    www .omega.com e-mail: info@omega.com U ser ’ s Guide DPF75, DPF76 AND DPF78 SERIES Rate/T otalizer Shop online at M1090 /071603 ...

  • Omega DPF76 - page 2

    Servicing Nor th America: USA: One Omega Drive, P.O. Box 4047 ISO 9001 Certified Stamford CT 06907-0047 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca For immediate technical or application assistance: USA and Canada: Sales Servic ...

  • Omega DPF76 - page 3

    T ABLE OF CONTENTS i QUICK SET-UP GUIDE ..................................................................................................... ii SPECIFICATIONS ............................................................................................................. 1 HOW TO MOUNT THE UNIT ........................................................ ...

  • Omega DPF76 - page 4

    QUICK SET -UP GUIDE 1) Determine what function the panel meter is to perform, e.g., rate display with alarm, batch control, totalizer, etc. 2) Ensure compatibility with sensor. The DPF70 panel meters require a pulse (or contact closure) input. The pulse must exceed 3 VDC and drop below 1 VDC. If necessary, as in the case of paddle wheel, turbine me ...

  • Omega DPF76 - page 5

    COUNTER STEP 2: DPF75, DPF76: A,) Reset to zero "rST 0" will allow us to count up. "SET Pr" counts down from the preset value. B.) To read the total in whole gallons, set the decimal point to the far right. C.) Since you do not want to sum or take the difference of two inputs, choose "A SEP B" (DPF75 only). D.) The fre ...

  • Omega DPF76 - page 6

    1 DESCRIPTION & SPECIFICA TIONS DESCRIPTION: The OMEGA DPF70 is a dual input (channel A and B) counter/ ratemeter, each with its own 5 digit dividing scale factor. The two 5 AMP preset relay outputs can be programmed by the user to apply to the "A" total counter, the "A" ratemeter, or the "B" counter or the net tot ...

  • Omega DPF76 - page 7

    2 HOW TO MOUNT THE UNIT The unit is designed to be mounted with a gasket providing a water tight seal. Two mounting brackets are provided to secure the unit to the panel. A panel less than .1" may distort if the clamps are screwed too tightly. Slide the body of the unit through the rubber gasket. Insert the unit into the panel. As shown in &qu ...

  • Omega DPF76 - page 8

    3 WIRING The rear terminal contains 12 screw terminals for connecting #14 to #28 gauge wire. The unit is controlled by a microprocessor and, therefore, an electrically "noisy" environment could cause operating problems. The input power line should not be common to power lines for motors, pumps, contactors, etc. The unit is designed to be ...

  • Omega DPF76 - page 9

    4 CONNECTING AC / DC POWER NOTE: Connect power only after other connections are finished. Do not touch the live AC power terminals! The unit has been designed with an isolated AC input. Thus, polarity is not a concern for the AC input. The chassis is plastic, therefore earth ground is not used. For D.C. operation, connect + 12V to pin 7 and - D.C. ...

  • Omega DPF76 - page 10

    5 WHA T CAN YOU VIEW? 1) "A sub B" - If you have selected the "A sub B" mode, pressing "view" shows: A) The net result of subtracting input B from input A. Pulses on input B will subtract (count down). Pulses on input A will add (count up) if "reset to 0" is selected, if "set to preset" is selected, ...

  • Omega DPF76 - page 11

    6 PRGM FACToR RUN MODE DP F A ##### DP F B ##### ENTER ENTER ENTER ENTER ENTER PRGM Hi CPS Lo CPS PRGM RUN MODE ENTER RATE METER (MR2) ONLY CouNT ENTER RST 0 SET PR PRGM ENTER DPLoC ENTER A NET B A SEP B PRGM ENTER ENTER A AD D B A SuB B PRGM Hi CPS Lo CPS PRGM ENTER RUN MODE ENTER RATE ENTER SCALE R P S PRGM Z3600 Z60 PRGM ENTER ENTER ENTER RUN MO ...

  • Omega DPF76 - page 12

    7 The following is a list of abbreviations as they appear on the display and front panel of the unit. ABBREVIATION DESCRIPTION SCALING FACTOR - For A and B Inputs. Each input has a separate 5 digit dividing scale factor. DECIMAL POINT FOR FACTOR A - Enter location of decimal point for scaling Factor A by pressing the button under the digit where th ...

  • Omega DPF76 - page 13

    8 LOCK - This portion of the menu allows you to: 1) lock the program (presets are still accessible) 2) lock all (presets and program are locked). LOCK PROGRAM - This will lock the program and allow the presets to be changed when the unit is in the lock mode. LOCK ALL - This will lock the program and the presets when the unit is in the locked mode. ...

  • Omega DPF76 - page 14

    9 CALCULA TING SCALE F ACT ORS There are two separate dividing scale factors, one for input "A" and one for input "B". The factor to enter is the number of pulses per the desired unit of measurement. The factor ranges from 0.0001 to 99999. The factor is the same for rate and count on input "A". Because the "units ...

  • Omega DPF76 - page 15

    10 REMARKS This section of the menu is used to set up the scaling factors for inputs A & B. This sets the decimal for factor A. Press the arrow key under the digit where the decimal is desired. To clear the decimal, press the arrow key fur- thest to the right (PRGM) . This is the scaling factor for input A. To change, press the arrow key under ...

  • Omega DPF76 - page 16

    11 PRESS DISPLAY REMARKS PRGM PRGM ENTER ENTER SETTING THE COUNTER STEP 2 DPF75 & DPF76 ONLY or or or or ENTER ENTER ENTER This section of the menu sets up the counter information. Press the PRGM key to choose RST 0 (reset to 0, count up) or SET PR (set to preset, count down), press the ENTER key to enter the displayed choice. This sets the dec ...

  • Omega DPF76 - page 17

    12 ENTER STEP 3 REMARKS PRESS DISPLAY This section of the menu is used to set up the rate information. Press the PRGM key to choose RPS (rate per second) or SCALE (RPM, RPH). Press ENTER to enter displayed choice. Press PRGM to choose ÷ 60 (RPM) or ÷ 3600 (RPH). Press ENTER to enter displayed choice. This sets the normalizing (averaging) factor. ...

  • Omega DPF76 - page 18

    13 DISPLAY REMARKS or Flashes followed by: Flashes followed by: or PRGM PRGM PRGM PRGM This section of the menu is used to set up the lockout type and code. LC PG = Locks program but presets and reset are accessible. LC ALL= Locks entire keypad. Press the PRGM button to toggle between choices; Press ENTER to enter displayed choice. After CODE flash ...

  • Omega DPF76 - page 19

    14 REMARKS DISPLAY PRESS PRGM PRGM PRGM PRGM PRGM STEP 5 SETTING THE RELAYS ENTER This section sets up the relay information. Press the PRGM key to choose A TOT (A assigned to total) or A RATE (A assigned to rate). Press ENTER when the desired choice is displayed. This is the duration (.1 to 99.9 sec) that relay A will remain energized. If 00.0 is ...

  • Omega DPF76 - page 20

    15 OUTPUT WIRING The following diagrams detail the connection of the relay and analog output options. Each relay consist of a form A contact (Normally Open). NPN transistor or Normally Closed contacts are available with solder jumpers (see Jumper Options). SOLENOID (Single Stage) Use Relay A (pins 1 & 2) for single stage preset. When the count ...

  • Omega DPF76 - page 21

    16 ANALOG OUTPUT OPTION Description: An optional 4-20mA (0-20mA) output has been added to the DPF70 series. The output can be programmed to track rate or total. This feature is available by adding suffix A to the part number. Connec- tions are via a 2 terminal pluggable screw connector. Connections: (see FIG. 1) PIN13 supplies the 12 to 18 VDC to p ...

  • Omega DPF76 - page 22

    17 INTERF ACE CARD RS 232/422 OPERA TION RS 232/422 SET-UP: All serial communication mode changes must be done through serial communications. Mode changes cannot be done through the front panel. To initialize the unit, place a jumper between pin 7(+12V )[bottom board] and pin 1(init) [DB -9 connector] on initial power up. The unit defaults to: 300 ...

  • Omega DPF76 - page 23

    18 SERIAL INPUT COMMANDS: To get a unit on line you must address it by its device number. This is done by typing DXX(S), XX= device number. The unit comes on line and echoes back DEVICE# XX. Insure that "DEVICE# XX:" is received before requests are sent. The unit is now ready to receive a command or string of commands separated by a space ...

  • Omega DPF76 - page 24

    19 COMPUTER HOOKUP: RS 232: When connecting the unit to a computer with RS 232 communication, only three connections are needed. These connections are: Receive data, Transmit data and Ground. The connections should be made as follows: DP -9 CONNECTOR COMPUTER Transmit data (pin 2) Receive data Receive data (pin 3) Transmit data Ground (pin 5) Groun ...

  • Omega DPF76 - page 25

    20 PROBLEM SOLUTIONS POSSIBLE CAUSES TROUBLESHOOTING GUIDE Power is applied to unit but the display does not light. Unit works, but occasionally the display freezes or skips counts. Input signal is connected but the unit does not count or display rate. Rate is displaying: r FFFFF. Relays are not activating properly. Counter resets before reaching 9 ...

  • Omega DPF76 - page 26

    21 NOTES ...

  • Omega DPF76 - page 27

    22 NOTES ...

  • Omega DPF76 - page 28

    23 NOTES ...

  • Omega DPF76 - page 29

    W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OM EGA ’ s W ARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA ’ s cust ...

  • Omega DPF76 - page 30

    Where Do I Find Ever ything I Need for Pr ocess Measurement and Control? OM EGA…Of Course! Shop online at www .omega.com TEMPERA TURE 䡺 ⻬ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies 䡺 ⻬ Wire: Thermocouple, RTD & Thermistor 䡺 ⻬ Calibrators & Ice Point References 䡺 ⻬ Recorders, Controllers ...

Produzent Omega Kategorie Calculator

Dokumente, die wir vom Produzenten des Geräts Omega DPF76 erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Omega
- Bedienungsanleitungen DPF76
- Produktkarten Omega
- Informationsbroschüren
- oder Energieetiketten Omega DPF76
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Omega DPF76.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Omega DPF76, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Omega DPF76. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Omega DPF76 ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Omega DPF76, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Omega DPF76 hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Omega DPF76 verständlich ist.

Eine gänzliche Bedienungsanleitung von Omega sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Omega DPF76 bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Omega DPF76 befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Omega DPF76, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Omega DPF76 - die uns die ersten Schritte während der Nutzung von Omega DPF76 erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Omega DPF76 hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Omega DPF76 im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Omega DPF76?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Omega DPF76 nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Omega DPF76 hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)