    765000 AL TERNA TING PRESSURE MA TTRESS SYSTEM WITH ACTIVE SENSOR TECHNOL OGY USER MANUAL Impor tant: Do not operate the Mattress System without rst reading and understanding this manual! Save this manual for future use. Note: The most current version of this manual can be found online at www.grahameld.com 765000-INS-LAB-RevB10 ...

  • Graham Field 765000 - page 2

    2 765000-INS-LAB-RevB10 CONTENTS INTRODUCTION ..................................................................................... 3 INTENDED USE OF THIS DEVICE ........................................................ 3 IMPORT ANT SAFETY PRECAUTIONS .................................................. 4 DANGER ....................................... ...

  • Graham Field 765000 - page 3

    765000-INS-LAB-RevB10 3 INTRODUCTION Use this manual for set-up and operation of the Lumex ® AltaDyne ® Sensor Plus 765000 Alternating Pr essur e Mattr ess System with Active Sensor T echnology . Read all instructions befor e using the mattr ess system. Save this manual for futur e r efer ence. INTENDED USE OF THIS DEVICE The intended use of the ...

  • Graham Field 765000 - page 4

    4 765000-INS-LAB-RevB10 IMPORT ANT SAFETY PRECA UTIONS The safety statements pr esented in this chapter r efer to the basic safety information that should be obser ved by those using this Mattr ess System. Ther e ar e additional safetystatementsinotherchaptersorsections,whichmay bethesameasorsimilarto ...

  • Graham Field 765000 - page 5

    765000-INS-LAB-RevB10 5 W ARNING: Always use close supervision when this product is used by , on, or near children or those who require close supervision. W ARNING: Use this product only as intended and described in this manual. Do not use attachments or accessories not recommended by Graham-Field. W ARNING: Never operate this product if: a) It has ...

  • Graham Field 765000 - page 6

    6 765000-INS-LAB-RevB10 W ARNING: Patient entrapment with bed side rails may cause injury or death. The bed frame and its components, including the mattress, bed side rails, head and foot board, bedding, and any accessories added to the bed, can all affect the risk of entrapment. Thorough patient assessment and monitoring are necessary to reduce th ...

  • Graham Field 765000 - page 7

    765000-INS-LAB-RevB10 7 EQUIPMENT SYMBOLS Manufacturer Indicates that this product is according to the degree of protection against electric shock for type BF equipment Functional earth (for UL only) Attention! Read the instructions! Disposal of Electrical & Electronic Equipment (WEEE): Do not treat this product as household waste. For more det ...

  • Graham Field 765000 - page 8

    8 765000-INS-LAB-RevB10 PRODUCT DESCRIPTION PUMP AND MA TTRESS SY STEM pumpandmattresssystem CPRvalves quickconnectors (mattress-pump connectors) mattress footend symbol pump PUMP: FRONT quickconnector front panel PUMP: REAR mainpower switch powercord receptacle fuse hanger filter ...

  • Graham Field 765000 - page 9

    765000-INS-LAB-RevB10 9 PUMP: FRONT P ANEL Fr ont panel of pump is shown below . Descriptions of but- tons and indicators follow . T uningbutton; indicatorsatleftofbutton Powerfailureindicator Alarmmutebutton Serviceindicator Power(ON/OFF)button; indicatoratrightofbutton Panellockbutton; indicat ...

  • Graham Field 765000 - page 10

    10 765000-INS-LAB-RevB10 Tuning button: Thetuningbuttonne-tunesthe mattr ess system's auto-adjusted pr essur e. Thr ee tuning settings ar e available: 0 System auto-adjusted pr essur e - One level softer than auto-adjusted pr essur e +  Onelevelrmerthanauto-adjustedpr essur e Note: The tuning function ...

  • Graham Field 765000 - page 11

    765000-INS-LAB-RevB10 11 Maxfir m button: The pump will automatically enter maxrmmodeever ytimethepoweristurnedon.Maxrm mode ensur es the pump is able to r each its maximum oper- atingpr essur e.Maxrmmodewilllastfor20minutes,and then the system will automatically enter auto-adjus ...

  • Graham Field 765000 - page 12

    12 765000-INS-LAB-RevB10 Low pr essur e indicator: The low pressur e indicator(yellowLED)illuminates,andthealarmsounds, whenanabnormallylowpr essur eleveloccurs.Ifthelow pr essur e indicator persistently illuminates inappr opriately: 1. Checktoensur eallconnectionsar epr ope ...

  • Graham Field 765000 - page 13

    765000-INS-LAB-RevB10 13 INST ALLA TION UNP ACKING 1. Car efullyr emoveallcomponentsfr omthecarton. 2. Inspectallcomponents. 3. Ifdamageisevidenttothecontents,pleasenotifythe carrierandyourGraham-Fieldequipmentpr ovider . PUMP AND MA TTRESS SETUP 1. Removeallshee ...

  • Graham Field 765000 - page 14

    14 765000-INS-LAB-RevB10 W ARNING: Locate all cords so that they will not be stepped on, tripped over , or otherwise subjected to damage or stress. W ARNING: Do not use a household extension cord if the electrical cord does not reach the power outlet. Use of an improper extension cord could result in re and electric shock. If an extension cord m ...

  • Graham Field 765000 - page 15

    765000-INS-LAB-RevB10 15 OPERA TION GETTING ST AR TED 1. T urnthemainpowerswitch to the ON position. 2. Pr essthefr ontpanelpower button. The power indicator will illuminate. Note: The mattress system will automatically enter maxfirm mode for the quickest inflation at set-up; when inflated, it will then automa ...

  • Graham Field 765000 - page 16

    16 765000-INS-LAB-RevB10 TRANSPORT CAP Incaseofpowerfailur eortrans - port:Disconnectthequickconnec - tor and cap the mattr ess with the transport cap to slow deation. Note: This system is equipped with cell-on-cell technology to provide added support and prevent bottoming out in the event of a power failure. Lower ...

  • Graham Field 765000 - page 17

    765000-INS-LAB-RevB10 17 MAINTENANCE CLEANING W ARNING: T o reduce the risk of increased bacterial growth, infection, illness, or injury from contamination, thoroughly clean and dry the mattress system before use and as necessary during use as follows: Pump :W ipethepumpwithaclean,dampclothandmild deter gent,and? ...

  • Graham Field 765000 - page 18

    18 765000-INS-LAB-RevB10 Fuse replacement tool needed: small screwdriver 1. Ifyoususpectablownfuse,discon - nect the plug fr om the wall outlet immediately . 2. Use a small scr ewdriver to r emove the cover of the fuse holder (fuse lo- cations ar e shown in pictur e at right). 3. Insertanewfuseofthecorre ...

  • Graham Field 765000 - page 19

    765000-INS-LAB-RevB10 19 SPECIFICA TIONS Pump Specication Power Supply (Note: see rating label on product) AC 120V 60 Hz 0.17A (for 120V system) Fuse Rating T1A, 250V Cycle time Fixed / 10 minutes Dimensions (L x W x H) 1 1.4" x 7.9" x 4.6" (29.1 x 20 x 11.7 cm) Weight 5 lb (2.2 kg) Environment T emperature Operation: 50°F to 104 ...

  • Graham Field 765000 - page 20

    20 765000-INS-LAB-RevB10 LIMITED WARRANTY GFHealthPr oducts,Inc.warrantstheLumex ® AltaDyne ® Sensor Plus 765000 Alternating Pr essur e Mattr ess Sys- tem with Active Sensor T echnology for a period of twelve monthsfordefectsinworkmanshipandmaterials.Ifa pr oductisdeemedtobeunderwarra ...

  • Graham Field 765000 - page 21

    765000-INS-LAB-RevB10 21 APPENDIX A: EMC INFORMA TION Guidance and Manufacturer’s De claration- Electromagnetic Emissions: This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. Emissions T est Compliance Electromagnetic Environment-Gui dance ...

  • Graham Field 765000 - page 22

    22 765000-INS-LAB-RevB10 Guidance and Manufact urer ’ s Declaration- Electromagn etic Immunity: This device is intended for use in the electromag netic environment specified below . The user of this device should make sure it is used in such an environment. Immunity T est IEC60601 test level Compliance Electromagnetic Environment-Guidance Conduct ...

  • Graham Field 765000 - page 23

    765000-INS-LAB-RevB10 23 a The ISM (indu strial, scientific and medica l) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz;13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 M Hz to 40,70 MHz. b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,5 GHz are intended ...

  • Graham Field 765000 - page 24

    USA Corporate Headquar ters: GF Health Products, Inc. 2935 Nor theast Parkway Atlanta, Georgia 30360 telephone: 800-347-5678, 770-368-4700 fax: 800-726-0601, 678-291-3232 www .grahamfield.com © 2010 GF Health Products, Inc. ...

Fabbricante Graham Field Categoria Camping Equipment

I documenti che otteniamo dal fabbricante del dispositivoGraham Field 765000 possono essere suddivisi in diversi gruppi. Questi sono tra gli altri:
- disegni tecnici Graham Field
- manuali per l’uso 765000
- schede prodotto Graham Field
- opuscoli
- o etichette energetiche Graham Field 765000
Tutti sono importanti, ma le informazioni più importanti dal punto di vista di utilizzo del dispositivo possono essere trovate nel manuale per l’uso Graham Field 765000.

Un gruppo di documenti denominato manuali per l’uso, è anche suddiviso in tipi più specifici, come: Manuali di installazione Graham Field 765000, manuali di manutenzione, brevi manuali o manuali utente Graham Field 765000. A seconda delle esigenze, si dovrebbe cercare il documento necessario. Nel nostro sito web, puoi visualizzare il manuale più popolare per l'uso del prodotto Graham Field 765000.

Manuale completo del dispositivo Graham Field 765000, come deve essere?
Il manuale per l’uso, denominato anche istruzioni per l’uso, o semplicemente il manuale è un documento tecnico progettato per aiutare ad utilizzare Graham Field 765000 dagli utenti. I manuali sono di solito scritti da uno scrittore tecnico, ma in un linguaggio accessibile a tutti gli utenti Graham Field 765000.

Il manuale per l’uso completo Graham Field, dovrebbe includere vari elementi di base. Alcuni di loro sono meno importanti, come ad esempio: copertina / pagina del titolo o pagina di autore. Tuttavia, la parte rimanente, dovrebbe fornire informazioni importanti dal punto di vista dell'utente.

1. Introduzione e la guida su come utilizzare il manuale Graham Field 765000 - All'inizio di ogni manuale, dovremmo trovare indizi su come utilizzare il documento. Dovrebbe contenere informazioni sulla posizione dell’indice Graham Field 765000, FAQ o i problemi più comuni : i punti che sono più spesso cercati dagli utenti di ogni manuale
2. Indice - elenco di tutti i suggerimenti per Graham Field 765000 che possono essere trovati nel documento corrente
3. Suggerimenti su come utilizzare le funzioni di base del dispositivo Graham Field 765000 - che ci dovrebbe aiutare nei primi passi durante l'uso Graham Field 765000
4. Troubleshooting - sequenza sistematica delle attività che vi aiuterà a diagnosticare e, successivamente, risolvere i problemi più importanti del Graham Field 765000