    MultiT emp ® III User Manual 18-1 106-33 Edition AD A P H A R M A C I A B I O T E C H ...

  • TRENDnet 18-1106-33 - page 2

    Reading this entire manual is necessary for full under- standing and safe use of this product. Should you have any comments on this manual, we will be pleased to receive them at: Pharmacia Biotech AB S-75182 Uppsala Sweden Pharmacia Biotech reserves the right to make changes in the specifications without prior notice. W arranty and Liability Pharma ...

  • TRENDnet 18-1106-33 - page 3

    Contents 1. Introduction ..................................................................... 3 2. Safety information ........................................................... 3 2.1 Safety precautions ................................................. 3 2.2 Safety standards .................................................... 4 2.3 Safety feature ...

  • TRENDnet 18-1106-33 - page 4

    2 ...

  • TRENDnet 18-1106-33 - page 5

    1. Introduction and 2. Safety information 1. Introduction MultiT emp ® III circulates constant temperature liquid to external electrophoresis equipment. The liquid temperature range is -10 to +90 °C. The unit contains KLEA 134a*, an ozone friendly HFC-type coolant. MultiT emp III both supplies coolant for high voltage techni- ques and provides te ...

  • TRENDnet 18-1106-33 - page 6

    1. Introduction and 2. Safety information 7. Do not use the circulator beyond its specifications or for other than its intended use. 8. Do not touch the stainless steel bath cover and tubings when the unit is used at a high working temperature. Under these conditions, the surface temperature can rise up to approximately 70 °C. MultiT emp III is de ...

  • TRENDnet 18-1106-33 - page 7

    1. Introduction and 2. Safety information General cut-off The heater , the compressor and the circulation pump can be switched off by a relay built into the unit. This general cut- off will be activated if a malfunction is detected by the safety sensors. See 7.4, Description of Safety Features. Protection against over-heating A built-in temperature ...

  • TRENDnet 18-1106-33 - page 8

    3. Installation 3. Installation 1. Check the contents against the packing list supplied. Inspect for any damage that may have occurred during transit. Report any damage immediately to your local Pharmacia Biotech representative and to the transport company concerned. 2. Place MultiT emp III in its working location, ensuring ade- quate space for ven ...

  • TRENDnet 18-1106-33 - page 9

    3. Installation Silicone tubing (inner diameter 8 mm) can be used over the entire working temperature range of the unit. PVC tubing can only be used in the temperature range of 10 to 60 °C. 1. Connect the inlet and outlet tubing to the nozzles on the rear of the circulator (Fig. 1). 2. Connect the other ends to the electrophoresis unit. The nozzle ...

  • TRENDnet 18-1106-33 - page 10

    3. Installation liquid pour into a large vessel using hydrostatic pressure. The bath contains 3 litres of liquid, plus the volume in tubing and connected equipment. The last millilitres can be removed by opening the screw on the front of the unit, (see Fig. 1). W arning! Do not drain hot water . Use protective gloves and clothing when draining the ...

  • TRENDnet 18-1106-33 - page 11

    4. Operation 4. Operation Fig. 2. Front panel of MultiT emp III. 1. Mode selector 6. Mains switch 2. Set value down 7. Control lamp cooling on/off 3. Set value up 8. Display 4. Enter 9. Heater control sign 5. Reset 10. Display Before switching on, confirm that the outlet and inlet nozzles are either firmly connected to external equipment or to each ...

  • TRENDnet 18-1106-33 - page 12

    4. Operation the external system and the level in the bath will fall. Add more liquid until the level is once again 2 cm below the top plate. If the safety circuit is triggered, please top up with more liqu- id and restart the unit using . Repeat this action if necessary . Note: Check the liquid level after the unit has been in use for a while. Whe ...

  • TRENDnet 18-1106-33 - page 13

    4. Operation 2. Use or to set the maximum allowed set tempe- rature. The factory pre-set value is 90.0 °C. 3. Confirm with before the display returns to the actu- al temperature value. Note: The new value is not saved until has been pres- sed. 1. Press to select the S mode. 2. Use or to set the working temperature. 3. Confirm with before the displ ...

  • TRENDnet 18-1106-33 - page 14

    4. Operation 5. Confirm with before the display returns to the actual temperature value. Note: The new value is not saved until has been pres- sed. 12 ...

  • TRENDnet 18-1106-33 - page 15

    5. T rouble shooting 5. T rouble shooting Message on display Cause Remedy The bath temperature has exceeded Pour in more liquid. the safety temperature limit. T oo low water level. Pour in more liquid. The winding of the pump motor is too hot. Check pump and wait a few minutes until the motor temperature has dropped. If this fault occurs often, ret ...

  • TRENDnet 18-1106-33 - page 16

    6. Maintenance 6. Maintenance W arning! Always disconnect MultiT emp III from the mains (Pull out the mains plug) before maintenance and/or service. Service should be carried out by authorised personnel only . T o maintain efficient operation of the compressor cooling unit, have the air grid and liquifier (heat exchanger) checked and cleaned by ser ...

  • TRENDnet 18-1106-33 - page 17

    6. Maintenance The circuit breaker (thermal fuses) for MultiT emp III are loca- ted on the rear of the unit. Note: If the circuit breaker has been triggered, there may be some delay before it can be released. Please push the circuit breaker button only; do not pull. Ratings for the circuit breaker: for 198 to 253 V version two circuit breakers 2 x ...

  • TRENDnet 18-1106-33 - page 18

    7. Detailed Description 7. Detailed Description MultiT emp III regulates temperature by balancing the effects of constant cooling and controlled pulses of heating. Constant cooling is provided by a built-in unit that includes a compressor , an air cooled liquifier and an evaporator (a coo- ling coil as heat exchanger) located in the chamber . This ...

  • TRENDnet 18-1106-33 - page 19

    7. Detailed Description Mains Switch. Starts the cooling compressor and the pump motor that circulates the temperature controlled liquid. Activates the heater if the temperature set is greater than the actual temperature. Change the temperature value. V alues are changed in an accelerating manner when either key is held down. Changes between showin ...

  • TRENDnet 18-1106-33 - page 20

    7. Detailed Description General cut-off The heater , the compressor and the circulation pump can be switched off by a relay built into the unit. This general cut- off will be activated if a malfunction is detected by the safety sensors. When the general cut-off switch is activated, the circulator can only be restarted after the cause of the indicat ...

  • TRENDnet 18-1106-33 - page 21

    7. Detailed Description Accurate Setting of T emperature Limit If a more accurate temperature limit is required, for example 50 °C, follow the procedure below . 1. Set the temperature limit switch to its highest value. 2. Set the working temperature to 50 °C according to Section 4.4. 3. Run the unit and wait until it reaches exactly 50 °C. 4. T ...

  • TRENDnet 18-1106-33 - page 22

    7. Detailed Description Note: Although this feature activates the general cut-off, it is not a dedicated ”safety-function” of the type listed above. When the error message shown above is dis- played, the sensor must be exchanged by service per- sonnel. Restart after power failure MultiT emp III automatically restarts after a power failure. Make ...

  • TRENDnet 18-1106-33 - page 23

    7. Detailed Description T o start MultiT emp III in 8 mode ON, follow the instructions below: 1. Check that all tubings are firmly connected and secured with hose clamps. 2. Switch the power on with the mains switch. The display flashes . 3. Press for start. The version number of the software is briefly shown on the display . The pump motor and com ...

  • TRENDnet 18-1106-33 - page 24

    8. T echnical specifiations 8. T echnical specifications T echnical specifications (ambient temperature 20 ° C) W orking temperature range -10 to 90 °C T emperature control constancy at 20 °C < ± 0.1 °C Range of temperature limit switch 20 to 95 °C Heater capacity , 230V 800 W 1 15V 800 W Cooling capacity , 1 15 V at 20 °C 240 W at 5 °C ...

  • TRENDnet 18-1106-33 - page 25

    9. Ordering information Designation Code No. MultiT emp III 100-120 V 18-1 102-77 MultiT emp III 220-240 V 18-1 102-78 Accessories Item Quantity Code No. Cooling T ubing, 8/12 mm, silicone 4 m 80-1 106-56 T ubing Connector Set female and male 4 18-1 104-26 Hose clamps 10 18-1 104-27 Insulation for cooling tubing 14/27 mm 8 m 80-1 1 16-1 1 3-way V a ...

  • TRENDnet 18-1106-33 - page 26

    ...

  • TRENDnet 18-1106-33 - page 27

    Short instructions Note: The following short instructions are intended as a guide to users who are fully familiar with safety procedures and operating instructions described earlier in this manual. 1. Check to make sure that all tubings are firmly connected and secured with hose clamps. 2. Check the liquid level and fill the bath if needed (the lev ...

メーカー TRENDnet カテゴリー Network Card

TRENDnet 18-1106-33のメーカーから受け取ることができるドキュメントは、いくつかのグループに分けられます。その一部は次の通りです:
- #BRANDの図面#
- 18-1106-33の取扱説明書
- TRENDnetの製品カード
- パンフレット
- またはTRENDnet 18-1106-33の消費電力シール
それらは全部重要ですが、デバイス使用の観点から最も重要な情報は、TRENDnet 18-1106-33の取扱説明書に含まれています。

取扱説明書と呼ばれる文書のグループは、TRENDnet 18-1106-33の取り付け説明書、サービスマニュアル、簡易説明書、またはTRENDnet 18-1106-33のユーザーマニュアル等、より具体的なカテゴリーに分類されます。ご必要に応じてドキュメントを検索しましょう。私たちのウェブサイトでは、TRENDnet 18-1106-33の製品を使用するにあたって最も人気のある説明書を閲覧できます。

関連する取扱説明書

TRENDnet 18-1106-33デバイスの取扱説明書はどのようなものですか?
取扱説明書は、ユーザーマニュアル又は単に「マニュアル」とも呼ばれ、ユーザーがTRENDnet 18-1106-33を使用するのを助ける技術的文書のことです。説明書は通常、全てのTRENDnet 18-1106-33ユーザーが容易に理解できる文章にて書かれており、その作成者はその分野の専門家です。

TRENDnetの取扱説明書には、基本的な要素が記載されているはずです。その一部は、カバー/タイトルページ、著作権ページ等、比較的重要度の低いものです。ですが、その他の部分には、ユーザーにとって重要な情報が記載されているはずです。

1. TRENDnet 18-1106-33の説明書の概要と使用方法。説明書にはまず、その閲覧方法に関する手引きが書かれているはずです。そこにははTRENDnet 18-1106-33の目次に関する情報やよくある質問、最も一般的な問題に関する情報を見つけられるはずです。つまり、それらはユーザーが取扱説明書に最も期待する情報なのです。
2. 目次。TRENDnet 18-1106-33に関してこのドキュメントで見つけることができる全てのヒントの目次
3. TRENDnet 18-1106-33デバイスの基本機能を使うにあたってのヒント。 TRENDnet 18-1106-33のユーザーが使い始めるのを助けてくれるはずです。
4. トラブルシューティング。TRENDnet 18-1106-33に関する最も重要な問題を診断し、解決するために役立つ体系化された手続き
5. FAQ。よくある質問
6. 連絡先。一人では問題を解決できない場合に、その国におけるTRENDnet 18-1106-33のメーカー/サービスへの連絡先に関する情報。

TRENDnet 18-1106-33についてご質問がありますか?

次のフォームを使用してください

見つけた説明書を読んでもTRENDnet 18-1106-33の問題を解決できない場合、下記のフォームを使用して質問をしましょう。ユーザーのどなたかがTRENDnet 18-1106-33で同様の問題を抱えていた場合、その解決方法を共有したいと考えるかもしれません。

画像のテキストを入力してください

コメント (0)