    Dealer: This manual MUST be given to the end user. User: Before using this cylinder, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com Operator’s Manual HomeFill ™ Oxygen Post Valve Cylinder HF2POST6 HF2POST9 HF2POSTD “Caution: Federal law restr ...

  • Invacare 1148114 - page 2

    HomeFill ™ 2 Part No 114811 4  WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UN DERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT A HEALTHCARE PROFESSI ONAL, DEALER OR A QUALIFIED TECHNICIAN BEFORE ATTEMPTING TO USE THIS EQUIPM E NT - OTHERWISE INJURY OR DAMAGE MAY RESU ...

  • Invacare 1148114 - page 3

    TABLE OF CONTENTS Part No 1148114 3 HomeFill ™ TABLE OF CONTENTS SPECIAL NOTES ............................................... 4 RECEIVING INSPECTION .................................. 4 INTENDED USE ................................................ 4 SECTION 1—GENERAL GUIDELINES .................. 5 SECTION 2—FEATURES ........................ ...

  • Invacare 1148114 - page 4

    SPECIAL NOTES HomeFill ™ 4 Part No 114811 4 SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could re sult in personal injury or property damage. Refer to the following table for definitions of the signal words. RECEIVING INSPECTION Remove product from package and inspect for damage. If there is an ...

  • Invacare 1148114 - page 5

    SECTION 1—GENERAL GUIDELINES Part No 1148114 5 HomeFill ™ SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - IMPORTANT SAF EGUARDS contains important information for the safe operation and use of this product. DO NOT use this prod uct or any available optional equipment without fi rst completely reading and understanding these instruct ions ...

  • Invacare 1148114 - page 6

    SECTION 1—GENERAL GUIDELINES HomeFill ™ 6 Part No 114811 4  WARNING DO NOT attempt to repair product. DO NOT immerse product in any kind of liquid. Store product in a clean area when not in use. Only use an Inv acare Co rporation HF2POST6BAG or HF2POST9BAG carrying bag. NEVER smoke in an area where oxygen is being administered. AL W A YS con ...

  • Invacare 1148114 - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1148114 7 HomeFill ™  WARNING NEVER oil or lubricate the cylinder fittings. NEVER use any penetrating oil or lubricant such as WD-40 or 3 in One. If connection difficulty is experien ced v erify that the cylinder fitting is being inserted straight down into the compressor coupler . If continued difficul t ...

  • Invacare 1148114 - page 8

    SECTION 2—FEATURES HomeFill ™ 8 Part No 114811 4 SECTION 2—FEATURES FIGURE 2.1 Product Diagram Cylinder Fill Port Toggle Valve ...

  • Invacare 1148114 - page 9

    SECTION 3—TYPICAL PRODUCT PARAMETERS Part No 1148114 9 HomeFill ™ SECTION 3—TYPICAL PRODUCT PARAMETERS DIMENSIONS : HF2POST6: Weight 4.0 lbs (1.81 kg) HF2POST6: Overall L ength 11.1 in. (2.8 cm) HF2POST6: Diameter 4.4 in. (11.1 cm) HF2POST9: Weight 5.0 lbs (2.27 kg) HF2POST9: Overall Length 14 in (35.5 cm) HF2POST9: Diameter 4.4 in (11.1 cm) ...

  • Invacare 1148114 - page 10

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 10 Part No 1148114 SECTION 4—INSPECTION AND COMPRESSOR INTERFACE NOTE: Refer to the HomeFill ™ Compr essor Owner’s Manual for compressor operating instructions. Cylinder Prefill Inspection  WARNING All cylinders MUST be inspected before attempting to fill; otherwise, injury or da ...

  • Invacare 1148114 - page 11

    SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1148114 11 HomeFill ™ External Examination NOTE: For this procedur e, r efer to FIGURE 4.1 on page 12. 1. Examine the outside of the cylinder for the following conditions and replace the cylinder if they exist: • Dents or dings • Arc burns • Oil or grease • Any other signs of damage ...

  • Invacare 1148114 - page 12

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 12 Part No 1148114 FIGURE 4.1 External Examination Checking Cylinder Pressure  WARNING NEVER use tools of any kind to conn ect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. DO NOT drop oxygen cylind ers. Use two hands when handling/transp ...

  • Invacare 1148114 - page 13

    SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1148114 13 HomeFill ™ 2. Ensure toggle v alve is open. Open the toggle v alve by rotating counterclockwise. 3. Examine the gauge on the connected conserver or regulator . P erform one of the following: • If the cylinder pressure is less than 1500 psig, fill the cylinder . Connect the cyli ...

  • Invacare 1148114 - page 14

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 14 Part No 1148114 FIGURE 4.2 Component Identification 4. Remove the post v alve cov er (located on the cylinder) and the connector fillport covers (located on the compressor) (FIGURE 4.3). FIGURE 4.3 Fillport Covers 5. On the compressor unit, mome ntarily push DOWN on the outer ring (sle ...

  • Invacare 1148114 - page 15

    SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1148114 15 HomeFill ™ NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing DOWN momentarily will reset the connector fill port (GREEN dots visible) to accept the cylinder fillport. F ...

  • Invacare 1148114 - page 16

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 16 Part No 1148114 NOTE: Refer to Cylinder Fill T ime s on page 20 for the length of time it will take to fill the cylinder . FIGURE 4.5 Connecting the Cylinde r to the Compressor Cylinder Compressor Cradle Outer Ring (Sleeve) Connector Fillport Power Switch Cylinder Fillport PULL UP DETA ...

  • Invacare 1148114 - page 17

    SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1148114 17 HomeFill ™ 11. After the cylinder is full , press the compressor powe r switch to the Off (O) position (FIGURE 4.6). 12. Grasp the cylinder assembly (FIGURE 4.6). 13. W ith the other hand, grasp the outer ring (sleeve) of the connector fillport and push DOWN (FIGURE 4.6). 14. Lif ...

  • Invacare 1148114 - page 18

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 18 Part No 1148114  WARNING The fillport cover on the co nnector fillport and the post valve fillport MUST be replaced after filling and whenever not in use. If either of th e fillport covers are missing contact your healthcare provider before using. 17. Place the fillport covers onto ...

  • Invacare 1148114 - page 19

    SECTION 5—MAINTENANCE Part No 1148114 19 HomeFill ™ SECTION 5—MAINTENANCE 1. After each use, clean exterior of the product with a dry , lint free cloth only . 2. Store product in a clean area free from grease, oil, and other sources of contamination. CAUTION DO NOT use cleaning solutions. DO NOT immerse product in any kind of liquid. All repa ...

  • Invacare 1148114 - page 20

    SECTION 6—CYLINDER FILL TIMES HomeFill ™ 20 Part No 1148114 SECTION 6—CYLINDER FILL TIMES NOTE: All filling times ar e a ppr oximate and may vary on environmental conditions. * Troubleshooting If post v alv e cylinder fails to function, try the following solutions. If problem cannot be corrected, consult your dealer/home care equipment provid ...

  • Invacare 1148114 - page 21

    NOTES Part No 1148114 21 HomeFill ™ NOTES ...

  • Invacare 1148114 - page 22

    LIMITED WARRANTY HomeFill ™ 22 Part No 1148114 LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the origina l purchaser/user of our products. This warranty gives you specific legal rights and you may also have othe ...

  • Invacare 1148114 - page 23

    LIMITED WARRANTY Part No 1148114 23 HomeFill ™ LIMITATIONS AND EXCLUSIONS : THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BE EN REMOVED OR DEFACED; PRODUC TS SUBJECTED TO NEGL IGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENA NCE, OR STORAGE; PRODUCTS MODIFIED WITHOUT INVACARE’S EXPRESS WRITTEN CONSENT ...

  • Invacare 1148114 - page 24

    Invacare Corporation www.invacare.co m USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707 All rights reserved . Trademarks are identified by the symbols ™ and ®. All trademarks are owned by or licensed to Invacare Corporation unless otherwise noted. © 2007 Invacare Corporation Part No 1148114 Rev B - 9 ...

メーカー Invacare カテゴリー Oxygen Equipment

Invacare 1148114のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 1148114の取扱説明書
- Invacareの製品カード
- パンフレット
- またはInvacare 1148114の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Invacare 1148114の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Invacare 1148114の取り付け説明書、サービスマニュアル、簡易説明書、またはInvacare 1148114のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Invacare 1148114の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Invacare 1148114デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがInvacare 1148114を使用するのを助ける技術的文書のことです。説明書は通常、全てのInvacare 1148114ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Invacareの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Invacare 1148114の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははInvacare 1148114の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Invacare 1148114に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Invacare 1148114デバイスの基本機能を使うにあたってのヒント。 Invacare 1148114のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Invacare 1148114に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるInvacare 1148114のメーカー/サービスへの連絡先に関する情報。