    Dealer: This manual MUST be given to the end user. User: Before using this cylinder, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com Operator’s Manual HomeFill ® Cylinder Oxygen 540 Valve HF2E540M9 “Caution:  Federal  law  restricts ? ...

  • Invacare HF2E540M9 - page 2

    HomeFill ® Cylinder 2 Part No 116315 6  WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UN DERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT. IF YOU ARE ...

  • Invacare HF2E540M9 - page 3

    TABLE OF CONTENTS Part No 1163156 3 HomeFill ® Cylinder TABLE OF CONTENTS SPECIAL NOTES ............................................... 4 RECEIVING INSPECTION .................................. 4 INTENDED USE ................................................ 4 SECTION 1—GENERAL GUIDELINES .................. 5 SECTION 2—FEATURES ................ ...

  • Invacare HF2E540M9 - page 4

    SPECIAL NOTES HomeFill ® Cylinder 4 Part No 116315 6 SPECIAL NOTES Signal  words  are  used  in  this  manual  and  apply  to  hazards  or  unsafe  practices  which  could  result  in  personal  injury  or  property  damage.  Refer  to  the  following  table  for ? ...

  • Invacare HF2E540M9 - page 5

    SECTION 1—GENERAL GUIDELINES Part No 1163156 5 HomeFill ® Cylinder SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GU IDELINES contains important information for the safe operation and use of this product.  DANGER This product is not intended as a life-sustaining or life-supporting device.  WARNING Only  personnel  ins ...

  • Invacare HF2E540M9 - page 6

    SECTION 1—GENERAL GUIDELINES HomeFill ® Cylinder 6 Part No 116315 6  WARNING DO  NOT  attempt  to  repair  product.  DO  NOT  immerse  product  in  any  kind  of  liquid.  Store  product  in  a  clean  area  when  not  in  use.  Only  use  an  Invacare  Cor ...

  • Invacare HF2E540M9 - page 7

    SECTION 1—GENERAL GUIDELINES Part No 1163156 7 HomeFill ® Cylinder  WARNING NEVER  oil  or  lubricate  the  cylinder  fittings.  NEVER  use  any  penetrating  oil  or  lubricant  such  as  WD ‐ 40  or  3  in  One.  If  connection  difficulty  is  experienced  ve r ...

  • Invacare HF2E540M9 - page 8

    SECTION 2—FEATURES HomeFill ® Cylinder 8 Part No 116315 6 SECTION 2—FEATURES FIGURE 2.1 Product Diagram Cylinder Fill Port Cylinder Valve 540 Connection Port (with Protective Cap) ...

  • Invacare HF2E540M9 - page 9

    SECTION 3—TYPICAL PRODUCT PARAMETERS Part No 1163156 9 HomeFill ® Cylinder SECTION 3—TYPICAL PRODUCT PARAMETERS DIMENSIONS : Weight 5.1 lbs (2.31 kg ) Overall Length 13.25 in (33.6 cm) Diameter 4.4 in (11.1 cm) STORAGE CONDITIONS TEMPERATURE: -40°F (-40°C) to 140°F (60°C) MAXIMUM HUMIDIT Y: 95% ...

  • Invacare HF2E540M9 - page 10

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 10 Part No 1163156 SECTION 4—INSPECTION AND COMPRESSOR INTERFACE NOTE:  Refer  to  the  HomeFill  ®  Compr essor  Owner’s  Manual  for  compressor  operating  instructions. Cylinder Prefill Inspection  WARNING All cylinders MUST be inspe ...

  • Invacare HF2E540M9 - page 11

    SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1163156 11 HomeFil l ® Cylinder External Examination 1. Examine  the  outside  of  the  cylinder  for  the  following  conditions  and  replace  the  cylinder  if  they  exist: •D e n t s  or  dings •A r c  burns •O i l  or ? ...

  • Invacare HF2E540M9 - page 12

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 12 Part No 1163156 Checking Cylinder Pressure  WARNING NEVER use tools of any kind to conn ect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur. DO NOT drop oxygen cylind ers. Use two hands when handling/transporting oxygen cylinders. ...

  • Invacare HF2E540M9 - page 13

    SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1163156 13 HomeFil l ® Cylinder Filling the Cylinder NOTE:  For  this  procedure,  ref e r  to  FIGURE  4.1  thru  FIGURE  4.6. 1. Check  the  cylinder  pressure  before  attempting  to  fill  cylinder .  Refer  to  Checking  ...

  • Invacare HF2E540M9 - page 14

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 14 Part No 1163156 FIGURE 4.2 Fillport Covers 5. On  the  compressor  unit,  momentarily  push  DOWN  on  the  outer  ring  (sleev e)  of  the  connector  fillport  to  reset  the  connector  (FIGURE  4.3). Connector F ...

  • Invacare HF2E540M9 - page 15

    SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1163156 15 HomeFil l ® Cylinder NOTE:  If  the  outer  ring  (sleeve)  is  in  the  UP  position  (GREEN  dots  not  visible),  the  connector  fillport  will  not  be  able  to  accept  the  cylinder  fillport. ? ...

  • Invacare HF2E540M9 - page 16

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 16 Part No 1163156 NOTE:  Refer  to  Cylinder  Fill  Ti m e s  on  page 20  for  the  length  of  time  it  will  take  to  fill  the  cylinder . FIGURE 4.4 Connecting the Cylinde r to the Compressor Cylinder Compressor Crad ...

  • Invacare HF2E540M9 - page 17

    SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1163156 17 HomeFil l ® Cylinder 11. After  the  cylinder  is  full,  press  the  compressor  pow er  switch  to  the  Off  (O)  position  (FIGURE  4.5). 12. Grasp  the  cylinder  assembly  (FIGURE  4.5). 13. W ith  the  oth ...

  • Invacare HF2E540M9 - page 18

    SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ® Cylinder 18 Part No 1163156  WARNING The fillport cover on the co nnector fillport and the post valve fillport MUST be replaced after filling and whenever not in use. If either of th e fillport covers are missing contact your healthcare provider before using. 17. Place  the  fillp ...

  • Invacare HF2E540M9 - page 19

    SECTION 5—MAINTENANCE Part No 1163156 19 HomeFil l ® Cylinder SECTION 5—MAINTENANCE 1. After  each  use,  cl ean  exterior  of  the  product  with  a  dry ,  lint  free  cloth  only . 2. Store  product  in  a  clean  area  free  from  grease,  oil,  and  other  sourc ...

  • Invacare HF2E540M9 - page 20

    SECTION 6—CYLINDER FILL TIMES HomeFill ® Cylinder 20 Part No 1163156 SECTION 6—CYLINDER FILL TIMES NOTE:  All  filling  times  are  appr oximate  and  may  vary  on  environmental  conditions. * Troubleshooting If  post  va l ve  cylind er  fails  to  function,  try  the  following ? ...

  • Invacare HF2E540M9 - page 21

    NOTES Part No 1163156 21 HomeFil l ® Cylinder NOTES ...

  • Invacare HF2E540M9 - page 22

    LIMITED WARRANTY HomeFill ® Cylinder 22 Part No 1163156 LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the origina l purchaser/user of our products. This warranty gives you specific legal rights and you may al so ...

  • Invacare HF2E540M9 - page 23

    LIMITED WARRANTY Part No 1163156 23 HomeFil l ® Cylinder LIMITATIONS AND EXCLUSIONS : THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BE EN REMOVED OR DEFACED; PRODUC TS SUBJECTED TO NEGL IGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENA NCE, OR STORAGE; PRODUCTS MODIFIED WITHOUT INVACARE’S EXPRESS WRITTE ...

  • Invacare HF2E540M9 - page 24

    Invacare Corporation www.invacare.co m USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707 All rights reserved . Trademarks are identified by the symbols ™ and ®. All trademarks are owned by or licensed to Invacare Corporati on or its subsidiaries unless otherwise noted. © 2009 Invacare Corporation Part ...

メーカー Invacare カテゴリー Oxygen Equipment

Invacare HF2E540M9のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- HF2E540M9の取扱説明書
- Invacareの製品カード
- パンフレット
- またはInvacare HF2E540M9の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Invacare HF2E540M9の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Invacare HF2E540M9の取り付け説明書、サービスマニュアル、簡易説明書、またはInvacare HF2E540M9のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Invacare HF2E540M9の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Invacare HF2E540M9デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがInvacare HF2E540M9を使用するのを助ける技術的文書のことです。説明書は通常、全てのInvacare HF2E540M9ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Invacareの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Invacare HF2E540M9の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははInvacare HF2E540M9の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Invacare HF2E540M9に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Invacare HF2E540M9デバイスの基本機能を使うにあたってのヒント。 Invacare HF2E540M9のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Invacare HF2E540M9に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるInvacare HF2E540M9のメーカー/サービスへの連絡先に関する情報。

Invacare HF2E540M9についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもInvacare HF2E540M9の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがInvacare HF2E540M9で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)