  • Quantum 9700 - page 1

    Quantum Composers, Inc PO Box 4248 Bozeman, MT 59772 Phone (406) 582-0227 Fax (406) 582-0237 www.quantumcomposers.com 9700 Series Pulse Generator User's Manual ...

  • Quantum 9700 - page 2

    10 Item Edit Mode ...................................................... 10 Numeric Input Mode ................................................. 10 Selecting Menu Items ................................................ 9 Overview ................................................................. 9 3 Operating Instructions ............................ ...

  • Quantum 9700 - page 3

    21 Replacing the Fuse ...................................................... 21 6 Maintenance ............................................................ 20 General ................................................................. 20 Computer Interface ...................................................... 20 Outputs .............................. ...

  • Quantum 9700 - page 4

    Table of Contents iii ...

  • Quantum 9700 - page 5

    1 INTRODUCTION This manual is designed to help you quickly and easily learn to use your Quantum Composers Model 9700 current pulse generator. The 9700 provides digitally controlled current pulses with currents from 0 - 25 Amps and pulsewidths from 100 µ s to 100ms. We also offer a long pulse version which holds up to 10A for 10 µ s - 500ms. Rathe ...

  • Quantum 9700 - page 6

    Custom Modifications The 9700 is microprocessor-controlled and employs a flexible and expandable user-interface. New features and operating modes often can be added without hardware modifications by merely modifying the software contained inside the 9700 - contact Quantum Composers for details. Safety Issues Normal use of test equipment exposes use ...

  • Quantum 9700 - page 7

    as the on-off switch, fuses, power transformer etc., any time the equipment is connected to an AC outlet, even if the equipment is turned off. 6. Never work alone. Someone should always be nearby to render aid if necessary. Training in CPR first aid is highly recommended. 3 ...

  • Quantum 9700 - page 8

    4 ...

  • Quantum 9700 - page 9

    2 CONTROLS AND CONNECTORS This chapter explains the function of each control and connector on the 9700. Front Panel The front panel contains the most frequently used controls and connectors. Display The 9700 features a backlit liquid crystal display, which displays menu selections and settings. The left half of the display shows the current menu se ...

  • Quantum 9700 - page 10

    Left Button For numeric items, this button is used to select which digit in a numeric entry is selected for editing. For other items, or when the left most digit is already selected for numeric items, pressing and holding this button down displays a brief description of the current item. Right Button For numeric items, this button is used to select ...

  • Quantum 9700 - page 11

    The charge level indicator illuminates when the capacitor bank is fully charged. The current monitor is an oscilloscope compatible reference voltage to monitor the current pulse, 1A = 100mV. External Trigger Trig In (BNC) -- external signal input to trigger the generation of pulses. Optoisolated input, 3-15V (>3mA). Pressing the RUN button in th ...

  • Quantum 9700 - page 12

    GPIB Port (optional) The GPIB (IEEE 488) interface is available as an option with the 9700. AC Input A standard EIC instrument-type cord should be used. The 9700 is supplied with a power cord that fits the type of socket used in your country. Input requirements: 100-240VAC, 1A 8 ...

  • Quantum 9700 - page 13

    3 OPERATING INSTRUCTIONS This chapter presents detailed instructions on the operation of the 9700. Overview The 9700 provides digitally controlled, single shot, current pulses. The current is provided by a 3700mF capacitor bank charged to approximately 44V. The ability of the instrument to maintain the requested current will be load dependent, limi ...

  • Quantum 9700 - page 14

    Selecting Menu Items The menu items in the 9700 are arranged in a set. Pressing the key will toggle between sets of menus, while the key will toggle between menu items within the current set. Hold down the Left arrow key to receive a short description of the current menu item. Pressing and holding the Right arrow key will display the valid range fo ...

  • Quantum 9700 - page 15

    Single Shot Mode During Single Shot operation, one pulse is generated each time the Run key is pressed. The rate at which pulses are produced is totally dependent on how fast the Run key is pressed. The delay of each of the outputs is relative to To (the internal clock). Channel Menus This set of menus is used to set enable and set the current, del ...

  • Quantum 9700 - page 16

    n :Wid - T n :Pulse Width Each of the two (2) channels has an independent pulse width setting which is set from the Tn:Wid menu, where n = 1, 2. Pulsewidths from 100 µ s to 100ms may be selected with a resolution of 200ns. *See custom features for Long Pulse option specifications. The pulse width is used to specify the duration of the output pulse ...

  • Quantum 9700 - page 17

    Shot Counter The 9700 includes a nine (9) digit shot counter. The maximum count displayed is 4,294,967,295 after which the count rolls over to zero. To manually reset the counter to zero, press the Left or Right arrow key. You will prompted to press the key to verify that you want to zero the counter. If you press any other key, the shot counter wi ...

  • Quantum 9700 - page 18

    14 ...

  • Quantum 9700 - page 19

    ____________________________________________________________________ 4 Computer Interface The 9700 comes standard with an RS232 serial interface; a GPIB (IEEE-488) interface is available as an option. All menu settings can be set and retrieved over the computer interface using a simple command language. Serial (RS232) Pinout The serial port is loca ...

  • Quantum 9700 - page 20

    Command Protocol All commands have the following syntax: $<name> <##>CR where P $ indicates a command follows. P <name> is the command name. No spaces are allowed between the $ and the <name> . The name is case sensitive. P <##> is an unsigned integer associated with the command. A space must be inserted between <na ...

  • Quantum 9700 - page 21

    where name = a valid command name. This will return an integer followed by a space, carriage return, and a line feed character (ASCII 10). Example - the following command would retrieve the current Cycles setting: $MODE ? The pulse generator will return the following: Echo Enabled: MODE ?<cr><lf> 100<cr><lf> Echo Disabled: 1 ...

  • Quantum 9700 - page 22

    Command List The following table summarizes all the commands: Sets the key click volume level. 0 - 9 VOLUME Sets the character used as a decimal point on the display. 0 = , (comma) 1 = . (period) Fmark, Enables/Disables the RS232 echo. The result string is always sent back. 0 = Disable 1 = Enable ECHO Resets shot counter. SHOT ? will return the num ...

  • Quantum 9700 - page 23

    4 SPECIFICATIONS DELAYS P CHANNELS Two (2) independent outputs, with digitally controlled delay and pulsewidth. P CURRENT 0 to 25A (Bank voltage 44V minimum) P RESOLUTION 10mA P ACCURACY 50mA P DELAY 0 to 99.9999998s P RESOLUTION 200ns P PULSEWIDTH 100 µ s to 100ms P RESOLUTION 200ns P ACCURACY 100ns + .0001 x delay P TIMEBASE 5MHz, 25PPM crystal ...

  • Quantum 9700 - page 24

    OUTPUTS (T1,T2) P IMPEDANCE 1.405 Ω P SLEW RATE > 2.5A/ µ s P OVERSHOOT < 100mA + 10% of pulse amplitude P AMPLITUDE 0-25A (loads 1 - 10 ohm) P PEAK CURRENT 25A per channel P DC CURRENT LIMIT 150mA (each channel) COMPUTER INTERFACE P RS232 9600 Baud. All instrument functions and settings may be controlled over the interface bus. GENERAL P ...

  • Quantum 9700 - page 25

    5 MAINTENANCE Except for fuse replacement, the 9700 should require no maintenance. Replacing the Fuse A pair of fuses are mounted in the power entry module. To replace fuses, follow the instructions given below: 1. Unplug the 9700 from power source and remove power cord from the back panel. 2. Remove the fuse holder by pushing down on the center ta ...

  • Quantum 9700 - page 26

    22 ...

  • Quantum 9700 - page 27

    6 CUSTOM FEATURES Long Pulse Option Specifications -10 µ s - 100ms @ 10A for the duration of the width. 23 ...

メーカー Quantum カテゴリー Metal Detector

Quantum 9700のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 9700の取扱説明書
- Quantumの製品カード
- パンフレット
- またはQuantum 9700の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Quantum 9700の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Quantum 9700の取り付け説明書、サービスマニュアル、簡易説明書、またはQuantum 9700のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Quantum 9700の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Quantum 9700デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがQuantum 9700を使用するのを助ける技術的文書のことです。説明書は通常、全てのQuantum 9700ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Quantumの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Quantum 9700の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははQuantum 9700の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Quantum 9700に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Quantum 9700デバイスの基本機能を使うにあたってのヒント。 Quantum 9700のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Quantum 9700に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるQuantum 9700のメーカー/サービスへの連絡先に関する情報。

Quantum 9700についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもQuantum 9700の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがQuantum 9700で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)