  • SRS Labs SR560 - page 1

    MODEL SR560 LOW-NOISE PR EAM PLIFI ER 1290-D Reamw ood Av enue Sunnyv ale, CA 94 089 U .S.A. (408) 7 44-9040 Copyr ight © 1989, 199 0, 1997, 1 999 Stanford Res earch Systems All Rights Reserved . Revi sion 2.8 (12/2006) ...

  • SRS Labs SR560 - page 2

    ...

  • SRS Labs SR560 - page 3

    TABLE OF CONTENTS i Introduction and Setup Instrument Overview 1 Preparation for Use 2 Line Voltage 2 Line Fuse 2 Line Cord 2 Ventilation 2 Power-Up 2 Repackaging for Shipment 3 Use in Biomedical Applications 3 Warning Regarding Use with Photomultipliers 3 Accessories Furnished 3 Environmental Conditions 3 Symbols 4 Specifications SR560 Low Noise P ...

  • SRS Labs SR560 - page 4

    TABLE OF CONTENTS ii Calibration and Repair Offset Adjustment 17 Calibration 17 Front End Replacement 17 SR560 Offset Adjustment Procedure 18 Battery Replacement 18 Fuse Replacement 18 Noise Contours 19 Input Voltage Noise 20 Dynamic Reserve 20 Appendix A Remote Programming A-1 Introduction A-1 Commands A-1 Appendix B Noise Sources and Cures B-1 In ...

  • SRS Labs SR560 - page 5

    INTRODUCTION AND SETUP 1 Figure 1: SR560 Block Diagram INSTRUMENT OVERVIEW The SR560 architecture is diagrammed above. The instrument provides DC- coupled low-noise amplification of single- ended and true differential input signals at gains of 1 to 50,000. Two configurable R-C filters are provided to selectively condition signals in the frequency r ...

  • SRS Labs SR560 - page 6

    INTRODUCTION AND SETUP 2 independent operation. Rear panel banana jacks provide access to the internal regulated power supplies (or batteries) for use as a bias source. PREPARATION FOR USE **********CAUTION********* * This instrument may be damaged if operated with the LINE VOLTAGE SELECTOR card set for the wrong applied AC input source voltage or ...

  • SRS Labs SR560 - page 7

    INTRODUCTION AND SETUP 3 ROLLOFF bypassed HIGH-PASS 0.03 Hz, +6 dB/oct LOW-PASS 1 MHz, -6 dB/oct GAIN MODE High Dynamic Reserve GAIN 20, calibrated LISTEN ON DEVICE ADDRESS As per SW601 Repackaging for Shipment The original packing materials should be saved for reshipment of the SR560. If the original packing materials are not available, wrap the i ...

  • SRS Labs SR560 - page 8

    4 ...

  • SRS Labs SR560 - page 9

    SPECIFICATIONS 5 SR560 LOW-NOISE PREAMPLIFIER SPECIFICATIONS CHART Inputs Single-ended or true differential Impedance 100 M Ω + 25 pF, DC-coupled Maximum Inputs 1 VDC before overload; 3 V peak to peak max AC coupled; protected to 100 VDC Maximum Output 10 Vpp Noise <4 nV/ √ Hz at 1 kHz CMRR >90 dB to 1 kHz, decreasing by 6 dB / octave (20 ...

  • SRS Labs SR560 - page 10

    SPECIFICATIONS 6 ...

  • SRS Labs SR560 - page 11

    OPERATION AND CONTROLS 7 Figure 2: SR560 Front Panel FRONT PANEL OPER ATING SUMMARY The operation of the SR560 Low-Noise Preamplifier has been designed to be as simple as possible. The effect of each keypress on the front panel is reflected in the change of a nearby LED. The front panel LED’s will remain lighted at all times unless dip switch SW6 ...

  • SRS Labs SR560 - page 12

    OPERATION AND CONTROLS 8 common with chassis ground by connecting the "AMP GROUND" and "CHASSIS GROUND" banana jacks on the rear panel of the SR560. When connected to AC power, the chassis of the unit is always connected to the grounding conductor of the AC power cord. The inputs are protected to 100 VDC but the DC input should ...

  • SRS Labs SR560 - page 13

    OPERATION AND CONTROLS 9 information on using filters with the amplifier in AC coupled mode. Gain Mode The allocation of gain throughout the instrument is set using the GAIN MODE pushbutton The Gain Mode is displayed by two indicator LED’s: HIGH DYNAMIC RESERVE and LOW NOISE. For a given gain setting, a HIGH DYNAMIC RESERVE allocates the SR560&ap ...

  • SRS Labs SR560 - page 14

    OPERATION AND CONTROLS 10 Figure 3: SR560 Rear Panel REAR PANEL OPERATING SUMMARY The SR560 rear panel is pictured in Figure 3. Various interface and power connectors are provided, along with fuses and charger status LEDs. AC Power Input The power entry module contains the receptacle for the AC line cord and fuse. The line fuse should be a 1 A slow ...

  • SRS Labs SR560 - page 15

    OPERATION AND CONTROLS 11 turn off and the yellow "MAINTAIN" LED will be on full brightness. Blanking Input The blanking input accepts a TTL-level signal and grounds the amplifier signal path after the front end for as long as the input is held high. The response time of the blanking input is typically "on" 5 µs after the risin ...

  • SRS Labs SR560 - page 16

    OPERATION AND CONTROLS 12 • AVOID DEEP DISCHARGE Recharge the batteries after each use. The two-step fast-charge / trickle-charge operation of the SR560 allows the charger to be left on indefinitely. ALWAYS recharge the batteries immediately after the BATT indicator LED on the SR560 turns red. Built-in protection circuitry in the unit removes the ...

  • SRS Labs SR560 - page 17

    CIRCUIT DESCRIPTION 13 DIFFERENTIAL LOW-NOISE FRONT END Two high-impedance inputs A and B allow the instrument to operate in either single- ended or true differential modes. Relays K103 and K104 allow the inputs A and B to be individually grounded, while K101 selects AC or DC coupling. Inversion of the inputs is provided by relay K105. The input ca ...

  • SRS Labs SR560 - page 18

    CIRCUIT DESCRIPTION 14 adjustment is provided by this stage. U405B, half of an AD7528 dual 8-bit DAC is used to provide a ±5 volt offset voltage at the non-inverting input of U402. The front panel offset control also sums at this junction, and provides an offset voltage of ±5 V that is buffered by U407D. Following amplifier U402 is the other half ...

  • SRS Labs SR560 - page 19

    CIRCUIT DESCRIPTION 15 regulator that provides twice the current of the LM317 negative battery regulator. Charging is controlled by changing the set voltage of the regulators based on battery charge status. Flip-flop U703 determines whether the charge regulators will be set to 15.5 volts for a quick charge or 13.8 volts for a trickle or "maint ...

  • SRS Labs SR560 - page 20

    CIRCUIT DESCRIPTION 16 BATTERIES AND P.E.M. The batteries used in the SR560 are of sealed lead-acid construction. There are three 12 V, 1.9 amp-hour batteries, two of which serve as the positive power supply, and one of which serves as the negative power supply. Powering the SR560 alone, battery life should be greater than 20 hours. The batteries s ...

  • SRS Labs SR560 - page 21

    CALIBRATION AND REPAIR 17 OFFSET ADJUSTMENT The SR560's front-panel offset adjustment provides an easy way for the user to null the amplifier's DC offset. Use the COUPLING pushbutton to light the GND LED. Now, regardless of the SOURCE setting, the input to the amplifier is grounded internally. Insert a small screwdriver through the front- ...

  • SRS Labs SR560 - page 22

    CALIBRATION AND REPAIR 18 SR560 OFFSET ADJUSTMENT PROCEDURE Required equipment: Digital volt meter Oscilloscope 4 BNC cables 1 BNC tee Function generator Small slotted screwdriver Phillips screwdriver Remove the bottom cover of SR560 to expose the component side of pc board. 1. Turn off the SR560. Hold down the "Reset" button, and turn th ...

  • SRS Labs SR560 - page 23

    CALIBRATION AND REPAIR 19 The other two fuses are in-line with the batte ries . The se fuse s w ill blow i f the rea r panel ± 12 VDC supplies are shorted, or if the unit sources or draws ex cessive current to o r fro m the ba tteri es. NOISE CONTOURS The noise contours (show n upper right) plot the noise figure as a function of source impedance a ...

  • SRS Labs SR560 - page 24

    CALIBRATION AND REPAIR 20 INPUT VOLTAGE NOISE The amplifier's input voltage noise approximates that of a 1000 Ω resistor (about 4 nV/ √ Hz). For source impedances below 1000 Ω , the output noise will be dominated by the amplifier's input voltage noise. A typical amplifier has an input voltage noise vs. frequency as shown in the figu ...

  • SRS Labs SR560 - page 25

    CALIBRATION AND REPAIR 21 Gain Input Noise (nV/ √ Hz ) Maximum DR (dB 0) 1 60 0 2 40 0 5 25 6 10, LN 13 6 10, HDR 25 14 20, LN 11 6 20, HDR 25 20 50, LN 10 14 50, HDR 25 28 100, LN 4 14 100, HDR 20 34 200, LN 4 20 200, HDR 18 40 500, LN 4 28 500, HDR 15 48 1000, LN 4 34 1000, HDR 15 54 Gain Input Noise (nV/ √ Hz ) Maximum DR (dB 0) 2000, LN 4 4 ...

  • SRS Labs SR560 - page 26

    CALIBRATION AND REPAIR 22 ...

  • SRS Labs SR560 - page 27

    APPENDIX A A - 1 REMOTE PROGRAMMING Introduction The SR560 is equipped with a standard DB- 25 RS-232C connector on the rear panel for remote control of all instrument functions. The interface is configured as listen-only, 9600 baud DCE, 8-bit, no parity, 2 stop bits, and is optically isolated to prevent any noise or grounding problems. Up to four S ...

  • SRS Labs SR560 - page 28

    APPENDIX A A - 2 ...

  • SRS Labs SR560 - page 29

    APPENDIX B B - 1 NOISE SOURCES AND CURES Noise, random and uncorrelated fluctuations of electronic signals, finds its way into experiments in a variety of ways. Good laboratory practice can reduce noise sources to a manageable level, and the lock-in technique can be used to recover signals, which may still be buried in noise. Intrinsic Noise Source ...

  • SRS Labs SR560 - page 30

    APPENDIX B B-2 ways which these noise sources work their way into an experiment. Capacitive Coupling A voltage on a nearby piece of apparatus (or operator) can couple to a detector via a stray capacitance. Although C stray may be very small, the coupled in noise may still be larger than a weak experimental signal. Capacitive Noise Coupling To estim ...

  • SRS Labs SR560 - page 31

    APPENDIX B B-3 Resistive Coupling (or ‘Ground Loops’) Currents through common connections can give rise to noise voltages. Resistive Coupling Here, the detector is measuring the voltage across the experiment, plus the voltage due to the noise current passing through the finite resistance of the ground bus. This problem arises because we have us ...

  • SRS Labs SR560 - page 32

    APPENDIX B B-4 ...

  • SRS Labs SR560 - page 33

    SR560 COMPONENT PARTS LIST C-1 Front Panel Parts List REF. SRS part# VALUE DESCRIPTION D 1 3-00884-306 RED LED, Rectangular D 2 3-00012-306 GREEN LED, Rectangular D 3 3-00012-306 GREEN LED, Rectangular D 4 3-00012-306 GREEN LED, Rectangular D 5 3-00012-306 GREEN LED, Rectangular D 6 3-00012-306 GREEN LED, Rectangular D 7 3-00012-306 GREEN LED, Rect ...

  • SRS Labs SR560 - page 34

    SR560 COMPONENT PARTS LIST C-2 D 48 3-00004-301 1N4148 Diode J 1 1-00035-130 20 PIN DIL Connector, Male N 1 4-00651-425 270X9 Resi stor Network SIP 1/4W 2% (Common) N 2 4-00651-425 270X9 Resi stor Network SIP 1/4W 2% (Common) N 3 4-00651-425 270X9 Resi stor Network SIP 1/4W 2% (Common) N 4 4-00336-425 270X5 Resi stor Network SIP 1/4W 2% (Common) N ...

  • SRS Labs SR560 - page 35

    SR560 COMPONENT PARTS LIST C-3 C 112 5-00023-529 .1U Cap, Monolythic Ceramic, 50V, 20%, Z5U C 113 5-00100-517 2.2U Capacitor, Tantalum, 35V, 20%, Rad C 114 5-00100-517 2.2U Capacitor, Tantalum, 35V, 20%, Rad C 115 5-00100-517 2.2U Capacitor, Tantalum, 35V, 20%, Rad C 116 5-00005-501 150P Capacitor, Ce ramic Disc, 50V, 10%, SL C 201 5-00019-501 68P ...

  • SRS Labs SR560 - page 36

    SR560 COMPONENT PARTS LIST C-4 C 319 5-00033-520 47U Capacitor, Electrolytic, 16V, 20%, Rad C 320 5-00033-520 47U Capacitor, Electrolytic, 16V, 20%, Rad C 321 5-00031-520 220U Capacito r, Electrolytic, 16V, 20%, Rad C 322 5-00031-520 220U Capacito r, Electrolytic, 16V, 20%, Rad C 323 5-00232-520 470U Capacito r, Electrolytic, 16V, 20%, Rad C 324 5- ...

  • SRS Labs SR560 - page 37

    SR560 COMPONENT PARTS LIST C-5 C 804 5-00225-548 .1U AXIAL Capacitor, Cerami c, 50V,+80/-20% Z5U AX C 805 5-00225-548 .1U AXIAL Capacitor, Cerami c, 50V,+80/-20% Z5U AX C 806 5-00225-548 .1U AXIAL Capacitor, Cerami c, 50V,+80/-20% Z5U AX C 807 5-00225-548 .1U AXIAL Capacitor, Cerami c, 50V,+80/-20% Z5U AX C 808 5-00023-529 .1U Cap, Monolythic Ceram ...

  • SRS Labs SR560 - page 38

    SR560 COMPONENT PARTS LIST C-6 D 705 3-00011-303 RED LED, T1 Package D 706 3-00062-340 KBP201G/BR-81D Integrat ed Circuit (Thru-hole Pkg) D 707 3-00226-301 1N5822 Diode D 708 3-00226-301 1N5822 Diode D 709 3-00306-340 LM385BZ-2.5 Integrated Circuit (Thru-hole Pkg) D 710A 3-00 391-301 MBR360 Diode D 710B 3-00 391-301 MBR360 Diode D 711 3-00198-301 1 ...

  • SRS Labs SR560 - page 39

    SR560 COMPONENT PARTS LIST C-7 Q 703 3-00374-329 MTP20P06 Voltage Reg., TO-220 (TAB) Package Q 801 3-00376-329 MTP5N05 Voltage Reg., TO-220 (TAB) Package R 2 4-00616-453 49.9 Resistor, 2W, 1% R 101 4-00306-407 100M Resistor, Metal Film, 1/8W, 1%, 50PPM R 102 4-00306-407 100M Resistor, Metal Film, 1/8W, 1%, 50PPM R 103 4-00030-401 10 Resistor, Carbo ...

  • SRS Labs SR560 - page 40

    SR560 COMPONENT PARTS LIST C-8 R 212 4-00600-40 7 15.8K Res istor, Meta l Film , 1/8W , 1%, 50PPM R 213 4-00600-40 7 15.8K Res istor, Meta l Film , 1/8W , 1%, 50PPM R 214 4-00030-40 1 10 Resistor , Carbon F ilm , 1/4W , 5% R 215 4-00165-40 7 200 Resist or, Meta l Film , 1/8W , 1%, 50PPM R 216 4-00030-40 1 10 Resistor , Carbon F ilm , 1/4W , 5% R 21 ...

  • SRS Labs SR560 - page 41

    SR560 COMPONENT PARTS LIST C-9 R 403 4-00325-40 7 845 Resist or, Meta l Film , 1/8W , 1%, 50PPM R 404 4-00165-40 7 200 Resist or, Meta l Film , 1/8W , 1%, 50PPM R 405 4-00030-40 1 10 Resistor , Carbon F ilm , 1/4W , 5% R 406 4-00030-40 1 10 Resistor , Carbon F ilm , 1/4W , 5% R 407 4-00030-40 1 10 Resistor , Carbon F ilm , 1/4W , 5% R 408 4-00317-4 ...

  • SRS Labs SR560 - page 42

    SR560 COMPONENT PARTS LIST C-10 R 707 4-00042-401 15K Resistor, Carbon Film, 1/4W, 5% R 708 4-00376-407 2.87K Resistor, Metal Film, 1/8W, 1%, 50PPM R 709 4-00169-407 249 Resistor, Metal Film, 1/8W, 1%, 50PPM R 710 4-00058-401 220K Resistor, Ca rbon Film, 1/4W, 5% R 711 4-00035-401 10M Resistor, Carbon Film, 1/4 W, 5% R 712 4-00035-401 10M Resistor, ...

  • SRS Labs SR560 - page 43

    SR560 COMPONENT PARTS LIST C-11 U 101 3-00371-340 DG444 Integrated Circuit (Thru-hole Pkg) U 102 3-00382-340 OPA37 Integrated Circuit (Thru-hole Pkg) U 103 3-00090-340 LF411 Integrated Circuit (Thru-hole Pkg) U 104 3-00385-340 74HC4053 Integrated Circuit (Thru-hole Pkg) U 105 3-00297-340 LT1028 Integrated Circuit (Thru-hole Pkg) U 106 3-00246-340 N ...

  • SRS Labs SR560 - page 44

    SR560 COMPONENT PARTS LIST C-12 U 801 3-00112-329 7805 Voltage Reg., TO-220 (TAB) Package U 802 3-00307-340 LM2940T-10 Integrated Circuit (Th ru-hole Pkg) U 803 3-00090-340 LF411 Integrated Circuit (Thru-hole Pkg) U 804 3-00250-340 PS2501A-2 Integrated Circuit (Thru-hole Pkg) Z 0 0-00014-002 6J4 Power Entr y Hardware Z 0 0-00017-002 TRANSCOVER Powe ...

  • SRS Labs SR560 - page 45

    SR560 COMPONENT PARTS LIST C-13 Z 0 0-00666-050 23" #18 WHITE Wire #18 UL1007 Stripped 3/8x3/8 No Tin Z 0 0-00667-050 20" #18 WHITE Wire #18 UL1007 Stripped 3/8x3/8 No Tin Z 0 1-00124-178 4 PIN Connector Housing, Plug Z 0 1-00125-179 4 PIN Connector Housing, Rece ptacle Z 0 1-00126-176 .062" DIAM Terminal, Male Z 0 1-00127-177 .062&q ...

  • SRS Labs SR560 - page 46

    SR560 COMPONENT PARTS LIST C-14 ...

Fabbricante SRS Labs Categoria Stereo Amplifier

I documenti che otteniamo dal fabbricante del dispositivoSRS Labs SR560 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici SRS Labs
- manuali per l’uso SR560
- schede prodotto SRS Labs
- opuscoli
- o etichette energetiche SRS Labs SR560
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso SRS Labs SR560.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione SRS Labs SR560, manuali di manutenzione, brevi manuali o manuali utente SRS Labs SR560. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto SRS Labs SR560.

Manuali per l’uso simili

Manuale completo del dispositivo SRS Labs SR560, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare SRS Labs SR560 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti SRS Labs SR560.

Il manuale per l’uso completo SRS Labs, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale SRS Labs SR560 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice SRS Labs SR560, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per SRS Labs SR560 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo SRS Labs SR560 - che ci dovrebbe aiutare nei primi passi durante l'uso SRS Labs SR560
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del SRS Labs SR560
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio SRS Labs SR560 in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su SRS Labs SR560?

Utilizza il modulo sottostante

Se non hai risolto il problema con SRS Labs SR560 con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con SRS Labs SR560 è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)