Konica Minolta PULSOX-300の取扱説明書

32ページ 1.51 mb
ダウンロード

ページに移動 of 32

Summary
  • Konica Minolta PULSOX-300 - page 1

    OXYGEN SA TURA TION MONITOR PULSOX -300 ® Instruction Manual This instrument must be used according to the instructions of the doctor . ...

  • Konica Minolta PULSOX-300 - page 2

    Safety Symbols W arnings and precautions noted in this manual and the instrument are indicated by the following markings, designed to prevent accidents caused by erroneous handling of the equipment. This indicates text consisting of a warning or precaution relating to safety . Please read the text carefully and use the equipment safely . This indic ...

  • Konica Minolta PULSOX-300 - page 3

    1 Foreword Safety Precautions T o ensure correct use of this instrument, read the following points carefully and adhere to them. After you have read this manual, keep it in a safe place where it can be referred to anytime a question arises. W ARNING (Failure to adhere to the following points may result in death or serious injury .) The instrument i ...

  • Konica Minolta PULSOX-300 - page 4

    2 Foreword CAUTION (Failure to adhere to the following points may result in injury or damage to the in- strument or other property .) Do not use batteries other than those specified by KONICA MINOL T A SENSING, INC. When installing batteries in the instrument, make sure that they are correctly oriented according to the mark. If alkali uid from t ...

  • Konica Minolta PULSOX-300 - page 5

    3 Foreword Contents Safety Symbols Safety Precautions ............................................................. 1 Foreword ............................................................................4 Safety Notes ........................................................................... 4 Package Contents...................................... ...

  • Konica Minolta PULSOX-300 - page 6

    4 Foreword Foreword Safety Notes CAUTION: U.S.Federal law restricts this device to sale by or on the order of a physician. The instrument is designed for measurement of the oxygen saturation (SpO 2 ) of arterial blood and the pulse rate. Do not use it for any other purposes, such as warning of sleep apnea and breathing abnormalities. The Oxygen Sat ...

  • Konica Minolta PULSOX-300 - page 7

    5 Foreword Package Contents Before using the instrument, make sure that the following items are present. 1) PULSOX-300 main body × 1 2) Wristband WB-300 × 1 3) AAA-size alkaline battery × 1 4) Instruction manuals • PULSOX-300 instruction manual × 1 • Separate manual (Using manual for Probes) × 1 ✻ The probe is optional. About Probe The f ...

  • Konica Minolta PULSOX-300 - page 8

    6 Foreword System Diagrams and Accessories The instrument consists of the following items. 05,3/ 8 Optional accessories Standard accessories Wristband WB-300 AAA-size alkaline battery Finger Clip Probe SR-5C Spot Check Probe SP-5C Monitor Probe LM-5C Personal Probe SD-5C Extension cable EC-300 Standard accessories Optional accessories ...

  • Konica Minolta PULSOX-300 - page 9

    7 Foreword Names of Parts PULSOX-300 Power button Probe connector Wristband Battery cover Used to start/end measurement. Measurement results and messages are displayed. Connect the probe to this connector . Used to set the instrument on your wrist. The cover can be opened by turning it approximately 90 degrees. Display (LED) ...

  • Konica Minolta PULSOX-300 - page 10

    8 Foreword Notes on Use • This instrument should be used under the following operat - ing conditions. • T emperature of 0 to 40°C (32 to 104°F), relative humidity of 30 to 85%, with no condensation. • Atmospheric pressure of 700 to 1060 hPa (altitude of – 400 to 3000m). • This is a precision instrument. T o avoid the possibility of it b ...

  • Konica Minolta PULSOX-300 - page 11

    9 Foreword Differenc e f ro m t he SaO 2 value to displayed value <T able 1> <T able 2> SaO 2 HbCO SaO 2 Methemoglobin 1% 5% 10% 1% 5% 10% 50 50 70 70 90 90 100 100 Measured value is always lower than th e a ctual value. Displayed value is sometimes higher than th e a ctual value if SaO 2 is around 50%. -0.1 -0.7 -1.5 +0.2 +1.3 +3.2 -0. ...

  • Konica Minolta PULSOX-300 - page 12

    10 Foreword Notes on Storage • This instrument should be stored under the following storage conditions. • T emperature of –10 to 60°C (14 to 140°F), relative humid- ity of 10 to 95%, with no condensation. • Atmospheric pressure of 700 to 1060 hPa (altitude of –400 to 3000m). • When storing the instrument: • Do not store the instrume ...

  • Konica Minolta PULSOX-300 - page 13

    11 Foreword Disposal Method • When disposing of the used battery , insulate the polarity ter - minals with insulating tape etc. If the terminals of the battery come into contact with metal objects, heat generation, explo- sion or re may result. • Make sure that the battery is either disposed of or recycled correctly in accordance with local ...

  • Konica Minolta PULSOX-300 - page 14

    12 Preparations Preparations Before starting measurements using this instrument, follow the procedure given below to ensure that measurements are stable. 1) Attaching the Wrist Band Attach the Wrist Band WB-300 (supplied as a standard accessory) to the instrument. Since the WB-300 is made of elastic material, it allows for exibility and comfort ...

  • Konica Minolta PULSOX-300 - page 15

    13 Preparations 4 Pass the other end of the band outward through band xture B. ◆  T ake care not to twist the band. 2) Connecting the probe Connect the probe’s connector plug into the Probe Connector on the instrument. ◆ Make sure that probe’s connector plug is oriented correctly . ◆ Do not connect it in with excessive force. Doing ...

  • Konica Minolta PULSOX-300 - page 16

    14 Preparations 3) Installing the battery Procedure 1 T urn the battery cover clockwise approximately 90 degrees to open the cover . ◆  When turning the cover , keep the flat part of your finger on the cover ’ s catch. This facilitates opening the cover with minimum force. 2 Insert the battery into the battery compartment according to the po ...

  • Konica Minolta PULSOX-300 - page 17

    15 Measurement Measurement 1) Starting the Measurement 1 Attach the probe to a nger of an adult patient. 2 Press the Power button. ● The Power is ON, all the elements on the display are displayed for about 2 seconds, and then the display changes to the measurement display . ◆  Until the measurement value has been calculated and displayed, ...

  • Konica Minolta PULSOX-300 - page 18

    16 Measurement 2) Reading the display When the instrument is turned ON, all the elements of the LED will light up. The pulse level, pulse rate and SpO 2 value are displayed during measurement. Pulse rate (P .R.) If an error occurs, an error message mainly about P .R. will be displayed. Battery indicate Flashing when battery is nearly empty . Pulse ...

  • Konica Minolta PULSOX-300 - page 19

    17 Measurement 3) Ending the Measurement About Auto Power-Off function The instrument has an auto power-off function that operates independently . If no probe is connected to the PULSOX ("C" blinks) or no probe is attached to the patient ("L" blinks) continuously for more than 2 minutes, the Power is automatically switched off. ...

  • Konica Minolta PULSOX-300 - page 20

    18 Check Points T rouble Shootings Error Messages The instrument display error messages when the error occurs. Check and correspond to them correctly . %RROR-ESSAGES #AUSE 3OLUTION s  #ONNECTTHEPROBETOTHE 05,3/8PROPERLY7HENUSING THEOPTIONAL%XTENSION#ABLE %#CONNECTTH ...

  • Konica Minolta PULSOX-300 - page 21

    19 Check Points s  #HECKTHATTHEPROBEISPROPERLY ATTACHEDTO THEPATIENT)FT HEPROBE ISPROPERLYATTACHEDBUTTHIS ME SSAG ES TILL AP PEAR S THE CIR CUL ATIO N INT HEARE ABEING MEASU REDMA YNOT BE GOOD -OV ETH EPRO BETO AN AREA ...

  • Konica Minolta PULSOX-300 - page 22

    20 Check Points %RROR-ESSAGES #AUSE 3OLUTION s  !TTACHTHEPROBETOTHEBODY CORRECTLY)FTHISMESSAGEIS DISPLAYEDWHENTHEUNITISATTACHED TOTHEPATIENTCORRECTLYTHE MEASUREMENTCONDITIONSARE REGARDEDASBLOODCIRCULATIONOR OTHERCIRCULATORYPROBLEMS? ...

  • Konica Minolta PULSOX-300 - page 23

    21 Check Points %RROR-ESSAGES #AUSE 3OLUTION 4HISMESSAGEINDICATESTHATTHE MEASUREDVALUEISOUTSIDETHE ACCURACYGUARANTEEDRANGE s 4HE3P/  VALUEISBELOW THELOWERL IMIT3P/   OFTHEMEASUREMENT RANGE4HEMEASURED VALUEWILLSTILLBE? ...

  • Konica Minolta PULSOX-300 - page 24

    22 Check Points Maintenance and Inspection • Before using the instrument, make sure that there is no dam - age to the instrument, no damage or wire-breakage in the probe cable and the instrument operates correctly and safely . Cleaning • Dampen a soft cloth with neutral detergent or water , wipe the instrument with it, then wipe off carefully w ...

  • Konica Minolta PULSOX-300 - page 25

    23 Appendix Specication ■ Dual-wavelength pulse type (665nm/880nm) arterial blood oxygen satura- tion measuring instrument • Name: Oxygen Saturation Monitor • Model name: PULSOX-300 ■ Functions: • Measuring range: SpO 2 : 0 to 100% SpO 2 Pulse rate: 30 to 230 bpm • Accuracy: SpO 2 : ±2% SpO 2 (70 to 100% range, 1S.D.) Pulse rate: ± ...

  • Konica Minolta PULSOX-300 - page 26

    24 Appendix Probes (SR-5C, SP-5C, LM-5C, SD-5C) ■ LED-SPD probe ■ T ype • Finger Clip Probe SR-5C • Spot Check Probe SP-5C • Monitor Probe LM-5C • Personal Probe SD-5C ✻ It can be used with one of Extension Cable EC-300 ■ Components that touch the body Cover: Polycarbonate Wrist band: Polyethylene terephthalate, nylon 6 Rating namep ...

  • Konica Minolta PULSOX-300 - page 27

    25 Appendix Appendix Calculation Method for Displayed V alues <Pulse level meter> The pulse level meter displays the transmitted light amount ratio calculated by the following equation. The pulse level is indicated in eight levels (0 to 15%, full scale if over 10%). <Oxygen saturation (SpO 2 )> Updates the moving average of the oxygen s ...

  • Konica Minolta PULSOX-300 - page 28

    26 Appendix EMC Guidance Guidance and manufactures declaration - electromagnetic emissions Guidance and manufactures declaration - electromagnetic immunity The Oxygen Saturation Monitor Model PULSO X-300 is intended for use in the electromagnetic environment specified below. The customer or the user of the Model PULSO X-300 should assure tha ...

  • Konica Minolta PULSOX-300 - page 29

    27 Appendix Guidance and manufactures declaration - electromagnetic immunity The Oxygen Saturation Monitor Model PULSO X-300 is intended for use in the electromagnetic environment specified below. The customer or the user of the Model PULSO X-300 should assure that it is used in such anenvironment. Immunity test IEC 60601 T est level Compliance ...

  • Konica Minolta PULSOX-300 - page 30

    28 Appendix ■ Spectral Absorption of Hb and HbO 2 Absorption coefficient Wa velength Red light Infrared light Measurement Principle This instrument measures the oxygen saturation (SpO 2 ) in arte - rial blood and pulse rate continuously in a non-invasive method by applying optical principles. With this instrument, SpO 2 is dened by the followi ...

  • Konica Minolta PULSOX-300 - page 31

    29 Appendix Relation between Oxygen Saturation and Partial Pressure The relation between oxygen saturation (SpO 2 ) and oxygen partial pressure (PaO 2 ) is shown in the graph below. SpO 2 is the oxygen saturation as measured by pulse oximeters. ■ Oxygen Saturation vs. Oxygen Partial Pressure The curve of the above graph may shift to the right or ...

  • Konica Minolta PULSOX-300 - page 32

    AJF AKK Printed in Japan 9222-1732-34 © 2007-2009 KONICA MINOL T A SENSING, INC. Manufacturer 3-91, Daisen-nishimachi, Sakai-ku, Sakai-shi, Osaka 590-8551, Japan ...

メーカー Konica Minolta カテゴリー Oxygen Equipment

Konica Minolta PULSOX-300のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- PULSOX-300の取扱説明書
- Konica Minoltaの製品カード
- パンフレット
- またはKonica Minolta PULSOX-300の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Konica Minolta PULSOX-300の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Konica Minolta PULSOX-300の取り付け説明書、サービスマニュアル、簡易説明書、またはKonica Minolta PULSOX-300のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Konica Minolta PULSOX-300の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Konica Minolta PULSOX-300デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがKonica Minolta PULSOX-300を使用するのを助ける技術的文書のことです。説明書は通常、全てのKonica Minolta PULSOX-300ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Konica Minoltaの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Konica Minolta PULSOX-300の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははKonica Minolta PULSOX-300の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Konica Minolta PULSOX-300に関してこのドキュメントで見つけることができる全てのヒントの目次
3. Konica Minolta PULSOX-300デバイスの基本機能を使うにあたってのヒント。 Konica Minolta PULSOX-300のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Konica Minolta PULSOX-300に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるKonica Minolta PULSOX-300のメーカー/サービスへの連絡先に関する情報。