Energizer PORTABLE GAS DETECTION INSTRUMENTの取扱説明書

50ページ 0.35 mb
ダウンロード

ページに移動 of 50

Summary
  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 1

    1 MINI-SA PORT ABLE GAS DETECTION INSTRUMENT ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 2

    2 INTRINSICALL Y SAFE GAS LEVEL DETECTOR SCOTT INSTRUMENTS EXTON, P A 19341 R US C 39NO CERTIFIED BY UNDERWRITERS LABORA TORIES INC. ONL Y AS TO INTRINSIC SAFETY FOR USE IN HAZARDOUS LOCA TIONS CLASS I, DIV , 1, GROUPS A, B, C, & D. CERTIFIED TO CANADIAN CSA ST ANDARD C22.2 NO. 157-92 W ARNING : SUBSTITUTION OF COMPONENTS MA Y IMP AIR INTRINSIC ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 3

    3 SCOTT INSTRUMENTS MINI-SA PORT ABLE GAS DETECTION INSTRUMENT Content s 1. 0 INTRODUCTION ................................................................. 4 2.0 OPERA TIONAL ENVIRONMENT ........................................ 4 3.0 GENERAL INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ................ 5 3. 1 UNP ACKI ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 4

    4 SCOTT INSTRUMENTS MINI-SA PORT ABLE GAS DETECTOR 1.0 INTRODUCTION The SCOTT INSTRUMENTS MINI-SA Portable Gas Detector is a self contained instrument designed to detect and provide audible, visual, and tactile indication of the presence of oxygen or toxic gas in the air . The toxic gas concentration is monitored and shown in the graphic display in ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 5

    5 3.0 GENERAL INSTRUCTIONS 3.1 UNP ACKING Remove the SCOTT INSTRUMENTS MINI-SA instrument from its packaging and examine it for external damage. If any damage is found, notify your SCOTT INSTRUMENTS distributor , sales representative, or the SCOTT IN- STRUMENTS Service Department. 3 . 2 REGISTRA TION Fill out and mail the postage paid Registration ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 6

    6 4.0 INSTRUMENT DESCRIPTION THE SCOTT INSTRUMENTS MINI-SA INSTRUMENT is designed for use as a personal monitor . The MINI-SA is equipped with a snap mounting clip for ease of use. A break-away neck strap and rubber boot are also available. All operations of the instrument are controlled by two push-button actuators which are large enough to accomm ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 7

    7 4 . 1 ALARM LEGENDS The following alarm legends appear in the graphic display: “ALARM” flashing legend indicates one of the following: a ) Gas concentration which exceeds the preset level b ) A T ime Weighted A verage (TW A ) Alarm c) A Short T erm Exposure Level (STEL) Alarm. For toxic gas, the ppm, TWA, or STEL level detected and the “ALA ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 8

    8 4.2 AUDIBLE ALARMS The SCOTT INSTRUMENTS MINI-SA has an audible alarm which is acti- vated when any of the following conditions occurs: - The preset gas alarm level is exceeded (ALARM) - The T ime Weighted A verage (TW A) is exceeded - The Short T erm Exposure Limit (STEL) is exceeded - The battery is exhausted - The sensor is missing or incompat ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 9

    9 4.3 VIBRA TION ALARM An internal vibrating motor activates whenever the Audible Alarm sounds. The V ibration Alarm operates as a secondary alarm feature in high noise environ- ments where the audible alarm may be dif ficult to hear . The Vibration Alarm is reset the same as the Audible Alarm as described in Section 4.2. The V ibration Alarm will ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 10

    10 5. 0 SWITCH FUNCTIONS AND DISPLA YS All operations on the MINI-SA Portable Detector are controlled by two push- buttons located on the bottom of the unit. The operations vary depending on the operating mode and the graphic display appearing above the two switches. The unit sounds a brief audible “beep” whenever either pushbutton is de- press ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 11

    11 6.0 OPERA TING INSTRUCTIONS Before using this instrument, read and understand ALL instructions for ALL operating modes described in this Section. Prior to use, the instrument should be checked for proper calibration as described in this Section. W ARNING BEFORE EACH DA Y’S USAGE, THE SENSITIVITY OF THE GAS SENSOR MUST BE TESTED ON A KNOWN CONC ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 12

    12 6. 1 INSTRUMENT TURN ON / OFF T o turn the instrument ON, depress either pushbutton. Three successive “beeps” will sound. The Red LED will flash and the vibration alarm will pulse for each “beep.” The graphic display will show one of the following screens while the instrument initializes. The sensor installed in the unit defines which di ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 13

    13 sensor type has been changed or if the previous calibration was not suc- cessful, the bottom line of the display will show “CALIBRA TION REQUIRED”. For T oxic Gases 0      MENU  For O 2 20.9% O 2      MENU  In the normal Operating Mode, the following information is displayed: - Gas concentration in either ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 14

    14 6.2 MENU SELECTION From the Operating Mode screen, there are six menus that can be selected by the user . They are “ZERO CALIBRA TION,” “SP AN CALIBRA TION,” “PEAK/ HOLD,” “ALARM SETPOINT ,” “TW A/STEL,” and “CONFIGURA TION.” De- press the pushbutton below the MENU prompt on the display and then one of the following scree ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 15

    15 PEAK / HOLD NEXT SELECT T o go to the PEAK/HOLD menu, depress the “SELECT” pushbutton. T o go to the next menu, depress the lef t pushbutton below the “NEXT” com- mand and one of the following screens will be displayed: <gas> ALARM SETPOINT For T oxic Gases NEXT SELECT O 2 ALARM SETPOINT For O 2 NEXT SELECT T o go to the ALARM SETP ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 16

    16 CONFIGURA TION NEXT SELECT T o go to the CONFIGURA TION menu, press the “SELECT” pushbutton. T o go to the next menu, depress the lef t pushbutton below the “NEXT” com- mand and the ZERO CALIBRA TION for toxic gas or AUT O CALIBRA TION for oxygen screen will again be displayed. The MINI-SA will scroll repeatedly through these Menus each ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 17

    17 6.3 ZERO CALIBRA TION MENU The MINI-SA instrument can be Auto Zeroed at any time. An Auto Zero should only be performed in an area known to be free of toxic gas or any known cross sensitive materials. See Section 8.0 for SCOTT INSTRUMENTS sup- plied calibration gas cylinders. When performing an auto-zero with an oxy- gen unit, be sure the area c ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 18

    18 6. 4 SP AN CALIBRA TION MENU FOR TOXIC GAS ONL Y Note: The MINI-SA software requires that an “AUTO ZERO” be performed as described in Section 6.3 before the instrument will allow a “SP AN CALIBRA- TION”. Failure to do so will display the following screen when attempting a “SP AN CALIBRA TION” of the instrument: AUTO ZERO REQUIRED BEF ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 19

    19 Calibration cylinders supplied by SCOTT INSTRUMENTS contain 25 PPM H 2 S or 50 PPM CO. T o use another certified concentration, depress the left pushbutton below the “CHANGE” command. The one of the following screens will be displayed: <gas> SENSOR For T oxic Gases SP AN CALIBRA TION GAS = <_> ppm <CALGAS> UP DOWN Factory p ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 20

    20 This screen will remain on until the instrument detects calibration gas. If no calibration gas is detected after thirty (30) seconds, the instrument will sound a single beep and return to the operating mode. Depressing the “EXIT” pushbutton will also return the instrument to normal operation with no calibration having occurred. The last succ ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 21

    21 Attach the hose of the Adapter to the calibration gas cylinder and begin flowing gas to the cap at a flow rate of 500 cc/min. (300 cc/min. for C l 2 and ClO 2 ). When the instrument detects the gas, the display will change to one of the following screens: <gas> SENSOR For T oxic Gases SP AN CALIBRA TION <time> seconds The display wil ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 22

    22 Note: SCOTT INSTRUMENTS supplies T eflon®-lined tubing with all chlorine and ammonia calibration kits. Sticky gases such as chlorine and ammonia require this tubing be used for calibration. Using tub- ing of another type may inhibit the automatic calibration of a MINI-SA and lead to improper readings or a false failed sensor message. All other ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 23

    23 6. 6 PEAK / HOLD MENU The MINI-SA can be set to display either the lowest and highest gas reading (PEAK) or just the highest toxic gas or lowest oxygen reading (HOLD) during use. T o enter the PEAK/HOLD menu, depress the “SELECT” pushbutton when the PEAK/HOLD menu is displayed as described in Section 6.2. The following screen will appear: SE ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 24

    24 4 For T oxic Gases MAX = <_> ppm      MENU  20 For O 2 MAX = 21.5 % MIN = 20.9 %      MENU  Current and peak readings update continuously while this screen is dis- played. If the PEAK menu is accessed from a peak reading display , the option “SHOW PEAK” will read “HIDE PEAK” instead, as follows: ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 25

    25 NOTE: After an Alarm Condition occurs and the Audible Alarm is si- lenced as instructed in Section 4.0, the instrument will automatically change to the “SHOW PEAK” display to inform the user of the actual level responsible for the alarm condition. T o return to the standard operating screen, depress the pushbutton under “HIDE PEAK,” whic ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 26

    26 6.7 ALARM SETPOINT MENU The “ALARM SETPOINT” can be adjusted by the user in the ALARM SETPOINT menu within the following ranges: C O = 5 to 250 ppm Cl 2 = 0.1 to 10.0 ppm ClO 2 = 0.10 to 2.00 ppm HCN = 0.1 to 50.0 ppm H 2 S = 5 to 100 ppm NH 3 = 5 to 100 ppm NO 2 = 0.2 to 5.0 ppm O 2 LO = 0.0 to 21.0% O 2 HI = 21.0 to 25.0% PH 3 = 0.10 to 5. ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 27

    27 the “OK” pushbutton to return to the Operating Mode. T o change the setpoint, depress the “CHANGE” pushbutton. One of the following screens will be displayed: <gas> ALARM SETPOINT For T oxic Gases HI ALARM = <_ >ppm UP DOWN CHANGE O 2 ALARM For O 2 HI ALARM = 23.5% LO ALARM = 19.5% LO HI For toxic gases, choose “UP” or ? ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 28

    28 CHANGE O 2 ALARM For O 2 LO ALARM = 19.5 % UP DOWN T o raise or lower this setpoint, select the “UP” or “DOWN” command. For example, choose “DOWN” and the following screen will appear: CHANGE O 2 ALARM For O 2 LO ALARM = 19.5 % DOWN SET The % value will flash while this screen is displayed. Each time the “DOWN” pushbutton is depr ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 29

    29 Depressing the “OK” pushbutton will display one of the following screens: <gas> ALARM For T oxic Gases NON-LA TCHING LA TCH OK O 2 ALARM For O 2 NON-LA TCHING LA TCH OK The st atus of the alarms will be displayed as “NON-LA TCHING” or “LA TCHING”. S tatus can be changed by depressing the left pushbutton. Depress the “OK” pu ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 30

    30 The STEL (Short T erm Exposure Level) function measures the level of toxic gas during a fifteen (15) minute interval and alarms when the preset STEL value in parts per million is exceeded. The alarm levels are as follows: STEL TWA Ammonia 35 p pm 25 p pm Carbon monoxide 100 ppm 25 ppm Chlorine 1 p p m 0.5 ppm Chlorine Dioxide 0.3 ppm 0.1 ppm Hyd ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 31

    31 Depressing the “RESET” pushbutton will reset the TWA value, the STEL value, and time all to “0”. The instrument will automatically go back into the Operating Mode and begin collecting measurement data. The “DISABLE” pushbutton turns off the TWA/STEL functions. If the TWA/ STEL functions are off, one of the following screens will be d ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 32

    32 6.9 CONFIGURA TION MENU The CONFIGURA TION menu of the SCOTT INSTRUMENTS MINI-SA pro- vides the user with the following two (2) optional features: • BA TTERY SA VER mode can extend operational battery life to sev- eral hundred hours by turning off the instrument graphic display dur- ing normal non-alarm operation. • P ASSWORD PROTECTION menu ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 33

    33 If an alarm condition occurs while in the BA TTER Y SA VER mode, the graphic display will come on automatically and remain on until the alarm is reset. After the alarm indication clears, the graphic display will remain on for thirty (30) seconds and then the display will go blank as described above. If the BA TTER Y SA VER mode had already been ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 34

    34 has been entered, the instrument will display the following screen: P ASSWORD PROTECTION FOR ADV ANCED MENUS TURN CHANGE ON P ASSWORD The MINI-SA is shipped from the factory with password protection off. The “TURN ON” pushbutton will bring up the following display: P ASSWORD PROTECTION LOCK OUT ACCESS TO ADV ANCED MENUS? YES NO Pressing “N ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 35

    35 Use the “UP” and “SET” push-buttons to enter the correct password. Enter- ing the incorrect p assword will display an INCORRECT P ASSWORD mes- sage and allow the user to enter the password again. When the correct password has been entered, the instrument will display the following screen: P ASSWORD PROTECTION FOR ADV ANCED MENUS TURN CHA ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 36

    36 VERIFY NEW P ASSWORD 5678 UP SET Note: If the pushbutton is not depressed within fifteen (15) seconds, the instrument will beep once and automatically go back to the oper- ating mode and the password will be unchanged. If you wish to change the password shown in the VERIFY NEW P ASS- WORD screen, depress the “CHANGE” pushbutton. The previous ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 37

    37 7.0 ERROR MESSAGES W ARNING A SCOTT INSTRUMENTS MINI-SA INSTRUMENT THA T DISPLA YS ANY ERROR MESSAGE MUST BE REMOVED FROM SERVICE IM- MEDIA TEL Y AND THE F AILURE DIAGNOSED AND REP AIRED BY AUTHORIZED PERSONNEL. DO NOT ASSUME THA T A PROBLEM HAS CORRECTED ITSELF IF A PREVIOUSL Y DISPLA YED ERROR MESSAGE DOES NOT APPEAR A SECOND TIME. During cali ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 38

    38 8.0 CALIBRA TION GASES Contact your SSCOTT INSTRUMENTS distributor , sales representative, or the SCOTT INSTRUMENTS Service Center for information regarding avail- ability of Calibration Gases and accessories for the SCOTT INSTRUMENTS MINI-SA Portable Gas Detection Instrument. Calibration eguipment from SCOTT INSTRUMENTS: Calibration Kits ( Incl ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 39

    39 9.0 SPECIFICA TIONS HYDROGEN SULFIDE SPECIFICA TIONS Range: 0 to 100 ppm (for H 2 S 100) 0 to 500 ppm (for H 2 S 500) Response Time: T90 < 30 seconds Operating T emperature Range: -40 to +50 o C (-40 to +122 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 15 to 90% RH, Non-Condensing Recommended S ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 40

    40 CARBON MONOXIDE SPECIFICA TIONS Range: 0 to 500 ppm (for CO 500) 0 to 999 ppm (for CO 999) Response Time: T90 < 25 seconds Operating T emperature Range: -20 to +50 o C (-4 to +122 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 15 to 90% RH, Non-Condensing Recommended S torage T emperature: 0 to 2 ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 41

    41 SULPHUR DIOXIDE SPECIFICA TION Range: 0.0 to 99.9 ppm Response Time: T90 < 35 seconds Operating T emperature Range: -20 to +50 o C (-4 to +122 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 15 to 90% RH, Non-Condensing Recommended S torage T emperature: 0 to 20 o C (32 to 68 o F) Sensor Accuracy: ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 42

    42 PHOSPHINE SPECIFICA TIONS Range: 0.00 to 9.99 ppm Response Time: T90 < 30 seconds Operating T emperature Range: -20 to +50 o C (-4 to +122 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 15 to 90% RH, Non-Condensing Recommended S torage T emperature: 0 to 20 o C (32 to 68 o F) Sensor Accuracy: ±3 ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 43

    43 CHLORINE SPECIFICA TION Range: 0.0 to 99.9 ppm Response Time: T80 < 60 seconds Operating T emperature Range: -20 to +50 o C (-4 to +122 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 15 to 90% RH, Non-Condensing Recommended S torage T emperature: 0 to 20 o C (32 to 68 o F) Sensor Accuracy: ±3% o ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 44

    44 AMMONIA SPECIFICA TIONS Range: 0 to 200 ppm Response Time: T90 < 150 seconds Operating T emperature Range: -25 to +30 o C (-13 to +86 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 15 to 90% RH, Non-Condensing Recommended S torage T emperature: 0 to 20 o C (32 to 68 o F) Sensor Accuracy: ±3% of ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 45

    45 CHLORINE DIOXIDE SPECIFICA TION Range: 0.00 to 2.00 ppm Response Time: T80 < 60 seconds Operating T emperature Range: -20 o to +50 o C (-4 o to +122 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 15 to 90% RH, Non-Condensing Recommended S torage T emperature: 0 to 20 o C (32 to 68 o F) Sensor Acc ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 46

    46 OXYGEN SPECIFICA TIONS Range: 0.0 to 25.0% Response Time: T95 = 15 seconds Operating T emperature Range: -20 to +50 o C (-4 to +122 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 0 to 99% RH, Non-Condensing Recommended S torage T emperature: 0 to 20 o C (32 to 68 o F) Sensor Accuracy: ±3% of Applie ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 47

    47 HYDROGEN CY ANIDE SPECIFICA TION Range: 100 ppm Response Time: T90 < 200 seconds Operating T emperature Range: -20 o to +50 o C (-4 o to +122 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 15 to 90% RH, Non-Condensing Recommended S torage T emperature: 0 to 20 o C (32 to 68 o F) Sensor Accuracy: ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 48

    48 NITROGEN DIOXIDE SPECIFICA TIONS Range: 150 ppm Response Time: T90 < 25 seconds Operating T emperature Range: -20 to +50 o C (-4 to +122 o F) Graphic Display T emperature Range: -10 to +60 o C (+14 to +140 o F) Operating Humidity: 0 to 90% RH, Non-Condensing Recommended S torage T emperature: 0 to 20 o C (32 to 68 o F) Sensor Accuracy: ±3% o ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 49

    49 NOTES ...

  • Energizer PORTABLE GAS DETECTION INSTRUMENT - page 50

    50 251 Welsh Pool Road EXTON, P A 19341 T elephone: (610)363-5450 • Fax: (610)363-0167 40014802 Rev . D 08/21/03 CN #3181 ® ...

メーカー Energizer カテゴリー Smoke Alarm

Energizer PORTABLE GAS DETECTION INSTRUMENTのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- PORTABLE GAS DETECTION INSTRUMENTの取扱説明書
- Energizerの製品カード
- パンフレット
- またはEnergizer PORTABLE GAS DETECTION INSTRUMENTの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Energizer PORTABLE GAS DETECTION INSTRUMENTの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Energizer PORTABLE GAS DETECTION INSTRUMENTの取り付け説明書、サービスマニュアル、簡易説明書、またはEnergizer PORTABLE GAS DETECTION INSTRUMENTのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Energizer PORTABLE GAS DETECTION INSTRUMENTの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Energizer PORTABLE GAS DETECTION INSTRUMENTデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがEnergizer PORTABLE GAS DETECTION INSTRUMENTを使用するのを助ける技術的文書のことです。説明書は通常、全てのEnergizer PORTABLE GAS DETECTION INSTRUMENTユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Energizerの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Energizer PORTABLE GAS DETECTION INSTRUMENTの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははEnergizer PORTABLE GAS DETECTION INSTRUMENTの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Energizer PORTABLE GAS DETECTION INSTRUMENTに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Energizer PORTABLE GAS DETECTION INSTRUMENTデバイスの基本機能を使うにあたってのヒント。 Energizer PORTABLE GAS DETECTION INSTRUMENTのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Energizer PORTABLE GAS DETECTION INSTRUMENTに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるEnergizer PORTABLE GAS DETECTION INSTRUMENTのメーカー/サービスへの連絡先に関する情報。

Energizer PORTABLE GAS DETECTION INSTRUMENTについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもEnergizer PORTABLE GAS DETECTION INSTRUMENTの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがEnergizer PORTABLE GAS DETECTION INSTRUMENTで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)