    Model 5081-P-FF/FI F OUNDA TION ® Fieldbus T wo-Wire pH/ORP T ransmitter Instruction Sheet PN 51A-5081P-FF-FI/rev .B October 2007 ESSENTIAL INSTRUCTIONS READ THIS P AGE BEFORE PROCEEDING! Y our purchase from Rosemount Analytical, Inc. has resulted in one of the finest instruments available for your particular application. These instruments have be ...

  • Emerson 5081-P-FF/FI - page 2

    MODEL 5081-P-FF/FI SPECIFICA TIONS SPECIFICA TIONS - GENERAL Case: Cast aluminum containing less than 6% magnesium. NEMA 4X (IP65), NEMA 7 (explosion-proof) Epoxy-polyester paint, Neoprene O-ring seals Dimensions: 6.3 in. x 6.9 in. x 6.4 in. (160 mm x 175 mm x 161 mm); diameter 6.1 in (155 mm) Conduit Openings: 3/4 in. FNPT Reference Impedance: T r ...

  • Emerson 5081-P-FF/FI - page 3

    3 MODEL 5081-P-FF/FI WIRING SENSOR WIRING AND SET -UP Wire sensor per Figure 1. Refer to the sensor instruction manual for more details. If your sensor has an integral preamplifier or you are using a remote preamplifier , you will need to change the pream- plifier location in the program menu. Please see Figure 14. Select “trAnS” for integral p ...

  • Emerson 5081-P-FF/FI - page 4

    4 MODEL 5081-P-FF/FI WIRING FIGURE 2. Exploded View of Model 5081-P-FF/FI FIGURE 3. Wiring Diagram — Sensor with Preamplifier DWG . NO. REV . 45081P30 A ...

  • Emerson 5081-P-FF/FI - page 5

    5 WIRING THROUGH A JUNCTION BOX WITH REMOTE PREAMPLIFIER MODEL 5081-P-FF/FI WIRING FIGURE 4. Wiring diagram — sensor Model 381+ with preamplifier Wire sensor as shown in Figure 5. Refer to the sensor instruction manual for more details. FIGURE 5. Wiring Through a Junction Box with Remote Preamplifier DWG . NO. REV . 45081P28 A DWG . NO. REV . 450 ...

  • Emerson 5081-P-FF/FI - page 6

    MODEL 5081-P-FF/FI INST ALLA TION 6 INST ALLA TION UNP ACKING AND INSPECTION Inspect the shipping container . If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unp ack the container . Be sure all items shown on the packing list are present. If items are missing, notify Rosemount Analyt ...

  • Emerson 5081-P-FF/FI - page 7

    7 FIGURE 8. FM Intrinsically Safe Inst allation Label ...

  • Emerson 5081-P-FF/FI - page 8

    8 FIGURE 7. FMRC Explosion-Proof Inst allation MODEL 5081-P-FF/FI HAZARDOUS AREA INST ALLA TION HAZARDOUS AREA INST ALLA TION ...

  • Emerson 5081-P-FF/FI - page 9

    9 FIGURE 9. FM Intrinsically Safe Inst allation (1 of 2) ...

  • Emerson 5081-P-FF/FI - page 10

    10 FIGURE 9. FM Intrinsically Safe Inst allation (2 of 2) ...

  • Emerson 5081-P-FF/FI - page 11

    11 FIGURE 10. CSA Intrinsically Safe Inst allation Label ...

  • Emerson 5081-P-FF/FI - page 12

    12 FIGURE 1 1. CSA Intrinsically Safe Inst allation (1 of 2) ...

  • Emerson 5081-P-FF/FI - page 13

    FIGURE 1 1. CSA Intrinsically Safe Inst allation (2 of 2) 13 ...

  • Emerson 5081-P-FF/FI - page 14

    FIGURE 12. A TEX Intrinsically Safe Inst allation Label 14 ...

  • Emerson 5081-P-FF/FI - page 15

    FIGURE 13. A TEX Intrinsically Safe Inst allation (1 of 2) 15 ...

  • Emerson 5081-P-FF/FI - page 16

    FIGURE 14. A TEX Intrinsically Safe Inst allation (2 of 2) 16 ...

  • Emerson 5081-P-FF/FI - page 17

    17 W/O AGENCY APPROVAL REVISIO NS NO T PE RMIT TED REV REV REV REV REV REV CERTIFIED BY THIS DOCUMENT IS °C Tamb= 70 FM A 5/6/04 J. FLOCK 5/6/04 J. FLOCK .140 .125 1.30 ±. 0 2 2X.650 ±.0 1 5 2.56 ±. 0 2 2.180 .120 ±. 01 5 Ø .125 R .25 OR SUITABIL ITY FOR DIV. 2. MAY IMPAIR INTRINSIC SAFET Y 5081-P-FI-67 MODEL GRPS. E, F & G GRPS. A,B,C,D, ...

  • Emerson 5081-P-FF/FI - page 18

    18 UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE I N INCHES REMOVE B URRS & SHA RP EDGES .020 MAX MACHINED F ILLET RADII .020 MAX FINISH MATERIAL PROJEC T ENGR A PVD SIZE DWG NO. REV TITLE SCALE TYPE CHECKED DRAWN THIS DWG CREATE D IN DATE DESCRIPTION BILL OF M ATERIAL QTY ITEM PART NO. RELEASE DATE ECO NO. REV LTR ECO DESCRIPTION BY DATE REVISION ...

  • Emerson 5081-P-FF/FI - page 19

    B FM ANY FM APPROVED ASSOCIATED APPARATUS ANY FM APPROVED ASSOCIATED APPARATUS ANY FM APPROVED ASSOCIATED APPARATUS ANY FM APPROVED TERMINATOR ANY FM APPROVED INTRINSICALLY SAFE APPARATUS ANY FM APPROVED TERM INA TOR ANY FM APPROVED TERMINA TOR ANY FM APPROVED INTRINSICALLY SAFE APPARATUS ANY FM APPROVED TERMINATOR ANY FM APPROVED TERMINA TOR ANY F ...

  • Emerson 5081-P-FF/FI - page 20

    20 IIC/ A,B,C,D,E,F,G IIB/ C,D,E,F ,G GROUPS 360 17.5 5 5081-P-FI 10 2.52 380 17.5 5 5081-P-FI FISCO PARAMETERS 5081-P-FI TABLE III MODEL NO. 10 5.32 Vmax (Vdc) Li (uH) Ci (nF) Pmax (W) Imax (mA) SUPPLY / SIGNAL TERM INALS TB 1-15, 16 La, Lt OR Lo; OR Lc/Rc (La/Ra OR Lo/Ro) AND Li/Ri (La/Ra OR L o/Ro) GROUPS GAS OUTPUT PA RAMETERS 25 21.69 (mH) La ...

  • Emerson 5081-P-FF/FI - page 21

    21 CLASS I, DIV. 2, GRPS A,B,C & D MODEL SA R LR34186 ROSEMO UNT ANALYT ICAL R WARNING-EXPLOSION H AZARD-DO NOT DISCONNECT WHILE CIRCUIT IS LIVE UNLESS AREA IS KNOWN TO BE NON- HAZARDOUS. WARNING-EXPLOSION HAZARD-SUBSTITU- TION OF COMPONENTS M AY IMPAIR SUITABILITY FOR CLASS I, DIV 2. CLASS I, GRPS B,C & D CLASS II, GRPS E, F & G CLASS ...

  • Emerson 5081-P-FF/FI - page 22

    22 Rosemount Analytical, Uniloc Division 2400 Ba rranca P kwy Irvine, CA 92606 UNLESS OTHERWISE SPECI FIED DIME NSIO NS AR E IN IN CHES REMOVE BUR RS & SHARP E DGES .020 M AX MACHINED FILLET RADII .020 MAX FINI SH MATE RIAL PROJE CT ENGR A PVD SIZE DWG NO. REV TITLE E P Y T E L A C S This document contains information proprietary to Rose mou nt ...

  • Emerson 5081-P-FF/FI - page 23

    23 9241502-01/ A DIMENSIONS ARE IN INCHES REMOVE BURRS & SHA RP EDGES .020 MAX MACHIN ED FILLET RADII .0 20 MAX - ITEM PART NO DESCRIPTION QT Y SCALE SIZE REV DW G NO BILL OF M ATERIAL MATERIAL FINIS H DRAW N CHECKED PROJECT ENGR APVD NOTES: UNLESS OTHERW ISE SPECIFIED REV ECO NO RELEASE DATE LTR ECO REVISIONS DESCRIPTION BY DATE CHK This docum ...

  • Emerson 5081-P-FF/FI - page 24

    24 Rosemount Analytical, Uniloc Division 2400 B arran ca Pkw y Irvin e, C A 92606 UN LESS OTH ERW ISE SPEC IFIE D DIMENSIONS ARE IN INCHES REMOVE BURR S & SHARP EDGES .020 M AX MACHINED FIL LET RADII .020 MAX FINISH MATE RIAL PROJECT ENGR AP VD SIZE DWG NO. REV TITLE SCALE TYPE This document c ontains information proprietary to Rosemount Analyt ...

  • Emerson 5081-P-FF/FI - page 25

    25 MODEL 5081-P-FF/FI INFRARED REMOTE CONTROLLER INFRARED REMOTE CONTROLLER (IRC) - KEY FUNCTIONS The infrared remote controller is used to calibrate and program the transmitter and to display diagnostic mes- sages. See figure below for a description of the function of the keys. Hold the IRC within 6 feet of the transmitter , and not more than 15 d ...

  • Emerson 5081-P-FF/FI - page 26

    26 MODEL 5081-P-FF/FI ST ART -UP - pH ST ART -UP CONFIGURA TION AND CALIBRA TION — pH T o perform a two-point buffer calibration, follow these step s: 1. Place the pH sensor in the first buffer solution. 2. On the remote, press CAL, ENTER, ENTER. 3. Use the arrow buttons to select the correct buffer value. Press ENTER. 4. Rinse the sensor and pla ...

  • Emerson 5081-P-FF/FI - page 27

    27 MODEL 5081-P-FF/FI ST ART -UP - ORP ST ART -UP CONFIGURA TION AND CALIBRA TION — ORP T o perform a one point standardization, follow these step s. 1. Place the ORP sensor in the ORP buffer solution. 2. On the Remote press CAL, ENTER, ENTER 3. Use the arrow buttons to adjust the value. Press ENTER 4. Press RESET MENU Sub-menu PROMPT Diag Messag ...

  • Emerson 5081-P-FF/FI - page 28

    MODEL 5081-P-FF/FI MENU MNEMONICS 5081-P DISPLA Y MANUAL TEXT MEANING DBMJcsBuF CALIbrAtE Calibration menu Tue S td S tandardize sensor cG2! 0! cG3 bF1 / bF2 Buffer 1 / Buf fer 2 TMPQF SLOPE Sensor slope Ipme HoLd Set hold GBVMu F AULt Set fault eQo dPn Set output dampening eJBHopTujd dIAGnoS tic Diagnostic submenu sPGGTu rOFFS t Reference offset J ...

Produzent Emerson Kategorie Satellite Radio

Dokumente, die wir vom Produzenten des Geräts Emerson 5081-P-FF/FI erhalten, können wir in mehrere Gruppen teilen. Unteranderem in:
- technische Zeichnungen Emerson
- Bedienungsanleitungen 5081-P-FF/FI
- Produktkarten Emerson
- Informationsbroschüren
- oder Energieetiketten Emerson 5081-P-FF/FI
Jede von ihnen ist wichtig, jedoch finden wir die wichtigsten Informationen für den Nutzer des Geräts in der Bedienungsanleitung Emerson 5081-P-FF/FI.

Die Dokumentengruppe, die als Bedienungsanleitungen bezeichnet wird, wird ebenfalls in detaillierte Arten geteilt, solche wie: Montageanleitungen Emerson 5081-P-FF/FI, Wartungsanleitungen, Kurzanleitungen oder Benutzeranleitungen Emerson 5081-P-FF/FI. Abhängig vom Bedarf, sollten Sie das Dokument finden, das Sie brauchen. In unserem Service können Sie sich die populärste Bedienungsanleitung des Produkts Emerson 5081-P-FF/FI ansehen.

Ähnliche Bedienungsanleitungen

Die komplette Bedienungsanleitung des Geräts Emerson 5081-P-FF/FI, wie sollte sie aussehen?
Die Bedienungsanleitung, auch bezeichnet als Benutzerhandbuch, oder einfach nur „Anleitung”, ist ein technisches Dokument, das dem Benutzer bei der Nutzung von Emerson 5081-P-FF/FI hilfreich sein soll. Die Bedienungsanleitungen werden in der Regel von technischen Schriftstellern geschrieben, aber in einer Sprache, die für alle Nutzer von Emerson 5081-P-FF/FI verständlich ist.

Eine gänzliche Bedienungsanleitung von Emerson sollte einige Grundelemente enthalten. Ein Teil von ihnen ist nicht so wichtig, wie z.B.: die Titelseite oder Autorenseiten. Die restlichen von ihnen jedoch, sollten Informationen liefern, die für den Nutzer von enormer Wichtigkeit sind.

1. Einführung und Hinweise, wie man sich in einer Bedienungsanleitung von Emerson 5081-P-FF/FI bewegt - Am Anfang jeder Bedienungsanleitung sollten wir Hinweise bezüglich der Nutzungsart eines bestimmten Ratgebers finden. In ihr sollten sich Informationen über die Lokalisierung des Inhaltsverzeichnisses von Emerson 5081-P-FF/FI befinden, FAQ oder über oft auftretende Probleme – also Stellen, die von den Benutzern in jeder Bedienungsanleitung am meisten gesucht werden
2. Inhaltsverzeichnis - Index aller Ratschläge bezüglich Emerson 5081-P-FF/FI, die wir im aktuellen Dokument finden
3. Ratschläge zur Nutzung der Grundfunktionen des Geräts Emerson 5081-P-FF/FI - die uns die ersten Schritte während der Nutzung von Emerson 5081-P-FF/FI erleichtern sollten
4. Troubleshooting - geordneter Tätigkeitslauf, der uns bei der Diagnose und als nächstes bei der Lösung wichtiger Probleme mit Emerson 5081-P-FF/FI hilft
5. FAQ - häufig gestellte Fragen
6. Kontaktdaten Informationen darüber, wo man Kontakt zum Produzenten / Service von Emerson 5081-P-FF/FI im bestimmten Land suchen kann, wenn es nicht gelingt, das Problem selbst zu lösen.

Haben Sie eine Frage bezüglich Emerson 5081-P-FF/FI?

Nutzen Sie das untere Formular

Wenn Sie mit Hilfe der gefundenen Bedienungsanleitung Ihr Problem mit Emerson 5081-P-FF/FI nicht gelöst haben, stellen Sie eine Frage, indem Sie das untere Formular nutzen. Wenn einer der Nutzer ein ähnliches Problem mit Emerson 5081-P-FF/FI hatte, ist es möglich, dass er mit Ihnen die Lösung teilen möchte.

Text vom Bild übertragen

Kommentare (0)