    Dealer: This manual MUST be given to the end user. User: BEFORE using this product, read this manual and save for future reference. Operator’s Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Perfecto 2 ™ Series Oxygen Concentrators without SensO 2 ® , Oxygen Concentrators with SensO 2 ...

  • Invacare IRC5PO2AW - page 2

    WARNING Perfecto 2 ™Series 2 Part No 1150713  WARNING DO  NOT  use  this  product  or  any  av ailable  optional  equipment  without  first  completely  reading  and  understanding  these  instructions  and  any  additional  instructional  material  such  as  owner ’ ...

  • Invacare IRC5PO2AW - page 3

    TABLE OF CONTENTS Part No 1150713 3 Perfecto 2 ™Series ACCESSORIES .................................................. 2 SPECIAL NOTES ............................................... 4 LABEL LOCATION ............................................ 6 SECTION 1—GENERAL GUIDELINES ..................... 7 Radio Frequency Interference .. ............... ...

  • Invacare IRC5PO2AW - page 4

    SPECIAL NOTES Perfecto 2 ™Series 4 Part No 1150713 SPECIAL NOTES Signal  w ords  are  used  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  following  table  for ? ...

  • Invacare IRC5PO2AW - page 5

    SPECIAL NOTES Part No 1150713 5 Perfecto 2 ™Series CAUTION “Caution:  Statutory  law  can  restrict  this  device  to  sal e  by  or  on  order  of  a  physician,  or  any  other  practitioner  licens ed  by  the  law  of  the  governmental  agency  in  which ...

  • Invacare IRC5PO2AW - page 6

    LABEL LOCATION Perfecto 2 ™Series 6 Part No 1150713 LABEL LOCATION Serial Number Lab el is located on the resonator intake assembly Specification Label is located on the back near the bottom Perfecto 2 ™ FOR OPERATING INSTRUCTIONS or ALARMS (See Operator’s Manual) DANGER - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room ...

  • Invacare IRC5PO2AW - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1150713 7 Perfecto 2 ™Series SECTION 1—GENERAL GUIDELINES In  order  to  ensure  the  safe  installation,  assembly  and  operation  of  the  Pe r f e c t o 2  concentrator  these  instructions  MUST  be  followed.  WARNING This  section  conta ...

  • Invacare IRC5PO2AW - page 8

    SECTION 1—GENERAL GUIDELINES Perfecto 2 ™Series 8 Part No 1150713 Operating Information For  optimum  performance,  Inv acare  recommends  that  each  concentrator  be  on  and  running  for  a  minimum  of  30  minutes  at  a  time.  Shorter  periods  of  operation  may ...

  • Invacare IRC5PO2AW - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1150713 9 Perfecto 2 ™Series Radio Frequency Interference This  equipmen t  has  been  tested  and  found  to  comply  with  EMC  limits  specified  by  IEC/EN  60601 ‐ 1 ‐ 2.  These  limits  are  designed  to  provide  a  reasonable ? ...

  • Invacare IRC5PO2AW - page 10

    SECTION 2—FEATURES Perfecto 2 ™Series 10 Part No 1150713 SECTION 2—FEATURES Oxygen Outlet Flowmeter Oxygen Purity Indicator Lights / Fault and Power Indicator Lights Power Switch Circuit Breaker Elapsed Time Meter Cabinet Filter REAR VIEW *HF Outlet Fitting Power Cord *NOTE:  This  outlet  fitting  is  to  be  used  on ...

  • Invacare IRC5PO2AW - page 11

    SECTION 3—HANDLING Part No 1150713 11 Perfecto 2 ™Serie s SECTION 3—HANDLING The  concentrator  should  AL W A YS  be  kept  in  the  upright  position  to  prevent  cabinet  damage  while  being  transported. If  the  concentrator  is  to  be  reshipped  by  common  c ...

  • Invacare IRC5PO2AW - page 12

    SECTION 4—TECHNICAL DESCRIPTION Perfecto 2 ™Series 12 Part No 1150713 SECTION 4—TECHNICAL DESCRIPTION The  Invacare  Per f e c t o 2  concentr ator  is  used  by  patients  with  respiratory  disorders  who  require  supplemental  oxygen.  The  device  is  not  intended  to  susta ...

  • Invacare IRC5PO2AW - page 13

    SECTION 5—TYPICAL PRODUCT PARAMETERS Part No 1150713 13 Perfecto 2 ™Serie s SECTION 5—TYPICAL PRODUCT PARAMETERS Alternating Current Type BF equipment Unit runnin g Unit not runnin g Attention - Consider Accompanying Documents DO NOT smoke Class II, Double Insulated Not Protected Electrical Requirements: 230 VAC ± 10% (253 VAC/20 7 VAC), 50 ...

  • Invacare IRC5PO2AW - page 14

    SECTION 5—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™Series 14 Part No 1150713 Flow Range: 0.5 to 5 L/min (maximum ). For flowrates less than 1 L/min, we recommend th e use of the Invacare Pediatric Flowmeter Accessory (IRCPF16AW). Low Flow Alarm: 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). Average Power Consumption ...

  • Invacare IRC5PO2AW - page 15

    SECTION 5—TYPICAL PRODUCT PARAMETERS Part No 1150713 15 Perfecto 2 ™Serie s No AP/APG Not suitable for use in the presence of a flammable anaesthetic mixture. Standards and Regulatory Listing: IRC5PO2AW/IRC5PAW IEC/EN61000-3-2 IEC/EN61000-3-3, IEC/EN 60601-1 , A1, A2 IEC/EN 60601-1 -2 IRC5PO2AW Models Only ISO8359 MDD 93/42/EEC, Annex I and IX ...

  • Invacare IRC5PO2AW - page 16

    SECTION 5—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™Series 16 Part No 1150713 IRC5PO2AW INDICATORS IRC5PAW INDI CATORS LABEL SYMBOL O 2 PURITY INDICATOR LIGHTS (LED) SYSTEM OKAY O 2 over 85% (± 2%) GREEN Indicator Light O 2 Between 73% (± 3%) to 85% (±2%) YELLOW Indicator light A. YELLOW Solid B. YELLOW Flashing Sensor Failure Call a qualif ied ...

  • Invacare IRC5PO2AW - page 17

    SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 17 Perfecto 2 ™Serie s SECTION 6— OPERATING INSTRUCTIONS Introduction Yo u r  oxygen  concentrator  is  intended  for  individual  use.  It  is  an  electronically  operated  device  that  separates  oxygen  from  room  air .  It  p ...

  • Invacare IRC5PO2AW - page 18

    SECTION 6—OPERATING INSTRUCTIONS Perfecto 2 ™Series 18 Part No 1150713 Yo u r  oxygen  concentrator  will  perform  best  when  operated  under  the  conditions  outlined  in  the  Ty p i c a l  Product  P arameters  on  page 13.  Usage  in  environments  other  than  those ...

  • Invacare IRC5PO2AW - page 19

    SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 19 Perfecto 2 ™Serie s 3. Insert  a  flathead  screwdriver  in  the  plate  groove  on  the  top  edge  of  the  filter  access  door  and  gently  pry  the  filter  access  door  off  (FIGURE  6.2). 4. Pull  up  ...

  • Invacare IRC5PO2AW - page 20

    SECTION 6—OPERATING INSTRUCTIONS Perfecto 2 ™Series 20 Part No 1150713 10. After  assembly ,  ensure  that  oxygen  is  flowing  through  the  cannula. FIGURE 6.3 Humidifier Compartmen t Power Switch NOTE:  For  this  pr ocedure,  refe r  to  FIGURE  6.4. 1. Press  power  switch  to  O ...

  • Invacare IRC5PO2AW - page 21

    SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 21 Perfecto 2 ™Serie s Flowrate NOTE:  For  this  procedure,  refe r  to  FIGURE  6.5. NOTE:  Ta k e  care  not  to  set  the  flow  above  RED  ring.  An  oxygen  flow  gr eater  than  5  L/min  will  decr ease  the ...

  • Invacare IRC5PO2AW - page 22

    SECTION 6—OPERATING INSTRUCTIONS Perfecto 2 ™Series 22 Part No 1150713 SensO 2 Oxygen Purity Indicator - Model IRC5PO2AW This  feature  monitors  the  purity  of  oxygen  generated  by  the  oxygen  concentrator .  If  purity  falls  below  factory  preset  standards,  indicator  light ...

  • Invacare IRC5PO2AW - page 23

    SECTION 6—OPERATING INSTRUCTIONS Part No 1150713 23 Perfecto 2 ™Serie s FIGURE 6.6 Explanation of Oxygen Purity Indicator Lights - Model IRC5PO2AW Explanation of Indicator Lights - Model IRC5PAW NOTE:  For  this  procedure,  refe r  to  FIGURE  6.7. IRC5PAW Model Indicator Light Explanation RED  light  () ‐  T ...

  • Invacare IRC5PO2AW - page 24

    SECTION 7—MAINTENANCE Perfecto 2 ™Series 24 Part No 1150713 SECTION 7—MAINTENANCE  WARNING The  Invacare  concentrators  are  specifically  designed  to  minimize  routine  preventiv e  mainte nance.  Only  qualified  personnel  should  perform  preventiv e  maintenance  on  the  ...

  • Invacare IRC5PO2AW - page 25

    SECTION 7—MAINTENANCE Part No 1150713 25 Perfecto 2 ™Serie s 2. Clean  the  cabinet  filter  with  a  va c u u m  cleaner  or  wa s h  in  w arm  soapy  wa t e r  and  rinse  thoroughly . 3. Dry  the  filter  thoroughly  before  reinstallation. FIGURE 7.1 Cleaning the Cabinet Filter ...

  • Invacare IRC5PO2AW - page 26

    SECTION 7—MAINTENANCE Perfecto 2 ™Series 26 Part No 1150713 PREVENTIVE MAINTENANCE RECORD Model  No.  IRC  ______________  Serial  No.________________ On  Each  Inspection *NOTE:  Refer  to  Preventive  Ma intenance  section  of  Service  Manual. **NOTE:  Model  not  available  within  ...

  • Invacare IRC5PO2AW - page 27

    SECTION 8—TROUBLESHOOTING GUIDE Part No 1150713 27 Perfecto 2 ™Serie s SECTION 8—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Main Power Loss: Short beeps, long pause 1. Power cord not plugged in. 1. Insert plug in to outlet. Concentrator not operating, power switch On. 2. No power at outlet. 2 . Inspect house circuit break ...

  • Invacare IRC5PO2AW - page 28

    SECTION 8—TROUBLESHOOTING GUIDE Perfecto 2 ™Series 28 Part No 1150713 YELLOW or RED light Illuminated. Alarm: Continuo us On RED light only. * Only applies to IRC5PO2AW models. 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tubing, cannula or humidifier.* 2. Inspect for kinks or blockages. Correct, clean or replace ite ...

  • Invacare IRC5PO2AW - page 29

    SECTION 9—OPTIONAL ACCESSORIES Part No 1150713 29 Perfecto 2 ™Serie s SECTION 9—OPTIONAL ACCESSORIES The  following  optional  accessories  are  also  av ailable:  • Standard  Adult  Nasal  Cannula,  2.1  m  (7  feet) ‐ M3120 • Standard  Humidifier ‐ M5120 • PreciseRX™  P ed ...

  • Invacare IRC5PO2AW - page 30

    RECYCLING INFORMATION Perfecto 2 ™Series 30 Part No 1150713 RECYCLING INFORMATION This  product  has  been  supplied  from  an  environmentally  aw are  manufacturer  that  complies  with  the  Wa s t e  Electrical  and  Electronic  Equipment  (WEEE)  Directive  2002/96/CE. This  p ...

  • Invacare IRC5PO2AW - page 31

    LIMITED WARRANTY Part No 1150713 31 Perfecto 2 ™Serie s LIMITED WARRANTY United  Kingdom:  Invacare  Ltd,  South  Road,  Bridgend  Industrial  Estate,  UK ‐ Bridge nd  CF31  3PY Te l :  (44)  (0)1656  664  321,  Fax:  (44)  (0)1656  667  532 UK@inv acare.com  Ireland: Inv acare  ...

  • Invacare IRC5PO2AW - page 32

    0120 Invacare Corporation www.invacare.com USA One Invacare Way Elyria, Ohio USA 44036-2125 440-329-6000 800-333-6900 Technical Services 440-329-6593 800-832-4707 EU Representative Invacare International Sarl Route de Cité Ouest 2 1196 Gland Switzerland Tel: +41 22 354 60 10 Fax: +41 22 354 60 11 Part No 1150713 Invacare Corporation 2101 E. Lake M ...

メーカー Invacare カテゴリー Oxygen Equipment

Invacare IRC5PO2AWのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- IRC5PO2AWの取扱説明書
- Invacareの製品カード
- パンフレット
- またはInvacare IRC5PO2AWの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Invacare IRC5PO2AWの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Invacare IRC5PO2AWの取り付け説明書、サービスマニュアル、簡易説明書、またはInvacare IRC5PO2AWのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Invacare IRC5PO2AWの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Invacare IRC5PO2AWデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがInvacare IRC5PO2AWを使用するのを助ける技術的文書のことです。説明書は通常、全てのInvacare IRC5PO2AWユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Invacareの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Invacare IRC5PO2AWの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははInvacare IRC5PO2AWの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Invacare IRC5PO2AWに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Invacare IRC5PO2AWデバイスの基本機能を使うにあたってのヒント。 Invacare IRC5PO2AWのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Invacare IRC5PO2AWに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるInvacare IRC5PO2AWのメーカー/サービスへの連絡先に関する情報。

Invacare IRC5PO2AWについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもInvacare IRC5PO2AWの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがInvacare IRC5PO2AWで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)