  • National Instruments 6008 - page 1

    USER GUIDE AND SPECIFICATIONS NI USB-6008/6009 This guide describes how to use the National Instruments USB-6008/6009 data acquisition (D A Q) de vices and lists specif ications. Introduction The NI USB-6008/6009 provid es connection to eight analog inpu t (AI) channels, two analog outpu t (AO) channels, 12 digital input/output (DIO) channels, and ...

  • National Instruments 6008 - page 2

    NI USB-6008/6 009 User Guid e and Specifications 2 ni.com Figure 1. NI USB-6008/6009 Figure 2. NI USB-6008/6009 Back View Table 2. Differences Between the NI USB-6008 and NI USB-6009 Featur e NI USB-6008 NI USB-6009 AI Resolution 12 bits dif ferential, 11 bits single-ended 14 bits differential, 13 bits single-ended Maximum AI Sample Rate, Single Ch ...

  • National Instruments 6008 - page 3

    © National Instruments Corporation 3 NI US B-6008/6009 User G u ide and Specifications Safety Guidelines Caution Operate the hardware only as describe d in these operating instructions. The following section contains important safety information that you must follow when installing and using the NI USB-6008/6009. Do not operate the NI USB-6008/600 ...

  • National Instruments 6008 - page 4

    NI USB-6008/6 009 User Guid e and Specifications 4 ni.com Operate the device at or belo w the Measurement Cate gory I 1 . Measurement circuits are subjected to working v oltages 2 and transient stresses (ov ervoltage) from the circuit to which they are connected during measurement or test. Measurement cat egories establish standard impulse withstan ...

  • National Instruments 6008 - page 5

    © National Instruments Corporation 5 NI US B-6008/6009 User G u ide and Specifications Software Software support for t he NI USB-6008/6009 fo r Windows 2000/XP/Vista is provided by NI-DAQmx . The NI-D A Qmx CD contains example progra ms that you can use to get started programming with the NI USB-6008/ 6009. Refer to the NI-DA Qmx for USB Devices G ...

  • National Instruments 6008 - page 6

    NI USB-6008/6 009 User Guid e and Specifications 6 ni.com Hardware The following block diagram show s key functional components of the NI USB-6008/6009. Figure 3. Device Bl ock Diagram P1.<0..3> P0.<0..7> Digital I/O T erminal Block AI <0..7> AO 0 AO 1 Analog I/O T erminal Block +2.5 V/CAL +5 V/200 mA 8 Channel 12/14b ADC 12b D AC ...

  • National Instruments 6008 - page 7

    © National Instruments Corporation 7 NI US B-6008/6009 User G u ide and Specifications Setting Up Hardware Complete the f ollowing step s to set up the ha rdware: 1. Install combicon scr e w terminal blocks by inserting them into the combicon jacks. 2. Figure 4 illustrates the si gnal labels that ship i n the NI USB-6008/6 009 kit. Y ou can apply ...

  • National Instruments 6008 - page 8

    NI USB-6008/6 009 User Guid e and Specifications 8 ni.com Figure 5. Signal Label Application Diagram Note Once you label the screw terminal blocks, you must only insert them into the matching combicon jack, as indi cated by the ov erlay label on the NI USB-6008/6009 device. 4. Connect the wiring to the appropriate scre w terminals. 1 Overlay Label ...

  • National Instruments 6008 - page 9

    © National Instruments Corporation 9 NI US B-6008/6009 User G u ide and Specifications I/O Connector The NI USB-6008/6009 ship s with one detachable screw term inal block for analog signals and one detachable screw terminal block for digit al signals. These terminal blocks provide 16 connect ions that use 16 A WG to 28 AWG wire. T able 3 lists the ...

  • National Instruments 6008 - page 10

    NI USB-6008/6 009 User Guid e and Specifications 10 ni.com Table 4. Digital T erminal Assignments Module T erminal Signal 17 P0.0 18 P0.1 19 P0.2 20 P0.3 21 P0.4 22 P0.5 23 P0 6 24 P0.7 25 P1.0 26 P1.1 27 P1.2 28 P1.3 29 PFI 0 30 + 2.5 V 31 + 5 V 32 GND 32 31 30 29 28 27 26 25 24 23 22 21 20 19 18 17 ...

  • National Instruments 6008 - page 11

    © National Instruments Corporation 11 NI US B-6008/600 9 User Gu ide and Specifications Signal Descriptions Table 5 describes the signals available on the I/O connectors. Table 5. Signal Descriptions Signal Name Reference Dir ection Description GND — — Gro und —The reference point for the single-ended AI measurements, bias current return poi ...

  • National Instruments 6008 - page 12

    NI USB-6008/6 009 User Guid e and Specifications 12 ni.com LED The NI USB-6008/6009 device has a green LED next to the USB connector. When the device is connect ed to a USB port, the LED blinks steadily to indicate that the device is initialized and is receiving power from the connection. If the LED is not blinking, i t may mean that the device is ...

  • National Instruments 6008 - page 13

    © National Instruments Corporation 13 NI US B-6008/600 9 User Gu ide and Specifications PGA The progammable-gain amplifi er provides input gains of 1, 2, 4, 5, 8, 10, 16, or 20 when configured for differential measur ements and gain of 1 when configured for single-en ded measurements. The PGA gain is automatically calculated based on the voltage r ...

  • National Instruments 6008 - page 14

    NI USB-6008/6 009 User Guid e and Specifications 14 ni.com The dif ferential input mode can measure ± 20 V signals in the ± 20 V range. Ho wev er , the maximum voltage on any one pin is ± 10 V with respect to GND. For e xample, if AI 1 is + 10 V and AI 5 is –10 V , then th e measurement returned from the de vice is + 20 V . Figure 8. Example o ...

  • National Instruments 6008 - page 15

    © National Instruments Corporation 15 NI US B-6008/600 9 User Gu ide and Specifications Connecting Reference Singl e-Ended V oltage Signals To connect reference single-ende d volt age signals (RSE) to the NI USB-6008/6009, connect the positive voltage signal to the desired AI terminal, and the ground signal to a GND termin al. When no signals are ...

  • National Instruments 6008 - page 16

    NI USB-6008/6 009 User Guid e and Specifications 16 ni.com Analog Output The NI USB-6008/6009 has two independent AO chann els that can generate outputs from 0 –5 V. All updates of AO lines are software-timed. Analog Output Circuitry Figure 11 illustrates the analog outpu t circuitry for the NI USB-6008/6009. Figure 11. Analog Output Circuitry DA ...

  • National Instruments 6008 - page 17

    © National Instruments Corporation 17 NI US B-6008/600 9 User Gu ide and Specifications Minimizing Glitches on the Output Signal When you use a DAC to generate a wa veform, y ou may observe g litches in the output signal. These glitches are no rmal; when a DAQ switches from one voltage to an other, it produces g litches due to released charges. Th ...

  • National Instruments 6008 - page 18

    NI USB-6008/6 009 User Guid e and Specifications 18 ni.com Caution Exceeding the maximum input voltage ratin gs or maximum output ratings, which are listed in the Specifications section, can damage the D A Q de vice and the computer . National Instruments is not liable for any damage resulting from such signal connections. Source/Sink Information T ...

  • National Instruments 6008 - page 19

    © National Instruments Corporation 19 NI US B-6008/600 9 User Gu ide and Specifications 6. Select a static resistor v alue for your pull-up resistor th at is greater than or equal to the ideal resistance. 7. Re-connect the load circu it and the pull-up resistor . I/O Protection To protect the NI USB-6008/6009 ag ainst overvol tage, undervoltage, a ...

  • National Instruments 6008 - page 20

    NI USB-6008/6 009 User Guid e and Specifications 20 ni.com Reference and Power Sources The NI USB-6008/6009 creates an external reference and suppl ies a power source. All voltages are relative to COM unless otherw ise noted. + 2.5 External References The NI USB-6008/6009 creates a high-purity reference volt age supply for the ADC using a multi-sta ...

  • National Instruments 6008 - page 21

    © National Instruments Corporation 21 NI US B-6008/600 9 User Gu ide and Specifications Timing accuracy ........ .............. ............... 1 00 ppm of actual sample rate Input range Single-ended .............. ................. .... ± 10 V Differential ....................... ............... ± 20 V 1 , ± 10 V, ± 5 V, ± 4 V, ± 2.5 V, ± ...

  • National Instruments 6008 - page 22

    NI USB-6008/6 009 User Guid e and Specifications 22 ni.com Analog Output Analog outputs ...................... ................. .2 Output resolution .................. ..................12 bits Maximum update rate .................... .........150 Hz, software-timed Output range .................... ................... ....0 to + 5 V Output impeda ...

  • National Instruments 6008 - page 23

    © National Instruments Corporation 23 NI US B-6008/600 9 User Gu ide and Specifications Compatibility ... ................. .................... . TTL, LVTTL, CMOS Absolute maximum voltag e range ......... – 0.5 to 5.8 V with respect to GND Pull-up resistor ............. ................. ......... 4.7 k Ω to 5 V Power-on state ................ ...

  • National Instruments 6008 - page 24

    NI USB-6008/6 009 User Guid e and Specifications 24 ni.com Input high voltage ............................. ......2.0 V Input low voltage ..................... ...............0.8 V Bus Interface USB specification ............ ...................... .USB 2.0 full-speed USB bus speed .................... ................ ....12 Mb/s Power Requirement ...

  • National Instruments 6008 - page 25

    © National Instruments Corporation 25 NI US B-6008/600 9 User Gu ide and Specifications Safety Standards The NI USB-6008/6009 is designed to meet th e requirements of the following standards of saf ety for el ectrical equipment for measurement, control, and laboratory use: • IEC 61010-1, EN 61010-1 • UL 61010-1, CSA 61010-1 Note For UL and oth ...

  • National Instruments 6008 - page 26

    NI USB-6008/6 009 User Guid e and Specifications 26 ni.com Storage temperature (IEC 60068-2-1 and IEC 60068-2-2) ...... – 40 to 85 °C Storage humidity (IEC 60068-2-56) ........ .............. ............5 to 90% RH, non condensing Pollution Degree (IEC 60664) .... ............ 2 Electromagnetic Compatibility This product is designed to meet the ...

  • National Instruments 6008 - page 27

    National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more informat ion about National Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s. For patents coveri ng Natio ...

メーカー National Instruments カテゴリー Computer Drive

National Instruments 6008のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 6008の取扱説明書
- National Instrumentsの製品カード
- パンフレット
- またはNational Instruments 6008の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、National Instruments 6008の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、National Instruments 6008の取り付け説明書、サービスマニュアル、簡易説明書、またはNational Instruments 6008のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、National Instruments 6008の製品を使用するにあたって最も人気のある説明書を閲覧できます。

National Instruments 6008デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがNational Instruments 6008を使用するのを助ける技術的文書のことです。説明書は通常、全てのNational Instruments 6008ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

National Instrumentsの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. National Instruments 6008の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははNational Instruments 6008の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。National Instruments 6008に関してこのドキュメントで見つけることができる全てのヒントの目次
3. National Instruments 6008デバイスの基本機能を使うにあたってのヒント。 National Instruments 6008のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。National Instruments 6008に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるNational Instruments 6008のメーカー/サービスへの連絡先に関する情報。

National Instruments 6008についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもNational Instruments 6008の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがNational Instruments 6008で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)