    Agilent 1604 8A T est Leads Operation and Service Manual Fourth Edition Agilent T echnologi es Japan, Ltd. Agilent PN 160 48-9000 1 March 2 000 Print ed in: Japan ...

  • Agilent Technologies 16048A - page 2

    2 Notices The in format ion con tained in this docume nt is subjec t to ch ange withou t n otice. This docu ment contains p ropri etary infor mation that is p rotec ted by c opyright. All rights are rese rved. No p art of t his doc ument may be pho tocopi ed, rep roduced, or transl ated to a nother language wi thout the prior writ ten consent of t ...

  • Agilent Technologies 16048A - page 3

    3 • Ground The Instr ument T o avoid electri c shock hazard, t he inst rument chassi s and cabi net must be connecte d to a s afety ea rth gr ound by t he suppli ed power cable with ea rth blade. • DO NOT Operate In An E xplosi ve Atmosphere Do not oper ate the i nstr ument in the prese nce of fl ammable ga sses o r fumes. Operati on of any el ...

  • Agilent Technologies 16048A - page 4

    4 Wa r r a n t y This Ag ilent T echn ologie s inst rument product is war rante d against defec ts in materia l and workmans hip fo r a perio d corre sponding t o the indi vidual warranty periods of it s compone nt product s. Inst rument s are warr anted for a p eriod of o ne year . Fixtures and adapte rs ar e warrant ed for a peri od of 90 days . ...

  • Agilent Technologies 16048A - page 5

    5 Assistance Product ma intena nce agree ments an d other cus t omer a ssistan ce agreements are availa ble for Agilent T echnologi es produc ts. For any as sistance, conta ct your nearest Agilen t T echnologi es Sal es and Servi c e Off ice. Addres ses ar e provide d at t he back of t his manual . Safety S ymbol General definiti ons of safety s ym ...

  • Agilent Technologies 16048A - page 6

    6 ...

  • Agilent Technologies 16048A - page 7

    Contents 7 1. Operation Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1 T ypical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies 16048A - page 8

    8 Contents ...

  • Agilent Technologies 16048A - page 9

    9 1 Operation This oper ating no te prov ides co mplete i nformat ion on th e 16048A T est Le ads. The 16048A is s hown pic toriall y in Figur e 1-1, i ts physic al di mensions ar e given in T able 1-1, a nd typi cal char acteristics related to of fset error are give n in T able 1-2. T o order add ition al copies of th is operat ing not e, use the ...

  • Agilent Technologies 16048A - page 10

    10 Chapter 1 Operation Product Des cription Product Description The 16048A co nsists of a dir ect att achment, 4-terminal pair type fixture which is equip ped w ith f our BNC (m) termi nat ed-co axia l tes t lead s. Th ese te st lead s ar e used to a ttach use r -fabri cated test fi xture s. DC bias le vels of up to ±40 V can be applied to the 160 ...

  • Agilent Technologies 16048A - page 11

    Chapter 1 11 Oper ation Specifications Specifications T able 1-1 Specifications of the 16 048A Function: For use with 4 terminal-pair LCR Meters and Impedance Analy zers. Connector T ype: BNC male Cable Length: Approximately 1m Maximum V oltag e: ± 40V peak max. (AC+DC) W eight: 315 g Safety Stand ards: EN610 10-1:199 3 +A2:1995 IEC6 1010-1: 1990 ...

  • Agilent Technologies 16048A - page 12

    12 Chapter 1 Operation T ypica l Character istics T ypical Characteristics T able 1 -2 T ypical Ch aracteristic s NOTE f is the meas urem ent fr eque ncy in meg ahert z. The incr emen tal er rors c alcul ated from the equation i n the t able for measur ements at f requenci es above 1 MHz are additi ve. Figur e 1-2 Incremental erro r at freq. ³ 1 M ...

  • Agilent Technologies 16048A - page 13

    Chapter 1 13 Oper ation Compensation for Fixtur e Residual Impedance Error Compensation for Fixtur e R esidual Impedance Err or The 16048A ha s inherent stray c apacitan ce, residual in ductanc e, and res idual resist ance that af fect the accura cy of mea sured val ues. T o compe nsate for , or negate, t hese res idual s to minimiz e measur ement ...

  • Agilent Technologies 16048A - page 14

    14 Chapter 1 Operation Compensation for Fixture Residual Impedance Er ror ...

  • Agilent Technologies 16048A - page 15

    15 2 Service ...

  • Agilent Technologies 16048A - page 16

    16 Chapter 2 Serv ice Maintena nce Maintenance An explode d view of the 16 048A (for pa rts i dentifi catio n) is shown in Figure 2-1. Do not dis assemble any furt her than s hown. Maintenance consist s prin cipally o f cleani ng conta cts and r eplacing w orn or d amaged par ts. T ake spe cial car e when cleani ng conta cts. T o order parts, us e ...

  • Agilent Technologies 16048A - page 17

    Chapter 2 17 Service Maintenance Figur e 2-1 Parts Iden tificat ion ...

  • Agilent Technologies 16048A - page 18

    18 Chapter 2 Serv ice Maintena nce T able 2-1 Parts Id entification Reference Agilent Part No. Qty . Description Note 1 * 4 SLEEVE 2 * 4 NUT 3 * 4 NUT 4 * 4 W ASHER 5 * 4 W ASHER 6 2360-0 192 2 SCREW 7 * 1 COVER-BOTTOM 8 * 4 INSULA TOR 9 * 2 CONNECTOR-BNC 10 * 2 CONNECTOR-BNC 1 1 2200-0103 1 S CREW 12 219 0-0206 1 W ASHER 13 160 47-400 00 1 STO PPE ...

  • Agilent Technologies 16048A - page 19

    Chapter 2 19 Service Maintenance *: Not se parat ely repla ceable. 31 1250-01 18 4 CONNECTOR-BNC FEMALE 32 236 0-01 15 2 SCREW 1 thru 30 33 160 32-10021 1 P LA TE 34 219 0-0016 4 W ASHER 35 295 0-0001 4 NUT 36 160 32-10022 1 P LA TE 37 160 32-600 01 1 BNC BRA CKET 31thr u36 a. Agilent interna l -only par t number . Reference Agilent Part No. Qty . ...

  • Agilent Technologies 16048A - page 20

    20 Chapter 2 Serv ice Maintena nce ...

  • Agilent Technologies 16048A - page 21

    REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative. 11/29/99 U ...

Produzent Agilent Technologies Kategorie Power Supply

Dokumente, die wir vom Produzenten des Geräts Agilent Technologies 16048A erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Agilent Technologies
- Bedienungsanleitungen 16048A
- Produktkarten Agilent Technologies
- Informationsbroschüren
- oder Energieetiketten Agilent Technologies 16048A
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Agilent Technologies 16048A.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Agilent Technologies 16048A, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Agilent Technologies 16048A. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Agilent Technologies 16048A ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Agilent Technologies 16048A, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Agilent Technologies 16048A hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Agilent Technologies 16048A verständlich ist.

Eine gänzliche Bedienungsanleitung von Agilent Technologies sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Agilent Technologies 16048A bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Agilent Technologies 16048A befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Agilent Technologies 16048A, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Agilent Technologies 16048A - die uns die ersten Schritte während der Nutzung von Agilent Technologies 16048A erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Agilent Technologies 16048A hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Agilent Technologies 16048A im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Agilent Technologies 16048A?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Agilent Technologies 16048A nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Agilent Technologies 16048A hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)