    Dealer: This manual MUST be given to the end user of this product. User: BEFORE using this product, read this manual and save for future reference. User Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.com Perfecto 2 ™ Series Oxygen Concentrators without SensO 2 ®, Oxygen Concentrators with ...

  • Invacare IRC5PO2V - page 2

    WARNING Perfecto 2 ™ Series 2 Part No 114348 2  WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UN DERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. I ...

  • Invacare IRC5PO2V - page 3

    TABLE OF CONTENTS Part No 1143482 3 Perfecto 2 ™ Series Table of Contents ACCESSORIES .................................................. 2 SPECIAL NOTES ............................................... 4 LABEL LOCATION ............................................ 6 SECTION 1—GENERAL GUIDELINES .................. 7 To Reduce the Risk of Burns, El ...

  • Invacare IRC5PO2V - page 4

    SPECIAL NOTES Perfecto 2 ™ Series 4 Part No 114348 2 SPECIAL NOTES Signal  words  are  used  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  following  table  for ...

  • Invacare IRC5PO2V - page 5

    SPECIAL NOTES Part No 1143482 5 Perfecto 2 ™ Series CAUTION “Caution:  Federal  law  restricts  this  device  to  sale  or  rental  by  or  on  order  of  a  physician,  or  any  other  practitioner  licensed  by  the  law  of  the  State  in  which  he/sh ...

  • Invacare IRC5PO2V - page 6

    LABEL LOCATION Perfecto 2 ™ Series 6 Part No 114348 2 LABEL LOCATION FOR OPERA TING INSTRUCTIONS or ALARMS (See Operator’s Manual) DANGER - ELECTRIC SHOCK DANGER - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and away from areas where o xygen is being delivered. T extiles, oil and other ...

  • Invacare IRC5PO2V - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1143482 7 Perfecto 2 ™ Series SECTION 1—GENERAL GUIDELINES In  order  to  ensure  the  safe  installation,  assembly  and  operation  of  the  Pe r f e c t o 2  concentrator  these  instructions  MUST  be  followed.  WARNING This  section  cont ...

  • Invacare IRC5PO2V - page 8

    SECTION 1—GENERAL GUIDELINES Perfecto 2 ™ Series 8 Part No 114348 2 Operating Information For  optimum  performance,  Inv acare  recommends  that  each  concentrator  be  on  and  runni ng  for  a  minimum  of  30  minutes  at  a  ti me.  Shorter  periods  of  operation  ...

  • Invacare IRC5PO2V - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1143482 9 Perfecto 2 ™ Series Other  devices  may  experience  interference  from  ev en  the  low  levels  of  electromagnetic  emissions  permitted  by  the  above  standards.  To  determine  if  the  emissions  from  the  Pe r f e ...

  • Invacare IRC5PO2V - page 10

    SECTION 2—FEATURES Perfecto 2 ™ Series 10 Part No 1143482 SECTION 2—FEATURES Oxygen Outlet Flowmeter Oxygen Purity Indicator Lights / Fault an d Power Indicator Lights Power Switch Circuit Breaker Elapsed Time Meter Cabinet Filter REAR VIEW ACCESSORIES (NOT SHOWN): PreciseRX™ Pedi atric Humidifier/Flowmeter Accessory - IRCPF16 (Perfecto 2 5 ...

  • Invacare IRC5PO2V - page 11

    SECTION 3—HANDLING Part No 1143482 11 Perfecto 2 ™ Series SECTION 3—HANDLING The  concentrator  should  AL W A YS  be  kept  in  the  upright  position  to  prevent  cabinet  damage  while  being  transported.  If  the  concentrator  is  to  be  reshipped  by  common ? ...

  • Invacare IRC5PO2V - page 12

    SECTION 4—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™ Series 12 Part No 1143482 SECTION 4—TYPICAL PRODUCT PARAMETERS Electrical Requirements: 120 VAC ± 10% (132 VAC/10 8 VAC), 60 Hz Rated Current Input: 3 A Sound Level: 43 ± 2 dBA Average for Perfecto 2 Models and Perfecto 2 V Model 40 ± 2 dBA Average for Perfecto 2 W Models Altitude: All Perf ...

  • Invacare IRC5PO2V - page 13

    SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1143482 13 Perfecto 2 ™ Series Change in maximum recommended flow when back pressure of 7kPa is applied: 0.7 L/min Filters: Cabinet, Outlet HEPA and Compressor Inlet Safety System: Current overload or line surge shutdown. H igh temperature compressor shutdown. High Pressure Alarm w/compressor shutdow ...

  • Invacare IRC5PO2V - page 14

    SECTION 4—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™ Series 14 Part No 1143482 *NOTE:  Stated  concentration  levels  achieved  after  initial  warm ‐ up  period  (appr oximately  30  minutes). Placement: No closer than three inche s from any wall , furniture, draperies, or similar surfaces. Tubing: 7 ft cannula ...

  • Invacare IRC5PO2V - page 15

    SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 15 Perfecto 2 ™ Series SECTION 5— OPERATING INSTRUCTIONS Introduction Yo u r  oxygen  concentrator  is  intended  for  individual  use  indoors.  It  is  an  electronically  operated  device  that  separates  oxygen  from  room  air . ...

  • Invacare IRC5PO2V - page 16

    SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 16 Part No 1143482 Set Up Plug in Power Cord Plug  in  pow er  cord  to  an  electrical  outlet. Connect Humidifier (If So Prescribed)  WARNING DO  NOT  overfill  humidifier . DO  NOT  reverse  the  oxygen  input  and  output  connectio ...

  • Invacare IRC5PO2V - page 17

    SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 17 Perfecto 2 ™ Series FIGURE 5.2 Humidifer Bottle Adapter 5. Attach  the  humidifier  bottle  adapter  to  the  humidifier  bottle  by  turning  the  wing  nut  on  the  humidifier  bottle  counterclockwise  until  it  is  securel ...

  • Invacare IRC5PO2V - page 18

    SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 18 Part No 1143482 FIGURE 5.3 Humidifier Compartmen t Power Switch NOTE:  For  this  pr ocedure,  refe r  to  FIGURE  5.4. 1. Press  pow er  sw itch  to  on  position.  All  the  panel  lights  and  the  audible  alarm  will  c ...

  • Invacare IRC5PO2V - page 19

    SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 19 Perfecto 2 ™ Series Flowrate NOTE:  For  this  procedure,  refe r  to  FIGURE  5.5  on  page 20. 1. T urn  the  flowrate  knob  to  the  setting  prescribed  by  you r  physician  or  therapist. CAUTION Never  set  the  fl ...

  • Invacare IRC5PO2V - page 20

    SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 20 Part No 1143482 FIGURE 5.5 Flowrate Front Panel Indicators - Units with SensO 2 The  SensO 2  feature  monitors  the  purity  of  oxygen  generated  by  the  oxygen  concentrator .  If  purity  falls  below  factory  preset  standard ...

  • Invacare IRC5PO2V - page 21

    SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 21 Perfecto 2 ™ Series Initial Startup of the Concentrator NOTE:  Concentrator  may  be  used  during  the  initial  sta rt  warm ‐ up  time  (approximately  30  min.)  while  waiting  for  the  O 2  purity  to  rea c h  maximum. ...

  • Invacare IRC5PO2V - page 22

    SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 22 Part No 1143482 Initial Startup of the Concentrator NOTE:  Concentrator  may  be  used  during  the  initial  sta rt  warm ‐ up  time  (approximately  30  min).  O 2  purity  will  build  to  a  maximum  during  this  p ...

  • Invacare IRC5PO2V - page 23

    SECTION 6—MAINTENANCE Part No 1143482 23 Perfecto 2 ™ Series SECTION 6—MAINTENANCE  WARNING The  Invacare  concentrators  are  specifically  designed  to  minimize  routine  preventiv e  mainte nance.  Only  qualified  personnel  should  perform  preventiv e  maintenance  on  the  ...

  • Invacare IRC5PO2V - page 24

    SECTION 6—MAINTENANCE Perfecto 2 ™ Series 24 Part No 1143482 2. Clean  the  cabinet  filter  with  a  va c u u m  cleaner  or  wa s h  in  w arm  soapy  wa t e r  and  rinse  thoroughly . 3. Dry  the  filter  thoroughly  before  reinstallation. FIGURE 6.1 Cleaning the Cabinet Filter ...

  • Invacare IRC5PO2V - page 25

    SECTION 6—MAINTENANCE Part No 1143482 25 Perfecto 2 ™ Series FIGURE 6.2 Preventi ve Maintenance Record ON EACH INSPECTION MODEL NO. I R C SERIAL NO. PREVENTIVE MAINTENANCE RECORD Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS ...

  • Invacare IRC5PO2V - page 26

    SECTION 7—TROUBLESHOOTING GUIDE Perfecto 2 ™ Series 26 Part No 1143482 SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Beep......Beep. ..... Main Power Loss: 1. Power cord not plugged in. 1. Insert plug in to outlet. 2. No power at outlet. 2 . Inspe ...

  • Invacare IRC5PO2V - page 27

    SECTION 7—TROUBLESHOOTING GUIDE Part No 1143482 27 Perfecto 2 ™ Series YELLOW or RED light Illuminated. Alarm: Continuo us On RED light only. * Only applies to IRC5PO2. 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tubing, cannula or humidifier.* 2. Inspect for kinks or blockages. Correct, clean or replace item. Once ...

  • Invacare IRC5PO2V - page 28

    NOTES Perfecto 2 ™ Series 28 Part No 1143482 NOTES ...

  • Invacare IRC5PO2V - page 29

    NOTES Part No 1143482 29 Perfecto 2 ™ Series NOTES ...

  • Invacare IRC5PO2V - page 30

    NOTES Perfecto 2 ™ Series 30 Part No 1143482 NOTES ...

  • Invacare IRC5PO2V - page 31

    LIMITED WARRANTY Part No 1143482 31 Perfecto 2 ™ Series LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURE D AFTER JULY 4, 197 5. This warranty is extended only to th e original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer. This ...

  • Invacare IRC5PO2V - page 32

    LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, SOOT OR SMOKE-FILLED ENVI RONMENTS, OR OTHER THAN NORMAL APPLICATION, USE OR SERVICE, OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION'S EXPRESS WRITTEN CONSENT (INCLUDIN G, BUT NOT L ...

メーカー Invacare カテゴリー Oxygen Equipment

Invacare IRC5PO2Vのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- IRC5PO2Vの取扱説明書
- Invacareの製品カード
- パンフレット
- またはInvacare IRC5PO2Vの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Invacare IRC5PO2Vの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Invacare IRC5PO2Vの取り付け説明書、サービスマニュアル、簡易説明書、またはInvacare IRC5PO2Vのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Invacare IRC5PO2Vの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Invacare IRC5PO2Vデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがInvacare IRC5PO2Vを使用するのを助ける技術的文書のことです。説明書は通常、全てのInvacare IRC5PO2Vユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Invacareの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Invacare IRC5PO2Vの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははInvacare IRC5PO2Vの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Invacare IRC5PO2Vに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Invacare IRC5PO2Vデバイスの基本機能を使うにあたってのヒント。 Invacare IRC5PO2Vのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Invacare IRC5PO2Vに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるInvacare IRC5PO2Vのメーカー/サービスへの連絡先に関する情報。

Invacare IRC5PO2Vについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもInvacare IRC5PO2Vの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがInvacare IRC5PO2Vで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)