Summary
  • National Instruments SCXI -1125 - page 1

    CVI ™ , LabVIEW ™ , Natio nal Instru ments ™ ,n i . c o m ™ ,N I - D A Q ™ , and SCXI ™ are trademarks of National Instruments Corporation. Product a nd company names mentioned herein are tr ademarks or trade names of their respective companies. 370271A-01 © Copyright 2000 National Instruments Corp. All rights reserved. December 2000 C ...

  • National Instruments SCXI -1125 - page 2

    SCXI-1125 Calibration Procedure 2 ni.com Equipment and Other T est Requirements This section describes the equipment, software, documentation, and test conditions required for calibrat ing the SCXI-1125. T est Equipment Calibration requires a high-precisi on voltage source with at least 50 ppm accuracy, a multiranging 5 1/2 digit digital multimeter ...

  • National Instruments SCXI -1125 - page 3

    © National Instruments Corpor ation 3 SCXI-1125 Calibration Procedur e Documentation The NI-DAQ Function Refe rence Online Help fi le and NI-DAQ User Manual for PC Compa tibles c o n t a i nd e t a i l e di n f o r m a t i o no nu s i n gt h e NI-DAQ driver. The funct ion reference online h elp file includes detaile d information on the driver fun ...

  • National Instruments SCXI -1125 - page 4

    SCXI-1125 Calibration Procedure 4 ni.com Setting Up Y our Module Complete the following steps to set up yo ur module for testing : 1. Install the SCXI-1125 in sl ot 1 of the SCXI chassis. 2. Connect a 68-to-68-pin cable between the S CXI module and the E Series D A Q de vice installed in the host computer . 3. Configure the hardware with Measuremen ...

  • National Instruments SCXI -1125 - page 5

    © National Instruments Corpor ation 5 SCXI-1125 Calibration Procedur e 7. Connect the calibrator to the appro priate analog input, starting with channel 0. Refer to Figure 1 to determine which pins on the 96-pin front conn ector correspond to t he positiv e and negati ve inputs of the specified channel. Fo r example, the positive input for channel ...

  • National Instruments SCXI -1125 - page 6

    SCXI-1125 Calibration Procedure 6 ni.com 13. Call SCXI_Scale . Set the following parameters: • SCXIChassisID — The v alue assigned b y MAX • moduleSlot — 1 • channel — The same channel number from step 7 • SCXIgain — The same gain value from step 9 • TBgain — 1 • DAQboard — The device number for the D A Q device • DAQChann ...

  • National Instruments SCXI -1125 - page 7

    © National Instruments Corpor ation 7 SCXI-1125 Calibration Procedur e • calOp — ND_SELF_CALIBRATE • s etOfCalConstants — ND_USER_EEPROM_AREA • calRefVolts — 0.0 4. Call AI_Configure to put the E Series device in bipolar mode. Set t he following parameters: • deviceNumb er — The v alue assigned b y MAX • chan — 0 • inputMode ...

  • National Instruments SCXI -1125 - page 8

    SCXI-1125 Calibration Procedure 8 ni.com • filterMode — 1 • cutoffFrequency — 10,000 9. Connect the calibrator to the analog input you w ant to adjust. Refer to Figure 1 to determine the pins on the 96-pin f ront connector that correspond to the positiv e and neg ati ve inputs of the specif ied channel. For e xamp le, the positi ve input fo ...

  • National Instruments SCXI -1125 - page 9

    © National Instruments Corpor ation 9 SCXI-1125 Calibration Procedur e 16. Record binary1 with volt1 and binary2 with volt2 for later use. 17. Repeat steps 7 through 16 for the remaining gain v alues you want to adjust on this channel. 18. Repeat steps 6 through 17 for the remaining chan nels you want to adjust. Y ou hav e determined the new calib ...

  • National Instruments SCXI -1125 - page 10

    SCXI-1125 Calibration Procedure 10 ni.com V erifying Adjusted V alues After completing the adjustment s to your SCXI-1125 module, it is important that you verify the analog input operation by repeating t he steps listed in the Veri fying Gain and Offset Errors section of this docu ment. Reverifying after adjustments ensures that your SCXI-1125 mod ...

  • National Instruments SCXI -1125 - page 11

    © National Instruments Corpor ation 11 SCXI-1125 Calibration Procedur e Front and Rear Signal Connector Diagrams Figure 1. Front Signal Connector Pin Assignments CHSGND +5 V RSVD CH7+ CH6+ CH5+ CH4+ CH3+ CH2+ CH1+ NC NC NC NC RSVD CH0+ 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 AB RSVD CJTEMP RSVD RSVD ...

  • National Instruments SCXI -1125 - page 12

    SCXI-1125 Calibration Procedure 12 ni.com Figure 2. Rear Signal Connector Pin Assignments SERCLK DIGGND SLOT0SEL* DAQD*/A SERDATIN OUTREF MCH7+ MCH6+ MCH5+ MCH4+ MCH3+ MCH2+ MCH1+ MCH0+ AOGND SCANCLK SYNC SERDATOUT DIGGND MCH7 – MCH6 – MCH5 – MCH4 – MCH3 – MCH2 – MCH1 – MCH0 – AOGND 49 50 47 48 45 46 43 44 41 42 39 40 37 38 35 36 33 ...

  • National Instruments SCXI -1125 - page 13

    © National Instruments Corpor ation 13 SCXI-1125 Calibration Procedur e Specifications The following table contains the sp ecifications you should use w hen verifying and calibrating y our SCXI-1125 module. If you r device has been calibrated within the last year, the test point value should fall between the Upper Limit and Lower Limit valu es. Ta ...

  • National Instruments SCXI -1125 - page 14

    SCXI-1125 Calibration Procedure 14 ni.com 200 – 0.025 – 0.024750 – 0.02468642 – 0 .02481358 250 0.02 0.019800 0.01985126 0.019 74874 250 0 0.000000 0.000022 – 0.000022 250 – 0.02 – 0.019800 – 0.01974874 – 0.0 1985126 500 0.01 0.009900 0.00992663 0.009 87337 500 0 0.000000 0.000012 – 0.000012 500 – 0.01 – 0.009900 – 0.00987 ...

Fabricant National Instruments Catégorie Car Stereo System

Les documents que nous recevons du fabricant de l'appareilNational Instruments SCXI -1125 peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques National Instruments
- manuels d’utilisations SCXI -1125
- fiches produit National Instruments
- dépliants
- ou étiquettes-énergie National Instruments SCXI -1125
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation National Instruments SCXI -1125.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation National Instruments SCXI -1125, manuels d’entretien, brefs manuels ou manuels de l’utilisateur National Instruments SCXI -1125. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit National Instruments SCXI -1125.

Manuel d’utilisation complet de l’appareil National Instruments SCXI -1125, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser National Instruments SCXI -1125 par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs National Instruments SCXI -1125.

Le manuel d’utilisation complet National Instruments, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel National Instruments SCXI -1125 - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières National Instruments SCXI -1125, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lNational Instruments SCXI -1125 qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil National Instruments SCXI -1125 - qui devraient nous aider dans les premières étapes lors de l'utilisation National Instruments SCXI -1125
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de National Instruments SCXI -1125
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service National Instruments SCXI -1125 dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de National Instruments SCXI -1125?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec National Instruments SCXI -1125, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec National Instruments SCXI -1125 il est probable qu’il a envie de partager la façon de le résoudre.

Réécrire le texte de l'image

Commentaires (0)