    Model 6081-P Wireless pH/ORP T ransmitter Instruction Manual PN 51-6081P/rev .A November 2008 ...

  • Emerson 6081-P - page 2

    ESSENTIAL INSTRUCTIONS READ THIS P AGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its product s to meet many national and international standards. Because these instrument s are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal spe ...

  • Emerson 6081-P - page 3

    About This Document This manual contains instructions for inst allation and operation of the Model 6081-P Wireless pH/ORP T ransmitter . The following list provides notes concerning all revisions of this document. Rev . Level Date Notes A 1 1/08 This is the initial release of the product manual. The manual has been reformatted to reflect the Emerso ...

  • Emerson 6081-P - page 4

    QUICK ST ART GUIDE FOR MODEL 6081 WIRELESS pH TRANSMITTER D 1. Refer to Section 2.0 Installation for inst allation instructions. 2 . Wire the pH or ORP sensor to the transmitter . Refer to the sensor instruction sheet for details. 3. Once the connections are secure and verified, install the Power Module to power to the transmitter . 4. When the tra ...

  • Emerson 6081-P - page 5

    E Quick Start Menu English Francais Espanol >> Deutsch Italiano Portuguese >> Measure? pH Redox ORP Use Preamp in? Xmtr Sensor Diagnostics? On Off Measure update rate: 1min Temperature in? °C °F Setup Wireless Network? Yes No Wireless Network HART PollAddress Network ID: 0 0000 Network Join Key 1 of 4: 0 0000000 Network Join Key 2 of ...

  • Emerson 6081-P - page 6

    F FIGURE 1-1. MENU TREE FOR MODEL 6081 pH W ireless TRANSMITTER ...

  • Emerson 6081-P - page 7

    MODEL 6081-P pH/ORP T ABLE OF CONTENTS MODEL 6081-P pH/ORP WIRELESS TRANSMITTER T ABLE OF CONTENTS Section Title Page 1.0 DESCRIPTION AND SPECIFICA TIONS ................................................................ 1 1.1 Features and Applications ........................................................................................ 1 1.2 S pe ...

  • Emerson 6081-P - page 8

    MODEL 6081-P pH/ORP T ABLE OF CONTENTS T ABLE OF CONTENTS CONT’D ii 8.0 PROGRAMMING THE TRANSMITTER .................................................................. 25 8.1 General .................................................................................................................... 25 8.2 Changing S tart-up Settings .............. ...

  • Emerson 6081-P - page 9

    iii MODEL 6081-P pH/ORP T ABLE OF CONTENTS LIST OF FIGURES Number Title Page 2-1 W all Mounting Insatallation Model 6081 ................................................................... 8 2-2 Pipe Mounting Insatallation Model 6081 .................................................................. 9 2-3 Removing Rear Cover ....................... ...

  • Emerson 6081-P - page 10

    MODEL 6081-P pH/ORP T ABLE OF CONTENT iv This p age left blank intentionally ...

  • Emerson 6081-P - page 11

    MODEL 6081-P pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1 • HIGH ACCURACY AND RELIABILITY for monitoring applications • SELF-ORGANIZING NETWORK for high dat a reliability and network stability • INDUSTR Y LEADING WIRELESS SECURITY • COMP A TIBLE WITH 1420 WIRELESS GA TEW A Y and Emerson Pr ...

  • Emerson 6081-P - page 12

    MODEL 6081-P pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 2 1.3 SPECIFICA TIONS - WIRELESS Output: WirelessHART V7 T ransmit Rate: User selectable, 1/sec. to 1/60 min (via 1420 Wireless Gateway or AMS) Measurement up date rate: 1/sec. to 1/10 min Antenna: PBT/PC integrated omni-directional antenna Radio Frequency: 2.4 GHz DSSS T ransmission d ...

  • Emerson 6081-P - page 13

    3 MODEL 6081-P pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.5 Product Certifications T elecommunication Compliance All wireless devices require certification to ensure that they adhere to regulations regarding the use of the RF spectrum. Nearly every country requires this type of product certification. Emerson is working with govern- mental ...

  • Emerson 6081-P - page 14

    MODEL 6081-P pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.7 ASSET MANAGEMENT SOLUTIONS Asset Management Solutions (AMS) is software that help s plant personnel better monitor the performance of ana- lytical instruments, pressure and temperature transmitters, and control valves. Continuous monitoring means maintenance personnel can anticipat ...

  • Emerson 6081-P - page 15

    MODEL 6081-P pH/ORP SECTION 2.0 INST ALLA TION SECTION 2.0 INST ALLA TION 2.1 Considerations 2.2 Unp acking and Inspection 2.3 Pre-Inst allation Set Up 2.4 Mechanical Inst allation 2.5 Ground the T ransmitter 2.6 Power Module Inst allation 2.1 CONSIDERA TIONS The transmitter can be commissioned before or after inst allation. It may be useful to com ...

  • Emerson 6081-P - page 16

    6 2.3 PRE-INST ALLA TION SETUP 2.3.1 T emperature Element The Model 6081-P pH/ORP transmitter is compatible with sensors having Pt 100 and Pt 1000. Sensors from other manufacturers may have a Pt 1000 RTD. For Rosemount Analytical sensors, the type of temperature element in the sensor is printed on the tag att ached to the sensor cable. For the majo ...

  • Emerson 6081-P - page 17

    7 W ARNING Failure to follow these installation guidelines could result in death or serious injury . • Make sure only qualified personnel perform the installation. Explosions could result in death or serious injury . • Before connecting a 375 Field Communicator in an explosive atmosphere, make sure the instruments are inst alled in accordance w ...

  • Emerson 6081-P - page 18

    8 MODEL 6081-P pH/ORP SECTION 2.0 INST ALLA TION 2.4.2 Mounting on a Flat Surface. Figure 2-1. W all Mounting Installation Model 6081. Use Pipe/W all Mounting Bracket Kit, PN 23820-00 Note: PN 23820-00 mounting bracket kit includes mounting hardware for pipe mounting only . Wall mounting hardware to be provided by customer . Only use suitable faste ...

  • Emerson 6081-P - page 19

    9 MODEL 6081-P pH/ORP SECTION 2.0 INST ALLA TION 2.4.3 Pipe Mounting. The pipe mounting kit (PN 23820-00/01) accommodates 1-1/2 to 2 in. pipe. Figure 2-2. Pipe Mounting Installation Model 6081. Use Pipe/W all Mounting Bracket Kit, PN 23820-00 ...

  • Emerson 6081-P - page 20

    10 2.6 POWER MODULE INST ALLA TION The section describes the procedure for installation of the power module (p art number 00753-9220-0001). The power module should stored in a safe place with a controlled environment until the Model 6081 is ready for live operation. For first time installation of the power module, follow these step s: 1. Unscrew th ...

  • Emerson 6081-P - page 21

    4. With gentle pressure, insert the keyed connector on the power module into the receptacle (labeled Power Module Connection on the drawing). The power module seat s in the connection receptacle with an o-ring. 5. Confirm that the power module is fully inserted in the receptacle and properly aligned with the surrounding terminal block. 6 . Replace ...

  • Emerson 6081-P - page 22

    12 This p age left blank intentionally MODEL 6081-P pH/ORP SECTION 2.0 INST ALLA TION ...

  • Emerson 6081-P - page 23

    MODEL 6081-P pH/ORP SECTION 3.0 WIRING SECTION 3.0 SENSOR WIRING 3.1 GENERAL INFORMA TION pH and ORP sensors without preamp s manufactured by Rosemount Analytical can be wired directly to the Model 6081-P wireless transmitter . 3.2 SENSOR WIRING T o assist in sensor wiring, please refer to the one of the following resources: 1. Sensor Instruction S ...

  • Emerson 6081-P - page 24

    14 MODEL 6081-P pH/ORP SECTION 3.0 WIRING This p age left blank intentionally ...

  • Emerson 6081-P - page 25

    15 MODEL 6081-P pH/ORP SECTION 4.0 INTRINSICALL Y SAFE & EXPLOSION PROOF 15 SECTION 4.0 INTRINSICALL Y SAFE INST ALLA TION This p age lef t blank intentionally ...

  • Emerson 6081-P - page 26

    16 MODEL 6081-P pH/ORP SECTION 4.0 INTRINSICALL Y SAFE & EXPLOSION PROOF This p age left blank intentionally ...

  • Emerson 6081-P - page 27

    17 MODEL 6081-P pH/ORP SECTION 5.0 COMMISSIONING SECTION 5.0 COMMISSIONING 5.1 Network Communications 5.2 Device Network Configuration 5.3 V erify Operation 5.1 NETWORK COMMUNICA TIONS The Model 6081 will receive any HART communications from a handheld Field Communicator , or AMS. When using a Field Communicator , any configuration changes must be ...

  • Emerson 6081-P - page 28

    18 MODEL 6081-P pH/ORP SECTION 5.0 OPERA TION WITH REMOTE CONTROLLER 5.3 VERIFY OPERA TION Operation can be verified in three locations, at the device via the Local Display , using the 375 Field Communicator , or at the Gateway via the 1420 Wireless Gateway’s integrated web server . Local Display: During normal operation, the LCD will display the ...

  • Emerson 6081-P - page 29

    19 MODEL 6081-P pH/ORP SECTION 6.0 DISPLA Y AND OPERA TION SECTION 6.0 DISPLA Y AND OPERA TION 6.1. DISPLA Y The Model 6081 has a two-line display . Generally , the user can pro- gram the transmitter to show one of three displays. If the transmitter has been configured to measure ORP or Redox, similar displays are avail- able. Figure 6-1 shows the ...

  • Emerson 6081-P - page 30

    20 6.3 MENUS - pH The Model 6081 pH transmitter has four menus: CALIBRA TE, PROGRAM, SIM PV , and DISPLA Y . Under the Calibrate and Program menus are several sub-menus. For example, under CALIBRA TE, the sub-menus are T emperature and pH or ORP/Redox. Under each sub-menu are prompts. Under PROGRAM, the sub-menus for 6081 are, T emp, Measurement, S ...

  • Emerson 6081-P - page 31

    21 MODEL 6081-P pH/ORP SECTION 6.0 DISPLA Y AND OPERA TION 6.5 SECURITY 6.6.1 How the Security Code Works Use security codes to prevent accidental or unwanted changes to program settings, displays, and calibration. T wo three-digit security codes can be used to do the following… a. Allow a user to view the default display and information screens ...

  • Emerson 6081-P - page 32

    22 MODEL 6081-P pH/ORP SECTION 6.0 DISPLA Y AND OPERA TION This p age left blank intentionally 22 ...

  • Emerson 6081-P - page 33

    23 MODEL 6081-P pH/ORP SECTION 7.0 OPERA TION WITH MODEL 375 7.1 Note on Model 375 HART Communicator 7.2 Connecting the HART Communicator 7.3 Operation SECTION 7.0 OPERA TION WITH MODEL 375 7.1 Note on Model 375 HART Communicator The Model 375 HART Communicator is a product of Emerson Process Management, Rosemount Inc. This section contains selecte ...

  • Emerson 6081-P - page 34

    24 7.3 Operation 7.3.1 Off-line and On-line Operation The Model 375 Communicator features off-line and on-line communications. On-line means the communicator is connected to the transmitter in the usual fashion. While the communicator is on line, the operator can view meas- urement data, change program settings, and read diagnostic messages. Of f-l ...

  • Emerson 6081-P - page 35

    25 MODEL 6081-P pH/ORP SECTION 8.0 PROGRAMMING THE TRANSMITTER SECTION 8.0 PROGRAMMING THE TRANSMITTER 8.1 General 8.2 Changing St art-up Settings 8.3 Choosing and Configuring the Analytical Measurement 8.4 Choosing T emperature Unit s and Manual or Auto T emperature Compensation 8.5 Setting a Security Code 8.6 Making HART -Related Settings 8.7 Res ...

  • Emerson 6081-P - page 36

    26 MODEL 6081-P pH/ORP SECTION 8.0 PROGRAMMING THE TRANSMITTER 8.3 CHOOSING AND CONFIGURING THE ANAL YTICAL MEASUREMENT 8.3.1 Purpose This section describes how to do the following: 1. Configure the transmitter to measure pH, ORP , or Redox. 2. Determine the location of the preamp. 3. If pH was selected, there are additional selections and settings ...

  • Emerson 6081-P - page 37

    27 MODEL 6081-P pH/ORP SECTION 8.0 PROGRAMMING THE TRANSMITTER 8.3.3 Procedure to configure: Measurement. T o choose a menu item, move the cursor to the item and press ENTER. T o store a number or setting, press ENTER. 1. Press MENU. The main menu screen appears. Choose Program . 2. Choose Measurement . 3. Choose pH, Redox , or ORP . If you chose p ...

  • Emerson 6081-P - page 38

    28 MODEL 6081-P pH/ORP SECTION 8.0 PROGRAMMING THE TRANSMITTER 8.4 CHOOSING TEMPERA TURE UNITS AND MANUAL OR AUTOMA TIC TEMPERA TURE COMPENSA TION 8.4.1 Purpose This section describes how to do the following: 1. Choose temperature display units (°C or °F). 2. Choose automatic or manual temperature compensation. 3. Enter a temperature for manual t ...

  • Emerson 6081-P - page 39

    29 MODEL 6081-P pH/ORP SECTION 8.0 PROGRAMMING THE TRANSMITTER 8.5.2 Procedure: Setting a security code. 1. Press MENU. The menu screen appears. Choose Program . 2. Choose >> . 3. Choose Security . 4. Choose Calib or Config . a. If you chose Calib , enter a three-digit security code. b. If you chose Config , enter a three-digit security code. ...

  • Emerson 6081-P - page 40

    30 MODEL 6081-P pH/ORP SECTION 8.0 PROGRAMMING THE TRANSMITTER 8.6 MAKING HART RELA TED SETTINGS For more information refer to Sec. 1.0 8.7 RESETTING F ACTOR Y CALIBRA TION AND F ACTOR Y DEF AUL T SETTINGS 8.7.1 Purpose This section describes how to install factory calibration and default values. The process also clears all fault messages and retur ...

  • Emerson 6081-P - page 41

    31 MODEL 6081-P pH/ORP SECTION 8.0 PROGRAMMING THE TRANSMITTER 8.8 SELECTING A DEF AUL T SCREEN AND SCREEN CONTRAST 8.8.1 Purpose This section describes how to do the following: 1. Set a default screen. The default screen is the screen shown during normal operation. The 6081 allows the user to choose from a number of screens. Which screens are avai ...

  • Emerson 6081-P - page 42

    32 This p age left blank intentionally MODEL 6081-P pH/ORP SECTION 8.0 PROGRAMMING THE TRANSMITTER 32 8.9 CHOOSING A DISPLA Y TIMEOUT 8.9.1 Purpose The local transmitter screen will timeout to preserve power module life. The transmitter will continue to receive measurement inputs and continue to transmit to the wireless network as programmed. 8.9.2 ...

  • Emerson 6081-P - page 43

    33 MODEL 6081-P pH/ORP SECTION 9.0 CALIBRA TION – TEMPERTURE SECTION 9.0 CALIBRA TION — TEMPERA TURE 9.1 INTRODUCTION The Calibrate Menu allows the user to calibrate the pH, ORP (or redox), and temperature response of the sensor . 9.2 CALIBRA TING TEMPERA TURE 9.2.1 Purpose T emperature affects the measurement of pH in three ways. 1. The analyz ...

  • Emerson 6081-P - page 44

    34 MODEL 6081-P pH/ORP SECTION 9.0 CALIBRA TION – TEMPERTURE 9.2.2 Procedure 1. Remove the sensor from the process. Place it in an insulated container of water along with a calibrated ther- mometer . Submerge at least the bottom two inches of the sensor . S tir continuously . 2. Allow the sensor to reach thermal equilibrium. For some sensors, the ...

  • Emerson 6081-P - page 45

    35 MODEL 6081-P pH/ORP SECTION 10.0 CALIBRA TION – PH SECTION 10.0 CALIBRA TION — pH AND ORP 10.1 INTRODUCTION For pH sensors, two-point buffer calibration is st andard. Both automatic calibration and manual calibration are available. Auto calibration avoids common pitfalls and reduces errors. Its use is recommended. In auto calibration the 608 ...

  • Emerson 6081-P - page 46

    36 MODEL 6081-P pH/ORP SECTION 10.0 CALIBRA TION – PH FIGURE 10-1. Calibration Slope and Offset During automatic calibration, the transmitter also measures noise and drift and does not accept calibration dat a until read- ings are stable. Calibration dat a will be accepted as soon as the pH reading is constant to within the factory-set limit s of ...

  • Emerson 6081-P - page 47

    37 MODEL 6081-P pH/ORP SECTION 10.0 CALIBRA TION – PH 10.2 PROCEDURE — AUTO BUFFER CALIBRA TION 1. Obtain two buf fer solutions. Ideally , the buffer values should bracket the range of pH values to be measured. 2. Remove the pH sensor from the process liquid. If the process and buffer temperatures are appreciably dif fer- ent, place the sensor ...

  • Emerson 6081-P - page 48

    38 10.3 PROCEDURE — MANUAL TWO-POINT BUFFER CALIBRA TION 1. Obtain two buf fer solutions. Ideally , the buffer values should bracket the range of pH values to be measured. 2. Remove the pH sensor from the process liquid. If the process and buf fer temperatures are appreciably different, place the sensor in a container of t ap water at the buffer ...

  • Emerson 6081-P - page 49

    39 MODEL 6081-P pH/ORP SECTION 10.0 CALIBRA TION – PH 10.4 PROCEDURE — ST ANDARDIZA TION 1. The pH measured by the transmitter can be changed to match the reading from a second or referee instru- ment. The process of making the two readings agree is called st andardization. 2. During st andardization, the difference between the two values is co ...

  • Emerson 6081-P - page 50

    40 10.5 PROCEDURE — ENTERING A KNOWN SLOPE V ALUE. 1. If the electrode slope is known from other measurements, it can be entered directly into the transmitter . The slope must be entered as the slope at 25°C. T o calculate the slope at 25°C from the slope at temperature t°C, use the equation: slope at 25°C = (slope at t°C) Changing the slope ...

  • Emerson 6081-P - page 51

    41 MODEL 6081-P pH/ORP SECTION 10.0 CALIBRA TION – PH 10.6 ORP CALIBRA TION 10.6.1 Purpose 1. For process control, it is often important to make the measured ORP agree with the ORP of a st andard solu- tion. 2. During calibration, the measured ORP is made equal to the ORP of a standard solution at a single point. 10.6.2 Prep aration of ORP st and ...

  • Emerson 6081-P - page 52

    This p age left blank intentionally 42 MODEL 6081-P pH/ORP SECTION 10.0 CALIBRA TION – PH ...

  • Emerson 6081-P - page 53

    43 MODEL 6081-P pH/ORP SECTION 1 1.0 MAINTENANCE SECTION 1 1.0 MAINTENANCE 1 1.1 OVERVIEW This section gives general procedures for routine maintenance of the 6081-P pH/ORP transmitter and pH and ORP sensors. The transmitter needs almost no routine maintenance. Sensors require periodic inspection and cleaning. The calibration of the transmitter-sen ...

  • Emerson 6081-P - page 54

    44 MODEL 6081-P pH/ORP SECTION 1 1.0 MAINTENANCE 1 1.3 pH SENSOR MAINTENANCE 1 1.3.1 Frequency of Cleaning The frequency at which a sensor should be inspected and cleaned can be determined only by experience. If the process liquid coats or fouls the sensor , frequent cleaning may be necessary . If the process does not contain a high level of suspen ...

  • Emerson 6081-P - page 55

    45 1 1.3.3 Checking the Reference Electrode. Some processes contain subst ances, for example, sulfides, that poison the reference electrode. Poisoning alters the electrode potential. For example, sulfide poisoning converts the reference electrode from a silver/silver chloride elec- trode into a silver/silver sulfide electrode, causing a shift in po ...

  • Emerson 6081-P - page 56

    46 1 1.4 ORP SENSOR MAINTENANCE 1 1.4.1 Frequency of Cleaning The frequency at which an ORP sensor should be inspected and cleaned can be determined only by experience. If the process liquid coats or fouls the sensor , frequent cleaning may be necessary . If the process does not contain a high level of suspended solids, the need for regular cleanin ...

  • Emerson 6081-P - page 57

    47 1 1.6 POWER MODULE REPLACEMENT Expected power module (part number 00753-9220-0001) life is two years at reference conditions. This section describes the procedure for replacement of power module (part number 00753-9220-0001). The new power mod- ule should stored in a safe place with a controlled environment until the Model 6081 is ready for live ...

  • Emerson 6081-P - page 58

    48 This p age left blank intentionally 48 MODEL 6081-P pH/ORP SECTION 1 1.0 MAINTENANCE 1 1.6.1 Handling Considerations The power module with the wireless unit contains 2 “C” size primary lithium/thionyl chloride batteries. Each power module contains approximately 2.5 grams of lithium, for a tot al of 5 grams in each pack. Under normal conditio ...

  • Emerson 6081-P - page 59

    49 MODEL 6081-P pH/ORP SECTION 12.0 RETURN OF MA TERIALS SECTION 12.0 RETURN OF MA TERIAL 12.1 GENERAL. T o expedite the repair and return of instruments, proper communication between the customer and the factory is important. Call 1-949-757-8500 or 1-800-854-8257 for a Return Materials Authorization (RMA) number . 12.2 W ARRANTY REP AIR. The follo ...

  • Emerson 6081-P - page 60

    50 This p age left blank intentionally MODEL 6081-P pH/ORP SECTION 12.0 RETURN OF MA TERIALS ...

  • Emerson 6081-P - page 61

    51 W ARRANTY Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defect s in workman- ship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller . Consumables, pH electrodes, membranes, liquid junctions, electrolyte, O-rings, etc. are warranted to ...

  • Emerson 6081-P - page 62

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .r ...

Manufacturer Emerson Category Wireless Office Headset

Documents that we receive from a manufacturer of a Emerson 6081-P can be divided into several groups. They are, among others:
- Emerson technical drawings
- 6081-P manuals
- Emerson product data sheets
- information booklets
- or energy labels Emerson 6081-P
All of them are important, but the most important information from the point of view of use of the device are in the user manual Emerson 6081-P.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Emerson 6081-P, service manual, brief instructions and user manuals Emerson 6081-P. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Emerson 6081-P.

Similar manuals

A complete manual for the device Emerson 6081-P, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Emerson 6081-P by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Emerson 6081-P.

A complete Emerson manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Emerson 6081-P - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Emerson 6081-P, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Emerson 6081-P, that we can find in the current document
3. Tips how to use the basic functions of the device Emerson 6081-P - which should help us in our first steps of using Emerson 6081-P
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Emerson 6081-P