Summary
  • Philips M5066A - page 1

    HeartStart Defibrillator OWNER’S MANUAL Guid e to Set U p, O peration , Mai ntenan ce, an d Acce ssories M5066 A Editi on 7 ...

  • Philips M5066A - page 2

    Phil ips Medi cal Syste ms Inte ntio nall y blank . ...

  • Philips M5066A - page 3

    Phil ip s Me dic al Sy ste ms Th e Hear tSta r t Defibr ill ator M 5066A Pads Car tridge Han dle . Pul l the handl e to tur n on the Hea rtStart a nd r e mo v e the car trid ge’ s hard cover . Ready Light. Th is g r ee n li gh t tells you the readin ess of the Hea rtStart. Blin ki ng: st and by mo de (rea dy for use) Solid: in use Off: needs att ...

  • Philips M5066A - page 4

    Phil ips Medi cal Syste ms Inte ntio nall y blank . ...

  • Philips M5066A - page 5

    Phil ips Medi cal Syste ms Hear tStar t Def ibr illator M5066A QUICK REFERENC E ...

  • Philips M5066A - page 6

    Phil ips Medi cal Syste ms Inte ntio nall y blank . ...

  • Philips M5066A - page 7

    Phil ips Medi cal Syste ms Hear tStar t M5066A Automated Exter nal Def ibr illator Ed ition 7 IMPOR T A NT NO T E: It is impo rta nt to un derst and that su r vival r ates for sudde n cardiac arrest are directly rela ted to how s oon v ictim s receive treatm ent. For eve r y minu te o f delay , the c han ce o f sur vi val decl ines by 7% to 1 0%. D ...

  • Philips M5066A - page 8

    Phil ips Medi cal Syste ms Inte ntio nall y blank . ...

  • Philips M5066A - page 9

    Phil ips Medi cal Syste ms HEARTSTA RT M5066A About this e ditio n The informa tion in this g uide applies to the mod el M5066A He artStart Defib ri llator . Its techn ical cont ents apply to all m odels in the He artSt ar t HS1 family of de fibrilla tors, inc ludi ng th e HeartS tart, th e HeartSta rt OnS ite, and the Hea rtStart Fir st A id De fi ...

  • Philips M5066A - page 10

    Phil ips Medi cal Syste ms Inte ntio nall y blank . ...

  • Philips M5066A - page 11

    Phil ips M edica l Syst ems i CONTENT S 1 Introduction to the H eartStart Desc riptio n ...... ..... ..... ...... ..... ...... ........ ..... ...... ..... ...... ..... ...... ..... ...... ........ ..... ...... ... 1-1 Sud den Cardiac Arres t ..... ..... ...... ..... ..... ...... ........ ..... ...... ..... ...... ..... ...... ........ ..... . 1-1 I ...

  • Philips M5066A - page 12

    Ph ilip s M edica l Sy ste ms ii HEARTSTA RT M5066A APPEND ICES A Accessori es for the HeartStart B Gloss ary of term s C Gl ossa ry of sy mbol s/cont rols D Warni ngs a nd preca utions E T echn ica l info rm ation F Configuration G Testing and troubleshooting H Additional technica l i nformatio n required for European conformity ...

  • Philips M5066A - page 13

    Phil ips Medi cal Syste ms 1-1 1 1 Int rodu cti on to th e He ar tS tar t Descri ption The Hear tStart Defibril lator M5066A is pa rt of the Philips Hear tStar t HS1 fa mily of defibri llators. Sm all, light we ight, and batter y p ow ered, it is designe d for sim ple and r e liabl e oper atio n. Sudde n Cardia c Arrest The Hear tS tar t is used t ...

  • Philips M5066A - page 14

    1-2 Phil ips Medi cal Syste ms HEARTSTART M5066A T ra ining a nd pra ctice The HeartS tart is one part of a w ell- desi gned eme rge ncy r espon se pla n. An y emerg ency re spons e plan sh ould be und er the ov ersi ght of a ph ysicia n and sho uld incl ude tra inin g in car dio pulmo nary resus citat ion (CPR ). Philip s r ecommend s that y ou tr ...

  • Philips M5066A - page 15

    Phil ips Medi cal Syste ms 2-1 2 2 S ett ing up the He ar tStar t Package content s Che ck the cont ent s of the HeartSta rt Def ibri llator M5066A bo x to be su r e it contai ns: • 1 Hear tS tar t D efibrill at or • 1 batt ery M5070A • 1 Adul t SMAR T Pads Cartri dge M50 71A, conta ining one s et of a dhesive d efibrilla tion pads • 1 Ow n ...

  • Philips M5066A - page 16

    2-2 Phil ips Medi cal Syste ms HEARTSTART M5066A 3. In sert the cartri dge in to th e ca rtrid ge wel l on th e f r ont of the de fib ri lla tor . It shoul d click into place when p roperly sea ted. The green PULL h andle should be al l the wa y do wn. NO TE: T o pre v ent the pad s’ adhe siv e g el fr om drying out, do no t op en the ha r d co v ...

  • Philips M5066A - page 17

    2-3 Phil ips Medi cal Syste ms 2 6. Pla ce t he defib rillator in t he carr y ca se, pressing it firmly into p lace . Inse rt the Quic k Reference G uide, * f ace up , in the clea r pla stic wi ndow on th e i nsid e of the c ase. If you pur c h ased a spar e SMAR T Pads Cartridge or an Infant /Chi ld Pads Cart ridge , pla ce it in the storag e area ...

  • Philips M5066A - page 18

    2-4 Phil ips Medi cal Syste ms HEARTSTART M5066A Recommended a ccessories It is al wa ys a goo d ide a to have a spare batt er y and a spare p ads set . Ot her th ings that ar e us eful to k eep wi th th e HeartSta rt incl u de: • sci ssors — for c utting t he victim’ s c lothes if ne eded • dis posa ble gloves — to protec t the u ser • ...

  • Philips M5066A - page 19

    Phil ips Medi cal Syste ms 3-1 3 3 Using t he Hear tS tar t IM POR T ANT NO TE: Be sure to read the R emi nd ers se cti on a t th e en d o f this cha pter as we ll as t he wa rni ngs a nd pr ecautio ns in Appe ndix D . Ov erview If you t hink s omeone is in S C A, act quick ly and c almly . If s omeon e el se is availa ble , ask hi m or her to ca l ...

  • Philips M5066A - page 20

    3-2 Phil ips Medi cal Syste ms HEARTSTART M5066A STEP 1 : PULL the green handle T urn on th e Hear tS tar t by pu lli ng th e SMA R T Pad s Car t ridg e’ s green hand le . * Rem ov e the hard cover from the p ads ca rtrid ge and s et i t aside . Rem ain c alm and follow the He ar tStar t’ s inst ruction s. The Hear t Star t sta rts by di rectin ...

  • Philips M5066A - page 21

    3-3 Phil ips Medi cal Syste ms 3 STEP 2 : PLA CE the pads Pul l t he t ab at t he to p of t he pa ds cartri dge to peel of f the fi lm s eal. I nsid e ar e tw o adhe siv e pa ds on a plast ic li ner . Remov e the pad s fr om the cart ridg e. P eel on e pad o ff th e li ner . Pla ce the pad on the pati ent ’ s bar e sk in, exactly as shown in the ...

  • Philips M5066A - page 22

    3-4 Phil ips Medi cal Syste ms HEARTSTART M5066A STEP 3 : PR ESS the Shock button As soon as t he H ear tSt art det ects tha t the pa ds are a ttach ed t o th e pa tien t, it beg ins ana lyzin g the pa tien t’ s he ar t rhythm. I t te lls you that no one sh ould be touc hing the pat ient , a nd t he Cautio n l igh t be gins f lash ing a s a r emi ...

  • Philips M5066A - page 23

    3-5 Phil ips Medi cal Syste ms 3 T reat ing infants and childr en W A RNING: Most card iac arrests in child ren ar e not ca used by hear t proble ms. When r espond ing to cardia c arrest in an infa nt or child : • Provide in fant/child CPR while a bystan der c alls EMS and brings the Hea rtS tar t. • If no bystande r is available , provide 1 -2 ...

  • Philips M5066A - page 24

    3-6 Phil ips Medi cal Syste ms HEARTSTART M5066A If the patient is under 55 pounds or 8 ye ar s ol d, but yo u do NO T have an Infant/Child P ads Cartr idge: • DO NO T DEL A Y TREA TM ENT . • Remo v e all clot hing fr om the tors o , to bare both the ches t and the bac k. • Appl y the H eartStart usi ng the adu lt pad s cartrid ge, but pla ce ...

  • Philips M5066A - page 25

    3-7 Phil ips Medi cal Syste ms 3 Reminders • R emov e any medi cin e patch es and r esi dual ad hesi ve fr om the pat ient ’ s chest bef or e a ppl ying the pa ds. • Do no t allo w th e p ads to co nta ct o the r ele ct r ode s or meta l part s th at ar e in con tact w ith the p atien t. • A void placin g the pads directly over an implan te ...

  • Philips M5066A - page 26

    Phil ips Medi cal Syste ms HEARTSTART M5066A Notes ...

  • Philips M5066A - page 27

    Phil ips Medi cal Syste ms 4-1 4 4 After using the Hear t Star t After e ach use 1. Chec k th e o utsi de o f the Hear t Star t for sign s of d ama ge, dir t, or conta mination . If y ou see sign s of damage , cont act Philips for tec hnical suppo rt . If th e de fibrilla tor is d irty or co ntam inated , clean it ac cording t o the gui delin es in ...

  • Philips M5066A - page 28

    4-2 Phil ips Medi cal Syste ms HEARTSTART M5066A Follow your lo cal protocol w ith regard to prompt data trans fer for medi cal review af ter using th e He artStart. * Details ab out da ta trans fer and timi ng are provided in Event Revi ew do cumen tat ion. The i nf ormati on autom ati cal ly st or ed b y th e HeartSta rt includes a sum mar y of l ...

  • Philips M5066A - page 29

    Phil ips Medi cal Syste ms 5-1 5 5 Maintaining the Hear t Star t Routine M aintenance The Hear tSta rt is ver y si mple to ma intain . The defibri llator p erforms a self- test ever y day . In a ddition , a bat ter y i nser ti on se lf- test i s ru n wh enever a ba tter y is install ed in the device . The defib rillator’ s ex tensive autom atic s ...

  • Philips M5066A - page 30

    5-2 Phil ips Medi cal Syste ms HEARTSTART M5066A Cleaning the H ear tStar t The outsi de o f the Hea rt Star t an d its carr y ca se c an be clea ned w ith a soft clot h damp ened in so ap y water , chl orine bleac h (2 table spoons per qua r t or liter of water ), or amm onia- base d clean ers . Rem inder s: • Do not u se isopr opyl (rubbing) al ...

  • Philips M5066A - page 31

    Phil ips Medi cal Syste ms A-1 A A Acce sso r ies fo r the He ar t Sta r t Access ori es * for th e H eart Star t Defib rillato r a vailab le separa tely from y our Philips representative o r on- line at www .medical.p hilips .com/hea rt start includ e: • Batte ry (spar e recomme nded ) [REF: M5070A] •P a d s • Ad ul t S MAR T Pads Cart ri dg ...

  • Philips M5066A - page 32

    A-2 Phil ips Medi cal Syste ms HEARTSTART M5066A • T raining • Adult T raining Pa ds Cartridg e [REF : M5073A] • Adult T raining Re plac ement P ads [REF : M5093A] • Adult P ads Pla cemen t Guide [REF: M5090A] • Inf ant/Ch ild T ra ining Pa ds Cartridg e [REF : M5074A] • Inf ant/C hild T ra ini ng Replac ement P ads [R EF: M509 4A] • ...

  • Philips M5066A - page 33

    Phil ips Medi cal Syste ms B-1 B BG l o s s a r y o f t e r m s The terms lis ted in this Glossar y are defin ed in the context of the Philips Hea rtS tart De fibrilla tor and its use . AED Au tomated e xternal defibril lator (a s emi-au tomatic defibrillat or). AE D mode T he stan da rd tr e atment mod e fo r the He artSt art Def ibr illa tor . It ...

  • Philips M5066A - page 34

    B-2 Phil ips Medi cal Syste ms HEARTSTART M5066A fibrilla tion A dist urbance of the norma l hea rt rh ythm t hat result s in chaotic , dis organ ized activ ity tha t cann ot effecti vely pump bloo d. V entric ular fib rillation (fibrill ation in the l ow er c hamber s of the hea rt) is as sociat ed wi th sud den c ar diac ar r es t. HeartSta rt Ev ...

  • Philips M5066A - page 35

    B-3 Phil ips Medi cal Syste ms B On/Of f but to n sto ps the ba tt ery in sertion se lf -te st th at automa ti cal ly run s when a batte ry is ins ert ed. pads Se e “SM AR T p ads. ” pa tient care pa use A d efine d pau se to all ow pa tient a sse ssme nt, tre atmen t, a nd/or CPR . See “NSA pause” and “pr otocol pause. ” per iodic se l ...

  • Philips M5066A - page 36

    B-4 Phil ips Medi cal Syste ms HEARTSTART M5066A cartr idge. The pad s ar e ap pli ed t o the p ati ent’ s bar e ski n and us ed t o de te ct the patie nt’ s hear t rhythm a nd to transfer the defib rillation shock . Only H ear tStar t SMA R T Pads can be used with the Hea rtS tar t Defib rillat or . standb y mode The o peratin g mo de of the H ...

  • Philips M5066A - page 37

    Phil ips Medi cal Syste ms C-1 C C Gl ossar y of symbols/controls sym bol descri ption Pads cartrid ge ha ndl e. Gre en. Pulling t he ha ndle turns on the de fibrilla tor and open s pads car tridge for u se. Refer to operating ins tructio ns. On/Of f butt on. Gr een. Pr essing the On/ Off butto n when the de fibrilla tor is in standby m ode tu rns ...

  • Philips M5066A - page 38

    C-2 Phil ips Medi cal Syste ms HEARTSTART M5066A Cer tified by the Canad ian Standards A ssociation. Reference o rd er num ber . Expirat ion date . Lith iu m ma nganes e dio xi de b att ery . One batt ery in package. Do n ot c rus h the ba ttery . Do not expose th e bat tery to high hea t or open flames . Do no t incine rate th e batt er y . Do no ...

  • Philips M5066A - page 39

    C-3 Phil ips Medi cal Syste ms C This sid e up . T r ansportati on r equir ements ( ref er to asso ciated thermometer symbol). Storage requirements (refer to asso ciated therm ometer symbol). Envir onmen tal (temperature and r elative humidity) requirements. Thes e pad s a r e disp osab le a nd ar e for s ingl e pat ient u se only . Car tridge cont ...

  • Philips M5066A - page 40

    C-4 Phil ips Medi cal Syste ms HEARTSTART M5066A Serial number . Lot numb er . Fe de ral la w (U SA) r estr ic ts thi s de vi ce to s ale by or on the or d er of a ph ysic ian. Dispose o f in accordance with your coun try's r equirements. Pri nte d on re cy cl ed pa p er . sym bol descri ption ...

  • Philips M5066A - page 41

    Phil ips Medi cal Syste ms D-1 D D W ar nings and precautions It is impor t ant to underst and how to use you r Hea rtS tar t D efibrill ator sa fely . Pleas e r e ad thes e warning s and pr ecaution s c ar efu lly . A wa rning de scr ibes so met hing that c ould caus e seri ous pe rson al inj ury or death. A pr ecaut i on d escribe s some thi ng t ...

  • Philips M5066A - page 42

    D-2 Phil ips Medi cal Syste ms HEARTSTART M5066A pad s Do not allo w the pad s to cont act ot her elect r ode s or met al parts tha t ar e in con tact w ith the p atien t. Precautions de vic e handli ng The He artStart was desi gned to be stur dy and r eli able for man y diff er ent u se condi tio ns. How ev er , handling the Heart Start too r ou g ...

  • Philips M5066A - page 43

    Phil ips Medi cal Syste ms E-1 E E T ec hni cal infor m at ion HeartSt art Defibril lat or spec ific atio ns The s pecific ations provided in the following ta bles a re nomin al va lues. Addit ional in form at ion c an b e foun d in the T echni cal Refe rence Manual for He artSta rt Auto mated Exter nal De fibrillato rs, lo cated online at ww w .me ...

  • Philips M5066A - page 44

    E-2 Phil ips Medi cal Syste ms HEARTSTART M5066A Co ntrols and ind icato rs seali ng Drip pr oof per EN 60529 clas s IPx1. Solid Obj ects per EN 60529 clas s IP2x. ESD/EMI (rad iated an d imm unity ) See El ectr omagneti c Conf ormi ty ta bles . cate gor y spec ifications cate go ry spe cification s controls Green SMAR T Pa ds cartri dge handle Gre ...

  • Philips M5066A - page 45

    E-3 Phil ips Medi cal Syste ms E De fibr illa ti on Wa ve form cate gor y spec ifications wa vef orm parameters Biphas ic truncated ex ponential. W a v eform parameters a r e automati call y ad justed as a fu nction of patient defibr illation impe dance . In t he d iagram a t left, D is t he durati o n of ph ase 1 and E i s t he durati on of pha se ...

  • Philips M5066A - page 46

    E-4 Phil ips Medi cal Syste ms HEARTSTART M5066A ene rgy * (pe diatric d oses in dicated ar e based on C DC gro wt h charts for t he 50th per centile we ight s for bo ys.) Using HeartStart Adult SMAR T Pads: 150 J nomi nal ( ± 15% ) into a 50 ohm lo ad. Usin g Hea rt Start Infant /Child SMAR T Pa ds: 50 J nom inal ( ± 15%) int o a 50 o hm load. S ...

  • Philips M5066A - page 47

    E-5 Phil ips Medi cal Syste ms E ECG analysis system cat eg o ry sp ecific ations func tion Eval uates impe danc e of adhe siv e pads for pr oper co ntact with th e pati ent’ s skin, and ev aluat es the ECG r h ythm and sign al q uali ty t o det ermine if a shock is appr opriate. sh ockable rhythms V entr icular fibril lation (VF) a nd some ventr ...

  • Philips M5066A - page 48

    E-6 Phil ips Medi cal Syste ms HEARTSTART M5066A ECG analysis performance rhythm class ECG tes t samp le a siz e meets AHA r ecommend ations b f or adult defibr illation obser ved 90% one-sided performance lower c onfidence limit shoc kab le rh ythm — ventricular fibri llation 300 sen sitiv ity >90 % (meets AAMI DF80 re quir e ment) (87%) shoc ...

  • Philips M5066A - page 49

    E-7 Phil ips Medi cal Syste ms E Acc ess ori es sp ecif icat ion s Batt e ry M5070A Hea rtSt ar t Adul t SMAR T Pads M 5071A and Infant /Chil d SMAR T Pads M5072A catego r y specifica tions batt er y type 9 VDC , 4.2 A h, lithium man gane se dioxide. Dispo sable , long-life pr imar y cell. ca paci ty When new , a minim um of 200 shoc ks or 4 h ours ...

  • Philips M5066A - page 50

    E-8 Phil ips Medi cal Syste ms HEARTSTART M5066A Envir onmen tal c onsidera tions By co mplying w ith you r natio nal regu lations regarding di spos al of e lectric , ele ctr o nic, and ba ttery waste, yo u can mak e a positi v e cont ri butio n to our shar e d envi r onm ent. Suc h w aste can int r odu ce harm ful e le ment s i nto t he environ me ...

  • Philips M5066A - page 51

    Phil ips Medi cal Syste ms F-1 F FC o n f i g u r a t i o n Ov erview The Ph ilips H ear tStar t D efibri llator co mes wit h a f actor y default config uration desi gne d to meet t he ne eds of mo st u sers. Thi s c on figur ati on c an onl y be chan ged by an aut hori zed per son using HeartSt art Configu re PD A soft war e or Ev ent Re vie w s o ...

  • Philips M5066A - page 52

    F-2 Phil ips Medi cal Syste ms HEARTSTART M5066A Patient treatment pr otocol opti ons para meter se tt in gs default default de script ion “call E MS” voice re mi nde r ti mi ng • At po w er o n ( when the us er tu rns on the Hear tS tar t) • At po w er on and at the sta rt of th e f irst pati e nt car e pa use • A t the star t of the fir ...

  • Philips M5066A - page 53

    F-3 Phil ips Medi cal Syste ms F pr ot ocol pause timer (minut es) 0.5, 1. 0, 1.5, 2.0, 2 .5, 3.0 2.0 A 2-m inut e pr oto col pa us e fo r CP R auto matic ally start s after voice ins truction is given when a shock ser ies is com plet ed. A fter the prot ocol pause, the def ibr illat or r et urns to rh ythm a nalysis. If th e use r presses th e i-b ...

  • Philips M5066A - page 54

    F-4 Phil ips Medi cal Syste ms HEARTSTART M5066A NSA pause timer (minut es) 0.5, 1. 0, 1.5, 2.0, 2 .5, 3.0 2.0 A 2-min ute NSA paus e for CPR auto matic ally start s after voice ins truction is gi v en when no shock is advised (NSA). * If th e use r presses th e i-b utton for opti onal CPR coach ing, the Hea rtStart pro vi des coac hing for 5 cy cl ...

  • Philips M5066A - page 55

    F-5 Phil ips Medi cal Syste ms F CPR Coac hing adu lt ventila tion instru ction Y es, No Y es Opt ional CPR Coaching incl udes r escu e br ea ths at th e rate de term ined by the CPR Coaching compression :v entilation ratio f or adults when an adul t pa ds cartridg e is ins talled. NO TE: i f t his p ar a met er is c onf igur ed to NO, CPR Coachi n ...

  • Philips M5066A - page 56

    Phil ips Medi cal Syste ms HEARTSTART M5066A Notes ...

  • Philips M5066A - page 57

    Phil ips Medi cal Syste ms G-1 G G T esting and troubl eshooting Te s t i n g As long as a b atter y i s i nstalled, the Hear tS tar t Def ibrill ator a utoma tically t ests itsel f ev ery day and aler ts you if it finds a problem . The self-te st include s pads readiness test ing. In additio n, it ru ns a pa ds self -test e ach t ime a pads c ar t ...

  • Philips M5066A - page 58

    G-2 Phil ips Medi cal Syste ms HEARTSTART M5066A no s par e battery is a v ailab le, r e mov e the in sta lled ba ttery fo r fiv e second s, then reinser t it a nd run a batter y in ser tion self- test. If th e prob lem co ntinue s, do no t us e the Hea rt Star t. A tten d to the p ati ent, pro vidin g CPR if neede d, un til Emerg ency Me dical S e ...

  • Philips M5066A - page 59

    G-3 Phil ips Medi cal Syste ms G ... to sto p al l moti on • The pati en t is be ing mo v ed or jost led. • The en vir onment is dry and mov e me nt ar ound the patie nt is causi ng stat ic elect ricity to inte rfe r e with ECG anal ysis. • Radio or elec tr ical so ur ce s are interferin g with ECG an alysis. • Sto p CP R; do no t to uch th ...

  • Philips M5066A - page 60

    G-4 Phil ips Medi cal Syste ms HEARTSTART M5066A T roublesho otin g while th e Hear tS tart is no t in use (g r een R eady li gh t is not on) beha vior possib le cau se re co mmende d actio n chirp s or i-but ton f la shes • The bat tery pow e r is lo w or the SMA R T Pads cart ri dg e nee d s to be re pla ce d. • The defibri lla tor ma y ha ve ...

  • Philips M5066A - page 61

    Phil ips Medi cal Syste ms H-1 H H Additional technical inf or mation required f or European conformity Electromagne tic con formity Guidanc e and man uf actur e r’ s dec larati on: Th e HeartSt art is in tend ed f or u se i n the elect romagn etic environm ent spe cifi ed in th e tab les b elow . Th e c ustom er or u ser of the Hear tSta rt shou ...

  • Philips M5066A - page 62

    H-2 Phil ips Medi cal Syste ms HEARTSTART M5066A Elec tr omagnetic immunity imm unit y test IEC 60 601 tes t le vel comp liance level elec troma gnet ic envi r onme nt - guidanc e elect rosta tic di scha rge (ESD) IEC 61000-4 -2 ± 6 kV contac t ± 8 kV air ± 6 kV c ont act ± 8 kV ai r Ther e ar e no specia l r equi r ement s with r e spec t to e ...

  • Philips M5066A - page 63

    H-3 Phil ips Medi cal Syste ms H Reco mm end ed s epa ra tion dis ta nces be tw e en po rtabl e and mob il e RF co mm uni ca tio ns eq ui pme n t an d th e Hea rtSt art The He artSt art is inte nded for use i n an elec tr omagneti c en vi r onme nt in which ra diat ed R F di stur ba nces ar e co ntr ol led . T he cus tomer or the u ser of the Heart ...

  • Philips M5066A - page 64

    H-4 Phil ips Medi cal Syste ms HEARTSTART M5066A Impor tant warnings and reminders • Do not allo w the pad s to cont act ot her el ect r ode s or met al parts tha t ar e in con tact w ith the p atien t. • Be fore deliv erin g a shock, it is impor tant to d isconnec t the patie nt from oth er medic al elect ric al equ ipm ent, such as blood -fl ...

  • Philips M5066A - page 65

    Phil ips Medi cal Syste ms Intent iona lly blan k. ...

  • Philips M5066A - page 66

    Inte ntio nall y blank . Phil ips Medi cal Syste ms ...

  • Philips M5066A - page 67

    Phil ips Medi cal Syste ms Intent iona lly blan k. ...

  • Philips M5066A - page 68

    Phil ips Medi cal Syste ms Inte ntio nall y blank . ...

  • Philips M5066A - page 69

    Phil ips Medi cal Syste ms Intent iona lly blan k. ...

  • Philips M5066A - page 70

    PO WER T O SA VE A LIFE Phili ps Medical Systems is par t of Ro ya l Philips Electr onics Philips Me dical Sys tem s Uni t ed S t ates Philips Medical System s 2301 Fifth Av enue , Suite 200 Seattle, W A, USA 98121 (800) 263-3342 Cana da Philips Medical System s 281 Hillmount Road Markham, Onta rio L6C 2S3 (800) 291-6743 Europe, M iddle Ea st, a nd ...

メーカー Philips カテゴリー Blood Pressure Monitor

Philips M5066Aのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- M5066Aの取扱説明書
- Philipsの製品カード
- パンフレット
- またはPhilips M5066Aの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Philips M5066Aの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Philips M5066Aの取り付け説明書、サービスマニュアル、簡易説明書、またはPhilips M5066Aのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Philips M5066Aの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Philips M5066Aデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがPhilips M5066Aを使用するのを助ける技術的文書のことです。説明書は通常、全てのPhilips M5066Aユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Philipsの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Philips M5066Aの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははPhilips M5066Aの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Philips M5066Aに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Philips M5066Aデバイスの基本機能を使うにあたってのヒント。 Philips M5066Aのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Philips M5066Aに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き