos t- us e d ins t rument func ti ons or t o moni tor a n i nstrument ( moni tor mod e) w hil e it i s bei ng c ontrol l ed f rom re mote. W e’ l l use the S w i tc hbox m enu to s how y ou how to use the i ns trument menus . M enus are a v ai l able f o ...

  • Agilent Technologies E1300B - page 26

    H o w to O p e n / C lo se C ha nne ls How to S c an Cha n n e ls 2-6 Using the Fr ont Panel ...

  • Agilent Technologies E1300B - page 27

    How t o Dis play M oni tor Ty pe, Desc r i pt i on , or R eset M odul e H o w to Se le ct M o nito r M o d e Us ing the Front Panel 2- 7 ...

  • Agilent Technologies E1300B - page 28

    Mo n i t o r Mo d e M o n it o r m o d e d is p la y s t h e st a t u s o f a n in st r u m e n t while it is b e in g co n t r o lle d f r om re mote. Moni t or mode i s usef ul f or de bug g i ng programs . You can p l ace an i ns t r ument i n moni tor mode us i ng f r ont p anel m enus , or by ex ecuti ng the D IS P : MO N: S TAT O N com m and ...

  • Agilent Technologies E1300B - page 29

    1 Execut ing C o mma n d s F r o m t h e f ro n t p a n e l , y o u c a n t y pe a n d e x e c u t e I EEE 488. 2 C o m m o n Com m ands and S CPI Com m ands f o r the i ns trum ent pres en tl y s el ected by the S elect an in str u m en t menu. ( H ow ev er , y ou cannot ex ecute a comm and w hen t he d isp la y is r e q u e s t in g t h a t y o u ...

  • Agilent Technologies E1300B - page 30

    1 K e y De scri ptions Thi s secti on e x p l a i ns the f uncti on of each of the f ront p ane l ’ s ded i c ated k ey s . If a key i s not f uncti onal in a par tic ul ar si tua ti on, p r es s ing that k ey doe s nothi ng ex cep t to cause a beep . U sers of the op tional IBAS IC interp r et er s houl d r ef er t o their IBAS IC manual s et f ...

  • Agilent Technologies E1300B - page 31

    the cur s or to t he beg i nni ng of t he l i ne. Pre s s ing CTRL f ollo w e d b y th e le ft arr ow k ey mov es the curs or 4 charac ter s paces to the l ef t. E r as es the c harac ter at the presen t c urs or pos i ti on (f o r us er - entered data onl y ) . E r as es the c harac ter to the l ef t of the curs or (f or us er - entered data onl y ...

  • Agilent Technologies E1300B - page 32

    1 I n Case of D i ff i c ul t y Pr ob le m : Pr ob le m C au se/ Solu t ion : E r r o r - 113 u n d e fi n e d h e a d e r e r r o r o c c u r s a ft e r e n t e r in g da ta in respons e to a menu pr omp t. For som e com m ands us e d by the menus, the da ta entered i s ap pended to a co m m and header. For ex am pl e, i f y ou enter " 1 &quo ...

  • Agilent Technologies E1300B - page 33

    2 I ns trume nt M e n us T hi s secti on contains char ts s how i ng the structur e and content for al l f ront p ane l i ns t r ument menus . A l s o s how n i n the charts are the S CPI or Comm on Comm ands us e d a nd d escri ptions of m enu- c ontroll ed i ns tr ument op er at i ons . Thi s secti on contains the f oll ow i ng char ts : • Sy s ...

  • Agilent Technologies E1300B - page 34

    Syste m I nstrum e nt M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 Lev el 5 Lev el 6 U ser En t r y Co m m a n d ( s) U sed D escr i p t i o n S Y S T E M C O N F IG ? L A D D S V X I:C O NF :D L A D ? D is p la y s lo g ic a l a d d re s s e s o f m a infra m e ins trum e nts DE VI C E l o gi cal ad d r es s VXI : C ...

  • Agilent Technologies E1300B - page 35

    Syste m I nstrum e nt M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 Lev el 5 Lev el 6 U ser En t r y Co m m a n d ( s) U sed D escr i p t i o n ( cont i n u e d f r om pr e vi ou s page ) CONT ROL DT R RE A D car d n u m b er S YS T: COM M : S ER[ n ] : CON T: D T R? R e ad c u r r en t s et t i n g f o r D T R l i n e ...

  • Agilent Technologies E1300B - page 36

    Switc hb o x M e n u Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 U ser En t r y Co m m a n d ( s ) U s ed D escr i p t i o n S WI TC H M ONI TOR c a r d n u mbe r ‡ or AU TO DI S P : M O N: C AR D < c a r d_nu mbe r > ; S T A T O N M on i t or i n s t r u me nt ope r at i on s O P E N c ha nn e l lis t † O P E N (@ c h a n ...

  • Agilent Technologies E1300B - page 37

    Notes Us ing the Front Panel 2- 1 7 ...

  • Agilent Technologies E1300B - page 38

    Sca nning V o ltm e te r M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 U ser En t r y Co m m a n d ( s) U sed D escr i p t i o n V O L T MTR MO NIT O R c ha nn e l lis t † o r 0 fo r a uto D IS P :M O N :C HAN < c ha nne l _ l i s t > ;S T A T O N M onit or i ns t r ume nt op e r a ti ons V D C c ha nne l l i s ...

  • Agilent Technologies E1300B - page 39

    Sca nning V o ltm e te r M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 U ser En t r y Co m m a n d ( s) U sed D escr i p t i o n ( cont i n u e d f r om pr e vi ou s page ) U NS TR N c ha nne l l i s t † ME A S : S TR : U NS T? < c ha nne l _ li s t > Me a s ur e br idg e uns t r a ine d D IAG C O M P R E S c h ...

  • Agilent Technologies E1300B - page 40

    A gil ent E 1326B / E1411B 5 1/ 2 Di gi t M ul ti m ete r ( St andal on e) M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 User En t r y Co m m a n d ( s ) U sed D escr i p t i o n VOL T M T R M ON I T OR DI S P : M ON : S T A T ON Di spl ay i n s t r u m en t o p er at i o n s VDC M E AS : VO L T: DC? M e as u r e DC v ...

  • Agilent Technologies E1300B - page 41

    A gil ent E 1328A 4-Chann el D/ A C onv e rt er M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 User En t r y Co m m a n d ( s) U sed D es cr i p t i o n D / A MO NI TO R C HAN 1 D IS P :M O N :C HAN 1 ; S T A T O N Monit or ins t r ume nt op e r a ti ons on c ha nne l 1 C H A N 2 D I S P :M O N: C HA N 2 ;S T A T O N Mo ...

  • Agilent Technologies E1300B - page 42

    A gil ent E1330A Quad 8-B it Di gi tal Input / Out pu t M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 U s er En t r y Co m m a n d ( s) U sed D escr i p t i o n D IG _ I/ O MO N IT O R P O R T 0 D IS P :MO N :C HA N 0 ;S T A T O N Mo nito r ins trum e nt o p e ra tio n s o n p o rt 0 P O R T1 DI S P : M ON: C HAN 1; S ...

  • Agilent Technologies E1300B - page 43

    Notes Us ing the Front Panel 2- 2 3 ...

  • Agilent Technologies E1300B - page 44

    A gil ent E1332A 4-Chan nel Coun t er/ T otal i zer M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 Lev el 5 U ser En t r y C om m an d ( s) U sed D escr i p t i o n C O U NTE R MO NI TO R C HAN 1 D IS P :M O N :C HAN 1 ; S T A T O N M onit or i ns t r ume nt op e r a ti ons on c ha nne l 1 C H A N 2 D I S P :M O N :C H ...

  • Agilent Technologies E1300B - page 45

    A gil ent E1332A 4-Chan nel Coun t er/ T otal i zer M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 Lev el 5 U ser En t r y C om m an d ( s) U sed D escr i p t i o n ( cont i n u e d f r om pr e vi ou s page ) T IME IN T C H A N 1 T R IG :S O U R IMM;:ME A S 1 :T INT ? T im e inte rv a l m e a s u re m e nt o n c ha nn e ...

  • Agilent Technologies E1300B - page 46

    A gil ent E 1333A 3-Chann el Un i v e rs al C o unte r M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 L ev el 5 U ser En t r y Co m m a n d ( s ) Used D escr i p t i o n C OU NTE R M ONI TOR C H AN1 DI S P : M ON: C HAN 1; S T A T O N M on i t or i n s t r ume nt ope r at i on s on chann e l 1 C H AN2 DI S P : M ON: C H ...

  • Agilent Technologies E1300B - page 47

    A gil ent E 1333A 3-Chann el Un i v e rs al C o unte r M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 L ev el 5 U ser En t r y Co m m a n d ( s ) Used D escr i p t i o n ( cont i n u e d f r om pr e vi ou s page ) T IME IN T C H A N 1 T R IG :S O U R IMM;:ME A S 1 :T INT ? T im e inte rv a l m e a s ure m e nt o n c ha ...

  • Agilent Technologies E1300B - page 48

    Notes 2 - 2 8 U s ing t he F r ont P a n e l ...

  • Agilent Technologies E1300B - page 49

    Ch a p t e r 3 Us ing the Dis p lay T er m inal I nter f ace U s i n g t h i s C h a p t e r T h i s c h a p t e r s ho w s y ou ho w t o u s e t h e A g il e n t E 1300 B a n d A g il e n t E 1301B Mai nf r ames ’ Di s pl ay Ter mi nal Interf ac e ( termi nal i nterf ac e) to operate i ns trum en ts i n th e m ai nf r am e. The termi nal i nterf ...

  • Agilent Technologies E1300B - page 50

    1 Te rmin a l I n ter fa ce Fe a ture s F igu r e 3-2 sh o w s a t y p ic a l t e r m in a l in t e r fa c e d isp la y with it s fu n c t io n la b e ls acros s the b o ttom of the s creen. The f i rs t f i v e f un c ti on k ey s ( f1 thr oug h f5 ) s e l ec t i ns t r ument menu c hoi ces . F uncti on k ey s f6 throug h f8 pr ov i de menu contro ...

  • Agilent Technologies E1300B - page 51

    1 Usi n g M enu s A S y s tem Ins t rument menu and a v ar i ety of other i ns t rument menus (d ep end i ng on i ns tal l ed i ns truments ) are av ai l abl e f rom the term i nal i n terf ace. Thes e m en u s i nc orp or ate t he mos t us ed f unc ti ons but do not p r ov i de a cc es s to the compl ete f uncti onal ity of a n i nstrument. If a p ...

  • Agilent Technologies E1300B - page 52

    I f y o u m a ke a n in co r r e c t e n t r y in r e sp o n se t o a n in p u t p r o mp t , t h e b o t t o m lin e o f t h e T e x t O u t p u t A r e a w ill sho w a n e r r o r m e ssa ge . W h e n thi s ha pp ens, jus t s el ect tha t menu choi ce ag ai n ( f1 - f5 k e ys ) , r e - t yp e t h e c o rre c t i nf orm ati on , and pre s s Re t u ...

  • Agilent Technologies E1300B - page 53

    Us i ng t he Sy st em I ns tru ment M e nu T he S y stem Ins trument menu all ow s y ou to: • S et or r ead the s y stem G P IB addr es s • R e s et (r eboot ) t he mai nf r ame • D is p la y t h e lo gica l a d d r e ss e s o f in st a lle d in st r u m e n t s • D i s p l ay i nf or mati on about i ns tal l ed ins tr uments H o w to Se t ...

  • Agilent Technologies E1300B - page 54

    H o w to Re se t the Syste m No t e: Th e RESET m enu s e le ction is e q uiva le nt to ex e cut ing the DIAG:B OOT com m and which ha s the s a m e ef f ect as c y c li n g t h e m ai n f ram e’ s p o w er. Pressi n g RST_ IN ST from t he S yste m I nstrum e nt m e nu is the equiv a lent to se nd ing the *R S T c o mma nd t o t h e S ys t e m I ...

  • Agilent Technologies E1300B - page 55

    H o w to Disp la y Log ic a l A d d r e sse s a nd I n str um e nt I nfo r m a tio n E nt e r de v i c e ’s lo gi c a l a ddr e s s a nd pr e s s Return f o r i n di v idu a l ins tr um ent infor mation, or jus t e nte r one s pace a nd Re turn , fo r inform a tion on a ll intrum ents . (In t h i s c ase, 8 was en t ere d ) GPI B seco n d ar y ad ...

  • Agilent Technologies E1300B - page 56

    Us i ng the Other I ns tru ment M e nu s The i ns tr ument menus al low y ou to acces s t he mos t- us e d ins t rument func ti ons or t o moni tor a n i nstrument ( moni tor mod e) w hil e it i s bei ng c ontrol l ed f rom re mote. W e’ l l use the S w i tc hbox m enu to s how y ou how to use the i ns trument menus . M enus are a v ai l able f o ...

  • Agilent Technologies E1300B - page 57

    H o w to O p e n / C lo se C ha nne ls Ent er C h an nel Li st an d p r ess Return (e.g ., 102 f o r c h ann el 2 o n c ard # 1) S w itc hbox ins trument a t logical a dd re s s 3 2 (sec on d ary ad d r ess = 04) SC PI c o m m a n d u s ed : OPEN < cha nnel_list> SC PI c o m m a n d u s ed : C LOSE < channel_list> How to S c an Cha n n ...

  • Agilent Technologies E1300B - page 58

    How t o Dis play M odul e Ty pe , Descr i pti on , or R eset M odul e Ent er C ard Nu m b er an d p ress Re turn Ent er C ard Nu m b er an d p ress Re turn Ent er C ard Nu m b er an d p ress Re turn SC PI c o m m a n d u s ed : SY ST: C PON < car d_ number> SC PI c o m m a n d u s ed : SY ST:C D ES? < c ard_ number> SC PI c o m m a n d ...

  • Agilent Technologies E1300B - page 59

    Mo n i t o r Mo d e M o n it o r m o d e d isp la y s t h e st a t u s o f a n in st r u m e n t while it is b e in g co n t r o lle d f r om re mote. Moni t or mode i s usef ul f or de bug g i ng programs . You can p l ace an i ns t r ument i n moni tor mode us i ng ter mi nal i nter f ac e menus , or by ex ecuti ng the D IS P: MON: S TA T O N c o ...

  • Agilent Technologies E1300B - page 60

    Note E nabl i ng m oni tor mode s l ow s i nstrument op er ations . If the t i m i ng or s peed of i ns trum en t o perati o ns i s cri ti cal (su c h as m aki ng mul ti meter r eadi n g s at a pr e ci se ti me interv a l ), y ou s houl d not use m oni tor mode . Table 3 -1 s how s the s tatus annunci ator s that may ap p ear i n the bottom l i ne ...

  • Agilent Technologies E1300B - page 61

    1 Execut ing C o mma n d s F r o m t h e t e rm i n a l i n t er f a c e , y o u c a n t y pe a n d e x e c u t e I EEE 488.2 C o m m o n Com m ands and S CPI Com m ands f o r the i ns trum ent pres en tl y s el ected by the S elect an in str u m en t menu. ( H ow ev er , y ou cannot ex ecute a comm and w hen t he sc r e e n is r e q u e s t in g t ...

  • Agilent Technologies E1300B - page 62

    1 Ge ne ra l K e y De scrip tions Thi s secti on ex plai ns the f unc ti on of each of the t er mi nal i nte rf ace’ s m enu, m enu control, and ed i ti ng k e y s . If a k ey i s not f uncti onal in a par tic ul ar si tuat i on, pr es s i n g that k ey d oes no th i ng ex cept to c au s e a beep. M e nu a nd M e nu Co n t ro l K ey s Labe l menu ...

  • Agilent Technologies E1300B - page 63

    E r as es the c harac ter to the l ef t of the curs or (f or us er - entered data onl y ) . (Cl ear- to - end k ey . ) E r as es al l ch aracters f rom the pres ent c ur s or pos i ti on to the end of the i n put l i n e (f or us er- en tered d ata onl y ). S el ec ts the upp er-c as e al phabeti c c har acters or the c h arac ter s ho w n on th e ...

  • Agilent Technologies E1300B - page 64

    1 Using Su ppor t e d Te rmin a ls The D i s play Ter mi nal Interf ace s upports s ev er al p opul ar termi nal b rands and mo d e ls. T h is c h a p t e r will sho w y o u h o w to a c c e s s a ll o f t h e t e r m in a l in t e r fa c e f uncti ons d e s c ri bed p r ev i ous ly us ing y our s upp or ted ter mi nal . The Support ed Te rmi nal s ...

  • Agilent Technologies E1300B - page 65

    Us i ng t he H P 700/ 22 T h e H P 700/ 22 t e r m in a l e m u la t e s t h e D E C ® V T 100 ® o r V T 220 ® ter mi nal s. S ome f unc ti ons of the D i s p l ay Ter mi na l Int er f ace hav e been map p ed i nto k ey s w ith other l abel s . A key boar d map i s p r ov i ded f or each of the emul ati on m odel s . U s e thes e k ey board maps ...

  • Agilent Technologies E1300B - page 66

    VT220 ® Key M ap The f unction k e y s that ar e nor mal l y l abel ed f6 thr oug h f14 are now l abel ed: Note Be c a u s e t h e H P 700/ 22 k e y b o a r d h a s n in e fu n c t i o n k e ys in t h e c e n t e r o f t h e k ey b o a rd, f 4 i s m a ppe d t w i c e The sy m bol s s how n i n the upp er lef t corne r of key each ar e now mapp e d ...

  • Agilent Technologies E1300B - page 67

    Us i ng the WY SE  WY - 3 0  Wi th the W Y S E  WY-3 0  ter mi nal , s ome f unc ti ons of the D i s p l ay Ter mi nal Interf ace hav e been as s i g ned to k ey s w i th other l abel s . U s e thi s k ey board map to hel p l oc ate the s e f uncti ons . The sy m bol s s how n i n the upp er lef t corne r of key each ar e now mapp e d w ...

  • Agilent Technologies E1300B - page 68

    What “ Not Su pported” Me a n s St r ict ly s p e a kin g, a t e r min a l is n o t su p p o r t e d if it h a s n o t b e e n r igo r o u sly t e st e d w i th th e Di s pl ay Term i nal Interf ac e. Th ere are s ev er al H P ter mi nal s w hi ch m ay be c o m pati bl e w i th the termi nal i nterf ace. Ter m i nal s s uc h as the DEC ® VT 10 ...

  • Agilent Technologies E1300B - page 69

    If y ou now see the “S elect an i nstrument” menu l abel s : G o to the b eg i nni ng of thi s ch apter and try the m enu s . or Turn the m ai nf rame of f and then on ag ai n . Us i n g a Ter mi n al Wit hout M e nu s Y o u c a n st ill c o n t r o l in s t r u m e n t s in st a lle d in y o u r m a in fr a me w it hout us i ng the t e r m in ...

  • Agilent Technologies E1300B - page 70

    To g et a l i s t o f the l o g i cal ad dr es s es u s ed i n y our mai nf rame, s end the S CPI comma n d V X I: CONF: DLA D? to th e S y s tem In s tr um en t. Then to determi ne w hat i nstrum ent i s at eac h l og i c al add ress , s end the com mand VX I : C O N F : D L I S ? n fo r e a c h lo gic a l a d d r e s s in t h e list ( w he r e n ...

  • Agilent Technologies E1300B - page 71

    1 I n Case of D i ff i c ul t y Pr ob le m : Pr ob le m C au se/ Solu t ion : E r r o r - 113 u n d e fi n e d h e a d e r e r r o r o c c u r s a ft e r enter ing da ta in respons e to a menu pr omp t. For som e com m ands us e d by the menus, the da ta entered i s ap pended to a co m m and header. For ex am pl e, i f y ou enter " 1 " as ...

  • Agilent Technologies E1300B - page 72

    Notes 3-24 Us in g th e Dis pl ay T er minal Inter face ...

  • Agilent Technologies E1300B - page 73

    2 I ns trume nt M e n us T hi s secti on contains char ts s how i ng the structur e and content for al l te rmi nal i nter f ace i ns trum en t m enu s . Al s o s how n i n the ch arts are the S CPI or Com m on Comm ands us e d a nd d escri ptions of m enu- c ontroll ed i ns tr ument op er at i ons . Thi s secti on contains the f oll ow i ng char t ...

  • Agilent Technologies E1300B - page 74

    Syste m I nstrum e nt M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 Lev el 5 Lev el 6 U ser En t r y Co m m a n d ( s) U sed D escr i p t i o n S Y S T E M C O N F IG ? L A D D S V X I:C O NF :D L A D ? D is p la y s lo g ic a l a d d re s s e s o f m a infra m e ins trum e nts DE VI C E l o gi cal ad d r es s VXI : C ...

  • Agilent Technologies E1300B - page 75

    Syste m I nstrum e nt M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 Lev el 5 Lev el 6 U ser En t r y Co m m a n d ( s) U sed D escr i p t i o n ( cont i n u e d f r om pr e vi ou s page ) CONT ROL DT R READ car d n u m b er S YS T: COM M : S ER[ n ] : CON T : D T R? R e ad cu r r en t se t t i n g f o r D T R l i n e S ...

  • Agilent Technologies E1300B - page 76

    Switc hb o x M e n u Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 U ser En t r y Co m m a n d ( s ) U s ed D escr i p t i o n S WI TC H M ONI TOR c a r d n u mbe r ‡ or AU TO DI S P : M O N: C AR D < c a r d_nu mbe r > ; S T A T O N M on i t or i n s t r u me nt ope r at i on s O P E N c ha nn e l lis t † O P E N (@ c h a n ...

  • Agilent Technologies E1300B - page 77

    Notes Us ing the Di s play Ter minal Inter face 3-29 ...

  • Agilent Technologies E1300B - page 78

    Sca nning V o ltm e te r M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 U ser En t r y Co m m a n d ( s) U sed D escr i p t i o n V O L T MTR MO NIT O R c ha nn e l lis t † o r 0 fo r a uto D IS P :M O N :C HAN < c ha nne l _ l i s t > ;S T A T O N M onit or i ns t r ume nt op e r a ti ons V D C c ha nne l l i s ...

  • Agilent Technologies E1300B - page 79

    Sca nning V o ltm e te r M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 U ser En t r y Co m m a n d ( s) U sed D escr i p t i o n ( cont i n u e d f r om pr e vi ou s page ) U NS TR N c ha nne l l i s t † ME A S : S TR : U NS T? < c ha nne l _ li s t > Me a s ur e br idg e uns t r a ine d D IAG C O M P R E S c h ...

  • Agilent Technologies E1300B - page 80

    A gil ent E 1326B / E1411B 5 1/ 2 Di gi t M ul ti m ete r ( St andal on e) M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 User En t r y Co m m a n d ( s ) U sed D escr i p t i o n VOL T M T R M ON I T OR DI S P : M ON : S T A T ON Di spl ay i n s t r u m en t o p er at i o n s VDC M E AS : VO L T: DC? M e as u r e DC v ...

  • Agilent Technologies E1300B - page 81

    A gil ent E 1328A 4-Chann el D/ A C onv e rt er M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 User En t r y Co m m a n d ( s) U sed D es cr i p t i o n D / A MO NI TO R C HAN 1 D IS P :M O N :C HAN 1 ; S T A T O N Monit or ins t r ume nt op e r a ti ons on c ha nne l 1 C H A N 2 D I S P :M O N: C HA N 2 ;S T A T O N Mo ...

  • Agilent Technologies E1300B - page 82

    A gil ent E1330A Quad 8-B it Di gi tal Input / Out pu t M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 U s er En t r y Co m m a n d ( s) U sed D escr i p t i o n D IG _ I/ O MO N IT O R P O R T 0 D IS P :MO N :C HA N 0 ;S T A T O N Mo nito r ins trum e nt o p e ra tio n s o n p o rt 0 P O R T1 DI S P : M ON: C HAN 1; S ...

  • Agilent Technologies E1300B - page 83

    Notes Us ing the Di s play Ter minal Inter face 3-35 ...

  • Agilent Technologies E1300B - page 84

    A gil ent E1332A 4-Chan nel Coun t er/ T otal i zer M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 Lev el 5 U ser En t r y C om m an d ( s) U sed D escr i p t i o n C O U NTE R MO NI TO R C HAN 1 D IS P :M O N :C HAN 1 ; S T A T O N M onit or i ns t r ume nt op e r a ti ons on c ha nne l 1 C H A N 2 D I S P :M O N :C H ...

  • Agilent Technologies E1300B - page 85

    A gil ent E1332A 4-Chan nel Coun t er/ T otal i zer M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 Lev el 5 U ser En t r y C om m an d ( s) U sed D escr i p t i o n ( cont i n u e d f r om pr e vi ou s page ) T IME IN T C H A N 1 T R IG :S O U R IMM;:ME A S 1 :T INT ? T im e inte rv a l m e a s u re m e nt o n c ha nn e ...

  • Agilent Technologies E1300B - page 86

    A gil ent E 1333A 3-Chann el Un i v e rs al C o unte r M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 L ev el 5 U ser En t r y Co m m a n d ( s ) Used D escr i p t i o n C OU NTE R M ONI TOR C H AN1 DI S P : M ON: C HAN 1; S T A T O N M on i t or i n s t r ume nt ope r at i on s on chann e l 1 C H AN2 DI S P : M ON: C H ...

  • Agilent Technologies E1300B - page 87

    A gil ent E 1333A 3-Chann el Un i v e rs al C o unte r M e nu Me nu L e ve l s a nd C o nte nt Lev el 1 Lev el 2 Lev el 3 Lev el 4 L ev el 5 U ser En t r y Co m m a n d ( s ) Used D escr i p t i o n ( cont i n u e d f r om pr e vi ou s page ) T IME IN T C H A N 1 T R IG :S O U R IMM;:ME A S 1 :T INT ? T im e inte rv a l m e a s ure m e nt o n c ha ...

  • Agilent Technologies E1300B - page 88

    Notes 3-40 Us in g th e Dis pl ay T er minal Inter face ...

  • Agilent Technologies E1300B - page 89

    Ch a p t e r 4 Us i ng th e Mainf r ame Us i n g t h i s Ch ap t er T hi s chap ter show s how to use the mai nf r ame’ s Pacer f uncti on, how to chang e the prim ar y G PIB addr es s , and how to s y nc hroni z e i nternal and ex ter nal i ns trum en ts us i ng the m ai nf r am e’ s E v ent In an d Tri g g er O ut ports. Thi s c hap ter a lso ...

  • Agilent Technologies E1300B - page 90

    • S O U R :PU LS : CO U N s ets the number of Pacer cy c l es. S peci fy from 1 to 8388607 c y c le s o r sp e c ify I N F fo r a c o n t i nuous output. • S O U R :PU LS : P E R sets the pe riod of each Pacer cy cl e. Y ou c an s peci f y p e r io d s fr o m 500n s t o 8.3 s e c o n d s . • TR IG : S OU R s ets the tri g g er s ou rc e. The ...

  • Agilent Technologies E1300B - page 91

    1 Chan gin g t h e Prima ry GP I B A d dr ess You c an s et the mai nf r am e’ s pri m ar y G PIB addr es s to any i nteg er v al u e betw een 0 and 3 0 . Th e addr es s i s s et to 9 at the f ac tory . ( S ee Chapter 2 f or i ns t r ucti ons on s e tting /read ing the G P IB add r ess f rom the f ront p ane l . ) The f ol l ow i ng com m and s e ...

  • Agilent Technologies E1300B - page 92

    Ex am ple : Sy nchr oniz ing an In te r nal In strum e n t to an E x te r n al In s tru m e n t Thi s ex ampl e u s es th e m ai nf r ame’ s Tr i g O ut and E v ent In ports to s y nch roni z e an ex t er nal m ul ti m eter to a mul ti p lex er ins t al l ed i n the mai nf rame. Connecti ons are sh o w n i n Fi g ure 4 -3 . Th e m ul ti meter’ ...

  • Agilent Technologies E1300B - page 93

    mul tim eter to an i nterna l m ul tipl ex er. Connecti ons are show n i n Fi g ur e 4 - 4 . Thi s method sy nchroni z es the com puter to t he i ns tr uments and r eli es on the co m puter to enter each readi n g and adv ance to th e n ex t chann el i n th e s can l i st. The s eq uenc e o f op erati o n i s : 1. I N I T ( l in e 50) c lo s e s c ...

  • Agilent Technologies E1300B - page 94

    1 M a in f ram e D at a M e mory When p ow e r i s app li ed or the sy s tem reboote d (D IAG : B O O T comm and) , mai nf rame memory i s a utomati c al ly conf i g ured to p r ov i de a p r ed ef i ned amount of mem o ry f or any i ns tal l ed i ns truments that req ui r e mem or y s p ace. For ex amp l e, each m ul tim eter i ns tr ument w i thi ...

  • Agilent Technologies E1300B - page 95

    Ex am ple : Storing and R etriev ing Data F r om M ainf r am e M e m or y . T h is ex am pl e sh o w s ho w to u s e mai nf r ame m emory to s tor e 1 5 readi n g s m ade us i ng a n A gi le n t E 1326A M u l t i m e t e r . A ft e r t h e r e a d in gs a r e st o r e d , t h e y a r e r e t r ie ve d by the c omputer and d i s p l ay ed . 10 R EAL ...

  • Agilent Technologies E1300B - page 96

    a d d r e ss . Y o u w ill t h e n kn o w t h e s t a r t in g a d d r e ss , a n d ( fr o m t h e … NR A M:CR E < size > com mand) the l en g th of the N RA M s eg ment. Ex a mple: A ll ocat ing an NRA M s egm e nt and lo cati ng it. Thi s ex ampl e sho w s h o w t o a l lo c a t e a s m a ll 128 b y t e N R A M s e gm e n t . I n a d d it ...

  • Agilent Technologies E1300B - page 97

    Us i ng : DOWNload and : U Pl oad? t o A c ces s Dat a T h e c om m a nd DI AG: DOW N l oa d < a ddr e ss >, < d a ta_blo ck > is used to store data i n to th e N RAM s eg men t. The c om mand DI A G: U PL o a d? < a ddr e ss >, < by t e _ c ount > i s us ed to retr i ev e d ata f r om th e NRA M s eg ment. The addr es s pa ...

  • Agilent Technologies E1300B - page 98

    Ex ampl e: St oring and Retriev ing data us ing DOW Nl oad a nd UPL o ad. define var iable s for DO WN load an d UPL o ad 90 DIM Ch ar s $[ 64] ,Ch ar s_b ac k $[ 80] 100 IN TE G E R Wo r d s ( 1: 16) , By t es ( 1: 32) , W o r d s _ b ac k ( 1: 16) , B yte s_ b a ck(1 :3 2 ) cr eate s tr in g o f ch ar acter s 110 Char s$= "123456789012345678 ...

  • Agilent Technologies E1300B - page 99

    Ch a p t e r 5 Do w nlo a d ing De v ic e Dr iv e r s Ab out thi s C h a pte r Thi s c hapter d es cri b es the pr oc edure f or us i ng d ow nl oad abl e d ev i ce dri v er s w it h t h e A gile n t E 1405 C o m m a n d M o d u le . T h i s fu n c t i o n a lit y w a s a d d e d s o t h a t SCP I c a p a b ilit y f or n e w r e gist e r b a se d d ...

  • Agilent Technologies E1300B - page 100

    Figu r e 5-1. Driv er and Do cument ati on U s age 5-2 Dow nlo ading Dev ice Dr iv er s GPIB bus GPIB bus GPIB GPIB bus ...

  • Agilent Technologies E1300B - page 101

    1 M e mory C o nfig ura tion Bef or e a ttemp ti ng to d ow nload any d e v i c e dr i v ers y ou s houl d und er s tand how memory i s af f ected w hen y ou speci f y a si ze f or one or more t y pe s of R A M. There are thr ee ty pes of R AM that y ou c an al l ocate i n the m ai nf r am e: • RA M di s k (RDI S K ) • N o n - v o la t ile R A ...

  • Agilent Technologies E1300B - page 102

    1 Download P rogram C o nfig ura tion I f y o u w ill no t b e u sin g t h e d e fa u lt co n figu r a t io n s fo r d o w nlo a d in g, y o u w ill ne e d to ed i t the conf i g ur ation f i l e t o match y our sy s t em c onf i g ur ati on. If the de f a ul t v al ues s how n below ar e corr ect f or y our s etup, y ou can pr oceed to the app rop ...

  • Agilent Technologies E1300B - page 103

    sp e cify a file n a m e h e r e , t h e d r iv e r d o w nlo a d e r will log t o t h e s cr e e n a n d t o t h e sp e c ifie d file . DRIV E R FIL E = s p e c ifie s t h e d r iv er file o r file s t o d o w nlo a d . T h e d e fa u lt is t o dow nl oad a l l d ev ic e dr i v er fi l es f ound i n the director y s peci f i ed by D I REC TO RY = ...

  • Agilent Technologies E1300B - page 104

    1 Download ing Drivers in M S- DO S System s Th e de v i c e dri v er do w nl o a d pro g ram V XIDLD. E XE prov i ded o n t h e di s k w i th t h e d r i v e r file s fo r u se wit h a n R S-232 i n t e r fa c e m u s t b e r u n fr o m M S-D O S. I t w ill se t u p t h e t h e r e q u ir e d d e v ice d r iv e r m e m o r y a n d a n y ot h e r m ...

  • Agilent Technologies E1300B - page 105

    1 Download ing Drivers in GPI B System s with IB A S IC Th e de v i c e dri v er do w nl o a d pro g ram A U TOS T prov i ded o n the di s k w i th t h e dr iv e r f il es f or us e w i th G PIB m us t be r un f r om IB AS IC ( Instrument Bas i c) . It w ill se t u p t h e t h e r e q u ir e d d e v ice d r iv e r m e m o r y a n d a n y ot h e r m ...

  • Agilent Technologies E1300B - page 106

    1 Download ing Drivers in GPI B System s w ith BA S I C Th e de v i c e dri v er do w nl o a d pro g ram V XIDLD_ GE T pro v i ded on the di s k w i th the d r i v e r f il es f or us e w i th G PIB m us t be r un f r om an BAS IC othe r t han I BA SI C . I t w ill se t u p t h e t h e r e q u ir e d d e v ic e d r iv e r m e m o r y an d a n y o t ...

  • Agilent Technologies E1300B - page 107

    1 Download ing M u lt ip le D rive rs The dr i v er dow nl o ader s of tw are automati c al l y chec k s f or the ex i s ten c e of other d r iv e r s whe n it is r u n . I f t h e r e a r e d e v ic e d r iv e r s p r e s e n t , it w ill ab o r t t h e p r o c e ss and i nf or m y ou that y ou mus t f i rst c l ea r t he other d ev ic e dr i v e ...

  • Agilent Technologies E1300B - page 108

    1 M a nua l ly Downloadin g a Dri v erdow n manu al D ow nl oad pr og rams ar e s u ppl i ed f or us e w i th the s y s tem s etups des c ri b ed earl i er i n thi s c hapter. I f y ou hav e a s y st em s etup that does not al l ow the us e of one of the s upp l i e d d ow nl oad p r og rams (f or i ns t ance, i f y ou are us i ng a M a c in t o sh ...

  • Agilent Technologies E1300B - page 109

    M a nu a lly D o wn lo a d in g Over GPIB M a n u a lly d o w nlo a d in g a d r iv e r o v e r G P I B is fa ir ly s t r a igh t fo r w a r d . T h is di s cus s i on as s u m es that the dow nl o adabl e dev i c e dri v er has been s upp l i ed by A gile n t . D r iv e r s su p p lie d b y A gile n t a r e fo r m a t t e d so t h a t y o u ju st ...

  • Agilent Technologies E1300B - page 110

    If y ou ar e g oi ng to us e any othe r setting , y ou mus t set up the ap p r op riate setti ng s i n the S y stem Ins t r ument us i ng the f oll ow i ng comm ands COMM: SE R [ n ] : R EC:B AUD <r a t e > sets B A UD r ate COMM: SE R [ n ] : REC: SB I TS <b i t s > se ts num be r of s t op b i t s DI AG : C O M M STOR saves settin g s ...

  • Agilent Technologies E1300B - page 111

    h a n d sh a k e m o d e is e n a b le d , t h e Sy st e m I n st r u m e n t w ill no t t r a n sm it ch a r a c t e r s w hen ei ther the CTS ( Cl ear to S end ) or the D S R (D ata S et Read y ) l i nes are not as s er ted. Thi s ac ts to pace the S y s tem Ins tr um en t o utput. NOTE T h e E 1405 C o m m a n d M o d u le R S-232 i n t e r fa c ...

  • Agilent Technologies E1300B - page 112

    Transmitt ing Us ing a C AT Command O n H P-U X sy s t ems y ou can us e the ca t co m m and to trans f er the dev i ce dr i v er . The appr opr i ate d ev i c e f i l e m us t ex i st. Al l s hel l com m and s are as s umed to be ex ec ut ed f r om ei ther the /b i n/ s h or /bi n/ k s h s hel l . 1 . S tar t a pr ocess tha t op ens the de v i c e ...

  • Agilent Technologies E1300B - page 113

    Figu r e 5-3. M anually Dow n loadi ng a Dev i ce Dri v er Dow nl oadi ng Dev ice Dr i ver s 5-15 ...

  • Agilent Technologies E1300B - page 114

    5-16 Dow nlo ading Dev ice Dr iv er s ...

  • Agilent Technologies E1300B - page 115

    Ch a p t e r 6 C ont r oll ing Ins tr um e nts Us ing G PI B Ab out thi s C h a pte r Thi s chap ter show s how to contr ol i nstruments i n the mai nf rame f rom an e x t e r n a l c o m pu t e r u s i n g I EEE 488.2 C o m m o n C o m m a n d s a n d t h e G P I B i nter f a ce. T hi s inc l ud es how to moni t or ins t rument status, interr up t ...

  • Agilent Technologies E1300B - page 116

    2 St at us S y st em Str u ct ur e The i ns tr um en t s tatus s tructure m on i tors i mportant ev ents f or an i ns trum ent s uc h as w hen an error occ u rs or w hen a readi n g i s av ai l abl e. Al l i ns trum en ts hav e t he f ol l ow i ng s tatus g r oups and r e g is te r s w i thi n t hos e g r oups: • S ta tus B y te S t atus Gr oup ? ...

  • Agilent Technologies E1300B - page 117

    The St atu s By te Regi st er A s s h o wn in F igu r e 4- 1, t h e St a t u s Byt e r e gis t e r i s t h e h i gh e s t - le ve l r e gi st e r in the status s tr uctur e. Thi s regi s ter contai ns bits w hi c h s ummariz e inf or mati on f r om the othe r s tatus g r oups. NOTE The b i ts i n the o th er s tatus g r oup reg i s ters m u s t be ...

  • Agilent Technologies E1300B - page 118

    Tabl e 4 - 1 s how s eac h o f the S tatus By te r eg i s ter bi ts and des c r i bes the ev ent that w ill se t e a c h b it . Table 4-1. St atu s B yte R egi s ter Bi t Numb er Decimal W eight Des cript ion 0 1 Ins trument S pec i f ic (not us ed by mos t i n s truments) 1 2 Ins trument S pec i f ic (not us ed by mos t i n s truments) 2 4 Ins tru ...

  • Agilent Technologies E1300B - page 119

    The f ol l o w i ng p rog ram reads th e s y s tem i ns trum en t’ s S tatus B y te reg i s ter us i ng the GP IB S er i al Pol l co m m and. 10 P= SPOLL( 70900) R ead S tatus B y t e R eg ist e r u si ng S er ial P oll, pl ac e w e i g hted sum in P 20 PR IN T P Pr in t weig h ted su m 30 END Serv ice Request En able R egist er The S erv i c e R ...

  • Agilent Technologies E1300B - page 120

    Sta nda rd E ve nt Sta tus Regi st er The S tandard E v ent S tatus Reg i s ter i n the S tandar d E v ent s tatu s g roup moni tor s the i ns t rument status ev ents s how n i n Tabl e 4 -2 . W hen one of t hese ev ents occ urs , i t s ets a correspondi ng bi t i n the S tandar d E v ent S tatus Reg i s ter. NOTE The S tandar d E v ent S tatus Reg ...

  • Agilent Technologies E1300B - page 121

    u n m a ske d , a n G P I B se r v ic e r e q u e st ( SR Q ) w ill be ge n e r a t e d . ( " I n t e r r u p t in g t h e E x ter nal Com puter" , l ater i n th i s c hapter con tai ns an ex ampl e p rog ram w hi ch demons trates th i s s equence). Note th at th e S tand ard E v ent S tatus Reg i s ter bi ts th at are not u nm ask ed s t ...

  • Agilent Technologies E1300B - page 122

    Table 4- 3. Operation Status Group - C ondition Register Bi t Numb er Decimal W eight Des cript ion 0 1 C a lib r a t in g 1 2 Se t t lin g 24 R a n g i n g 3 8 S w eepi ng 4 1 6 Measuri ng 53 2 W a i t i n g f o r T R G 66 4 W a i t i n g f o r A R M 7 1 2 8 Co rrec ti n g 8 2 56 Inter rupt ac k now le dg ed (S y stem Ins t rument) 9 -1 2 I ns t r ...

  • Agilent Technologies E1300B - page 123

    B i ts i n the Op er ati on S tatus Gr oup E v ent reg i s ter w hi c h are se t can be deter mi ned w i th the c omm an d: STAT u s : O PER at i o n : EVEN t ? Thi s c om m and returns th e deci m al w ei g hted s u m of the s et bi t(s ). Cl e ar i n g t he O per ati on E v ent Regi s ter B i ts B i ts i n the O per ati on S tatus Gr oup E v ent ...

  • Agilent Technologies E1300B - page 124

    SUB I n t r _r e s p B= SPOLL( 70900) OUTP UT 70900; "ST AT :O PER : EV EN ?" ENTE R 70900; E OUTP UT 70900; "DIAG:I N T R : R ESP?" ENTE R 70900; R . . . SUB EN D 1 Clearin g St at us Th e * CLS com m and c l ears al l statu s r eg i s ters (S tandar d E v ent S tatus R eg i s ter , S tandar d O p erati o n S tatus E v ent R eg ...

  • Agilent Technologies E1300B - page 125

    Ex ample: Int er r upt ing w hen an Er r or Occu r s Thi s prog r am s how s h o w to i nterr upt an ex ternal com puter w henev er an err or o c cu r s fo r t h e in st r u me n t b e in g p r o gr a m m e d whic h , in t h is e x am p le , is a mu l ti meter at s ec on dary add r es s 0 3 . 10 O PTIO N B ASE 1 ! A rra y n u m b e ri n g s t a rt ...

  • Agilent Technologies E1300B - page 126

    1 Synchr oni z i ng a n E x t er nal Com puter a n d I n stru me nts Th e *O P C ? a nd *O P C comma nd s ( op er a tion comp lete co mma nd s) a ll ow y ou to mai ntai n s y nc hroni z at i on be tw een a n ex ter na l c om pute r a nd an i ns tr ument. The * OP C? query pl ac es an AS CII c harac ter 1 i nto the i ns tr um en t’ s ou tput queue ...

  • Agilent Technologies E1300B - page 127

    Ex a m ple : Synchronizing an Ex ternal C omputer and Two Instr uments using the * OPC c om m a nd. Thi s ex ampl e u s es th e * O PC c om mand and s eri al pol l to s y nc hroni z e an ex ternal com puter and tw o i n s tr um ents ( D AC at s ec on dary add r es s 0 9 ; S can ni ng Vol tm eter at s ec o n dary ad dr es s 0 3 ). Th e adv antag e t ...

  • Agilent Technologies E1300B - page 128

    6-1 4 C ontr olling Instruments U s ing GPIB ...

  • Agilent Technologies E1300B - page 129

    Ch a p t e r 7 Sy s tem Ins tr um ent C o m ma nd R ef er enc e Ab out Thi s C h a pte r Th i s c hapt e r de s cri bes the St a nd a rd Co mma nd s fo r Pro gra mma bl e Ins t rum ents ( S CPI) com m an d s et an d the IEEE 488.2 Co mmo n Co mma nd s fo r the S y s tem Ins trument. The S y s tem In s tr um en t i s part of th e Ag i l ent E 1300/ ...

  • Agilent Technologies E1300B - page 130

    C om m a nd Sepa rator A col on ( : ) al w ay s s ep ar a tes one c om m and f rom the nex t l ow er l ev el c om m and as s how n bel ow : RO U T e : SCAN : MODE ? Col ons sepa r ate t he r oot comm and f rom the sec ond l ev el c om m and ( R O U T e :SC A N ) a n d t h e se c o n d le v e l fr o m t h e t h ir d le v e l ( SC A N :M O D E ? ) . ...

  • Agilent Technologies E1300B - page 131

    ( e . g., 123, 123E 2, - 123, - 1.23E 2, . 123, 1.23E - 2, 1. 23000E - 01) . Sp e c ia l cas es i ncl ud e M IN, MAX, and INF i ni ty . The Comm ents s e cti on w i thi n t h e C o m ma n d R e fe r e n c e will st a t e w h e t h e r a n u m e r ic p a r a m e t e r c a n a lso b e s p e c i fie d in h e x , o c t a l, a n d / o r b in a r y . # H ...

  • Agilent Technologies E1300B - page 132

    1 SC PI C om m and Re fe re nc e T h is se ct ion d e sc r ib e s t h e SCP I co m m a n d s fo r t h e Sy st e m I n st r u me n t . Com m ands ar e l i s ted al p habeti cal l y by s u b s y s tem an d al so w i thi n eac h su b s y st e m . A c o m m a n d gu id e is p r in t e d in t h e t o p m a r gin o f e a c h p a ge . T h e g ui de i nd i ...

  • Agilent Technologies E1300B - page 133

    DI A G nost ic The D IA G nosti c s ubs y s tem al l ow s c ontrol ov er t he S y s tem Ins tr ument’ s i nt er nal pr oc es s o r s y s tem (: B O O T, an d : INTerr upt), th e al l ocati on and co n tents of U s er RAM, and, di s c v o l um e RAM (: NRAM, an d : R D IS k ), a n d al l oc a t i on o f the bu i l t- i n s eri al i nt e rf ac e (: ...

  • Agilent Technologies E1300B - page 134

    : B OOT: C OL D DIA Gnos t i c: BOOT : COL D caus es the S y stem Ins trument to r e s tar t ( r e-boot). Conf ig ur ati ons store d i n non-v ol ati le m emory and R S - 232 co n figu r a t io n s a r e reset to th ei r d ef aul t s tates : • D R AM , NR AM , and R D IS k m emory s eg m ents are cl ear ed • S eri al I n t e rf ac e para m ete ...

  • Agilent Technologies E1300B - page 135

    : B OOT[ :WA RM ] D IA Gnos ti c: BOOT [ : W A RM] caus es the S y stem Ins tr ument to r e s tart (r e-boot) us i ng the curr ent conf i g uration s t ore d in non- v ol a ti l e mem ory . The ef fec t i s the same a s cy cl i ng po w e r . Comments • T h e Sy st e m I n st r u me n t go e s t h r o u gh it s p o w e r - u p se lf t e st s. • ...

  • Agilent Technologies E1300B - page 136

    : C OM M unica te :SER ial[ 0 ] [ :OW N e r] ? DI A Gnos t i c: COMMuni cat e: S ERia l [ 0] [ : OW Ner] ? R etur ns the current " ow ner " o f t h e b u ilt - in se r ia l in t e r fa c e . T h e v alu e s r e t u r n e d will be ; " SY ST " , " I B A S" , o r "N O N E ". Comments • R elated Com m ands : D ...

  • Agilent Technologies E1300B - page 137

    : DO W Nl oad: CHECk ed [ : M A DDress] DIA Gnos ti c: CH E C ked: DOW Nlo ad[ : MA DDres s ] < addre s s >, < data > wr i t e s dat a i nto a non- v ol a ti l e U ser R A M seg ment start i ng at addr e ss u s i ng erro r co rrecti on. The U s er R AM seg m ent i s al l o c ated by the DIAG : N RA M: CR E ate or D IAG : DR A M: CR E at ...

  • Agilent Technologies E1300B - page 138

    By te Format E a ch b y t e se n t w ith t h is c o m m a n d is e x p e ct e d t o b e in t h e fo llo w ing fo r m a t : B i t # 7 654 3210 Co n tr o l B it C h eck B its D ata B its • Cont r ol Bi t i s u s ed to i ndi c a te the s er i al dri v er i nf or m ati on s uch as cl ear, r e s e t , o r e n d o f t r a n sm iss io n . T h i s b it i ...

  • Agilent Technologies E1300B - page 139

    : DO W Nl oad: CHECk ed : SA DDre ss D I AG n o s t i c :C H E C k e d :D O WN l o a d :S AD D r e s s < addre ss >, < data > wr i t e s dat a to non-v ol a ti l e U s er R AM at a s i ng l e ad d r ess speci f i ed by addr e ss u s i ng erro r co rrecti on. It can al s o w ri te to dev i ces w i th reg i s ters i n the A1 6 ad dr es s ...

  • Agilent Technologies E1300B - page 140

    By te Format E a ch b y t e se n t w ith t h is c o m m a n d is e x p e ct e d t o b e in t h e fo llo w ing fo r m a t : B i t # 7 654 3210 Co n tr o l B it C h eck B its D ata B its • Cont r ol Bi t i s u s ed to i ndi c a te the s er i al dri v er i nf or m ati on s uch as cl ear, r e s e t , o r e n d o f t r a n sm iss io n . T h i s b it i ...

  • Agilent Technologies E1300B - page 141

    :D O W N lo a d [ : M A DDress] DIA Gnos ti c: DOW Nlo ad[ : MA DDres s ] < addre ss >, < data > wr i t e s dat a int o a non-v olati l e U s er R AM seg ment starting at addr e ss . The U s er R AM s eg m ent i s al l oc ated by the DIAG : N RA M: CRE ate co m m an d. Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e ...

  • Agilent Technologies E1300B - page 142

    Ex am ple L o ad in g Dy na m ic C on f igu r ation in f orm a tio n in to an allo c at e d R A M seg m e n t . DIAG: NRA M:CR E 6 A llo ca te a seg m en t o f u s er RAM D I A G :BO O T :W A R M R e-b o o t system to co m plete allo catio n DIAG: NRA M: ADDR ? qu er y s t ar ti ng ad dr ess ent er val u e t o var i ab l e X g et s tar tin g a d d ...

  • Agilent Technologies E1300B - page 143

    Ex am ple Dow nloading D ata to a Sing le A ddr e s s L oc ation T h is p r o gr a m d o w n lo a d s a n a r r a y w it h t h e d a t a 1, 2, 3, 4, 5 t o r e gis t e r 32 o n a d e v ice wit h lo gic a l a d d r e s s 40 in V X I b u s A 16 a d d r e s s s p a c e . DIM Dn l d _d at a( 1: 5) Di m ens i on c ontr ol l er ar r ay DAT A 1,2, 3,4, 5 R ...

  • Agilent Technologies E1300B - page 144

    : DRA M :CREate D I A Gn o s t i c:DRA M: CREat e < si z e >< n um _drive rs > crea tes a non- v olati l e R AM ar ea f or l oad i ng in s trument dri v er s . D I A Gn o s t i c:DRA M : CREat e 0 re mov es the R A M seg ment w hen the s y stem i s re-booted . Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default ...

  • Agilent Technologies E1300B - page 145

    : DRIVer: L OA D < d riv e r_bloc k > D I AG n o s t i c :D R I Ve r :L O AD < drive r_bl o c k > lo a d s t h e in st r u m e n t d r iv e r co n tai ned i n th e dri v er _ b l ock i n to a prev i ou s l y created D R AM seg m ent. Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units d r iver _b lo ck ar ...

  • Agilent Technologies E1300B - page 146

    :D R I V e r :LI S T [ :t y p e ] ? DIA Gnos ti c: DRIV er : L IST [ : ty pe] ? l i sts al l dr i v ers f r om the s peci f i ed ta bl e f ound on th e s y s tem . If no par ameter i s s p ec i f i ed, al l d r i v er tabl es are s earch ed and the da ta f rom each dr i v er t abl e i s s e pa ra ted f rom the other s by a s emi col on. Pa ram e te ...

  • Agilent Technologies E1300B - page 147

    :I N T e r r u p t :AC T i va t e DIA G nos ti c: INT er rupt: A CT i v ate < mode > enabl es an i nterrup t on th e V XI b a ck p la n e in t e r r u p t lin e sp e c ifie d b y D I A G :I N T :SE T [n ] t o b e a c kn o w le d ge d . Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units m ode bool ean 0 | 1 | O ...

  • Agilent Technologies E1300B - page 148

    • O N o r 1 sp e cify t h a t in t e r r u p t h a n d lin g is t o b e s e t u p fo r t h e sp e c ifie d in t e r r u p t lin e . O F F o r 0 in d ic a t e t h a t in t e r r u p t h a n d lin g o f t h e sp e c ifie d l i ne i s to b e done b y the oper ati ng sy stem . • Re lated Comm ands : D I A G :I N T :A C T , D I A G :I N T :P R I o r ...

  • Agilent Technologies E1300B - page 149

    :I N T e r r u p t :P R I o r it y [ n ] DIA Gnos ti c: INT err upt: PR Iority [ n] [ < le ve l >] giv e s a p r io r it y le v e l t o t h e V X I in t e r r u p t lin e sp e c ifie d b y [n ] in … P R I o r it y [n ]. Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units level numeric 1 t hroug h 7 | MIN| M A ...

  • Agilent Technologies E1300B - page 150

    : I N T e rrup t: RESPonse ? DIA Gnos ti c: INT err upt: RE SPons e? R eturns the i nterrup t ack now l edg e res po n s e (S TAT U S /I D w ord) f rom th e h i g he s t pri ori ty V XI i n t e rru pt l i ne. Comments • The v alu e returned i s the r es pons e f r om th e i nterr upt ack now l edg e cy cl e (S TATUS / I D w o rd) o f a dev i c e ...

  • Agilent Technologies E1300B - page 151

    : NRA M :A DDRe ss ? DIA Gnos ti c: NRA M: A DDRess ? R et ur ns the s tar ti ng ad dr e s s of the non- v ol at i l e U s e r R A M s e gm e n t a llo c a t e d u sin g D I A G :N R A M :C R E a t e . Comments • D IAG : NR AM : CR E Ate d oes not al l oc ate the RA M s eg ment un ti l af ter a s ubs e que nt r e-boot. To g et accurate r e s ul t ...

  • Agilent Technologies E1300B - page 152

    :N R AM :C R E a t e ? D I A Gn o s t i c:NRA M : CREat e ? [ MIN | MA X] R e tur ns the curr ent or all ow abl e (M IN | M AX ) s i z e of the U ser non-v ol a ti l e R AM s eg ment. Comments • D IAG : NR AM : CR E d oes not al l ocate dr i v er R A M until a subs e q uent re -boot. T o g e t accurat e r es ul ts, ex ecute D IA G : N R AM : CR E ...

  • Agilent Technologies E1300B - page 153

    :P O K E DIA Gnos t ic: POK E < addre s s >, < w idth >, < data > w r it e s da t a (number of bi ts g i v en by wid t h ) s tarti ng at addr e ss . Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units addr es s n u m e r ic 0 t o 16,777, 215 ( # H FFFFFF) none wid th numeric 8 | 1 6 | 3 2 none dat a ...

  • Agilent Technologies E1300B - page 154

    : RDISk : CREa te DIA Gnos ti c: RDISk: CR E ate < si z e > All oca te s me mor y for a R A M d isc v olume . The R AM d i s c v ol ume i s d ef i ned f or us e onl y by the IBAS IC op tion. Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units size num er i c 0 to av ai l abl e R AM or MIN | MAX none Comments ? ...

  • Agilent Technologies E1300B - page 155

    : UPL oad[ : M A DDres s ] ? D I A G n o sti c : U P L o a d [ : M A D D r e ss] ? < addre ss >, < byt e_count > R eturns the number of by tes s p eci f i ed by by t e _ c ount , s tar ti ng at addr e ss . Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units addr es s n u m e r ic 0 t o 16,777, 215 ( # H F ...

  • Agilent Technologies E1300B - page 156

    : UPl oad: SA DDr es s ? D I A G n ost i c : U P L o a d : SA D D r e ss? < addre ss >, < byt e_count > R etur ns the number of by tes s p eci f i ed by by t e _ c ount , a t addr e ss . Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units addr es s n u m e r ic 0 t o 16,777, 215 ( # H FFFFFE ) none by t e ...

  • Agilent Technologies E1300B - page 157

    INI T i a t e T he INITi ate com m and subs y stem control s the i ni tiati on of the t r i g g e r s y stem fo r o n e o r m o r e t r igge r c y c le s. I N I T ia t e e n a b le s w hile A BO R t d isa b le s t h e tr i g g er sy s tem. T he TR IG g er comm and subs y stem c ontrol s t he beha v i or of the t r igge r sy s t e m w hile it is e n ...

  • Agilent Technologies E1300B - page 158

    [ SOU Rce] Th e S y st em In s tr um en t c o ntai ns a P ac er w hi ch prod uc es TTL l ev el pul s es. The S O U R CE comm and subs y s tem c ontrols the number a nd p er iod of t hese pul s e s . The outp ut of t he Pa cer i s a v ai l able at the r e ar -panel B NC connector label ed “Pac er O ut” . Sub syste m Syn t a x [SO U R c e ] :P U ...

  • Agilent Technologies E1300B - page 159

    : PUL Se:PERi od S OURce: PU L Se: PER iod < peri o d > sets the per i od of the pul se(s ) to be g e ner ate d by t he P acer. Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units pwe i od n u m e r ic 500E - 9 t o 8. 388607 o r MIN | MAX sec ond Comments • The r esol ution of per iod is 500E - 9 s e c o n d ...

  • Agilent Technologies E1300B - page 160

    ST A T u s The S TATus s ubs y stem c om mands ac ces s the co ndi ti o n , ev ent, and enabl e reg i s ters in th e Op erati o n S tatus g roup and the Q ues ti onabl e D ata g r oup. Sub syste m Syn t a x ST A T u s :O P E R a t io n : CON Di ti o n ? :E N A B le < even t> :E N A B le ? [ : EV EN t ] ? :P R E Se t : Q U E S tionabl e : CON ...

  • Agilent Technologies E1300B - page 161

    • Bi t 8 i s the onl y bi t us ed i n the condi ti on r eg i ster ( by t he S y stem Ins trument) , there f or e , i t is t he onl y bit w hi c h needs to be unmas k e d i n the ev ent reg i s ter . S p eci f y i n g the " b i t w ei g ht" f or the even t unmask s t he b i t . T h e b it we igh t is 256 a n d c a n b e s p e c ifi e d i ...

  • Agilent Technologies E1300B - page 162

    Ex am ple Readi ng t he Ev ent Regis ter STAT : O PER: EVEN ? qu er y if bit( s ) is set ent er st at em en t : PRESe t ST A T us: P R E Se t s ets eac h b i t i n the enabl e r eg i s ter (stan dard operati on s tatus g r o u p ) t o ’0 ’. Ex am ple Pres ett i ng the Ena ble Regis ter STAT : PRES pr eset en ab le r eg is ter : QUEST ionab le T ...

  • Agilent Technologies E1300B - page 163

    SY S T em The S YS TE M com m and subs y stem f or the S y s t em Ins trument p r ov i des f or : • C o n figu r a t io n o f t h e R S- 232 in t e r fa c e • Con trol and acces s o f the S y st em In s tr um en t’ s real ti m e c l ock /c al endar ( SY ST : T I M E , SY ST : T I M E ? , SY ST :D A T E , SY ST : D A T E ? ) . • Ac ce s s t ...

  • Agilent Technologies E1300B - page 164

    : C OM M unica te : G PI B: A DDRe ss SY S T em : COMMunica te: G PI B: A DDRes s < addre s s > s ets th e pri m ary add res s of t he Instrument’ s G PIB por t. Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units addr es s numer i c mus t round to 0 to 3 0 none Comments • The v al ue o f < addr e ss > ...

  • Agilent Technologies E1300B - page 165

    • Se r ia l c o m m u n i c a t io n s e t t i n gs fo r t h e A gile n t E 1324A D a t a c o m m i nter f a ce c an be s tored in i ts on- board non-v ol a ti l e E E R O M onl y af ter the DIAG : CO MM : S E R [ n ] : S TO R e co m m an d i s ex ecuted. Thes e s etti n g s are us ed at pow er - u p and D IAG : B O O T[ : W A R M] . • D I A G ...

  • Agilent Technologies E1300B - page 166

    : C OM M unica te :SER ial[ n ] :C O N T r o l :D T R ? S Y S T e m :C O M M u n i c a t e :S E R i a l [ n ] :CO N T r o l :D T R ? ret u rns th e c urre n t s ett i ng fo r D T R lin e c o n t r o l. Ex am ple Checking th e s et ti ng of DTR contro l . SYST:CO M M :SER0:C ON T:DT R? ent er st at em en t statem en t en ter s th e str in g "O ...

  • Agilent Technologies E1300B - page 167

    : C OM M unica te :SER ial[ n ] :C O N T r o l :R T S ? S Y S T e m :C O M M u n i c a t e :S E R i a l [ n ] :CO N T r o l :R T S ? r eturns the c ur rent s etti ng fo r R T S lin e c o n t r o l. Ex am ple Checking th e s et ti ng of RT S con tro l. SYST:CO M M :SER0:C ON T:RTS? ent er st at em en t statem en t en ter s th e str in g "O N &q ...

  • Agilent Technologies E1300B - page 168

    : C OM M unica te :SER ial[ n ] [ :RE C e ive ] :BI T S S Y S Tem : COMMunica te: SERial [ n] [ : R EC eive] : BITS < bits > S ets the number of bi ts to b e us ed to tr ans mi t and recei v e data. Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units bits numer i c 7 | 8 | MIN| M AX none Comments • Attempti n ...

  • Agilent Technologies E1300B - page 169

    : C OM M unica te :SER ial[ n ] [ :RE C e ive ] : PA CE [ : PR OTocol ] S Y S Tem : COMMuni cat e: S ERia l [ n] [ : REC eive] : PA CE [ : PROT ocol ] < pr ot ocol > ena bl es or dis a bl es recei v e pa ci ng (XO N/XO FF ) pr otocol . Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units pr ot oc ol di s crete X ...

  • Agilent Technologies E1300B - page 170

    : C OM M unica te :SER ial[ n ] [ :RE C e ive ] :P AC E :T H R e s h o l d :S T AR t S Y S Tem : COMMuni cat e: S ERia l [ n] [ : REC eive] : PA CE : THR es ho l d: S TA Rt < c h ar_co u n t > co n figu r e s t h e in p u t b u ffe r le v e l a t w hic h t h e sp e c ifie d in t e r fa c e may s end the XO N c har acter (AS CII 1 1 16 ), ass ...

  • Agilent Technologies E1300B - page 171

    : C OM M unica te :SER ial[ n ] [ :RE C e ive ] :P AC E :T H R e s h o l d :S T O P S Y S Tem : COMMuni cat e: S ERia l [ n] [ : RECeiv e] : PA CE : THR es ho l d: S T OP < c h ar_co u n t > co n figu r e s t h e in p u t b u ffe r le v e l a t w hic h t h e sp e c ifie d in t e r fa c e may s end the XO FF charac ter (AS CI I 1 3 16 ) , d e- ...

  • Agilent Technologies E1300B - page 172

    : C OM M unica te :SER ial[ n ] [ :RE C e ive ] : PA Ri t y : CHECk S Y S Tem : COMMunica te: SERial [ n] [ : R EC eive] : PA R ity : CHEC k < check_ cn t r l > co n t r o ls whe t h e r o r n o t t h e p a r it y bit in r e c e iv e d se r ia l d a t a fr a me s will be co n sid e r e d sign ific a n t . Pa ram e te rs Pa rameter Name Parame ...

  • Agilent Technologies E1300B - page 173

    • The f ol l ow i ng ta bl e defi nes each v al ue of type : V alu e D ef inition EV EN I f … P A R it y :C H E C K is O N , t h e r e c e iv e d p a r it y b it mu s t mai ntai n ev en par i ty . The trans mi tted par i ty b it w ill ma in t a in e v e n p a r it y . ODD I f … P A R it y :C H E C K is O N , t h e r e c e iv e d p a r it y bi ...

  • Agilent Technologies E1300B - page 174

    : C OM M unica te :SER ial[ n ] [ :RE C e ive ] : PA R ity [ : TY PE] ? S Y S Tem : COMMuni cat e: S ERia l [ n] [ : REC eive] : PA Rity [ : T Y PE] ? retur ns the ty pe o f pari ty c h e c k e d a n d g en e ra t e d. Ex am ple W hat ty pe of pari ty checking is s et? SYST:CO M M :SER0:PAR? as k f or par it y ty pe ent er st at em en t re t u rn s ...

  • Agilent Technologies E1300B - page 175

    : C OM M unica te :SER ial[ n ] [ :RE C e ive ] :S BI T s ? S Y S Tem : COMMunica te: SERia l[ n ] [ : REC eiv e] : S BITs ? [ MIN | MA X] ret u rn s : • The cur re nt s top bit s e tting i f no p ar ameter i s sent. • T h e m a x imu m a llo w a b le se t t in g if M A X is s e n t . • The mi ni mum al low abl e s etti ng i f M IN i s s e nt ...

  • Agilent Technologies E1300B - page 176

    : C OM M unica te :SER ial[ n ] :T RAN sm it : PA CE [ : PR OTocol ] S Y S Tem : COMMuni cat e: S ERia l [ n] : TR A Ns mi t: PA CE[ : PROT o col ] < pr ot ocol > ena bl es or dis a bl es the tr ansm i t p aci ng ( XO N/XO FF ) pr ot ocol . Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units pr ot oc ol di s cr ...

  • Agilent Technologies E1300B - page 177

    • Re lated Comm ands : SY ST : T I M E , SY ST : T I M E ? , SY ST : D A T E ? • *R ST C onditi on: *R S T does n ot c hang e the s etti ng of the cal endar . Ex am ple S etting th e sy s tem Date SY ST:DA TE 1991,09,08 set S E P 8, 1991 :D AT E ? S YS T e m : D AT E ? [ M I N | M AX , M I N | M AX , M I N | M AX ] retu rn s : • W hen no p ar ...

  • Agilent Technologies E1300B - page 178

    :T I M E SY ST e m : T I M E < hour >, < m i nut e >, < sec ond > s e t s t h e E 1300/ E 1301 mai nf rame’ s i nternal cl ock . Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units hour numeric m us t r ound t o 0 t o 2 3 none m i nut e numer i c mus t r ound to 0 to 5 9 none s e cond num eri c mu ...

  • Agilent Technologies E1300B - page 179

    TRI G g e r The TR IG g er comm and subs y s tem c ontrols the behav ior of the tr i g g er sy s tem o n c e it is in it ia t e d ( se e I N I T ia t e c o m m a n d su b sy st e m ) . T h e t r igge r c o m m a n d s ubs y stem c ontrols : • The del ay betw een tri g g er and f i rs t Pac er pul s e ( TR IG : D E Lay ) • An i m m edi at e s of ...

  • Agilent Technologies E1300B - page 180

    [: I MMe d i a t e ] T R IGg er: IMMedi a te w ill ca u se a t r igge r c y cle t o o c cu r im m e d ia t e ly , p r o v id e d that the tri g g er s y stem has been i ni ti ated (INITi ate). Comments • R elated Com m ands : A B O R t, INITi at e • *R ST C onditi on: T hi s comm and is an e v ent a nd ha s no * R S T cond i ti on. Ex am ple Tr ...

  • Agilent Technologies E1300B - page 181

    Comments • The f ol l o w i ng tabl e ex p l ai ns the pos si bl e cho i ces . Para meter V a l ue S o urce o f T rigger BU S G r oup E x ecute Tr i g g er ( G E T ) bus c om m and, *TR G comm on c om m and, or TR IG g er comma n d . E X Ternal “E v ent In” s i g nal at rear p anel BN C connector, or TR IG g er comm and. H O L D O n ly t h e ...

  • Agilent Technologies E1300B - page 182

    VXI The VXI command subs y stem pr ov i des f or: • Dete rm i ni n g th e n um b e r, ty pe , and l og i c al addre s s of th e de v i c e s ( in s t r u m e n t s ) i n s t a lle d in t h e E 1300/ E 1301 m a in fr a m e . • D ir e c t a cc e ss t o V X I b u s A 16 r e giste r s w ith in d e v ic e s in st a lle d in t h e Mai nf r ame. Sub s ...

  • Agilent Technologies E1300B - page 183

    :C O N F i g u r e :D LI S t ? VX I : C O N F : D L I S ? [ < l o g i c al _addr >] r etur ns i nf ormati on about the d ev i c e sp e c ifie d b y lo g ica l_ad d r . R e sp o n se d a t a is in t h e for m: n1, n2, n3 , n4 , n5 , n6 , c1 , c2 , c3 , c4 , c5 , s 1, s 2, s 3, s 4 Wh ere the f i el ds abo v e ar e def i ned as : n field s I n ...

  • Agilent Technologies E1300B - page 184

    Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units lo g ical_ad d r n u m e r ic 0- 255 ( o r n o t h i n g) none Comments • Whe n lo g ica l_ad d r is n o t s p e c ifie d , V X I :C O N F :D L I S? r e t u r n s i nf or mati on f or each of the de v i c es i ns tal l e d, s e pa r ated by semi col ons. If the Co ...

  • Agilent Technologies E1300B - page 185

    :C O N F i g u r e :H I E R a r c h y ? V XI: CONF: HIER ? R etur ns c urr ent hierar chy conf i g ur ati on i nf ormati on about the s el ec ted l og i cal add ress . Th e i ndi v idual f i el ds o f the r es pons e ar e co m m a s epar at ed . If the inf ormation about the sel e cted log i c al add r ess is not a v ai l able f r om th e des ti na ...

  • Agilent Technologies E1300B - page 186

    • Car d s w hi c h are p ar t of a c om bi ned i ns t rument suc h as a s w i t chbox or s c a n n i ng v ol tm e t e r a l w ay s re t u rn th e s am e m an u f ac t u re r’ s c om m e n t s as th e f i rs t c ard i n t h e i ns tru m en t . I n f orm ati on i n th e ot h er f i el ds co r r e sp o n d t o t h e c a r d fo r w hic h t h e L o ...

  • Agilent Technologies E1300B - page 187

    reg i s ters ) w hi c h ar e pr es en t o n the dev i ce. -1 in di cates that the dev i c e has no A 1 6 me mo r y . • A 24 memo ry offs et : a n in t e ge r b e t w e e n - 1 a n d 16777215 i n c l u s iv e . Indi cates th e bas e add r es s f or any A2 4 reg i s ters w hi ch are pr es en t o n the dev i c e. - 1 i ndi c ates that the dev i c e ...

  • Agilent Technologies E1300B - page 188

    :C O N F i g u r e : I NFor mat i on : A L L ? V X I: C ONF: INF: A L L ? R et urns the s tati c i nf ormati on about al l l og i c al add ress es . The i nf or mati on i s returned i n th e o rder s peci f i ed i n the respons e to VXI: CONF : LAD D ?. The i nf or mati on about mul ti pl e l og i cal a dddre s s es w ill b e s e m i c o l o n s e ...

  • Agilent Technologies E1300B - page 189

    • Re lated Comm ands : VX I: W R ITE , V XI: R E G : R E AD ? Ex am ple Read from o ne of a d ev ice’ s co nfigura ti on regis t ers VXI :READ? 8, 0 r ea d I D r eg ister o n d evice a t L og i ca l A ddr e s s 8 ent er st at em en t en ter va lu e f r o m d evice r eg ister :REGiste r :REA D? V XI : REG: RE A D ? < r eg i s ter > returns ...

  • Agilent Technologies E1300B - page 190

    :REGiste r :WRI T e V XI: R EG : W R IT e ? < r eg ister > ,< d ata > w ri tes to the s p eci f i ed 1 6 b i t reg i s ter at the s el ec ted l og i cal add ress ( s ee V XI: S E Lect). Th e data i s a 1 6 bi t v al u e sp e c ifie d a s a n u m e r i c v a l u e in t h e r a n ge o f - 32768 t o 32767 o r 0 t o 65535. T h e reg i s ter ...

  • Agilent Technologies E1300B - page 191

    : SEL e ct V XI: SEL ect < lo g ical _addr > sp e c ifie s t h e lo gic a l a d d r e ss w hic h is t o b e u se d b y many s ubs eq uent com m ands i n the VX I s ubs y stem. Pa ram e te rs Pa rameter Name Paramet er Type Ra n ge o f Va l u e s Default Units lo g ical_ad d r numer i c 0 throug h 2 5 5 none Comments • The * R S T def au l t ...

  • Agilent Technologies E1300B - page 192

    :W R I T e V XI: W RIT e < lo g ic al_addr >, < re g i s te r_addr >, < data > al l ow s ac c es s to th e enti r e 6 4 b y te A 1 6 reg i s ter ad d res s s pace f or the dev i ce s peci f i ed by lo g ical_ad d r . S i n c e the VXIbus s y stem i s b y te-addr es s ed, w hi l e the r eg i s ters are 1 6 b i ts w i d e, r eg i s ...

  • Agilent Technologies E1300B - page 193

    1 C o mmon C o mma n d Re fe re nc e T h i s s e c t i o n de s c ri b e s t h e I EEE- 488.2 C o m m o n C o m m a n d s t h a t c a n b e u se d t o p r o g r am instr ume n ts in th e ma i nf ra me . Comma n d s ar e li st ed by co mman d g r oups i n the s umm ary tabl e bel o w , and al p habeti cal l y in th e res t of thi s s ec ti on . E x ...

  • Agilent Technologies E1300B - page 194

    *C L S Cl ear S tatus Comm and . Th e * CLS com m an d c l ears al l statu s r eg i s ters (S tand ard E v ent S tatus R eg i s ter , S tand ard O per ati on E v ent S tatus Reg i s ter, Q ues ti on ab l e D ata E v ent R eg i s ter ) and the er ror queue f or an i ns trument. Thi s cle a r s t h e co r r e sp o n d in g su mm a r y bit s ( b it s ...

  • Agilent Technologies E1300B - page 195

    * ESE? S tandar d E v ent S tatus E nabl e Q uer y . R e turns the w ei g hted s um of al l en abl ed (unmas k ed) bi ts i n the S tandar d E v ent S tatus R eg i s ter . Ex am ple 10 O UT P UT 70900; "* ESE? " S end s st at us enable quer y 20 ENTE R 70900; A Places r espo n s e in var iable 30 PR IN T A Pr in ts r espo n s e 40 END *E S ...

  • Agilent Technologies E1300B - page 196

    *I D N ? Identi ty . Retur ns the d ev i ce i d enti ty . Th e res pons e con s i s ts o f the f ol l o w i ng f o ur f i el ds (f i el ds are s eparated by c o m m as ) : • Manufac tur er • Model N u m ber • S eri al Num b er (r eturns 0 i f not av ai l abl e) • Fi rm w are Rev i s i on (re t u rn s 0 i f no t av ai l ab l e) T h e * I D N ...

  • Agilent Technologies E1300B - page 197

    NOTE The S y s tem Ins trum en t n o l ong er i m pl ements the *LRN? co m m and. A t t e m p t in g t o h a v e t h e Sy st e m I n s t r u m e n t e x e cu t e t h is c o m ma n d will g e n e r a t e a n e r r o r - 113 “ U n d e fin e d h e a d e r ” . *O P C O p er at i on Com pl ete. Caus e s an i nstrument t o s et bi t 0 ( O p er at i o ...

  • Agilent Technologies E1300B - page 198

    *R C L < sta t e num b er > R ecal l s tored s tate. R ecal l s a s to red s tate f rom m emory and con f ig ur es th e i ns trum en t to th at s tate. S tates are s tored us i ng the * S A V c om mand. Ex am ple O UT P UT 70900; "* R CL 4" R eca lls in str um en t s tate num be r 4 *R MC < name_ s tr ing > R emo v e Indiv i d ...

  • Agilent Technologies E1300B - page 199

    *S R E ? S tatus R e g i s ter E nabl e Q uer y . R eturns th e w ei g h ted s um of al l enab l ed (unmas k ed) ev ents (thos e enabl ed to as s er t S R Q ) i n the S tatus B y te R eg i s ter. Ex am ple 10 O UT P UT 70900; "* SR E ?" S end s S tatus R eg i s t e r E n able quer y 20 ENTE R 70900; A Places r espo n s e in var iable 30 P ...

  • Agilent Technologies E1300B - page 200

    1 GP I B M e ssa ge Re fe re nc e T h i s s e c t i o n de s c ri b e s I EEE- 488.1 d e fin e d m e s sa ge s a n d t h e i r a ffe c t o n i ns trum en ts i ns tal l ed i n th e m ai nf r am e. The ex am pl es s how n are s pec i f i c al l y f o r H P 9000 Se r ie s 200/ 300 c o m p u t e r s u s in g B A SI C la n gu a ge . A n y I EEE- 4 8 8 c ...

  • Agilent Technologies E1300B - page 201

    Dev i ce Cl ear ( DCL ) or S el ec t ed D ev i ce Clea r (S D C ) D CL c l ears al l i n s tr um en ts i n th e m ai nf r am e. S D C cl ears a s peci f i c i ns trum en t. The p ur p os e of D CL or S DC is to p r ep ar e one or mor e i ns t ruments to re cei v e and ex ecu te c om mands (us ual l y * RS T) . DCL or S D C do the f ol l ow i n g to ...

  • Agilent Technologies E1300B - page 202

    Re m ote S ets th e GPIB rem o te enab l e l in e (R E N) tr ue w hi c h pl aces an i n s trument i n the remote s tate. Comments • T h e R E M O T E 709ss ( ss = se c o n d a r y a d d r e ss) co m m a n d p la c e s t h e i ns trum en t i n the remo te s tate. Th e RE MO TE 7 c om mand, does not, by i tsel f , pl ace the i ns trum ent i n the r ...

  • Agilent Technologies E1300B - page 203

    1 C o mma n d Q u ick Re fe re nc e T h e f o l l o w i n g t a b l e s s u m m a r i z e S C P I a n d I EEE 488.2 C o m m o n ( * ) c o m m a n d s fo r t h e A gile n t E 1300/ E 1031 M a i n fr a m e Sy s t e m I n s t r u m e n t . S CPI C ommands Qui ck R eference Com mand Des cripti on AB ORt [ IMMedi ate] Abor t Pac er output. D IAG nosti c ...

  • Agilent Technologies E1300B - page 204

    S CPI C ommands Qui ck R eference Com mand Des cripti on :N R A M : AD D Ress ? Returns s tarting address of the U ser non- v ola ti le RAM. : CRE ate < s i ze > | MI N| MAX Cr eates a User non- v o l ati le RAM seg ment. : CRE ate? [ MI N | MA X] Retur ns the c urr ent or al l ow abl e s iz e of U s er NVRAM. : P EEK ? < addr e ss> | M ...

  • Agilent Technologies E1300B - page 205

    S CPI C ommands Qui ck R eference Com mand Des cripti on SY ST e m : B EEPe r [ : IMMedi ate] S ound beeper (fi x ed dur ati on and tone). : C O MMuni c ate :G P I B : AD D Ress S ets the pri mary address of the c ommuni c ati ons port. : AD D Ress ? Retur ns GPIB address or mi n| max all ow ed v al ue. :SE R ia l[n ] :C O N T r o l : D TR O N| O F ...

  • Agilent Technologies E1300B - page 206

    S CPI C ommands Qui ck R eference Com mand Des cripti on TRIGg er : D E Lay < numeri c v al ue> S ets del ay bet w een t rig g er and f i r s t Pac er puls e. : D E LA y ? [ MIN| MA X] R eturns c ur rent tr i g g er del ay or MIN| MA X al l ow abl e v al ue. [ : I MMedi ate] S et s tri g g er s ou rc e f or ti me r/ pac er. :SL O P e [N E G A ...

  • Agilent Technologies E1300B - page 207

    IEEE 488. 2 Comma n Co mma nds Qu ick Reference Ca t eg o ry Co mm a n d Tit l e General *I D N? Identi f ic ati on Quer y *RS T Reset Command *T ST? Self Te st Q u er y Ins t rume nt Sta tus *C LS Cl ear S tatus C ommand *E SE < mask> S tandard E v ent S tatus Enabl e Reg is t er Command *ES E? S tandard E v ent S tatus E nabl e Q uery *ES R ...

  • Agilent Technologies E1300B - page 208

    7– 80 C ommand Qu ick R efer ence ...

  • Agilent Technologies E1300B - page 209

    A ppe ndi x A S p ec if ic atio ns Ma i n f r a m e Sp ecificati ons Pace r ( 5 0% duty cycle) : P r o gr a m m a b le in t e r v a ls : 500 n se c t o 8.389 s e c w it h 500 n se c r e so lu t io n . A ccur acy : F ir st p u l se a ft e r t r igge r : 0. 01% o f p r o gr a m m e d t i m e + 600 t o 850 n s e c . Addi ti on a l pul s e s : 0 . 0 1 ...

  • Agilent Technologies E1300B - page 210

    Size: i nch es mm H ei g ht w i thout f ee t 6 . 9 7 1 7 7 H e igh t wit h fe e t 7.44 189 W id t h 16.75 426 D e p t h 20.1 510 D e p t h w it h t e r m i n a l b lo c k s 22.38 569 We ig ht: E1300B E1301B N e t 7.4 k g 7.8 k g Max per m odu l es 1 . 3 k g 1 . 3 k g Pow e r: L in e v o lt a ge : 115 o r 230 V a c @ 50 t o 400 H z F u s e d a t : 3 ...

  • Agilent Technologies E1300B - page 211

    SC PI C onfor mance I n fo rm a ti on T h e A g il e n t E 1300 / 1301 B c on f o r m s t o S C P I - 1990 . 0 I n d o c u m e n t a t i o n p r o d u c e d p r i o r t o J u n e 1990, t h e s e SC P I c o m m a n d s a r e la b e le d a s T M SL c o m m a n d s. The f ol l ow i ng ta bl e s l i s t al l the S CPI conf ormi ng , app r ov ed , and n ...

  • Agilent Technologies E1300B - page 212

    M u lt im e t e r C o m m a nd s T h e fo ll o win g t a b le s a p p l y t o t h e A gile n t E 1326A a n d E 1326B . Table A -3. M ult imet er SCPI-1990. 0 C onfir med Commands AB ORt CA Libr at i on :Z E R O :A U T O : AUT O? :V A L u e CO N Fig u re :F R E Sis t a n ce :R E Sist a n ce : TE MPe ratu re : VO LTag e :A C [:D C ] CO N Fig u re ? F ...

  • Agilent Technologies E1300B - page 213

    Table A-5. M ultimeter Non-SC PI C ommands CA Libr at i on : LFRequenc y : LFRequenc y ? :ST R a in CO N Fig u re :ST R a in : QUAR t e r :H B E N d in g :H P O isso n :F B E N d in g : FPOis s on :F BP o isso n :Q T E N sio n :Q C O M p r e ssio n :U N ST r a in e d DI S P l a y :M O N it o r :C H A N n e l :C H A N n e l? [:ST A T e ] [:ST A T e ...

  • Agilent Technologies E1300B - page 214

    C o unt e r C o m m a nds T h e fo l lo w in g t a b l e s a p p ly t o t h e A gi le n t E 1332A 4 C h a n e l C ounter /Tota l i z er a n d t h e A gile n t E 1333A 3 C h a n n e l U n iv e r sa l C ounte r. Table A -6. Ag ilen t E 1332A SCPI- 1990.0 C onfir med C ommands AB ORt CO N Fig u re :F R E Q u e n cy :P E R io d :P W I D t h :N W ID t h ...

  • Agilent Technologies E1300B - page 215

    Table A -8. Ag ilen t E 1333A SCPI- 1990.0 C onfir med C ommands AB ORt FE TCh? CO N Fig u re :F R E Q u e n cy :P E R io d :P W I D t h :N W ID t h CO N Fig u re ? FO RMat [:D A T A ] INITiate [ : IMM edi ate] INPut :ATTenuation : ATTenuat i on? :C O U P lin g :C O U P lin g? : FILTer [:L P A Ss] [:ST A T e ] [:ST A T e ]? : IMPedanc e : IMPedanc ...

  • Agilent Technologies E1300B - page 216

    D/ A Co nv e r ter C ommands T h e fo l lo win g t a b le s a p p ly t o t h e A gi le n t E 1328A 4 C h a n n e l D / A C o n v e r t e r . Table A -10. Ag ilen t E 1328A SC PI- 1990. 0 C onfir med Commands CA Libr at i on : S TATe : S TATe? SY ST e m : E RRor? : VERSi on? ST A T u s : QUES t i o n a b l e : C ON Di t i on ? [ : EV EN t] ? :E N A ...

  • Agilent Technologies E1300B - page 217

    D i g i ta l I/O Co mm an d s T h e fo llo win g t a b le s a p p ly t o t h e A gi le n t E 1330A Q u a d 8- b it D i git a l I / O Modul e. Table A -12. Ag ilen t E 1330A SC PI- 1990. 0 C onfir med Commands ST A T u s : Q U E S ti onabl e : C ON Di t i on ? [ : EV EN t] ? :E N A B le :E N A B le ? :O P E R a t io n : C ON Di t i on ? [ : EV EN t] ...

  • Agilent Technologies E1300B - page 218

    Syste m I ns trum e nt C ommands Tab le A -14. Sy s tem Ins tr ument S C PI- 1990. 0 C onfir med Commands AB ORt INITiate [ : IMM edi ate] [ S OURc e ] :P U L Se : C OUN t : C OUN t? :P E R io d :P E R io d ? ST A T u s : Q U E S ti onabl e : C ON Di t i on ? [ : EV EN t] ? :E N A B le :E N A B le ? :O P E R a t io n : C ON Di t i on ? [ : EV EN t] ...

  • Agilent Technologies E1300B - page 219

    Table A-17. Syst em I nstr ument Non- SC PI C om m a nds D IAG nosti c :A U T st a r t :A U T o s t a r t ? :C H E C ks u m : C O MMuni c ate :SE R ia l [:O W N e r ] [:O W N e r ]? : B OOT :C O L D [:WA R M ] : U PLoad? : DOW N l o a d : IN Terr upt :A C T : S E Tup( n) : S E Tup( n)? :P R I o r it y ( n ) :P R I o r it y ( n ) ? :WA I T ? :J SR : ...

  • Agilent Technologies E1300B - page 220

    A - 12 Specifi cati ons ...

  • Agilent Technologies E1300B - page 221

    A ppe ndi x B Erro r M e s s a ge s Usi n g T hi s A ppendi x T h is a p p e n d ix sh o ws ho w to r e a d a n in st r u m e n t ’ s e r r o r q u e u e , d iscu sse s t h e ty pes o f co m mand l ang uag e - rel ated err or mes s ag es , and p rov i des a tabl e of al l of the S y s tem In s tr um ent’ s er r or mes s ag es and thei r p robab ...

  • Agilent Technologies E1300B - page 222

    1 Error Type s Neg ati v e er r or number s are l ang uag e- related a nd categ ori z ed as s how n bel ow . Pos i ti v e er ror num bers are i ns trum en t s p ec i f i c and f or the S y s tem Ins trum ent are s um m ari z ed i n T a b l e B - 2 . Fo r oth e r i ns tru m ent s , ref er to t h e i r ow n u s e r’ s m a n u a l f or a de s c ri p ...

  • Agilent Technologies E1300B - page 223

    Error Mes s ages an d Caus es Co de M e s s a ge Cau s e -101 Inv al i d character U nrec og niz ed c harac t er in s pec i fi ed paramet er. - 1 0 2 S y nt ax err or Command i s mis si ng a s pac e or c omma betw een pa ram et ers - 1 0 3 Inv ali d se parator Command parameter i s s epar ated by som e c har ac ter other than a c omma. - 1 0 4 D at ...

  • Agilent Technologies E1300B - page 224

    Error Mes s ages an d Caus es Co de M e s s a ge Cau s e 2110 C o n fig e r r o r 10, I n s u f fi cie n t sy st e m m e m o r y Too many i nstruments i nstal l ed for the amount o f R A M in st a lle d in t h e m a in fr a m e . C a n n o t con figur e in st r u m e n t s. O n ly t h e sy st e m in st r u m e n t is st a r t e d . 2111 C o n fig e ...

  • Agilent Technologies E1300B - page 225

    1 Star t -up Err o r M e ssa g e s S tart-up err ors are m os t of ten g enerated j us t af ter the mai nf rame i s pow ered -u p or r e - booted ( D IAG : B O O T comm and) . If y ou hav e an Ag i l ent E 1301B, o r a n A gile n t E 1300B wit h a t e r m in a l c onnected to t he D is p l ay Ter mi nal Interf ace ( b u i lt - in R S-232 o n ly ) , ...

  • Agilent Technologies E1300B - page 226

    B- 6 E rro r M e ssa g e s ...

  • Agilent Technologies E1300B - page 227

    A ppe ndi x C C onne c ting and C o nf ig u r ing a Dis p lay T er m ina l Usi n g t hi s A ppen di x Thi s a pp end ix show s y ou how to conf ig ure the mai nf r ame and a suppor t ed termi nal to operate w i th th e Di s pl ay Ter mi nal Interf ac e. U s i ng the D i s p l ay Termi nal I nterf ace i s di s c us s ed i n Chapter 3 . • O v e r v ...

  • Agilent Technologies E1300B - page 228

    Figu r e C–1 Connect in g a Ter min al to th e M ainfr ame C - 2 C onne c ting and C onfiguring a Display Terminal Agilent E1300B/E1301B Mainframe Agilent Agilent Agilent Agilent E1300B/E1301B Mainframe Agilent E1300B/E1301B Mainframe Agilent Agilent 24542H Agilent 24542H Agilent 10833A, B or C to Connect to GPIB ...

  • Agilent Technologies E1300B - page 229

    1 C o nfig urin g a Te rmin a l f or the Mai n f r a m e We’ l l f i rs t s et the ter m i nal ’ s seri al c om muni cati on parameters to m atch the mai nf rame’ s def aul t setti ng s . If the mai nf r am e i s n ew and i ts f actory def au l t v a lu e s a r e s t ill se t , t h e t e r m in a l w ill be r e a d y to u se . I f t h e s e t ...

  • Agilent Technologies E1300B - page 230

    Try i n g it Tur n on th e m ai nf r am e w hi l e w atch i ng the termi nal ’ s di spl ay . Af ter the mai nf rame f i ni s hes i t s s elf -tes t, t he ter mi nal shoul d d i s p l ay " S el e ct an i ns t r ument" . If not, the mai nf rame’ s c om m uni cation par ameter s are not s et to the d ef aul t v al ues . G o back to “ R ...

  • Agilent Technologies E1300B - page 231

    E a ch S ET Menu w ill ha v e two or m ore choice s E nter Card Num b er p ress Return (0 fo r b ui lt -in, 1-7 for a p l ug -in ) T he s e tting is now in v ola tile R A M stor e a g e. See th e “ Ho w t o St o re Int erf ac e Sett i n g s” Menu C ha rt for non- vola tile s tora ge w hich m ainta ins se ttings thr ough power cy c l e s. P r e ...

  • Agilent Technologies E1300B - page 232

    How t o S t ore the S eri a l In terfa ce Configura t i on E nter Card Num b er p ress Return . Card N um b er 0 f or built - in s tor e s s e tting s into non_ v ola t ile R A M. Ca rd Nu m b er 1-7 f o r A g il en t E1324A sto res set t i n g s in t o i t s on - bo a r d E E R OM ) C - 6 C onne c ting and C onfiguring a Display Terminal ...

  • Agilent Technologies E1300B - page 233

    A ppe ndi x D S en din g B in a r y D at a Ov e r R S -232 A bo ut t his A pp endix Thi s ap p end i x desc ri bes the pr oced ur e f or sendi ng p ure bi nary d a ta ov er a n R S- 232 in t e r fa c e . T h e fo r m a t t in g d e s c r i b e d is u s e d in t h e DI A G: DOW N : C HEC: M A DD, DI A G: DOW N : C HEC: S ADD, a n d D I A G :D R I V ...

  • Agilent Technologies E1300B - page 234

    1 Se nd in g Bi na ry Dat a Ov er RS - 232 T h e R S-232 in t e r fa c e d iffe r s fr o m t h e G P I B in t e r fa c e i n t h a t t h e r e is n o d e v ice ad d re s s ing bui l t into the i nterf ac e de f ini tion. D e v i c e add r ess i ng m us t be d one on t o p o f t h e R S- 232 fu n c t io n s. T h is a d d r e ss in g is d o n e t h r ...

  • Agilent Technologies E1300B - page 235

    3 . S end " S I S YS TE M" and a c arri ag e return to g et the attenti on of the Sy st e m I n st r u m e n t . 4 . S end < CTR L-C> to c l ear the s y stem . 5 . S end " * RS T" and a carri ag e r eturn to put the S y s tem In s tr um en t i n a know n s tate. The p r og ram mus t then send the bi nary d a ta. Thi s bl o ...

  • Agilent Technologies E1300B - page 236

    D-4 Sending Binary Data Over RS-232 ...

  • Agilent Technologies E1300B - page 237

    In dex ! 3 - C h a n n e l U n ive r s a l C o u n t e r Menu (f ront panel ), 2 - 26 - 2- 27 M e nu ( t e r m ina l i n t e r face) , 3- 38 - 3- 39 4 - C h a n n e l C o u n t e r / T o t a lize r Menu (f ront panel ), 2 - 24 - 2- 25 M e nu ( t e r m ina l i n t e r face) , 3- 36 - 3- 37 4 - Channel D /A C onv erter Menu (f ront panel ), 2 - 21 Me ...

  • Agilent Technologies E1300B - page 238

    *S AV , 7-70 *S R E , 7-70 *S R E ? , 7-71 *S T B? , 7-71 *T R G , 7-71 *T ST ?, 7-71 *W AI , 7-71 Common Comma nds func ti onal g roupi ng s , 7 - 65 COMMu ni c at e: G P IB :AD D R e ss , 7- 36 :AD D R e ss ?, 7-36 C O M M u n ic a t e : S E R ia l[ 0 ] : OWN er, 7 -7 : OWN er?, 7 -8 C O M M u n ic a t e : S E R ia l[ n ] :C O N T r o l:D T R , 7 ...

  • Agilent Technologies E1300B - page 239

    D I AG :P E E K? , 7-24 DI A G: POK E, 7 - 2 5 DI A G: R DI S : A DD? , 7 - 2 5 DI A G: R DI S : C RE, 7 - 2 6 DI A G: R DI S : C RE?, 7 - 2 6 DI A G: UPL : S A DD? , 7 - 2 8 DI A G: UPL [ : MA DD] ? , 7 - 2 7 D R IV: L IS T: R O M?, 7- 18 D R IV : LIS T ?, 7- 18 DI A Gn o s t i c : DRIV e r: L I S T: R A M ?, 7 - 1 8 DI A Gn o s t i c : DRIV e r: ...

  • Agilent Technologies E1300B - page 240

    F Fi l e s : VX IDL D. C F G, 5 - 4 For ma t Common Comm and, 7 - 1 S CPI Co mma n d, 7 - 1 Fr ont panel f eatures , 2- 1 menu tutori al , 2 - 2 m e nu s, 2- 2 G G E T (g ro up ex ec u te tri g g er), 7 - 7 2 G o t o lo c a l ( G T L ) , 7 - 7 2 G rou p ex ec ute tr i g g er (G E T) , 7 - 72 G T L ( g o t o lo c a l) , 7 - 7 2 H H i nts, prog rammi ...

  • Agilent Technologies E1300B - page 241

    tutori al , 2 - 2 Menu ( termi nal i nt erf ac e) 3-Cha n n e l U n i v er sal Cou n t e r , 3-38 - 3- 39 4-Cha n n e l Cou n t e r /T o t a l i ze r , 3- 36 - 3- 37 Q uad 8 - Bi t D i g i tal Input/ O u tput, 3 - 3 4 4 - Channel D /A C onv erter , 3 -3 3 5 1/2 D ig i t M u l time t e r , 3-32 con t r o l key s, 3- 14 i nst r u m e n t , 3-25 ke y ...

  • Agilent Technologies E1300B - page 242

    S S A, termi nal i nterf ac e c o mma n d, 3 - 2 1 S c an c hannel s (f ront panel ), 2 - 5 S w i tc hbox , te rmi nal i nterf ac e, 3 -8 S c anni ng Vol t meter Menu (f ront panel ), 2 -1 8 - 2 - 19 Vol tmeter Menu (termi nal i nterf ac e), 3 - 3 0 - 3 - 3 1 SCP I C o m m a n d s, 7-1 F o r m a t , 7-1 R e fer e n ce, 7-4 S D C (s el ec ted dev i ...

  • Agilent Technologies E1300B - page 243

    SY ST : C O M M : SE R [ n ] :R E C :P A C E : T H R : ST O P , 7 - 4 3 SY ST : C O M M : SE R [ n ] :R E C :P A C E : T H R : ST O P ? , 7 - 43 SY ST :C O M M :SE R [n ]:R E C :P A R :C H E C , 7 -44 SY ST :C O M M :SE R [n ]:R E C :P A R :C H E C ? , 7 -44 SY ST : C O M M : SE R [ n ] :R E C :P A R [ :T Y P E ? ], 7 -4 6 SY ST : C O M M : SE R [ ...

  • Agilent Technologies E1300B - page 244

    Inde x -8 ...

Manufacturer Agilent Technologies Category Computer Hardware

Documents that we receive from a manufacturer of a Agilent Technologies E1300B can be divided into several groups. They are, among others:
- Agilent Technologies technical drawings
- E1300B manuals
- Agilent Technologies product data sheets
- information booklets
- or energy labels Agilent Technologies E1300B
All of them are important, but the most important information from the point of view of use of the device are in the user manual Agilent Technologies E1300B.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Agilent Technologies E1300B, service manual, brief instructions and user manuals Agilent Technologies E1300B. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Agilent Technologies E1300B.

A complete manual for the device Agilent Technologies E1300B, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Agilent Technologies E1300B by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Agilent Technologies E1300B.

A complete Agilent Technologies manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Agilent Technologies E1300B - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Agilent Technologies E1300B, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Agilent Technologies E1300B, that we can find in the current document
3. Tips how to use the basic functions of the device Agilent Technologies E1300B - which should help us in our first steps of using Agilent Technologies E1300B
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Agilent Technologies E1300B
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Agilent Technologies E1300B in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Agilent Technologies E1300B?

Use the form below

If you did not solve your problem by using a manual Agilent Technologies E1300B, ask a question using the form below. If a user had a similar problem with Agilent Technologies E1300B it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)