    Owner’s Operator and Maintenance Manual DEALER: This manual MUST be given to the user of the product. USER: B E F OR E u s i n g t h i s product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com MA90 Series MA90Z Alternating Rotation with On Demand ...

  • Invacare MA90Z - page 2

    SPECIAL NOTES MA90 Series 2 Part No 1148139  WARNING DO NOT use this product or any availa ble option al equipmen t without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manu als or instruction sheets supplied with this pr oduct or optional equipmen t. I ...

  • Invacare MA90Z - page 3

    TABLE OF CONTENTS Part No 1148139 3 MA90 Series TABLE OF CONTENTS SPECIAL NOTES ................................................................................ 2 REGISTER YOUR PR ODUCT ............................................................... 4 TYPICAL PRODUCT PARAMETERS .................................................... 5 Electrical Param ...

  • Invacare MA90Z - page 4

    TABLE OF CONTENTS MA90 Series 4 Part No 1148139 TABLE OF CONTENTS Low Pressure LED ............. ............................................. .................................... .................................19 Displaying the Total Run Ti me (MA95Z) .............. ............................................. .................. ..........19 Po ...

  • Invacare MA90Z - page 5

    TYPICAL PRODUCT PARAMETERS Part No 1148139 5 MA90 Series TYPICAL PRODUCT PARAMETERS Electrical Parameters Performance Parameters MA90Z MA95Z INPUT VOLTAGE AC: 90 V INPUT FREQUENCY: 60 Hz CURRENT: 1 A 2 A MAXIMUM POWE R CONSUMPTION: 30 ± 10 W 150 ± 10 W CIRCUIT PROTECTION: Dual fused, 250 V, 1 A fast blow fuses Dual fused, 250 V, 5 A fast blow fus ...

  • Invacare MA90Z - page 6

    TYPICAL PRODUCT PARAMETERS MA90 Series 6 Part No 1148139 Mechanical Parameters Environmental Parameters MA90Z MA95Z CONTROL UNIT DIMENSIONS (L X W X H): 15” x 6½” x 11” CONTROL UN IT WEIGHT: 15 lbs CONTROL UNIT POWER CORD: 10 - 14 Feet Long, Hospital Grade CONTROL UNIT CO NNECTION: Quad Flow - Four ¼” Single Quick Coupling Triple Flow - T ...

  • Invacare MA90Z - page 7

    LABEL LOCATION Part No 1148139 7 MA90 Series LABEL LOCATION DANGER-EXPLOSION HAZARD: DO NOT use in the presence of flammable anesthetics. CAUTION: Equipment should be co nnected to a properly grounded receptacle (3-prong). Risk of El ectrical shock. DO NOT remove back. Disconnect air hose before administering CPR. NOTE:  This  label  is ? ...

  • Invacare MA90Z - page 8

    SECTION 1—GENERAL GUIDELINES MA90 Series 8 Part No 1148139 SECTION 1—GENERAL GUIDELINES  WARNING SECTION 1 - GENERAL GUIDELIN ES contai ns important inform ation for the safe operation and use of this product. Contraindications ALW AYS  consult  the  patient’ s  physician  before  using  the  MA90Z  and  MA95Z ...

  • Invacare MA90Z - page 9

    SECTION 1—GENERAL GUIDELINES Part No 1148139 9 MA90 Series Fire Hazard  DANGER Smoking DO NOT SMOKE while using this device. This system uses room air for circulation through the mattress. A cigarette can burn a hole in the bed surface and cause damage to the mattress. Also , patient clothing, bed sheets, etc. may be combustible and cause a fi ...

  • Invacare MA90Z - page 10

    SECTION 2—OPERATION MA90 Series 10 Part No 1148139 SECTION 2—OPERATION CAUTION The control unit and mattress on the MA90Z serie s are designed to be used as a system. DO NOT replace matt res ses or control units with other mode ls or other brands. Otherwise, damage to the system may occur . Co ntact your supplier to get the correct replac ement ...

  • Invacare MA90Z - page 11

    SECTION 2—OPERATION Part No 1148139 11 MA90 Series FIGURE 2.1 Installing the Mattre ss Replacement System Installing the Side Rails  WARNING Patient entrapment with bed side rails ma y cause injury or death. Mattress MUST fit bed frame and side rails snugly to prevent pati ent entrapment. Follow the manufacturer’s instructions. Monitor pati ...

  • Invacare MA90Z - page 12

    SECTION 2—OPERATION MA90 Series 12 Part No 1148139 Connecting the Hose CAUTION Ensure that the hose connecti ng the control unit to the mattress is routed such that it cannot be stepped on, kinked, squeezed or otherwise damaged. NOTE:  For  this  proced ure,  ref er  to  FIGURE 2.3. 1. Locate  the  hose  at  the  ...

  • Invacare MA90Z - page 13

    SECTION 2—OPERATION Part No 1148139 13 MA90 Series FIGURE 2.3 Connecting the Hose Connecting the Power Cord  WARNING DO NOT alter plug to fit a non-conforming o utlet. Inst ead, have an electrician install a properly grounded outlet. Failure to use the co rrect plug and outlet can result in a potential safety hazard. CAUTION Ensure that the po ...

  • Invacare MA90Z - page 14

    SECTION 2—OPERATION MA90 Series 14 Part No 1148139 Using the Front Panel NOTE:  For  this  proced ure,  ref er  to  FIGURE 2.4. FIGURE 2.4 Using the Front Panel Power Button 1. T o  turn  the  control  unit  on  or  off,  press  and  release  the  Po we r  button  () . NOTE:  Onc e ? ...

  • Invacare MA90Z - page 15

    SECTION 2—OPERATION Part No 1148139 15 MA90 Series Turn Button (MA90Z) 1. Press  the  Tu r n  button  ()  to  select  one  of  three  rotation  modes  or  to  set  the  unit  to  No  Tu r n  (Static)  mode.  NOTE:  The  rotation  modes  ar e  right,  left  or ? ...

  • Invacare MA90Z - page 16

    SECTION 2—OPERATION MA90 Series 16 Part No 1148139 3. Use  the  Select  button  to  move  the  pointer  to  the  times  or  attributes  av ailable  for  the  selected  function. 4. Use  the  Firmer  or  Softer  buttons  to  change  the  times  or  attributes. Hold ...

  • Invacare MA90Z - page 17

    SECTION 2—OPERATION Part No 1148139 17 MA90 Series Programming the Automatic Wi reless Fowler Transmitter 1. Ensure  the  transmitter  is  inside  the  mattress  pocket  located  on  the  right  side  of  the  mattress  at  the  head  end. 2. Ensure  the  mattress  is  in  a ...

  • Invacare MA90Z - page 18

    SECTION 2—OPERATION MA90 Series 18 Part No 1148139 Mode Button (MA90Z) 1. Press  the  Mode  button  ()  to  select  the  Max  Inflate  Mode  or  the  Low  Air  Loss  mode. •M a x  Inflate  Mode ‐ In  this  mode,  the  mattress  inflates  rapidly  to  maximum ? ...

  • Invacare MA90Z - page 19

    SECTION 2—OPERATION Part No 1148139 19 MA90 Series Lock/Alarm Silence Button 1. Press  the  Lock/Alarm  Silence  button  ()  to  select  one  of  the  following  modes:  •L o c k  Mode ‐  Pressing  this  button  locks  out  all  control  unit  functions,  including  ...

  • Invacare MA90Z - page 20

    SECTION 2—OPERATION MA90 Series 20 Part No 1148139 Placing the Patient on the Mattress 1. Press  the  Mode  button  or  Max  Inflate  button  to  turn  on  the  control  unit  to  maximu m  flow .  NOTE:  In  this  mode,  the  Max  Inflate  LED  lights  up. 2. After  ...

  • Invacare MA90Z - page 21

    SECTION 2—OPERATION Part No 1148139 21 MA90 Series 4. Raise  or  lower  the  bed  to  match  the  gurney  height.  Refer  to  the  owner ’ s  manual  provided  with  the  bed. 5. When  the  mattress  has  reached  maximum  firmness,  perform  one  of  the  fo ...

  • Invacare MA90Z - page 22

    SECTION 2—OPERATION MA90 Series 22 Part No 1148139 FIGURE 2.5 Preparing for CPR Procedure Connector Cover Control Unit Connectors MA95Z Hose Connector DETAIL “A” Control Unit RED CPR Connector (MA90Z Only) Mattress DETAIL “B” DETAIL “D” MA90Z Hose Connector Tab DETAIL “C” ...

  • Invacare MA90Z - page 23

    SECTION 3—MAINTENANCE AND TROUBLESHOOTING Part No 1148139 23 MA90 Series SECTION 3—MAINTENANCE AND TROUBLESHOOTING Cleaning the System  WARNING Before cleani ng or disass embling the MA90 series, ch eck the underside of the mattress folds for sharp object s such as scissors, needles, etc. These objects should be removed and discarded before ...

  • Invacare MA90Z - page 24

    SECTION 3—MAINTENANCE AND TROUBLESHOOTING MA90 Series 24 Part No 1148139 CAUTION Laundry workers sh ould al ways wash their hands befo re working with clean bedding. DO NOT overload the machine. DO NOT use chlorine bleach because it may damage the fa bric coating. High air temperatures will damage the fabrics and vo id the Invacare warranty. i. P ...

  • Invacare MA90Z - page 25

    SECTION 3—MAINTENANCE AND TROUBLESHOOTING Part No 1148139 25 MA90 Series Changing the Filter NOTE:  For  this  proced ure,  ref er  to  FIGURE 3.1  on  page 25. Clean  the  filter  every  5  months  or  whenever  it  is  dirty .  1. Remove  the  two  screws  securing  the  ...

  • Invacare MA90Z - page 26

    SECTION 3—MAINTENANCE AND TROUBLESHOOTING MA90 Series 26 Part No 1148139 No power Control unit off Check power source and turn on unit Power cord disconnected Connect power cord to the power source No power in the power source Check power source has power power and turn it On Power outage Wait till the powe r source has power Blown fuse Replace b ...

  • Invacare MA90Z - page 27

    SECTION 3—MAINTENANCE AND TROUBLESHOOTING Part No 1148139 27 MA90 Series NOTES ...

  • Invacare MA90Z - page 28

    LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BE EN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MA NUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purch aser who purchases this product when new and unused from Invacare or a dealer. This warranty is not extended to any other person or entity and is ...

メーカー Invacare カテゴリー Camping Equipment

Invacare MA90Zのメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- MA90Zの取扱説明書
- Invacareの製品カード
- パンフレット
- またはInvacare MA90Zの消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、Invacare MA90Zの取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、Invacare MA90Zの取り付け説明書、サービスマニュアル、簡易説明書、またはInvacare MA90Zのユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、Invacare MA90Zの製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

Invacare MA90Zデバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがInvacare MA90Zを使用するのを助ける技術的文書のことです。説明書は通常、全てのInvacare MA90Zユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

Invacareの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. Invacare MA90Zの説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははInvacare MA90Zの目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。Invacare MA90Zに関してこのドキュメントで見つけることができる全てのヒントの目次
3. Invacare MA90Zデバイスの基本機能を使うにあたってのヒント。 Invacare MA90Zのユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。Invacare MA90Zに関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるInvacare MA90Zのメーカー/サービスへの連絡先に関する情報。

Invacare MA90Zについてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもInvacare MA90Zの問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがInvacare MA90Zで同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)