Summary
  • Invacare Compressor - page 1

    User: Befo re using the product , read this manual and save for future reference. Dealer: This m anual must be given to the end user. For more inform ation reg arding Inva care pro ducts, parts, and serv ices, please v isit www .invacare.com Operator’s Manual HomeFill ™ II Compressor ...

  • Invacare Compressor - page 2

    WARNING HomeFill ™ II Compresso r 2 Part No 1100873  WARNIN G DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERS TANDING THIS MANUAL. IF Y OU ARE UNAB LE TO UNDERSTAND THE DANGERS, WA RNINGS, CAUTIONS, AND INS TRUCTIONS, CONTACT INVACARE TECHNICA L SERVICES BEFORE ATTEMPTING TO USE THIS EQUIPMENT - OTHERWISE, SERIOUS INJURY OR PROP ...

  • Invacare Compressor - page 3

    TABLE OF CONTENTS TABLE OF CONTEN TS HomeFill ™ II Compresso r 3 Part No 1100873 SPE CI AL NO TES ...... ....... .... ...... ...... ....... ...... ..... 4 TYP IC AL PRO DU CT ....... ...... .... ...... ....... ...... ..... 5 PARA ME TE RS ...... ....... .... ...... ...... ...... ....... ...... . 5 LAB EL L OCA TIO N ...... ...... .... ...... .... ...

  • Invacare Compressor - page 4

    SPECIAL NOTES HomeFill ™ II Compresso r 4 Part No 1100873 SPECIA L NO TES Check all parts for shipping damage. In case of damage, DO NOT use. Contact Inv acare for further instruction. Signal words ar e used in this manual and apply to hazards or unsafe practices whi ch could result i n personal injury or property d amage. Refer to t he followi n ...

  • Invacare Compressor - page 5

    Part No 1100873 5 HomeFill ™ II Compresso r TYPICAL PRODUCT PARA METERS TYPICAL PRODUC T PARAME TERS - CO MPRESSOR Regulato ry Listing : Double Insulated P roduct ETL Certified Complying w ith UL73 and UL1097 (U S) ETL Certified Complying w ith CAN/CSA C22.2 No. 68 (Canada) For household use only. Operating tempe rature: 50 to 95° F (10 to 35° ...

  • Invacare Compressor - page 6

    HomeFill ™ II Compresso r 6 Part No 1100873 LABE L LOCA TIO N DEALER IN FORMATION NOTE: Invacare recomme nds leaving a full cylinder o f o xygen with the patient after setting up the HomeFill II compressor . Before leaving a HomeFill II Compressor with a user , the following checklist MUST be completed: ❑ Make sure the pressure test o n c ylind ...

  • Invacare Compressor - page 7

    HOMEFILL II INFORMATION HomeFill ™ II Compresso r 7 Part No 1100873 HOMEF ILL II INF ORMATION HomeFill I I is an accessory device to t he oxygen concentrator that enables the patient to refill custom portable oxyg en cylinders for personal use. NOTICE The HomeFill II co mpressor is for use with specially designed Invacare Platinum™5, 10 and XL ...

  • Invacare Compressor - page 8

    SECTION 1—IMPORTANT SAFETY INSTRUCTIONS Part No 1100873 8 HomeFill ™ II Compresso r SECTI ON 1— IMPO RTANT SAFETY INSTRU CTIO NS  WARNING SECTION 1 - IMPORTANT SAFETY INSTRUCTIONS contains important info rmation for the safe operation an d use of this product. DO NOT use this product or any available o ptional equipmen t without first comp ...

  • Invacare Compressor - page 9

    SECTION 1—IMPORTANT SAFETY INSTRUCTIONS HomeFill ™ II Compresso r 9 Part No 1100873 DO NOT allo w oil from y our hands or other sources to come into contact with the compressor couplin g or bottle fittings. These substances may become flam mable in the presence of oxygen and cause injury . A v oid tou ching the bottle fitting whenever poss ible ...

  • Invacare Compressor - page 10

    SECTION 1—IMPORTANT SAFETY INSTRUCTIONS Part No 1100873 10 HomeFill ™ II Compresso r DO NOT store filled oxygen cylinders near a furnace, hot w ater tank, in the trunk of a car or other high temperatur e area. Storage in areas such as thi s c an result in bu rsting of the cylinder or fire. The use of this devi ce is li mited to the oxy gen pati ...

  • Invacare Compressor - page 11

    SECTION 1—IMPORTANT SAFETY INSTRUCTIONS HomeFill ™ II Compresso r 11 Part No 1 100873 If any leakage of the o xygen cylinder i s detected DO NOT attempt to use the c ylinder . T urn the flow s elector knob on the regulator t o Off. If leakage persists, place the cyli nder outdoors and notify y our Home Health Care Provider or Service Representa ...

  • Invacare Compressor - page 12

    SECTION 2—INITIAL SETUP Part No 1100873 12 HomeFill ™ II Compresso r SECTI ON 2— INITI AL SETU P NOTE: Dealers Only - Refer to Dealer Informat ion on page 6 before performing this pr ocedure. Init ial Set Up of Co mpresso r Compressor Initial Setup Che cklist Before operat i ng the HomeFill II Compressor for the first time, complete the follo ...

  • Invacare Compressor - page 13

    SECTION 2—INITIAL SETUP HomeFill ™ II Compresso r 13 Part No 1 100873 Transporting the C ompressor  WARNING NEVER transport a compressor with a cyli nder connected to or stored on the compressor, otherwise injury or damage can occur. CAUTION When transporting the compressor, be careful not to bump the compressor or the connectors into obstac ...

  • Invacare Compressor - page 14

    SECTION 2—INITIAL SETUP Part No 1100873 14 HomeFill ™ II Compresso r 4. Plug the opposite end of the power cord i nto the w all outlet. FIGURE 2.2 Installing the Power Cord Connecti ng the C ompressor t o the Conc entrator NOTE: For this procedur e, refer to F I GURE 2.3 on page 15. 1. Insert one end of the interconnect hose into the outlet fit ...

  • Invacare Compressor - page 15

    SECTION 2—INITIAL SETUP HomeFill ™ II Compresso r 15 Part No 1 100873 FIGURE 2.3 Connecting the Compressor to the Concentrator ...

  • Invacare Compressor - page 16

    SECTION 3—CHECKLIST AND CYLINDER INSPECTION Part No 1100873 16 HomeFill ™ II Compresso r SECTI ON 3— CHECK LIST A ND CYLIND ER IN SPECTIO N Compres sor Op eration C hecklist Each time the HomeFill II Compressor is us ed to fill a cylinder , complete the following checkli st: ❑ Ensure the concentrator has been On for at least thi rty minutes ...

  • Invacare Compressor - page 17

    SECTION 3—CHECKLIST AND CYLINDER INSPECTION HomeFill ™ II Compresso r 17 Part No 1 100873 FIGURE 3.1 Cylinder Prefill Inspection DO NO T REMO VE THIS PRODUCT LABEL Not for Commer cial Resale Inv acare Corporation 2101 E. Lake Mar y Blvd. Sanford, FL 32773 1-800-832-4707 P ar t Number 1125090 Re v B - 11/04 O XYGEN 2 This cylinder is to b e used ...

  • Invacare Compressor - page 18

    SECTION 3—CHECKLIST AND CYLINDER INSPECTION Part No 1100873 18 HomeFill ™ II Compresso r The Hydrost atic Test ing Date Aluminum cyli nders MUST under go te sting ev ery fiv e y ears.  WARNING DO NOT fill cylinders that have not been tested in the past five years. Otherwise, serious personal injury may result. Contact your dealer for replace ...

  • Invacare Compressor - page 19

    Please Seal with Tape Before Mailing ...

  • Invacare Compressor - page 20

    REGISTER YOUR PRODUCT! The benefits of registering: 1. Safeguard your investment. 2. Ensure long term maintenance and servicing of your purchase. 3. Receive updates with product information, maintenance tips, and industry news. 4. Invacare can contact you or your provider, if servicing is needed on your product. 5. It will enable Invacare to improv ...

  • Invacare Compressor - page 21

    1. Method of purchase: (check all that apply) ❏ Medicare ❏ Insurance ❏ Medicaid ❏ Other 2. This product was purchased for use by: (check one) ❏ Self ❏ Parent ❏ Spouse ❏ Other 3. Product was purchased for use at: ❏ Home ❏ Facility ❏ Other 4. I purchased an Invacare product because: ❏ Price ❏ Features (list features) 5. Who ...

  • Invacare Compressor - page 22

    ...

  • Invacare Compressor - page 23

    SECTION 4—COMPRESSOR OPERATION HomeFill ™ II Compresso r 23 Part No 1 100873 SECT ION 4—COMPR ES SOR OPER ATION Connectin g/Disconn ectin g the C ylinder to/from th e Comp ressor Conn ectin g t he Cyl inder to t he Comp re ss or  WARNING NEVER use tools of any kind to con nect/disconnect the cylinder and the compressor. Otherwise, severe i ...

  • Invacare Compressor - page 24

    SECTION 4—COMPRESSOR OPERATION Part No 1100873 24 HomeFill ™ II Compresso r CAUTION DO NOT connect the outlet (cannula) connection to the HomeFill II compressor, otherwise the conserving cylinder will not work properly. 4. Momentarily push DOWN on the outer r ing (sleeve) of the connector fi llport until GREEN dots are vi sible to reset the con ...

  • Invacare Compressor - page 25

    SECTION 4—COMPRESSOR OPERATION HomeFill ™ II Compresso r 25 Part No 1 100873 8. Pull UP on the outer ri ng (sleev e) of the connec tor fillport while pushing DOWN on the cylinder/regulator assembly to couple the cylinder fillport into the connector fillport (FIGURE 4.2). NOTE: The cylinder is pr operly connecte d when an audible “click” is ...

  • Invacare Compressor - page 26

    SECTION 4—COMPRESSOR OPERATION Part No 1100873 26 HomeFill ™ II Compresso r FIGURE 4.2 Connecting the Cylinder to the Compre s sor ...

  • Invacare Compressor - page 27

    SECTION 4—COMPRESSOR OPERATION HomeFill ™ II Compresso r 27 Part No 1 100873 Disconnecti ng the Cy linder from the Compre ssor  WARNING NEVER use tools of any kind to con nect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. NOTE: For this procedur e, refer to FIGURE 4.3. 1. Press the c ompressor ...

  • Invacare Compressor - page 28

    SECTION 4—COMPRESSOR OPERATION Part No 1100873 28 HomeFill ™ II Compresso r 5. When the cylinder fillport is di sconnected from the connector fillport, release the outer ri ng (sleeve) o f the connector fillport and use two hands to r emove the cylinder from the compressor cradle.  WARNING The fillport cover on the connector fillport and the ...

  • Invacare Compressor - page 29

    SECTION 4—COMPRESSOR OPERATION HomeFill ™ II Compresso r 29 Part No 1 100873 Turni ng the Comp resso r On NOTE: For this procedur e, refer to FIGURE 4.5 on page 30. 1. Make sure the concentrato r i s On. Refer to the concentrator Owner ’ s Manual. CAUTION The co ncentrator flow rate to the patient MUST be set at maximum: 3 L/min. or less for ...

  • Invacare Compressor - page 30

    SECTION 4—COMPRESSOR OPERATION Part No 1100873 30 HomeFill ™ II Compresso r oxygen level r eaches 90%. If light does not go Off within 10 minutes, r efer to T r oubleshoot ing/ Cylin der Fill times on page 34. NOTE: If the A TTENTION (RED) light and audi ble alarm are on. Refer to T roubleshooting/ Cylinder Fi ll times on page 34. 6. Remov e th ...

  • Invacare Compressor - page 31

    SECTION 4—COMPRESSOR OPERATION HomeFill ™ II Compresso r 31 Part No 1 100873 Settin g the Flow Knob to the Prescri bed Settin g NOTE: For this procedur e, refer to FIGURE 4.6. 1. Attach the nasal cannula to the outlet fitting of the cylinder . 2. T urn the flow knob to the L/min. setting prescribed by yo ur physi ci an or t herapist.  WARNIN ...

  • Invacare Compressor - page 32

    SECTION 5—INDICATOR LIGHTS Part No 1100873 32 HomeFill ™ II Compresso r SECTI ON 5— INDIC ATO R LIGHT S INDICATOR LIGHT COL OR COMPRESSOR STATUS EFFECT None Compressor is Off O 2 flowing to user if concentrator is on O 2 BELOW NORMAL - YELLOW Compressor is On and warming up (0-3 minutes after power is On) Cylinder filling not started O 2 flow ...

  • Invacare Compressor - page 33

    SECTION 5—INDICATOR LIGHTS HomeFill ™ II Compresso r 33 Part No 1 100873 ATTENTION - RED (Audible Alarm) Compressor is On. Cylinder NOT filling Causes: Full cylinder a ttached when unit turned On - replace with empty cylin der or Internal failure - con tact de aler Compressor does not begin to fill or stops filling Unit m ust be turned Off and ...

  • Invacare Compressor - page 34

    SECTION 6—TROUBLESHOOTING/ CYLINDER FILL TIMES Part No 1100873 34 HomeFill ™ II Compresso r SECT ION 6—TROUBL ES HOOTIN G/ CYLIND ER FI LL TIM ES TROUBLESHOOTING PROBLEM SOLUTION No indicator lig hts on when compressor turned on. 1. Turn compressor Off. 2. Check that power cord is properly plugged into compressor and wall outlet . 3. Turn the ...

  • Invacare Compressor - page 35

    SECTION 6—TROUBLESHOOTING/ CYLINDER FILL TIMES HomeFill ™ II Compresso r 35 Part No 1 100873 O 2 BELOW NORMAL (YELLOW) light stays On 1. Ensure the concentrator has warmed up for at least 20 minutes.If it has not been 20 minutes, turn the compressor Off until the concentrator warms up. If the YELLOW light still remains on proceed to STEP 2. 2. ...

  • Invacare Compressor - page 36

    SECTION 6—TROUBLESHOOTING/ CYLINDER FILL TIMES Part No 1100873 36 HomeFill ™ II Compresso r NOTE: All filling times ar e approximate and may vary on envir onmental conditions. Regulator is On but O 2 is not being delivered 1. Refill cyl inder 2. Check that cannula is straight and not pinched. If damag ed, replace. 3. Call home health care provi ...

  • Invacare Compressor - page 37

    SECTION 7—MAINTENANCE HomeFill ™ II Compresso r 37 Part No 1 100873 SECT ION 7—MAIN TENAN CE Clea ning the Compress or Filter  WARNING Push the power switch to t he Off (O) position and unplug the compressor before cleaning. DO NOT operate the compress or without the filter installed. CAUTION DO NOT unplug by pulling on co rd. To unplug, g ...

  • Invacare Compressor - page 38

    LIMITED WARRANTY Part No 1100873 38 HomeFill ™ II Compresso r Cabi net  WARNING Push the power switch to t he Off (O) position and unplug the compressor before cleaning. To avoid electrical shock , DO NOT remove cabinet. 1. Clean the cabinet with a mi ld household cleaner and non-abrasive cloth or spong e. LIMITE D WARRAN TY NOTE: THE WARRANTY ...

  • Invacare Compressor - page 39

    HomeFill ™ II Compresso r 39 Part No 1 100873 For Mod el Nos . IOH2M6 and IOH2M9, Invacare warrants the HomeFill II regulator (NOT i ncludi ng the cyli nder) to be f ree from defects in materi als and workmans hip for a period of six (6) yea rs from date of p urchas e from Invacare or a dealer, with a copy o f the seller’s invoi ce requir ed fo ...

  • Invacare Compressor - page 40

    (INCLUDING, BU T NOT LIMITED TO, MODIFICATION THROUGH THE U SE OF UNA UTHORIZED PARTS OR ATTACHMENTS) OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY C OMPONENT WITHOU T THE SPECIFIC CONSENT OF INVACAR E CORPORATION OR TO PRODUCTS DAMAG ED BY CIRCUM STANCES BEYOND INVACARE CORPOR ATION'S CONTROL. THE FOREGOING EX PRESS W ARRANTY IS EXC ...

Fabbricante Invacare Categoria Air Compressor

I documenti che otteniamo dal fabbricante del dispositivoInvacare Compressor possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Invacare
- manuali per l’uso Compressor
- schede prodotto Invacare
- opuscoli
- o etichette energetiche Invacare Compressor
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Invacare Compressor.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Invacare Compressor, manuali di manutenzione, brevi manuali o manuali utente Invacare Compressor. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Invacare Compressor.

Manuali per l’uso simili

Manuale completo del dispositivo Invacare Compressor, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Invacare Compressor dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Invacare Compressor.

Il manuale per l’uso completo Invacare, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Invacare Compressor - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Invacare Compressor, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Invacare Compressor che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Invacare Compressor - che ci dovrebbe aiutare nei primi passi durante l'uso Invacare Compressor
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Invacare Compressor
5. FAQ - domande più frequenti
6. Dettagli di contatto Informazioni su dove cercare il contatto con il fabbricante / servizio Invacare Compressor in un dato paese, se il problema non può essere risolto da noi stessi.

Hai una domanda su Invacare Compressor?

Utilizza il modulo sottostante

Se non hai risolto il problema con Invacare Compressor con l'aiuto del manuale che hai trovato, fai una domanda utilizzando il modulo sottostante. Se un utente ha avuto un problema simile con Invacare Compressor è probabile che voglia condividere il modo di risolverlo.

Riscrivi il testo dall’immagine

Commenti (0)