  • Emerson 1058 - page 1

    1056 PR OFIB US DP ADDENDUM For use with the 1056 units inst alled with Communications option –DP (1056-0X-2X-3X-DP) Instruction Manual Addendum PN 51-1056DP/rev .D March 2012 1056 ...

  • Emerson 1058 - page 2

    ESSENTIAL INSTRUCTIONS READ THIS P AGE BEFORE PROCEEDING! Y our instrument purchase from Rosemount Analytical, Inc. is one of the finest available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that it s performance is directly related to the ...

  • Emerson 1058 - page 3

    QUICK ST ART GUIDE Profibus 1. Refer to Model 1056 Instruction Manual, PN 51-1056 for installation instructions. 2. Ensure that the M12 interconnect cable is installed. See Section 3.0 for wiring instructions. Make, output, and power connections. 3. Once connections are secured and verified, apply power to the analyzer . 4. When the analyzer is pow ...

  • Emerson 1058 - page 4

    This manual contains instructions for inst allation and operation of the Model 1056 Dual-Input Intelligent Analyzer with Profibus DP digit al communications. The following list provides notes concerning all revisions of this document. Rev . Level Date Notes A 1/08 This is the initial release of the product manual. The manual has been reformatted to ...

  • Emerson 1058 - page 5

    i MODEL 1056DP T ABLE OF CONTENTS 1056 Profibus DP Instruction Manual Addendum For use with the 1056 units inst alled with –DP Communications ordering option 1056-0X-2X-3X-DP T ABLE OF CONTENTS QUICK ST ART GUIDE QUICK REFERENCE GUIDE T ABLE OF CONTENTS Section Title Page 1.0 DESCRIPTIONS AND SPECIFICA TIONS ...................................... ...

  • Emerson 1058 - page 6

    THIS P AGE LEFT BLANK INTENTIONALL Y ii ...

  • Emerson 1058 - page 7

    1.1 GENERAL SPECIFICA TIONS 1.1.1 Profibus Slave (St ation_T ype =0) Pure DP Device (FMS_supp =0) 1.1.2 T en Baud Rates Supported: 1.1.3 Adding slave address via Profibus DP Enabled (Set_Slave_Add_Supp =1) Automatic transmission rate recognition supported (Auto_Baud_supp =1) Modular device (Modular_S t ation =1) Max number of modules (Max_Module =2 ...

  • Emerson 1058 - page 8

    MODEL 1056DP SECTION 2.0 INTRODUCTION This page lef t blank intentionally 2 ...

  • Emerson 1058 - page 9

    3 SECTION 2.0 INTRODUCTION T O PROFIBUS ADAPTOR AND PROFIBUS NETWORKS MODEL 1056DP SECTION 2.0 INTRODUCTION 2.1 INTRODUCTION T O MODEL 1056 WITH PROFIBUS 2.2 OVERVIEW OF PROFIBUS-DP 2.3 INTRODUCTION T O PROFIBUS - GENERAL 2.1 INTRODUCTION TO MODEL 1056 WITH PROFIBUS Profibus configurations of Model 1056 (1056-0X-2X-3X-DP) allow digital communicatio ...

  • Emerson 1058 - page 10

    4 2.3 INTRODUCTION TO PROFIBUS - GENERAL Profibus is an open F IELDBUS standard complying with the harmonized European st andard EN 50170. It permits devices from different manufacturers to communicate with one another without the need for specialized interfaces. It may be used equally well for high speed, time critical data transmission or for tra ...

  • Emerson 1058 - page 11

    5 SECTION 3.0 INST ALLA TION AND WIRING 3.1 PROFIBUS BOARD 3.2 WIRING DIRECTL Y T O A PROFIBUS NETWORK 3.3 WIRING TO A PROFIBUS NETWORK WITH AN M12 CABLE ASSEMBL Y MODEL 1056DP SECTION 3.0 INST ALLA TION AND WIRING This section outlines recommended methods of wiring Model 1056 units which include a Profibus DP digital com- munications board. The co ...

  • Emerson 1058 - page 12

    6 2 . Secure the purple Profibus cable with a cable gland fittings and nut from the outside of the enclo- sure to ensure a proper seal. Note that the cable gland fittings (PN 23554-00 - Cable Gland Kit, Quantity 5) does not require the securing nut inside the enclosure to properly install the fitting. Simply thread the fitting into the grounding pl ...

  • Emerson 1058 - page 13

    7 3. Remove the 4-lead terminal block which is installed on the Profibus slide-in board. This photo shows the 4-lead terminal block installed onto the board. 4. Wire the two leads of the purple Profibus cable (red and green) to the A and B positions of the 4-lead removable terminal block. A #0 Philips head screw driver is required to open and close ...

  • Emerson 1058 - page 14

    8 6 . Ensure that the 10-lead ribbon cable is properly connected from the Profibus board to the 10-pin shrouded connector labeled “DIG I/O” on the main printed circuit board. Once power is wired to the unit (as shown in photo), the Profibus-configured Model 1056 is ready for power up and communica- tion on a Profibus network. 3.3 WIRING TO A PR ...

  • Emerson 1058 - page 15

    9 MODEL 1056DP SECTION 3.0 INST ALLA TION AND WIRING 2 . Secure the M12 cable assembly with the metal threaded nut from the outside of the enclosure to ensure a proper seal. 3. Remove the 4-lead terminal block which is installed on the Profibus slide-in board. Wire the four leads of the cable assembly to the 4-lead removable terminal block. The lea ...

  • Emerson 1058 - page 16

    10 MODEL 1056DP SECTION 3.0 INST ALLA TION AND WIRING Ensure that the 10-lead ribbon cable is properly connected from the Profibus board to the 10-pin shrouded connector labeled “DIG I/O” on the main printed circuit board (see photo). 5. Connect the Model 1056 Profibus unit to a Profibus network cable (purple) using standard Profibus connectors ...

  • Emerson 1058 - page 17

    11 MODEL 1056DP SECTION 4.0 PROGRAMMING PROFIBUS SECTION 4.0 PROGRAMMING PROFIBUS 4.1 CHANGING PROFIBUS ADDRESS A T THE DEVICE Properly wired and configured Profibus Model 1056 units can be accessed by the Profibus master af ter a GSD file is downloaded to the Master . See Appendix A for details on the GSD file available from Rosemount Analytical. ...

  • Emerson 1058 - page 18

    This page lef t blank intentionally 12 MODEL 1056DP SECTION 4.0 PROGRAMMING PROFIBUS ...

  • Emerson 1058 - page 19

    5.1 COMMUNICA TIONS ARCHITECTURE- OVERVIEW The PROFIBUS-DP communications profile specifies the operating characteristics of a serial fieldbus intended for the interconnection of distributed digital devices at the field level. T o achieve this, PROFIBUS designates these devices as either masters or slaves: • Masters, or active stations, control t ...

  • Emerson 1058 - page 20

    This page lef t blank intentionally 14 MODEL 1056DP SECTION 5.0 DEVICE CONNECTION TO PROFIBUS NETWORK ...

  • Emerson 1058 - page 21

    15 MODEL 1056DP SECTION 6.0 PROFIBUS ACCESSORIES SECTION 6.0 PROFIBUS ACCESSORIES 6.1 PROFIBUS NETWORK ACCESSORIES 6.2 MODEL 1056 PROFIBUS ACCESSORIES 6.3 SP ARE P ARTS T ABLE Accessories are available to support connection of Profibus devices to a Profibus network and Profibus master unit. 6.1 PROFIBUS NETWORK ACCESSORIES. Common Profibus network ...

  • Emerson 1058 - page 22

    16 MODEL 1056DP SECTION 6.0 PROFIBUS ACCESSORIES 6.3 Sp are part s t able Part Number Description Image 9120705 PROFIBUS M12 T -connector 9120706 PROFIBUS M12 T erminating resistor 9200344 PROFIBUS M12 Bus cable, 2-position, 1 m 9200345 PROFIBUS M12 Bus cable, 2-position, 0.5 m 9120707 PROFIBUS M12 Bus plug connector , male, straight, 5-position, s ...

  • Emerson 1058 - page 23

    17 MODEL 1056DP SECTION 7.0 PROFIBUS DIAGNOSTICS, F AUL TS, W ARNINGS SECTION 7.0 PROFIBUS DIAGNOSTICS, F AUL TS, W ARNINGS 7.1 PROFIBUS DIAGNOSTICS 7.2 PROFIBUS F AUL T AND W ARNINGS 7.3 TROUBLESHOOTING PROFIBUS F AUL T AND W ARNINGS 7.1 PROFIBUS DIAGNOSTICS Profibus-configured Model 1056 units of fer a number of useful diagnostics to assist in st ...

  • Emerson 1058 - page 24

    18 7.2 PROFIBUS F AUL T AND W ARNINGS Profibus-related Fault and W arning conditions will be automatically detected and indicated on the Model 1056 display . The main display of the Model 1056 includes a Fault and W arning banner on the lower portion of the display that will intermittently flash if any Fault or W arning condition exists. Profibus-r ...

  • Emerson 1058 - page 25

    19 7.3 TROUBLESHOOTING PROFIBUS F AUL T AND W ARNINGS Model 1056 provides on-screen User Help information and advice to directly assist the user in troubleshooting Fault and/or W arning conditions. Once the individual Fault or W arning condition is displayed, the user can access User Help for each Fault and W arning condition. MODEL 1056DP SECTION ...

  • Emerson 1058 - page 26

    20 NOTES: ...

  • Emerson 1058 - page 27

    A W orldwide Network of Sales and Service Emerson Process Management’s field sales of fices are your source for more information on the fill line of Rosemount Analytical prod- ucts. Field sales personnel will work closely with you to supply technical dat a and application information. For more information, please contact your nearest Emerson Proc ...

  • Emerson 1058 - page 28

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .rosemountanalytical.com 8 Emerson Process Management Rosemount Analytical Inc. 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .rosemountanalytical.com © ...

メーカー Emerson カテゴリー Medical Alarms

Emerson 1058のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 1058の取扱説明書
- Emersonの製品カード
- パンフレット
- またはEmerson 1058の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Emerson 1058の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Emerson 1058の取り付け説明書、サービスマニュアル、簡易説明書、またはEmerson 1058のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Emerson 1058の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Emerson 1058デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがEmerson 1058を使用するのを助ける技術的文書のことです。説明書は通常、全てのEmerson 1058ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Emersonの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Emerson 1058の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははEmerson 1058の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Emerson 1058に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Emerson 1058デバイスの基本機能を使うにあたってのヒント。 Emerson 1058のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Emerson 1058に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるEmerson 1058のメーカー/サービスへの連絡先に関する情報。