  • Emerson 5081T-FF/FI - page 1

    Model 5081T -FF/FI Instruction Sheet PN 51A-5081T -FF/FI/rev .A November 2005 Y our purchase from Rosemount Analytical, Inc. has result- ed in one of the finest instruments available for your p ar- ticular application. These instrument s have been designed, and tested to meet many national and interna- tional standards. Experience indicates that it ...

  • Emerson 5081T-FF/FI - page 2

    PHYSICAL SPECIFICA TIONS HOUSING: Epoxy-polyester painted over low-copper alu- minum. Neoprene O-rings on cover . 160.5 mm x 175.3 mm x 161.3 mm (6.3 in. x 6.9 in. x 6.4 in.) DIAMETER: 155.4 mm (6.1 in.) ELECTRICAL CONDUIT OPENINGS: 3/4 in. FNPT POWER SUPPL Y AND LOAD: A power supply voltage of 9 Vdc to 32 Vdc at 22 mA is required; Intrinsically Sa ...

  • Emerson 5081T-FF/FI - page 3

    3 3 MODEL 5081T -FF/FI SPECIFICA TIONS TRANSMITTER DISPLA Y DURING CALIBRA TION AND PROGRAMMING (FIGURE 1) 1. Continuous display of conductivity , % concentration, or custom reading. 2. Units: µS/cm, mS/cm, ppm, %, or none. 3. Current menu appears here. 4. Submenus, prompts, and diagnostic readings appear hear . 5. Commands available in each subme ...

  • Emerson 5081T-FF/FI - page 4

    4 U1 7 :! <[9:1 -E< / ! 922<U '"*#=22 /1[[! /9P<U <Q2U 4-SG U<[9:1 2D>[=T '""" 25.0C 12.00mA CAL key PROG key DIAG key HOLD key Model 5081T -FF/FI Process Display Screen µS/cm /-[5ES-U1 /1[[! /9P<U <1P<9S! " U1 7 :! -G6 U1 7 : G5<:[-A <1U>:! /V<= G12->[= PROGRAM Proc ...

  • Emerson 5081T-FF/FI - page 5

    MODEL 5081T -FF/FI DEF AUL T SETTINGS V ARIABLE NAME MNEMONIC F ACTOR Y SETTINGS CUSTOMER SETTINGS Program Menu T emperature UH OR Auto temperature compensation UDVUQ on ___________ Manual temperature UODP 25.0°C (overridden by auto) ___________ T emperature compensation algorithm /97: ( [LPHDS or P9P1 ) LInEAr ___________ Display GLTRND Y Measure ...

  • Emerson 5081T-FF/FI - page 6

    6 MODEL 5081T -FF/FI WIRING WIRING Wire sensor as shown below in Figure 2. Keep sensor wiring separate from power wiring. For best EMI/RFI protection, use shielded output signal cable in an earth-grounded metal conduit. Refer to the sensor instruction manual for more det ails. FIGURE 2. Wiring Model 5081T -FF/FI ...

  • Emerson 5081T-FF/FI - page 7

    7 7 MODEL 5081T -FF/FI WIRING WIRING THROUGH A JUNCTION BOX The sensor can be wired to the analyzer through a remote junction box (PN 23550-00). Wire the extension cable and sensor cable point-to-point. Refer to the sensor instruction manual for more details. Factory-terminated (PN 23747-00) and unterminated (PN 9200275) connecting cable are availa ...

  • Emerson 5081T-FF/FI - page 8

    8 MODEL 5081T -FF/FI INST ALLA TION INST ALLA TION UNP ACKING AND INSPECTION Inspect the shipping container . If it is damaged, cont act the shipper immediately for instructions. Save the box. If there is no apparent damage, unp ack the container . Be sure all items shown on the packing list are present. If items are missing, notify Rosemount Analy ...

  • Emerson 5081T-FF/FI - page 9

    9 ...

  • Emerson 5081T-FF/FI - page 10

    10 ...

  • Emerson 5081T-FF/FI - page 11

    11 ...

  • Emerson 5081T-FF/FI - page 12

    1 1 2 2 ...

  • Emerson 5081T-FF/FI - page 13

    13 ...

  • Emerson 5081T-FF/FI - page 14

    1 1 4 4 FM A 5/6/04 J. FLOCK 5/6/04 J. FLOCK .140 ±. 0 0 5 1.30 ±. 0 2 .650 ±. 0 15 .125 2.180 ±. 0 0 5 2.56 ±. 0 2 .120 ±. 0 15 Ø .125 R .25 5-6-04 89 33 4 4 FINISH:SILKSCREEN BLACK EPOX Y PAINT (BAKED). 5081-T-FI-67 MODEL 5081-T-FI HARDNES S BRINEL L 190 . BE ANNEALED & PASSIVATED. MAX IMUM STEEL .015+/-.005 THICK. MATERIAL TO 1 MATERI ...

  • Emerson 5081T-FF/FI - page 15

    1 1 5 5 5/6/04 J. FLOCK 5/6/04 J. FLOCK 70°C MA X CLASS III, DI V 1 CLASS II, DIV 1, GPS E-G CLASS I, DIV 1, GPS B-D HAZARDOU S AREA 3 R T D C O M 6 R C V S H L D 7 R C V C O M 8 R C V 9 D R V S H L D 1 0 D R V C OM 2 R T D S H L D 5 R T D I N 1 2 N C 1 3 N C 4 R T D S E N S E 1 R E S E R V E D 1 5 H T / F F ( - ) 1 6 H T / F F ( + ) 1 4 N C 1 1 D ...

  • Emerson 5081T-FF/FI - page 16

    1 1 6 6 FM B La, Lt OR Lo; OR Lc/Rc (La/Ra OR Lo/Ro) AND Li/Ri (L a/Ra OR Lo /Ro) 1. INSTALLATION SHOULD BE IN ACCORDANC E W ITH ANSI /ISA RP12.06 "INSTALLATION OF INTRIN SICALLY SAFE SYSTEMS FOR H AZARD OUS (C LASSIFIED) LOC ATIONS" (EXCE PT CHAPTER 5 FOR FISCO INSTALLATIONS) AND THE NATIONAL ELECTRICAL CODE (A NSI/NFPA 70) SECTIONS 504 ...

  • Emerson 5081T-FF/FI - page 17

    1 1 7 7 THIS DOCUMENT IS CERTIFIED BY REV REV REV REV REV REV REVISIONS NOT PERMITTED W/O AGENCY APPROVAL DIMENSIONS ARE IN INC HES REMOVE BURRS & SHA RP EDGES .020 MAX MACHINED FILLE T RADII .020 MAX - ITEM PART NO DESCRIPTION QT Y SCALE SIZE REV DW G NO BILL OF M AT ERIAL MATERIAL FINIS H DRAW N CHECKED PROJECT ENGR APVD NOTES: UNLESS OTHERW ...

  • Emerson 5081T-FF/FI - page 18

    1 1 8 8 Rosemount Analytical, Uniloc Divisi on 2400 Barran ca Pkwy Irvine, CA 92 606 UNLESS OTHERWISE SPECIFIED DIMENSIO NS ARE IN INCHES REMOVE BURRS & S HARP EDGES .02 0 MAX MACHINED F ILLET RADII .0 20 MAX FINISH MATERIAL PROJECT ENGR A PVD SIZE DWG NO. REV TITLE SCALE TYPE This document contains information propriet ary to Rosemount Analyti ...

  • Emerson 5081T-FF/FI - page 19

    1 1 9 9 9241514-01/ A DIME NSIONS ARE IN I NCHES REMOVE BURRS & SHA RP EDGES .020 MAX MACHINED FILLE T RADII .020 MA X - ITEM PART NO DESCRIPTION QT Y SCALE SIZE REV DW G NO BILL OF M AT ERIAL MATERIAL FINIS H DRAW N CHECKED PROJECT ENGR APVD NOTES: UNLESS OTHERW ISE SPECIFIED REV ECO NO RELEASE DATE LTR ECO REVISIONS DESCRIPTION BY DATE CHK Th ...

  • Emerson 5081T-FF/FI - page 20

    2 2 0 0 Rosemount Analytical, Uniloc Division 2400 Ba rranca Pk wy Irvine, CA 9 2 606 UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES REMOVE BURRS & SHARP EDGES .020 MAX MACHINED FILLET RADII .020 MAX FINISH MATER IAL PROJECT ENGR APV D SIZE DWG NO. REV TITLE SCALE TYPE This document contains information proprietary to Rosemount Analytical, ...

  • Emerson 5081T-FF/FI - page 21

    2 2 1 1 OVERVIEW This section gives general procedures for routine maintenance of the 5081-T transmitter . The transmitter needs almost no routine maintenance. TRANSMITTER MAINTENANCE Periodically clean the transmitter window with household ammonia or glass cleaner . The detector for the infrared remote controller is located behind the window at th ...

  • Emerson 5081T-FF/FI - page 22

    2 2 2 2 Replacement Part s for Model 5081-T -FF/FI T ransmitter Location in Shipping drawing PN Description Weight 1 23992-07 PCB stack consisting of the CPU (p art 3) and analog (part 4) boards, 1 lb/0.5 kg display board is not included, CPU and analog boards are factory- calibrated as a unit and cannot be ordered separately 2 23638-01 LCD display ...

  • Emerson 5081T-FF/FI - page 23

    2 2 3 3 QUICK-ST ART 1. On the Remote, press PROG , NEXT , ENTER. 2. Use the arrow buttons to select COnduc (conductivity), nAOH (Sodium Hydroxide 0-15%), HCL (Hydrochloric Acid 0-16%), H2SO4L (Sulfuric Acid 0-30%), H2SO4H (Sulfuric Acid 95-99.99%), or CuSt (custom curve) mode. Press ENTER. If you chose CuSt , continue with step 3. If you chose COn ...

  • Emerson 5081T-FF/FI - page 24

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .r ...

Produzent Emerson Kategorie TV Converter Box

Dokumente, die wir vom Produzenten des Geräts Emerson 5081T-FF/FI erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Emerson
- Bedienungsanleitungen 5081T-FF/FI
- Produktkarten Emerson
- Informationsbroschüren
- oder Energieetiketten Emerson 5081T-FF/FI
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Emerson 5081T-FF/FI.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Emerson 5081T-FF/FI, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Emerson 5081T-FF/FI. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Emerson 5081T-FF/FI ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Emerson 5081T-FF/FI, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Emerson 5081T-FF/FI hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Emerson 5081T-FF/FI verständlich ist.

Eine gänzliche Bedienungsanleitung von Emerson sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Emerson 5081T-FF/FI bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Emerson 5081T-FF/FI befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Emerson 5081T-FF/FI, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Emerson 5081T-FF/FI - die uns die ersten Schritte während der Nutzung von Emerson 5081T-FF/FI erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Emerson 5081T-FF/FI hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Emerson 5081T-FF/FI im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Emerson 5081T-FF/FI?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Emerson 5081T-FF/FI nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Emerson 5081T-FF/FI hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)