  • Emerson IP221 - page 1

    1 Contents Safety Information 2 1. Introduction 1 . 1 Sensors 3 Gap Sensors Hi-Sens 1 . 2 Interface detection 4 Attenuation method Reflection method 1 . 3 Sludge density detection 5 1 . 4 Head amplifier units 5 Normallay acting Inverse acting Summary 1 . 5 Control Units 6 2. Installation 2 . 1 Calibration 7 2 . 2 Sensor and head amplifier installat ...

  • Emerson IP221 - page 2

    2 Warning : If this equipment is used in a manner not specified by the manufacturer, the protection provided may be impaired. All installation and commissioning of this equipment must be carried out by electrically competent persons. Protection of permanently installed equipment : This equipment is regarded as permanently installed equipment. Ensur ...

  • Emerson IP221 - page 3

    3 1. Introduction The Mobrey Electropulse system is used for the detection of liquids or liquids with suspended / settled solids. It consists of three units, a sensor, a head amplifier unit, and a control room unit. 1.1 Sensors There are many different Mobrey sensors available, but they fall into two categories: a) Gap sensors Gap sensors feature t ...

  • Emerson IP221 - page 4

    4 1.2 Interface detection Attenuation method – Figure 1.3 Attenuation is the reduction in strength of the ultrasonic signal caused by its transmission through a liquid. Viscous liquids, emulsions and liquids containing solid particles have a greater attenuation than clear thin liquids. Usually the difference in attenuation between the two liquids ...

  • Emerson IP221 - page 5

    5 1.3 Solids Density Detection The Electropulse system can be used in conjunction with a 433S type sensor to provide blanket level detection in settling tanks, facilitating the control of automatic de-sludging. In addition the Electropulse system can be used to detect the solids density of a slurry in a pipe line. In this case the Mobrey pipe secti ...

  • Emerson IP221 - page 6

    6 The head amplifier transmits pulses when the sensor is not oscillating (its normal state), and stops transmitting in the alarm condition when the sensor is oscillating. Thus the normal light on the control unit indicates a wet Hi-Sens, or dry gap sensors. When the inverting amplifier is used the sensor is in its quiescent state for most of the ti ...

  • Emerson IP221 - page 7

    7 2. Installation 2.1 Calibration When installing the sensor, it should be remembered that it is measuring instrument, and should be treated with care. Before installation the sensor and control unit should be inspected for transit damage, the gap faces of gap sensors must be parallel. Before installation, the equipment should be calibrated, if pos ...

  • Emerson IP221 - page 8

    8 Calibration procedure by sensor type i) Hi-Sens Sensors 1 ) Ensure that the sensor is clean and dry , that nothing is touching the sensor body , and that the delay adjustment on the control unit is fully anti-clockwise. 2 ) Set the gain potentiometer in the head amplifier fully anti-clockwise, and the gain switch (if fitted) in the low gain posit ...

  • Emerson IP221 - page 9

    9 5 ) With the sensor in the upper liquid, increase the gain, using the switch if necessary , until the control unit changes to ‘Normal’. Note the position of the potentiometer. If the control unit does not change to normal then the end of the scale should be used as a reference. 6 ) The gain should be set half way between the settings found in ...

  • Emerson IP221 - page 10

    10 2.2 Sensor and head amplifier installation Make sure that there is room inside the tank for the sensor without anything touching it, and that the gap is not obstructed. At least an inch of clearance must be left around Hi-Sens senors. The head amplifier should have its cable glands downwards if possible to reduce the possible to reduce the possi ...

  • Emerson IP221 - page 11

    11 * DIN RAIL Mounting Clip available on request Figure 2.3 - Control unit mounting details ...

  • Emerson IP221 - page 12

    12 2.3 Control unit installation The control units should be installed in an area suitable to their specifications. They are designed to be used indoors, but could be installed outside using a suitable weatherproof enclosure. The front panel should be accessible to the operator. The standard control unit MEP*L is installed by undoing the two screws ...

  • Emerson IP221 - page 13

    13 ii) Head Amplifier Unit The head amplifier terminals A+ and B- should be connected to the cores of the control unit cable. In the case of the dual marine unit, lower sensor connections A+ and B- are connected to the two cores of one cable, and upper sensor connections, C- and D+ to the other. The screen should not be connected in any head amplif ...

  • Emerson IP221 - page 14

    14 Figure 2.6 - Control Unit Connections MEP 3L I.S. Relays Screen to be connected T o head amplifier MEP 3B I.S. Protective Earth Neutral Live Supply ...

  • Emerson IP221 - page 15

    15 2.5 Intrinsically safe installation Control units type MEP3* and head amplifier units to type MEP3** and ***S*P*H** are approved by A TEX as intrinsically safe equipment to II 1 G Ex ia IIC. Refer to safety instruction IP221/SI and local code of practice. The following points must be noted when installing these units:- i) The cable from the head ...

  • Emerson IP221 - page 16

    16 3. Fault Finding If the unit does not work as described in the text, this section suggests some simple checks which will rectify most faults. If the system still ceases to work or there is any doubt as to its working, the factory should be consulted. IT SHOULD BE NOTED THA T SOME OF THE TESTS BELOW VIOLA TE I.S. REQUIREMENTS AND SHOULD NOT BE A ...

  • Emerson IP221 - page 17

    17 4. Specification and Description 4.1 Sensors Sensor T ype 302S 30HS 312S 312SP 323S 352S 362S 366S 373S 393S 39HS 402S 420S 433S 442S HL*S Minimum T emperatur e ºC -70 0 -70 -70 -70 -70 -210 -210 -70 -70 0 -70 -70 -70 -70 -70 Maximum T emperatur e ºC 150 250 130 130 130 150 65 65 150 150 250 150 150 150 150 130 Maximum Pressure Kg/cm 2 56 50 2 ...

  • Emerson IP221 - page 18

    18 4.3 Control Units Rack Mounted Unit MEP*R Less than 6VA Fault relay: SPCO, Energises in fault mode Alarm relay: High current SPCO, Low current DPCO, Energised in normal mode releases on fault or alarm T erminal block on the rear of the rack unit. 19” rack unit to hold up to 14 controls. Alarm relay de-energises Fault relay energises Alarm rela ...

  • Emerson IP221 - page 19

    19 5. Recommended Spare Parts 5.1 Sensor and head amplifier Sensors and head amplifiers are supplied as integral items, and therefore no spare parts are available. In the case of sensor mounted head amplifiers, the sensor is attached to the head amplifier, and the two are not usually available separately . 5.2 Control Units i) Fuses Standard Contro ...

  • Emerson IP221 - page 20

    20 Appendix I - Part Numbers Please note that the existence of a part number does not imply the availability of a unit. 1. Sensors and Sensors with Head Amplifiers Fitting 0 Outside Fitting 2 Double Fitting 3 Inside Fitting 6 Inside Fitting M6 8 Sludge Pipe 9 Flange Head Amplifier T ype P Sensor Mounted U Adjacent Mounted Approval O None H A TEX L ...

  • Emerson IP221 - page 21

    21 3. Control Units Intrinsic Safety 3 Intrinsically Safe Mounting R Rack Mounted L Wall Mounted Mains Voltage 1 220/240 V olts 2 110/120 V olts Output Relay S Single Pole Changeover D Double Pole Changeover M E P 3 L / 1 D 4. Racks Intrinsic Safety 3 Intrinsically Safe Number of spaces for control units 01 to 14 Remaining spaces are blanked off M ...

  • Emerson IP221 - page 22

    22 Appendix II – List of illustrations Section Figure T itle Page 1. Introduction 1 . 1 Gap Sensor 3 1 . 2 Hi-Sens Sensor 3 1 . 3 Attenuation Method 4 1 . 4 Interface Method 4 1 . 5 Sludge Density Sensor 5 2. Installation 2 . 1 Head Amplifier 7 2 . 2 Adjacent head amplifier 8 2 . 3 Control unit mounting details 1 1 2 . 4 Wiring Summary 1 2 2 . 5 ...

  • Emerson IP221 - page 23

    23 Maintenance / Inspection This is limited to periodic inspection by a qualified person to ensure that the installation including wiring and equipment housing is safe. Clean only with a damp cloth, ensuring that no moisture enters control unit. Check unit for damage and if damaged do not use. ...

  • Emerson IP221 - page 24

    24 abcdef Instructions Leaflet IP221, Rev . AA November 2004 Level International: Emerson Process Management Mobrey Measurement 158 Edinburgh Avenue, Slough, Berks, SL1 4UE, UK T +44 1753 756600 F +44 1753 823589 ww w .mobrey .com The Emerson logo is a trade mark and service mark of Emerson Electric Co. Rosemount is a registered trademark of Rosemo ...

メーカー Emerson カテゴリー Fitness Electronics

Emerson IP221のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- IP221の取扱説明書
- Emersonの製品カード
- パンフレット
- またはEmerson IP221の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Emerson IP221の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Emerson IP221の取り付け説明書、サービスマニュアル、簡易説明書、またはEmerson IP221のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Emerson IP221の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Emerson IP221デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがEmerson IP221を使用するのを助ける技術的文書のことです。説明書は通常、全てのEmerson IP221ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Emersonの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Emerson IP221の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははEmerson IP221の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Emerson IP221に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Emerson IP221デバイスの基本機能を使うにあたってのヒント。 Emerson IP221のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Emerson IP221に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるEmerson IP221のメーカー/サービスへの連絡先に関する情報。

Emerson IP221についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもEmerson IP221の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがEmerson IP221で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)