Graham Field 750000 - page 1
750000 AL TERNA TING PRESSURE / L OW AIR L OSS MA TTRESS SYSTEM USER MANUAL Impor tant: Do not operate the Mattress System without rst reading and understanding this manual! Save this manual for future use. Info: The most current version of this manual can be found online at www.grahameld.com. 750000-INS-LAB-RevC11 ...
-
Graham Field 750000 - page 2
2 750000-INS-LAB-RevC11 CONTENTS INTRODUCTION..................................................................................... 3 INTENDEDUSEOFTHISDEVICE ........................................................ 3 CONTRAINDICA TION ...................................................................... 3 IMPORT ANTSAFETY ...
-
Graham Field 750000 - page 3
750000-INS-LAB-RevC11 3 INTR ODUCTION Use this manual for set-up and operation of the Lumex AltaDyne 750000 Alternating Pr essure / Low Air Loss Mattr ess System. Read all instructions before using the mattr ess system. Save this manual for future r eference. INTENDED USE OF THIS DEVICE The intended use of the Lumex AltaDyne 750000 Alternating P ...
-
Graham Field 750000 - page 4
4 750000-INS-LAB-RevC11 IMPORT ANT SAFETY PRECA UTIONS The safety statements pr esented in this chapter refer to the basic safety information that should be obser ved by those using the Alternating Pr essure / Low Air Loss Mattr ess System. There ar e additional safety statements in other chapters or sections, which may be the same as or similar ...
-
Graham Field 750000 - page 5
750000-INS-LAB-RevC11 5 WARNING To reducetheriskofburns,electrocution,r e,or personalinjury: W ARNING:Donotleavethisproductunattended whenpluggedin. W ARNING: Alwaysuseclosesupervisionwhenthis productisused b y, on,ornearchildrenor ...
-
Graham Field 750000 - page 6
6 750000-INS-LAB-RevC11 W ARNING:Patiententrapmentwithbedsiderails maycauseinjuryordeath.Thebedframeandits components,includingthemattress,bedsiderails, headandfootboard,bedding,andanyaccessories addedtothebed,canallaffectthe? ...
-
Graham Field 750000 - page 7
750000-INS-LAB-RevC11 7 EQUIPMENT SYMBOLS Attention! Consult accompanying documents (this manual) “BF”symbol: Indicates that this product is in accordance with the degree of protection against electric shock for type BF equipment Functional earth (for UL only) Class II (for UL only) Disposal of Electrical & Electronic Equipment (WEEE): D ...
-
Graham Field 750000 - page 8
8 750000-INS-LAB-RevC11 PR ODUCT DESCRIPTION PUMP AND MA TTRESS SY STEM pumpandmattresssystem CPRvalves mattress-pump connectors mattress footend symbol pump Pump: front quick connector ...
-
Graham Field 750000 - page 9
750000-INS-LAB-RevC11 9 Pump: rear Pump: top panel pressur eadjustmentknob normalpressur eindicator lowpressur eindicator mainpowerswitch staticfunctionswitch Pr essur e adjust knob : The pressur e adjust knob controls the air pr essure output. T urn it clockwise to increase output pr essure (rmer); turn it count ...
-
Graham Field 750000 - page 10
10 750000-INS-LAB-RevC11 Static function switch : Pr ess the static function switch to suspend alternating function, if needed; the pr essure inside the air tubes will then be adjusted to identical pr essure. Pr ess the static function switch again to change back to alternating mode. The static function switch toggles between static and alternat ...
-
Graham Field 750000 - page 11
750000-INS-LAB-RevC11 11 INST ALLA TION UNP ACKING 1. Befor e unpacking, check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment pr ovider . 2. Remove all loose packing fr om the car ton. 3. Car efully remove all components fr om the car ton. 4. Inspect all component ...
-
Graham Field 750000 - page 12
12 750000-INS-LAB-RevC11 PUMP AND MA TTRESS INST ALLA TION 1. Place the mattr ess on top of the bed frame; ensur e that the foot end symbol faces up at foot end as shown at right. 2. Fold open the wir e hangers on the back of the pump as shown at right. 3. Either hang the pump on the bed's foot end as shown at right, or place the pump on a ...
-
Graham Field 750000 - page 13
750000-INS-LAB-RevC11 13 W ARNING:Locateallcordssothattheywillnotbe steppedon,trippedover ,orotherwisesubjectedto damageorstress. W ARNING:Donotuseahouseholdextensioncordif theelectricalcorddoesnotreachthepoweroutlet. Useo ...
-
Graham Field 750000 - page 14
14 750000-INS-LAB-RevC11 OPERA TION 1. T urn the power switch to the ON position. 2. Adjust the pr essure adjust knob to the maximum position. Info: Set the pressure adjust knob to the maximum position for the quickest inflation at set-up; the mattress can then be adjusted to the desired pressure. 3. The pump will begin delivering air into the ...
-
Graham Field 750000 - page 15
750000-INS-LAB-RevC11 15 under the patient's buttock ar ea; your ngers should just t. If the t is too loose, decr ease the setting. If your ngers do not t well, incr ease the setting two notches rmer . W ait ten minutes and then check again. Y ou should only have to adjust this setting one time per patient. Please consult ...
-
Graham Field 750000 - page 16
16 750000-INS-LAB-RevC11 MAINTENANCE CLEANING Perform the following cleaning pr ocedures befor e use. Pump : W ipe the pump with a clean, damp cloth and mild deter gent, and keep it away fr om dust. Air dr y . s NOTICE:Donotusephenolicproductsorcorrosive orpowderedcleanserstocleanthepump. s NOTIC ...
-
Graham Field 750000 - page 17
750000-INS-LAB-RevC11 17 Fuse replacement tool needed: small screwdriver 1. If you suspect a blown fuse, disconnect the plug fr om the wall outlet immediately . 2. Use a small scr ewdriver to r emove the cover of the fuse holder (fuse locations ar e shown in pictur e at right). 3. Insert a new fuse of the correct rating (T1A/250V , VDE appr oved ...
-
Graham Field 750000 - page 18
18 750000-INS-LAB-RevC11 SPECIFICA TIONS Pump Specication Power Supply 120V System: AC 120V 60 Hz Info: see rating label on product Fuse Rating T1A, 250V Cycle time 8 minutes Air input >4 LPM Dimensions (L x W x H) 9.8" x 5.3" x 3.7" (25 x 13.5 x 9.5 cm) Weight 4 lb (1.8 kg) Environment T emperature Operation: 50°F to 104° ...
-
Graham Field 750000 - page 19
750000-INS-LAB-RevC11 19 WARRANTY GF Health Pr oducts, Inc. warrants the Lumex AltaDyne 750000 Alternating Pr essure / Low Air Loss Mattr ess System for a period of twelve months for defects in workmanship and materials. If a pr oduct is deemed to be under warranty , GF Health Products, Inc. shall pr ovide, at its option, (1) r eplacement of any ...
-
Graham Field 750000 - page 20
USA Corporate Headquar ters: GF Health Products, Inc. 2935 Nor theast Parkway Atlanta, Georgia 30360 telephone: 770-368-4700 fax: 770-368-2386 www .grahamfield.com © 2011, GF Health Products, Inc. ...