Manuale per l’uso Continental Scientific Refrigerators and Freezers

36 pagine 6.11 mb
Scarica

Vai alla pagina of 36

Summary
  • Continental Scientific Refrigerators and Freezers - page 1

    Operation, Installation and Instruction Manual for Reach-In Refrigerators and Freezers Cont inental Scient ific A Division of National Refrigeration & Air Conditioning Products, Inc. 539 Dunksferry Road · Bensalem, PA 19020-5908 215-244-1400 · 1-800-523-7138 · Fax: 215-244-9579 ...

  • Continental Scientific Refrigerators and Freezers - page 2

    Operators Manual Table of Contents Page RECEIVING YOUR NEW M ODEL........................................................................ 1 GENERAL INFORMATION AND IMPORTANT OPERAT ING FACTS ................. 1 UNCRATING YOUR NE W MODEL ...................................................................... 2 INSTALLATION AND LOCAT ION ............. ...

  • Continental Scientific Refrigerators and Freezers - page 3

    TROUBLESHOOTING AND SER VICING GUIDE .............................................. 22 W IRING DIAGRAMS .......................................................................................... 26 ...

  • Continental Scientific Refrigerators and Freezers - page 4

    RECEIVING YOUR NEW M ODEL Congratul ations on your r ecent purchase of Continental Scienti fic superior laboratory /pharmacy equipment products! W hen your shipment ar riv es, please thoroughly ex amine the shipping crate for any punctures, dents, or sig ns of roug h handling. It i s in y our best inter est to partiall y remov e or open the shippin ...

  • Continental Scientific Refrigerators and Freezers - page 5

    and sanitation, all models are li sted under the reexami nation ser vi ce of underw riter's l aboratories. UNCRA TING YOUR NEW MO DEL The shipping container shoul d remain on y our model as pr otection against dents or scratches whil e transporting it to the actual set- up location. Remov e the shipping container only at the last possible mome ...

  • Continental Scientific Refrigerators and Freezers - page 6

    efficiency. Important Note: To assure suf ficient air supply and circulation to the condensing unit, a minimum clearance of 12" above th e grill and 3" on each side and back of the cabinet must be prov ided (see figure 1) . If necessary, special v enting or air supply ducts must be installed by the instal ler for this purpose. Do not at a ...

  • Continental Scientific Refrigerators and Freezers - page 7

    approxi m ately 700 pounds of product ( 35 x 20) and assuming the refriger ator itself weig hs 300 pounds, the total combined wei ght of cabinet and product is approxi m ately 1000 pounds. Therefore, the floor in this ex ample must be capable of supporting up to 1000 pounds. INSTA LLING LEGS A ND LEVELING REA CH-IN M ODELS Your new reach-in model i ...

  • Continental Scientific Refrigerators and Freezers - page 8

    Your roll -in model comes with one cart r amp per door opening . T o install the ramp(s), simply open the cabinet door and positi on the ramp slots onto the screw s located along the cabi net bottom f ront. If the door wiper rubs along the ramp to the extreme where the door w ill not self -close , you will need to adjust the door (see "door ad ...

  • Continental Scientific Refrigerators and Freezers - page 9

    FIGURE 3 CA STERS MUST BE TIGHT TO CA BINE T BOTTOM INSTA LLI NG CONDENSA TE EVAPORA TOR No floor drains or plumbing connecti ons are req uired since al l models use an automatic condensate w ater ev aporating system. All models are supplied wi th an electric condensate evaporator pan and mounting bracket packed in the accessor y carton. To install ...

  • Continental Scientific Refrigerators and Freezers - page 10

    FIGURE 4 CONDENSA TE EVA PORA TOR INS TA LLATION REMO VA L OF DOORS A ND DOOR A DJUSTM ENT During installation, it may become necessary to remove the cabi net doors to facilitate passage through narrow doorways or hallw ays. To remove a door, pry off all hing e covers using a shar p tool or knife (see fig ure 5). Sw ing the door to the open-door po ...

  • Continental Scientific Refrigerators and Freezers - page 11

    FIGURE 5 DOOR A DJUSTME NT A ND REMOV A L REMO VA L OF GRILL To remove the grill, loosen all four (4) grill mounting screw s located on the backside of the grill one or two turns. Simply lift grill up off of its mo unting screw s and out. To replace the grill, line up the gr ill m ounting screws w ith the keyhole slots located on the cabinet body, ...

  • Continental Scientific Refrigerators and Freezers - page 12

    cleaning may result in a streaking or staining of the metal. Complete cleaning procedur es and precautions are l isted in the "per iodic cleani ng procedure" under the mai ntenance section. STA RT-UP PROCEDURE ELECTRICA L CONNECTI ONS To insure proper oper ation, your new model must be connected to an i ndivi dual circuit that can suppl y ...

  • Continental Scientific Refrigerators and Freezers - page 13

    the condensing unit is 208- 230 volts i n which the wi ring includes a neutral and a mechanical ground. This w iring should be connected to the appropr iate power source by a q ualified electrician and must conform to all local electr ical codes. SPECIA L VOLTA GE CONNECTIONS W hen models are ordered from the factory w ith special, optional v oltag ...

  • Continental Scientific Refrigerators and Freezers - page 14

    condensing unit, and electrical hook-up should be per f ormed by q ualified refriger ation personnel of a competent refrigeration company only. OPERA TION All cabinets must be g iven sufficient time to reach normal operating temper ature before placing any product inside. Refriger ators are desig ned to maintain an ideal cabinet temperature of 3.3` ...

  • Continental Scientific Refrigerators and Freezers - page 15

    and turn the needle to the desired temperature usi ng y our finger s. Replace the lens cover and recheck the calibrati on. DIGITAL THERMOMETER For a digital thermometer, note how many degrees the digital display needs to be increased or decreased. Pr ess twice directly on the myl ar display towards the rig ht center using y our finger. After pressi ...

  • Continental Scientific Refrigerators and Freezers - page 16

    FIGURE 6 THERMOMETER CA LIBRATION REFRIGERA TOR SYSTEM A ND ADJUSTMENT All self-contai ned refriger ators are desi gned and factory set to maintain an aver age cabinet temperature of 4`c. The temperatur e control is accessible f rom the top of the electrica l console bo x located on the cabinet to p behind the fron t grill (see figure 6). If an adj ...

  • Continental Scientific Refrigerators and Freezers - page 17

    compressor is not oper ating dur ing an off-cycle. Do not set the thermostat too cold where the cabine t tem perature will f all below 2`c becau se the evapo rator will becom e b locked by ice since th e comp ressor of f -cycle will be considerably shortened. This will result in loss of produc t stored within the cabinet and require servi ce to def ...

  • Continental Scientific Refrigerators and Freezers - page 18

    cabinet temperature. On l ow- temperature freez ers, place a screw driv er into the right ther mostat adjustment screw ( do not turn the left di fferent ial adjustment screw) and turn clockwise for a colder cabinet temperatur e or counterclockwise for a warmer cabinet temperature. Fur ther adjustments out of the factory desig n temperature rang e m ...

  • Continental Scientific Refrigerators and Freezers - page 19

    FIGURE 8 DEFROST TIMER SETTING EVA PORATOR A SSEMBLY All refrig erators and freezers have an easi ly accessible, easily serviceabl e, performance rated, f orced-air evaporator assembly w hich utilizes a plasticiz ed fin coil for ex tended life (see figur e 9). LOW-PROFILE, CEILIN G-MOUNT EVAPORATOR ASSEMBLY The low-pr of ile evapor ator system is c ...

  • Continental Scientific Refrigerators and Freezers - page 20

    FIGURE 9 LOW-PROFILE,CEILING- MOUNT EVA PORA TOR A SSEMBLY INTERIOR REA CH-IN A CCESSORIES The standard interior accessory packag e that is supplied from the f actory wi th y o ur reach-in r efrigerator or freezer consi sts of standard pil aster str ips w ith pilaster clips (four (4)clips per shelf), three ( 3 ) epoxy coated shelves per section, an ...

  • Continental Scientific Refrigerators and Freezers - page 21

    the shelf on the clips as show n in figure 10. The standard pil aster and clip ar e shown i n figure 10, and the optional heavy -duty pilaster and clip are shown in figure 11. Important Note: W hen loading shel ves with product, al low space between row s of product for proper air ci rculation, and do not load pr oduct to block back wall (leave at ...

  • Continental Scientific Refrigerators and Freezers - page 22

    FIGURE 11 OPTIONA L HEA VY-DUTY PILA STER FIGURE 12 OPTIONA L A NGLE PA N SLIDES SA FETY PRECA UTIONS The following saf ety precautions shoul d be followed when operati ng any appliances: ¬ Alway s disconnect the power cor d before attempting to w ork on or clean - 19 - ...

  • Continental Scientific Refrigerators and Freezers - page 23

    any eq uipment. ¬ Disconnect the power cord w hen the appliance will be idled for a long period of time. ¬ Do not attempt to serv ice this unit yourself as remov ing any cover s may cause exposure to dang erous voltage. ¬ Alway s route the power cor d so that it is not l ikely to be walked on or pinched by other appliances. Nev er use extension ...

  • Continental Scientific Refrigerators and Freezers - page 24

    contact with stai nless steel. 3. Tincture of iodine, or iron shoul d not come in contact with stainless steel. These solutions, whic h cause stainless steel to di scolor, should be ri nsed off immediately if contact occurs. 4. Gritty, hard abrasives will mar the f inish of stainless steel an d alumin um and are not recommended. SLIDING A ND HINGED ...

  • Continental Scientific Refrigerators and Freezers - page 25

    this cleaning process. Af ter cleaning , restore electrical ser vic e to your model . PA RTS AND SERVICE Al w a ys provide the cabi net model and serial number (l ocated on the data plate on the inside ri ght w all of the cabinet) w henever contacting the factory or your dealer regardi ng questi ons or when ordering parts. M O D E L # SERIA L # ___ ...

  • Continental Scientific Refrigerators and Freezers - page 26

    PROBLEM PROBABLE CA USE CORRECTION Condensing unit will not start - no hum. 1.LINE DISCONNECTED, SW ITCH OPEN. 2.FUSE REMOVED OR BLOW N. 3.OVERLOAD PROTECTOR BLOW N. 4.CONTROL "OFF" DUE TO COLD LOCATION. 5.CONTROL STUCK IN OPEN POSITION. 6.W I RING IMPROPER OR LOOSE. 1.CLOSE START OR DISCONNECT SW ITCH. 2.REPLACE FUSE. 3.DETERM INE REASON ...

  • Continental Scientific Refrigerators and Freezers - page 27

    PROBLEM PROBABLE CA USE CORRECTION Condensing uni t runs, but short cycl es on: 1.OVERLOAD PROTECTOR. 2.THERM OSTAT. 3.HIGH PRESSURE CUT-OUT DUE TO: (a) INSUFFICIENT AIR SUPPLY. (b) OVERCHARGE. (c) AIR IN SYSTEM . 4.LOW PRESSURE CUT-OUT DUE TO: (a) VALVE LEAK. (b) UNDERCHARGE. (c) RESTRICTION IN EXPANSION DEVICE. 1.SEE PROBLEM # 3. 2.DIFFERENTIAL M ...

  • Continental Scientific Refrigerators and Freezers - page 28

    PROBLEM PROBABLE CA USE CORRECTION Product zone temperature too hig h. 1.CONTROL SETTING TOO HIGH. 2.INADEQUATE AIR CIRCULATION. 3.DIRTY CONDENSER 1.ADJUST T-ST AT 2.REARRANGE PRODUCT LOAD TO IMPROVE AIR CIRCULATION. 3.CLEAN CONDENSER COIL Suction line frosted or sweating . 1.OVERCHARGE OF REFRIGERANT. 2.EVAPORATOR FAN NOT RUNNING. 3.EXPANSION VALV ...

  • Continental Scientific Refrigerators and Freezers - page 29

    WIRING DI A G RA M S - 26 - ...

  • Continental Scientific Refrigerators and Freezers - page 30

    - 27 - ...

  • Continental Scientific Refrigerators and Freezers - page 31

    - 28 - ...

  • Continental Scientific Refrigerators and Freezers - page 32

    - 29 - ...

  • Continental Scientific Refrigerators and Freezers - page 33

    - 30 - ...

  • Continental Scientific Refrigerators and Freezers - page 34

    - 31 - ...

  • Continental Scientific Refrigerators and Freezers - page 35

    - 32 - ...

  • Continental Scientific Refrigerators and Freezers - page 36

    - 33 - ...

Fabbricante Continental Scientific Categoria Refrigerator

I documenti che otteniamo dal fabbricante del dispositivoContinental Scientific Refrigerators and Freezers possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Continental Scientific
- manuali per l’uso Refrigerators and Freezers
- schede prodotto Continental Scientific
- opuscoli
- o etichette energetiche Continental Scientific Refrigerators and Freezers
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Continental Scientific Refrigerators and Freezers.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Continental Scientific Refrigerators and Freezers, manuali di manutenzione, brevi manuali o manuali utente Continental Scientific Refrigerators and Freezers. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Continental Scientific Refrigerators and Freezers.

Manuali per l’uso simili

Manuale completo del dispositivo Continental Scientific Refrigerators and Freezers, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Continental Scientific Refrigerators and Freezers dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Continental Scientific Refrigerators and Freezers.

Il manuale per l’uso completo Continental Scientific, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Continental Scientific Refrigerators and Freezers - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Continental Scientific Refrigerators and Freezers, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Continental Scientific Refrigerators and Freezers che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Continental Scientific Refrigerators and Freezers - che ci dovrebbe aiutare nei primi passi durante l'uso Continental Scientific Refrigerators and Freezers
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Continental Scientific Refrigerators and Freezers
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Continental Scientific Refrigerators and Freezers in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Continental Scientific Refrigerators and Freezers?

Utilizza il modulo sottostante

Se non hai risolto il problema con Continental Scientific Refrigerators and Freezers con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Continental Scientific Refrigerators and Freezers è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)