  • Fluke 116 - page 1

    ® September 2006 ©2006 Fluke Corporation. All rights reserved . 1 114, 115, 116, and 117 Digital Multimeters Calibration Information Introduction XW Warning To avoid electric shock or injury, do not perform the performance tests or calibration adjustment procedures unless qualified to do so. The information provided in this document is for the us ...

  • Fluke 116 - page 2

    114, 115, 116, and 117 Calibration Information 2 Safety Information "Warning" and "Caution" Statements A “ XW Warning" identifies hazardous conditions and actions that could cause bodily harm or death. A " W Caution" identifies conditions and actions that coul d damage the Meter, the equipment under test, or cau ...

  • Fluke 116 - page 3

    Digital Multimeters International Electrical Symbols 3 International Electrical Symbols Table 1 lists the international symbols that appear in this document and on the Meter. Table 1. Electrical Symbols Symbol Description Symbol Description B AC (Alternating Current) I Fuse F DC (Direct Current) T Double Insulated X Hazardous voltage W Important In ...

  • Fluke 116 - page 4

    114, 115, 116, and 117 Calibration Information 4 Accuracy Specifications Function Range Resolution Accu racy ± ([% of Reading] + [Cou nts]) Model [1] DC millivolts 600.0 mV 0.1 mV 0.5 % + 2 114, 115, 116, 117 DC Volts 6.000 V 60.00 V 600.0 V 0.001 V 0.01 V 0.1 V 0.5 % + 2 114, 115, 116, 117 DC, 45 to 500 Hz 500 Hz to 1 kHz Auto-V LoZ [2] True-rms ...

  • Fluke 116 - page 5

    Digital Multimeters Basic Maintenan ce 5 Accuracy Specifications (cont) Function Range Resolution Accu racy ± ([% of Reading] + [Cou nts]) Model [1] Hz (V or A input) [3] 99.99 Hz 999.9 Hz 9.999 kHz 50.00 kHz 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 % + 2 115, 116, 117 Notes: [1] Models listed in this column also refer to the “C” version of the m ...

  • Fluke 116 - page 6

    114, 115, 116, and 117 Calibration Information 6 HOLD MIN MA X RANGE 10 A FUSED COM V A Lo / H OFF Hz Hz AUTO-V LoZ Volt Alert A A V V mV VoltA le r t TRUE RMS MU L TIMETER 117 <.5 OK OK erc010f.emf Figure 1. Fuse Testing Replacing the Battery and Fuse XW Warning To avoid shock, injury, or damage to the Meter: • Remove test leads from the Mete ...

  • Fluke 116 - page 7

    Digital Multimeters Performance Tests 7 2. Remove the Meter from its holster. 3. Remove two screws from the case bottom. 4. Separate the case bottom from the case top. 5. Remove the fuse from its holder and replace with an 11 A, 1000 V, FAST fuse having a minimum interrupt rating of 17,000 A. Use only Fl uke PN 803293. 6. To re-assemble the Meter, ...

  • Fluke 116 - page 8

    114, 115, 116, and 117 Calibration Information 8 Table 2. Required Equipment (cont) Recommended Equipment Measu rement Function Accuracy 5500A Multi-product Calibrato r (or equivalent) Frequency (115, 116, and 117) 2 V, 50 kHz ± 0.025 % Fluke 80 AK K-type Thermocouple Adapter Accessory Temperature (116) K-type Thermocouple, mini-plug on b oth ends ...

  • Fluke 116 - page 9

    Digital Multimeters Performance Tests 9 2. Warm up the calibrator as required by its specifications. 3. Allow the temperature of the UUT to stabilize at room temperature (23 ° C ± 5 ° C [73 ° F ± 9 ° F] ). 4. Check the fuses and Battery, and replace them if necessary. Refer to “Testing the Fuses”, and “Replacing the Battery and Fuse”. ...

  • Fluke 116 - page 10

    114, 115, 116, and 117 Calibration Information 10 Table 3. DMM Performance Tests (cont) Display Reading Step Function Range Applied 114 [1] 115 [1] 116 [1] 117 [1] 18 6.000V 0V -0.002 to 0.002 19 6.000V 5V 4.973 to 5.027 20 60.00V 50V 49.73 to 50.27 21 600.0V 600V 596.8 to 603.2 22 L DC Volts 600.0V -600V -596.8 to -603.2 23 600.0 mV 6 mV, 45 Hz 5. ...

  • Fluke 116 - page 11

    Digital Multimeters Performance Tests 11 Table 3. DMM Performance Tests (cont) Display Reading Step Function Range Applied 114 [1] 115 [1] 116 [1] 117 [1] 41 VoltAlert Hi N/A N/A N/A Refer to steps 1 – 5 in the procedure below 42 VoltAlert Lo N/A N/A N/A Refer to steps 6 – 9 in the procedure below [1] If using a Fluke 9100 calibrator, the Calib ...

  • Fluke 116 - page 12

    114, 115, 116, and 117 Calibration Information 12 S T B Y 7 8 2 5 0 + / - 4 . . S C O P E O U T T R I G A U X A S E N S E A U X V N O R M A L V R T D T R U E R M S M U L T I M E T E R 1 1 7 V o l t A l e r t Center Meter top on Hi ter minal. Note that LED is RED Inser t Double Ba nana erc013f.emf Figure 4. VoltAlert Testing 8. Hold the Meter so tha ...

  • Fluke 116 - page 13

    Digital Multimeters Calibration Adjustment 13 2. Press g once to see the calibration counter. 3. Press g again to start the password entry. The Meter displays “ ?>>> ” 4. The Meter buttons indicated below represent the num bers 1 through 5 when entering or changing the password: K = 1 M = 2 q = 3 g = 4 Q = 5 5. Press 4 buttons to enter ...

  • Fluke 116 - page 14

    114, 115, 116, and 117 Calibration Information 14 Meter Buttons Used in the Calibration Steps When performing the Calibration Adjustment Procedure, the Meter buttons behave as follows. This may be of help determining why a calibration step is not accepted and for de termining the input value without referring to Table 4. Press and hold K to show th ...

  • Fluke 116 - page 15

    Digital Multimeters Calibration Adjustment 15 Notes Set the calibrator to Standb y prior to changing the function switch position and after completing adjustment of each function. If the calibration adjustment procedure is not properly completed, the Meter will not operate correctly. Table 4. Calibration Adjustment Steps Calibration Steps Rotary Sw ...

  • Fluke 116 - page 16

    114, 115, 116, and 117 Calibration Information 16 Table 4. Calibration Adjustment Steps (cont) Calibration Steps Rotary Switch Position 114 115 [1] 116 [1] 117 [1,2] Input Terminals Calibrator Source Value $ DC μ amps N/A N/A C/23 N/A + and COM 600 μ A, 0 Hz [1] Models listed in this column also refer to the “C” version of the model. For exam ...

  • Fluke 116 - page 17

    Digital Multimeters Replacement Parts 17 Replacement Parts Table 5 lists the Meter’s replacable parts identified in Figure 6. 6 12 10 23 20 17 21 9 7 5 8 1 3 4 2 11 15 4 PL 2 PL 13 19 14 18 22 16 erc14.emf Figure 6. Exploded View of Meter ...

  • Fluke 116 - page 18

    114, 115, 116, and 117 Calibration Information 18 Table 5. Replacable Parts List Item Description Part Number Qty. 1 LCD,FLUKE-11X,3.2V,T N,4-DIGIT,1/ 4-DUTY,1/3-BIAS,LEPTON 250 9955 1 2 CONNECTOR,ELASTOMERIC,.010 IN CT R,.218 IN HIGH,.090 IN THK,2.284 IN LONG,BULK 2534229 1 FLUKE-117-2006,BRACKET MASK, 117 2525608 1 FLUKE-117-2006-06,BRACKET MASK, ...

  • Fluke 116 - page 19

    Digital Multimeters Warranty 19 Table 5. Replaceable Parts List Item Description Part Number Qty. 19 BATTERY,PRIMARY,MNO2- ZN,9V,505MAH,6LR61,ALK ALINE, 17X26X48MM,BULK 614487 1 20 FLUKE 12-8004,SHOCK ABSORBER 878983 1 FLUKE-117-2005,TI LT STAND 2525594 1 21 FLUKE-117-2005,T ILT STAND, 11X China 2631071 1 FLUKE-117-2010,HOLSTER 252 5649 1 22 FLUKE- ...

  • Fluke 116 - page 20

    114, 115, 116, and 117 Calibration Information 20 ...

Fabricant Fluke Catégorie Thermometer

Les documents que nous recevons du fabricant de l'appareilFluke 116 peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Fluke
- manuels d’utilisations 116
- fiches produit Fluke
- dépliants
- ou étiquettes-énergie Fluke 116
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Fluke 116.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Fluke 116, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Fluke 116. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Fluke 116.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Fluke 116, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Fluke 116 par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Fluke 116.

Le manuel d’utilisation complet Fluke, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Fluke 116 - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Fluke 116, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lFluke 116 qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Fluke 116 - qui devraient nous aider dans les premières étapes lors de l'utilisation Fluke 116
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Fluke 116
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Fluke 116 dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Fluke 116?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Fluke 116, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Fluke 116 il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)