Summary
  • Emerson Process Management T-SERIES - page 1

    Micro Motion ® T -Series Sensor Inst allation Manual Inst allation Manual P/N 20002172, Rev . B July 2008 ...

  • Emerson Process Management T-SERIES - page 2

    ©2008, Micro Motion, Inc. All right s reserved. ELITE and ProLink are registered tr ademarks, and MVD and MVD Direct Connect are trademarks of Micro Motion, Inc., Boulder , Colorado. Micro Motion is a registered tra de name of Micro Motion, Inc., Boul der , Colorado. The Micro Motion and Emerson lo gos are trademarks and service ma rks of Eme rson ...

  • Emerson Process Management T-SERIES - page 3

    1 Before Y ou Begin Before Y ou Begin This manual describes ho w to install a Micro Motion ® T -Series sensor . The followi ng informat ion is provided in this manual: Customer service For technical assistance, phone the Micr o Motion Customer Service depa rtment: • In the U.S.A., phone 800-522-MASS (800-522-6277) (toll free) • In Canada and L ...

  • Emerson Process Management T-SERIES - page 4

    2 Before Y ou Begin Definitions The term MVD ™ transmitt er refers to the follo wing transmitter models: • Models 1500, 170 0, 2500, an d 2700 • Models 3500 and 3700 with sensor interf ace code 5 or 6 European installations This Micro Motion prod uct complies w ith all applicable European direct i ves when properly installed in accordance wit ...

  • Emerson Process Management T-SERIES - page 5

    3 Before Y ou Begin Figure 1 T -Series sensor with core processor Figure 2 T -Series sensor with integrally mounted Model 1700/2700 transmitter Approv al tag Flow direction arrow Core processor housing Calibration tag Process connection Approv al tag Flow direction arrow Core processor housing Calibration tag Process connection Model 1700/2700 tran ...

  • Emerson Process Management T-SERIES - page 6

    4 Determining a Location Figure 3 T -Series sensor with junction box Step 1 Determining a Location Choose a location for the sensor based on the requirements described in th is section. The foll o wing general guidelines can help you select an appropriate locatio n for the sensor . Full flow tubes For op timal performance, the sensor t ubes should ...

  • Emerson Process Management T-SERIES - page 7

    5 Determining a Locati on Figure 4 T -Series sensor ambient an d process temperat ure limits Hazardous area approv als may impose additional limits on ambient and process temperature. For the A TEX “T” rating, refer to the A TEX documentation shipped w ith the sensor or av ailable on the Micro Motion web site at www .micromotion.com . UL and CS ...

  • Emerson Process Management T-SERIES - page 8

    6 Determining a Location Maximum wiring distances If the transmitter is mounted remote ly from the sensor , the maximum di stance between the sensor and transmitter depends on cable type. See T able 2. Pipe run Micro Motion sensors d o not require a straight run of pipe upstream or downstream. Va lv e s After the sensor and transmitter ha ve been i ...

  • Emerson Process Management T-SERIES - page 9

    7 Orienting the Sensor Step 2 Orienting the Sensor The sensor will function proper ly in an y orientation if the sensor tube s remain f illed with process fluid. Micro Motion recommends orienting the sensor in a v ertical pipeline, with process flu id flo w ing upwa rd through the sensor (see Figure 5). Figure 5 Recommended se nsor orientation Self ...

  • Emerson Process Management T-SERIES - page 10

    8 Orientin g the Se nsor Figure 6 Eccentric redu cer • The gap between the ele ctronics housing and se nsor body shou ld be inspected period ically . Man ually clean this gap when necessary . • The ASME Bioprocess Equipment Standard (ASME BPE-19 97), section SC-3.12.1, includes recommendations re garding ho w to install process piping to ensure ...

  • Emerson Process Management T-SERIES - page 11

    9 Mounti ng the Sensor Step 3 Mounting the Sensor Use your commo n practices to minimize to rque and bendi ng load on process connections. T o reduce the risk of condensation or excessiv e moisture , the conduit opening should not point upward (if possible). The conduit opening of the junction box or core processo r can be rotated freely to facilit ...

  • Emerson Process Management T-SERIES - page 12

    10 Wiring Step 4 Wiring Hazardous area installations If you are installing the sensor in a hazardous location, verify that the hazardous classif ication information printed on the sensor tag matches the en vironm ent in which the sensor will be installed. Failure to comply with the requirements f or intrin sic safety in a hazardous area could resul ...

  • Emerson Process Management T-SERIES - page 13

    11 Wiring Figure 8 Micro Motion cable g land and he at shrink 4. For conn ection at the core processor housing, pr epare shielded cable as follo ws (for armored cable, omit steps d, e, f, and g): a. Strip 4 1/2 inches ( 114 mm) of c able jack et. b . Remove the clear wrap that is inside the cable jacket, and remo ve the filler material between the ...

  • Emerson Process Management T-SERIES - page 14

    12 Wiring Figure 10 Applying the heat shrink g. Position gland clamping inse rt so the interior end is flush with the heat shrink. h. Fold the cl oth shield or braid and drain wires o ve r the clamping insert and approximately 1/8 inch (3 mm) past the O-ring. Figure 11 Folding the cloth shield i. Install the gland body into the co re processor ho u ...

  • Emerson Process Management T-SERIES - page 15

    13 Wiring Figure 13 Connecting the wires at th e core processor 7. Reinstall and tighten the co re processor housing co ver . 8. Additional wiring instructions for the tran smitter can be found in the transmitter manual. Note: Never gr ound the 4-wire cable shield an d shield dr ain wir e(s) at the transmitter . Junction box to a 9-wire remote tr a ...

  • Emerson Process Management T-SERIES - page 16

    14 Grounding Step 5 Grounding The sensor can be grounde d via the piping if the joints in the pipeline are ground- bonded. If the sen sor is not grounded via the piping, connect a groun d wire to the in ternal or exter nal grounding scre w , which is located on the core processo r housing or junction box. Improper gr ounding can cause measurement e ...

  • Emerson Process Management T-SERIES - page 17

    15 Purge Fittings 4. Connect the supp ly of nitrogen or ar gon g as to the inlet pur ge connection or open inlet pur ge line. Leav e the outlet connec tion open. • Exercise caution to a voi d introducing dirt, moisture, rust, or other contaminants into the sensor case. • If the purge g as is heav ier than air (such as argo n), locate the inlet ...

  • Emerson Process Management T-SERIES - page 18

    16 Return Policy Return Policy Micro Motion procedu res must be followed when retu rning equipment. These procedures ensure le gal compliance with go vernment transpor tation agencies and help pro vide a safe working en vironment for Micro Motion employees. F ailure to follo w Micro Motion proced ures will result in your equipment being refused del ...

  • Emerson Process Management T-SERIES - page 19

    ...

  • Emerson Process Management T-SERIES - page 20

    Micro Motion Inc. USA Worldwide Headqu arters 7070 Winchester Circle Boulder , Colorado 80301 T +1 303-527-5200 +1 800-522-6277 F +1 303-530-8459 Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556 Micro Motion Japan Emerson Process Management 1-2-5, Higashi Shinagawa S ...

Fabricant Emerson Process Management Catégorie Stud Sensor

Les documents que nous recevons du fabricant de l'appareilEmerson Process Management T-SERIES peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Emerson Process Management
- manuels d’utilisations T-SERIES
- fiches produit Emerson Process Management
- dépliants
- ou étiquettes-énergie Emerson Process Management T-SERIES
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Emerson Process Management T-SERIES.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Emerson Process Management T-SERIES, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Emerson Process Management T-SERIES. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Emerson Process Management T-SERIES.

Manuel d’utilisation complet de l’appareil Emerson Process Management T-SERIES, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Emerson Process Management T-SERIES par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Emerson Process Management T-SERIES.

Le manuel d’utilisation complet Emerson Process Management, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Emerson Process Management T-SERIES - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Emerson Process Management T-SERIES, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lEmerson Process Management T-SERIES qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Emerson Process Management T-SERIES - qui devraient nous aider dans les premières étapes lors de l'utilisation Emerson Process Management T-SERIES
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Emerson Process Management T-SERIES
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Emerson Process Management T-SERIES dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Emerson Process Management T-SERIES?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Emerson Process Management T-SERIES, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Emerson Process Management T-SERIES il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)