    NI PXI/PCI-5114 Specifications 8-Bit 250 MS/s Digitizer Unless otherwise noted, the followi ng conditions were used for each specification: • All filter settings • All impedance selections • Sample clock set to 250 MS/s T ypical values are representati ve of an av erage unit operating at room temperature. Specifications are subject to change ...

  • National Instruments NI PXI/PCI-5114 - page 2

    NI PXI/PCI-5114 Specifications 2 ni.com Power .............. .................... ................ .................... .................... .............14 Software .................. .................... .................... .................... ................... .. 15 Environment .............. .................... .................... ....... ...

  • National Instruments NI PXI/PCI-5114 - page 3

    © National Instruments Corporat ion 3 NI PXI/PCI-5 114 Specifications Voltage Levels Full Scale (FS) Input Range and Programmable V ertical Of fset 50 Ω 1 M Ω — Range (V pk-pk ) V ertical Of fset Range (V) Range (V pk-pk ) V ertical Offset Range (V) 0.04 ±0.8 0.04 ±0.8 0.1 ±0.8 0.1 ±0.8 0.2 ±0.8 0.2 ±0.8 0.4 ±0.8 0.4 ±0.8 1 ±6.5 1.0 ? ...

  • National Instruments NI PXI/PCI-5114 - page 4

    NI PXI/PCI-5114 Specifications 4 ni.com Figure 1. NI 5114 Frequency Response (T ypical) Bandwidth and Transient Response Bandwidth (–3 dB) Range (V pk-pk ) Bandwidth Rise/F all T ime, T ypical — All ranges except 0.04 125 MHz 2.8 ns 0.04 100 MHz 3.5 ns Bandwidth Limit Filter 20 MHz Noise Filter — A C Coupling * Cutof f (–3 dB), T ypical 12 ...

  • National Instruments NI PXI/PCI-5114 - page 5

    © National Instruments Corporat ion 5 NI PXI/PCI-5 114 Specifications Specif ication Va l u e Commen ts Spectral Characteristics Spurious Free Dynamic Range with Harmonics (SFDR), T ypical Range (V pk-pk ) 10 MHz, –1 dBFS input signal. Includes the 2 nd through the 5 th harmonics. Measured from DC to 125 MHz. 20 MHz bandwidth limit filter of f. ...

  • National Instruments NI PXI/PCI-5114 - page 6

    NI PXI/PCI-5114 Specifications 6 ni.com Horizontal Sample Clock Specification Va l u e Comments Sources Internal, Onboard Clock (i nternal VCXO)* External, CLK IN (front panel SMB connector) * Internal Sample Clock is locked to the Reference Clock or deri ved from the onboard VCXO. Onboard Clock (Internal VCXO) Sample Rate Range Real-T ime Sampling ...

  • National Instruments NI PXI/PCI-5114 - page 7

    © National Instruments Corporat ion 7 NI PXI/PCI-5 114 Specifications Phase-Locked Loo p (PLL) Reference Clock External Sample Clock Sources CLK IN (front panel SMB connector) — Frequenc y Range 50 MHz to 250 MHz Di vide by n decimation av ailable where 1 ≤ n ≤ 65,535. For more informati on about Sample Clock and decimation, refer to the NI ...

  • National Instruments NI PXI/PCI-5114 - page 8

    NI PXI/PCI-5114 Specifications 8 ni.com CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) T r igger Reference (Stop) T rigger Specification Va l u e Comments Input V oltage Range Sine wav e: 0.65 V pk-pk to 2.8 V pk-pk (0 dBm to 13 dBm) Square wav e: 0.2 V pk-pk to 2 .8 V pk-pk — Maximum Input Overload 7 V rms with | Peaks | ...

  • National Instruments NI PXI/PCI-5114 - page 9

    © National Instruments Corporat ion 9 NI PXI/PCI-5 114 Specifications Analog Trigger (Edge, Window, and Hysteresis Trigger Types) Sources CH 0 (front panel BNC connector) CH 1 (front panel BNC connector) TRIG (front panel BNC connector) — T rigger Level Resolution 8 bits (1 in 256) T rigger Level Range CH 0, CH 1 TRIG (External Trigger) — 100% ...

  • National Instruments NI PXI/PCI-5114 - page 10

    NI PXI/PCI-5114 Specifications 10 ni.com TRIG (External T rigger , Front Panel Connector) PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connectors) T ypes Specific Line Any Line Specific Field — Standards SDTV : M-NTSC, B/G-P AL, SECAM, M-P AL EDTV : 480i/59.94 fps, 480i /60 fps, 480p/59.94 Fps, 480p/60 Fps, 576i/50 fps, 576p/ ...

  • National Instruments NI PXI/PCI-5114 - page 11

    © National Instruments Corporat ion 11 NI PXI/PCI-5 114 Specifications As an Input (Trigger) Destinations Start T rigger (Acquisition Arm) Reference (Stop) T rigger Arm Referen ce Advance T rigger — Input Impedance 150 k Ω — V IH 2.0 V — V IL 0.8 V — Maximum Input Overload –0.5 V to 5.5 V — Maximum Frequency 25 MHz — As an Output (E ...

  • National Instruments NI PXI/PCI-5114 - page 12

    NI PXI/PCI-5114 Specifications 12 ni.com TClk Specifications National Instruments TClk synchronizati on method and the NI-TClk driver are used to align the sample clocks on any number o f SMC-based modules in a chassis. For more inform ation about TClk synchronizati on, refer to the NI-TClk Synchronization Help , wh ich is located within the N I Hi ...

  • National Instruments NI PXI/PCI-5114 - page 13

    © National Instruments Corporat ion 13 NI PXI/PCI-5 114 Specifications W aveform Specifications Specif ication Va l u e Comments Onboard Memory Size 8 MB per Channel Standard 8 megasamples per channel — 64 MB per Channel Option 64 megasamples per channel 256 MB per Channel Option 256 megasamples per channel Minimum Record Length 1 Sample — Num ...

  • National Instruments NI PXI/PCI-5114 - page 14

    NI PXI/PCI-5114 Specifications 14 ni.com Calibration Power Specification Va l u e Comments Self-Calibrati on Self-calibration i s done on soft ware command. The calibration corrects for gain, of fset, compensated 1 M Ω attenuat or , triggering, and timing adjustment errors fo r all input ranges. — External Calibration (Factory Calibration) The e ...

  • National Instruments NI PXI/PCI-5114 - page 15

    © National Instruments Corporat ion 15 NI PXI/PCI-5 114 Specifications Software Specification Va l u e Comm ents Dri ver Software NI PXI-5114 : NI-SCOPE 2.9 or later . NI PCI-5114 : NI-SCOPE 3.1 or later . NI-SCOPE is an IVI-complian t dri ver that allo ws you to configure, control, and calib rate the NI 5114. NI-SCOPE provides application program ...

  • National Instruments NI PXI/PCI-5114 - page 16

    NI PXI/PCI-5114 Specifications 16 ni.com Environment NI PXI-5114 Note T o ensure that the NI PXI-5114 cools ef fectiv ely , follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit. The NI PXI-5114 is intended for indoor use only . Specif ication Va l u e Comments Operating T emperature 0 ºC to +55 ºC ...

  • National Instruments NI PXI/PCI-5114 - page 17

    © National Instruments Corporat ion 17 NI PXI/PCI-5 114 Specifications NI PCI-5114 Note T o ensure that the NI PCI-5114 cools ef fectiv ely , follo w the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PCI-5114 kit. Also , to maximize airflo w and extend the life of the de vice, leave an y adjacent PCI slots empty . ...

  • National Instruments NI PXI/PCI-5114 - page 18

    NI PXI/PCI-5114 Specifications 18 ni.com Safety , Electromagnetic Compatibility , and CE Compliance Specification Va l u e Comm ents Safety The NI 5114 meets the requirements of the following standards of safety for electri cal equipment for measurement, control, and laboratory use: • IEC 61010 -1, EN 6101 0-1 • UL 61010-1, CAN/CSA-C22.2 No. 61 ...

  • National Instruments NI PXI/PCI-5114 - page 19

    © National Instruments Corporat ion 19 NI PXI/PCI-5 114 Specifications Physical Specification Va l u e Comments Dimensions NI PXI-5114 NI PCI-5114 — 3U, One slot, PXI/cPCI Module 21.6 × 2.0 × 13.0 cm (8.5 × 0.8 × 5.1 in.) 35.5 × 2.0 × 11.3 cm (13.4 × 0.8 × 4.4 in.) We i g h t 455 g (16 oz) 421 g (1 4.8 oz) — Front Panel Connectors Labe ...

  • National Instruments NI PXI/PCI-5114 - page 20

    National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more inform ation about Natio nal Instruments trademarks. Other product and company names mentioned herein are trademarks or trad e names of their respective companies. For patents covering Nation ...

Fabricante National Instruments Categoría Computer Hardware

Los documentos del dispositivo National Instruments NI PXI/PCI-5114 que obtenemos del fabricante se pueden dividir en varios grupos. Entre ellos están::
- dibujos técnicos National Instruments
- manuales de instrucciones NI PXI/PCI-5114
- hojas de producto National Instruments
- folletos informativos
- o etiquetas energéticas National Instruments NI PXI/PCI-5114
Todos son importantes, pero lo más importante desde el punto de vista del usuario de un dispositivo lo encontraremos en el manual de instrucciones National Instruments NI PXI/PCI-5114.

Un conjunto de documentos determinado como manual de instrucciones se divide también en tipos más detallados, tales como: instrucciones de montaje National Instruments NI PXI/PCI-5114, instrucciones de servicio, instrucciones cortas o instrucciones de usuario National Instruments NI PXI/PCI-5114. Dependiendo de la situación debes buscar el documento que necesitas. En nuestra página puedes consultar el manual de instrucciones más popular del producto National Instruments NI PXI/PCI-5114.

Manuales de instrucciones parecidos

El manual de instrucciones del dispositivo National Instruments NI PXI/PCI-5114, ¿cómo debería ser?
El manual de instrucciones, también determinado como el manual de usuario o simplemente instrucciones, es un documento técnico que tiene como objetivo ayudar a los usuarios a utilizar National Instruments NI PXI/PCI-5114. Las instrucciones normalmente las realiza un escritor técnico pero en un lenguaje comprensible para todos los usuarios de National Instruments NI PXI/PCI-5114.

El manual de instrucciones completo de National Instruments debe contener unos elementos básicos. Una parte de ellos es menos importante, como por ejemplo: la portada / la página principal o las páginas de autor. Sin embargo, lo demás debe facilitarnos la información más importante desde el punto de vista del usuario.

1. Introducción y pistas sobre cómo orientarse con el manual National Instruments NI PXI/PCI-5114 - Al principio de cada manual se deben encontrar las indicaciones acerca de la manera de usar un manual de instrucciones. Debe contener información sobre dónde encontrar el índice National Instruments NI PXI/PCI-5114, preguntas frecuentes o problemas más comunes – son los apartados más buscados por los usuarios de cada manual de instrucciones.
2. Índice - listados de todos los consejos acerca de National Instruments NI PXI/PCI-5114 que encontraremos en el documento
3. Consejos de uso de las funciones básicas del dispositivo National Instruments NI PXI/PCI-5114 - que deben facilitarnos los primeros pasos durante el uso de National Instruments NI PXI/PCI-5114
4. Troubleshooting - una secuencia sistematizada de acciones que nos ayudará a diagnosticar los problemas más importantes con National Instruments NI PXI/PCI-5114
5. FAQ - las preguntas frecuentes
6. Datos de contacto Información acerca de cómo encontrar los datos de contacto del fabricante / servicio de National Instruments NI PXI/PCI-5114 en cada país si no somos capaces de solucionar el problema por nuestra cuenta.

¿Tienes una pregunta acerca de National Instruments NI PXI/PCI-5114?

Utiliza el formulario que se encuentra abajo

Si con la ayuda del manual de instrucciones encontrado no has podido solucionar tu problema con National Instruments NI PXI/PCI-5114, haz una pregunta utilizando el formulario que se encuentra abajo. Si alguno de los usuarios ha tenido un problema parecido con National Instruments NI PXI/PCI-5114 probablemente compartirá una forma de solucionarlo.

Copia el texto de la imagen

Comentarios (0)