    ® 80 Series V Digital Multimeter Calibration Manual PN 2102915 September 2004, Rev.1, 2/05 © 2004-2005 Fluke Corporation, All ri ghts reserved. Printed in USA All product names are trademarks of their respective companies. ...

  • Fluke 80 Series V - page 2

    Lifetime Limited Warranty Each Fluke 20, 70, 80, 170 and 180 Series DMM will be fr ee from defects in material and workmanship for its lifetime. As used herei n, “lifetime” is defined as seven years after Fluke discontinues manufacturing the product, but the warranty period shall be at least ten years from the date of purchase. This warranty do ...

  • Fluke 80 Series V - page 3

    i Table of Contents Title Page Introduction ........................................................................................................ 1 Contacting Fluke ................................................................................................ 1 Precautions and Safety Information ................................................ ...

  • Fluke 80 Series V - page 4

    80 Series V Calibration Manual ii ...

  • Fluke 80 Series V - page 5

    iii List of Tables Table Title Page 1. Electrical Symbols ................................................................................................. 4 2. Model 87 AC Voltage Function Specifications ..................................................... 5 3. Models 83 and 88 AC Voltage Function Specifications .................................. ...

  • Fluke 80 Series V - page 6

    80 Series V Calibration Manual iv ...

  • Fluke 80 Series V - page 7

    v List of Figures Figure Title Page 1. Opening the Meter, Battery and Fuse Replacement ............................................... 10 2. Removing LCD Mask to Access LCD ................................................................... 11 3. Testing the Current Input Fuses ....................................................................... ...

  • Fluke 80 Series V - page 8

    80 Series V Calibration Manual vi ...

  • Fluke 80 Series V - page 9

    1 Introduction XW Warning To avoid shock or injury: • Read “Precautions and Safety Information” before performing the verification tests or calibration adjustment procedures documented in this manual. • Do not perform the verification tests or calibration adjustment procedures described in this manual unless you are qualified to do so. • ...

  • Fluke 80 Series V - page 10

    80 Series V Calibration Manual 2 XW Warning To avoid possible electric shock or personal injury, inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity. Replace damaged test leads before using the Meter. Precautions and Safety Information In this manual, a Warning identifies conditions and actions that p ...

  • Fluke 80 Series V - page 11

    Digital Multimeter Precautions and Safety Information 3 • When making electrical connections, connect the common test lead before connecting the live test lead; when disconnecting, disconnect the live test lead before disconnecting the common test lead. • Do not use the Meter if it operates abnormally. Protection may be impaired. When in doubt, ...

  • Fluke 80 Series V - page 12

    80 Series V Calibration Manual 4 Electrical Symbols Electrical symbols used on the Meter and in this manual are explained in Table 1. Table 1. Electrical Symbols B AC (Alternating Current) J Earth ground F DC (Direct Current) I Fuse X Hazardous voltage. P Conforms to European Union directives W Risk of Danger. Important information. See Manual. $ C ...

  • Fluke 80 Series V - page 13

    Digital Multimeter Specifications 5 664 to 600 V Overvoltage Category IV. UL listed to UL61010-1. Licensed by TÜV to EN61010-1. Detailed Specifications For all detailed specifications: Accuracy is given as ± ([% of reading] + [number of least significant digits] ) at 18° C to 28° C, with relative humidity up to 90 %, fo r a period of one year a ...

  • Fluke 80 Series V - page 14

    80 Series V Calibration Manual 6 Table 4. DC Voltage, R esistance, and Conductance Function Specifications Accuracy Function Range Resolution Model 83 Model 87 Model 88 L 6.000 V 60.00 V 600.0 V 1000 V 0.001 V 0.01 V 0.1 V 1 V ± (0.1 % + 1) ± (0.1 % + 1) ± (0.1 % + 1) ± (0.1 % + 1) ± (0.05 % + 1) ± (0.05 % + 1) ± (0.05 % + 1) ± (0.05 % + 1) ...

  • Fluke 80 Series V - page 15

    Digital Multimeter Specifications 7 Table 7. Capacitance and Diode Function Specifications Function Range Resolution Accuracy E 10.00 nF 100.0 nF 1.000 µ F 10.00 µ F 100.0 µ F 9999 µ F 0.01 nF 0. 1 nF 0.001 µ F 0.01 µ F 0.1 µ F 1 µ F ± (1 % + 2) 1 ± (1 % + 2) 1 ± (1 % + 2) ± (1 % + 2) ± (1 % + 2) ± (1 % + 2) G 3.000 V 0.001 V ± (2 % ...

  • Fluke 80 Series V - page 16

    80 Series V Calibration Manual 8 Table 11. Electrical Charact eristics of the Terminals Function Overload Protection 1 Input Impedance (nominal) Common Mode Rejection Ratio (1 k Ω unbalance) Normal Mode Rejection L 1000 V rms 10 M Ω < 100 pF > 120 dB at dc, 50 Hz or 60 Hz > 60 dB at 50 Hz or 60 Hz F mV 1000 V rms 10 M Ω < 100 pF & ...

  • Fluke 80 Series V - page 17

    Digital Multimeter Basic Maintenance 9 Basic Maintenance XW Warning To avoid possible electric shock or personal injury: • Remove the test leads and any input signals before opening the case or replacing the battery or fuses. • Repairs or servicing covered in this manual should be performed only by qualified personnel. Cleaning the Meter XW War ...

  • Fluke 80 Series V - page 18

    80 Series V Calibration Manual 10 W Caution To avoid damaging the Meter, the gasket that is sealed to the bottom case, and is between the two case halves, must remain with the case bottom. The case top lifts away from the gasket easily. Do not damage the gasket or attempt to separate the case bottom from the gasket. 4. Hold the Meter display side u ...

  • Fluke 80 Series V - page 19

    Digital Multimeter Basic Maintenance 11 5. Place the PCA onto the top shield so that the screw holes align. 6. Place the bottom shield onto the PCA and secu re the assembly with five Phillips-head screws. Ensure that the shields are tigh tly attached. Properly fitted shields are required for the Meter to perform to specifications. LCD LCD Mask ama0 ...

  • Fluke 80 Series V - page 20

    80 Series V Calibration Manual 12 Replacing the Battery Replace the battery with a 9-V battery (NEDA A1604, 6F22, or 006P). XW Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator ( b ) appears. If the display shows “ bAtt ” the Meter will not funct ...

  • Fluke 80 Series V - page 21

    Digital Multimeter Required Equipment 13 MIN MAX RANGE REL Hz % AutoHOLD Peak MIN MAX 4½ DIGITS 1 Second ˚ C/ ˚ F OFF mA A mV V V A A mA COM V 400mA FUSED 10A MAX FUSED A TRUE RMS MULTIMETER 87 V LO LO MIN MAX RANGE REL Hz % AutoHOLD Peak MIN MAX 4½ DIGITS 1 Second ˚ C/ ˚ F OFF mA A mV V V A A mA COM V 400mA FUSED 10A MAX FUSED A TRUE RMS MUL ...

  • Fluke 80 Series V - page 22

    80 Series V Calibration Manual 14 Table 13. Required Equipment Equipment Required Charact eristics Recommended Model Calibrator AC Voltage Range: 0 - 1000 V ac Accuracy: ± 0.12 % Frequency Range: 60 - 20000 Hz Accuracy: ± 3 % DC Voltage Range: 0 - 1000 V dc Accuracy: ± 0.012 % Current Range: 350 µ A - 2 A Accuracy: AC (60 Hz to 1 kHz): ± 0.25 ...

  • Fluke 80 Series V - page 23

    Digital Multimeter Performance Tests 15 Basic Operability Tests Refer to the following sections to test the basic operability of the Meter. Testing the Fuses Refer to “Testing the Fuses”. Testing the Display Turn the Meter on while holding down D to view all segments of the display. Compare the display with the appropriate examples in Figure 4 ...

  • Fluke 80 Series V - page 24

    80 Series V Calibration Manual 16 Table 13. Display Features (cont.) Number Feature Indication L b The battery is low. XW Warning: To avoid false readings, which could lead to possible electric shock or personal inju ry, replace the battery as soon as the battery indicator appears. M A, µ A, mA Amperes (amps), Microamp, Milliamp V, mV Volts, Milli ...

  • Fluke 80 Series V - page 25

    Digital Multimeter Performance Tests 17 Testing the Pushbuttons To test the pushbuttons 1. Turn the Meter rotary knob to J . 2. Press each button and note that the meter responds with a beep for each button press. 3. Press and hold B a second time to exit MIN MAX mode. Testing Meter Accuracy Perform the accuracy test steps in Table 15. Table 15. Ac ...

  • Fluke 80 Series V - page 26

    80 Series V Calibration Manual 18 Table 14. Accuracy Tests (cont.) Display Reading Step Test Function Range 5500A Output 83 and 88 87 23 600 Ω 330 Ω ( Use 2 wire Comp) 1 328.5 to 331.5 329.1 to 330.9 24 6 k Ω 3.3 k Ω (Use 2 wire Comp) 1 3.286 to 3.314 3.292 to 3.308 25 60 k Ω 33 k Ω 32.86 to 33.14 32.92 to 33.08 26 600 k Ω 330 k Ω 3 ...

  • Fluke 80 Series V - page 27

    Digital Multimeter Performance Tests 19 Table 14. Accuracy Tests (cont.) Display Reading Step Test Function Range 5500A Output 83 and 88 87 47 Max = 5.896 to 6.104 48 L (87 and 88 only) Peak Min/Max 6 V dc 8 Vpp, 2 kHz Sq. Wave, DC offset 2 V Min = -1.898 to -2.102 49 0 ° C -1.0 to 1.0 -1.0 to 1.0 50 m L (87 and 88 only) Temperature 3 100 ° C 98. ...

  • Fluke 80 Series V - page 28

    80 Series V Calibration Manual 20 6. Set up the function generator as follows: Frequency 900 kHz Amplitude 8.3 V Burst Mode yes Burst Count 1 Burst Rate 100 Hz Burst Phase -90 Degrees 7. Set the oscilloscope for 0.5 V/DIV @ 0.5 µ s/DIV. 8. Trigger the waveform on channel 2. 9. Adjust the amplitude of the function generator to produce a 3 VP-P tria ...

  • Fluke 80 Series V - page 29

    Digital Multimeter Performance Tests 21 INDUCTIVE PICKUP SPARK PLUG SIDE OSCILLOSCOPE FUNCTION GENERATOR CH1 CH2 10 RESISTOR WIRE LOOP (14 AWG) OUTPUT RED aad08f.eps Figure 5. Setup for Inductive Pickup Test ...

  • Fluke 80 Series V - page 30

    80 Series V Calibration Manual 22 aad09f.wmf Figure 6. Waveform for Inductive Pickup Test ...

  • Fluke 80 Series V - page 31

    Digital Multimeter Calibration Adjustment 23 Calibration Adjustment The Meter features closed-case calibration adjustment using known reference sources. The Meter measures the applied reference source, calculates correction factors and stores the correction factors in nonvolatile memory. The following sections present the features a nd Meter pushbu ...

  • Fluke 80 Series V - page 32

    80 Series V Calibration Manual 24 Restoring the Default Password If the calibration password is forgotten, the de fault password (1234) can be restored using the following steps. 1. While holding down B , turn the rotary knob from OFF to VAC . The Meter displays “ Z CAL ”. 2. Remove the Meter's top case. Leave the PCA in the bottom case. ( ...

  • Fluke 80 Series V - page 33

    Digital Multimeter 25 Meter Buttons Used in the Calibration Steps The Meter buttons behave as follows when performing the Calibration Adjustment Procedure. This may be of help determini ng why a calibration step is not accepted and for determining the input value without referring to Table 16. A Press and hold to show the measured value. The measur ...

  • Fluke 80 Series V - page 34

    80 Series V Calibration Manual 26 Notes Set the calibrator to Standby prior to changing the function switch position and or after completing adjustment of each function. If the calibration adjustment proce dure is not completed correctly, the Meter will not operate correctly. Table 16. Calibration Adjustment Steps Input Value Function (Switch Posit ...

  • Fluke 80 Series V - page 35

    Digital Multimeter Service and Parts 27 Service and Parts Replacement parts are shown in Table 17 and Figure 8. To order parts and accessories, refer to “Contacting Fluke”. Table 17. 80 Series V Final Assembly Ref Des Description Part Number Qty AC72 Alligator Clip, Black 1670652 1 AC72 Alligator Clip, Red 1670641 1 BT1 Battery, 9 V 2139179 1 B ...

  • Fluke 80 Series V - page 36

    80 Series V Calibration Manual 28 H2-4 (3) BT1 MP390-391 MP6 MP22 MP10-11 MP5 Holster MP15 Alligator Clips TM3 TM2 TL75 Test Lead Set Tilt Stand MP8 TM1 W F2 F1 MP14 S2 MP31 U5 CR6 MP2 MP4 MP81 MP9 MP41 J1-2 MP27-30 J3 H5-9 (5) MP66 AC72 MP13 BT2 aom015c.eps Figure 8. 80 Series V Final Assembly ...

Fabricant Fluke Catégorie Universal Remote

Les documents que nous recevons du fabricant de l'appareilFluke 80 Series V peuvent être divisés en plusieurs groupes. Ceux-ci sont, entre autres:
- dessins techniques Fluke
- manuels d’utilisations 80 Series V
- fiches produit Fluke
- dépliants
- ou étiquettes-énergie Fluke 80 Series V
Tous sont importants, mais les informations les plus importantes du point de vue de l'utilisation de l'appareil se trouvent dans le manuel d’utilisation Fluke 80 Series V.

Un groupe de documents appelé manuels d’utilisation est également divisé en types plus spécifiques, tels que: Manuels d’installation Fluke 80 Series V, manuels d’entretien, brefs manuels ou manuels de l’utilisateur Fluke 80 Series V. Selon vos besoins, vous devriez chercher le document dont vous avez besoin. Sur notre site, vous pouvez voir le manuel le plus populaire d’utilisation du produit Fluke 80 Series V.

Manuels d’utilsiation similaires

Manuel d’utilisation complet de l’appareil Fluke 80 Series V, quelle devrait-elle être?
Le manuel d’utilisation, également appelé le mode d’emploi, ou tout simplement le manuel, est un document technique destiné à aider à utiliser Fluke 80 Series V par les utilisateurs. Des manuels sont généralement écrits par un rédacteur technique, mais dans un langage accessible à tous les utilisateurs Fluke 80 Series V.

Le manuel d’utilisation complet Fluke, devrait inclure plusieurs éléments de base. Certains d'entre eux sont moins importants, tels que: la couverture / page de titre ou pages d'auteur. Cependant, la partie restante, devrait nous fournir des informations importantes du point de vue de l'utilisateur.

1. Introduction et des conseils sur la façon d'utiliser le manuel Fluke 80 Series V - Au début de chaque manuel, nous devrions trouver des indices sur la façon d'utiliser le document. Il doit contenir des informations sur l'emplacement de la table des matières Fluke 80 Series V, FAQ ou des problèmes les plus fréquents - les points qui sont les plus souvent recherchés par les utilisateurs de chaque manuel
2. Table des matières - index de tous les conseils pour lFluke 80 Series V qui peuvent être trouvés dans le document courant
3. Conseils sur la façon d'utiliser les fonctions de base de l’appareil Fluke 80 Series V - qui devraient nous aider dans les premières étapes lors de l'utilisation Fluke 80 Series V
4. Troubleshooting - séquence systématique des activités qui nous aideront à diagnostiquer et ensuite résoudre les principaux problèmes de Fluke 80 Series V
5. FAQ - questions fréquemment posées
6. Détails du contact Informations sur l'endroit où chercher le contact avec le fabricant / service Fluke 80 Series V dans un pays donné, si le problème ne peut être résolu par nous-mêmes.

Avez-vous une question à propos de Fluke 80 Series V?

Utiliser le formulaire ci-dessous

Si vous n’avez pas résolu votre problème avec Fluke 80 Series V, avec l'aide du manuel que vous avez trouvé, posez une question en utilisant le formulaire ci-dessous. Si un utilisateur a eu un problème similaire avec Fluke 80 Series V il est probable qu’il a envie de partager la façon de le résoudre.