    Model TCL T otal Chlorine Analyzer Instruction Manual PN 51-TCL-1056/rev .D May 201 1 1.00 ppm 25.0 ° c ...

  • Emerson TCL - page 2

    W ARNING ELECTRICAL SHOCK HAZARD Making cable connections to and servicing this instrument require access to shock haz- ard level voltages which can cause death or serious injury . Be sure to disconnect all hazardous voltage before opening the enclosure. Relay contact s made to separate power sources must be disconnected before servic- ing. Electri ...

  • Emerson TCL - page 3

    ...

  • Emerson TCL - page 4

    CAUTION SENSOR/PROCESS APPLICA TION COMP A TIBILITY Wetted materials may not be comp atible with process com position and operating conditions. Application comp at ibility is entirely the responsibility of the user . DANGER HAZARDOUS AREA INST ALLA TION Installations near flammable liquids or in haz- ardous area locations must be carefully eval- ua ...

  • Emerson TCL - page 5

    MODEL TCL T ABLE OF CONTENTS MODEL TCL T OT AL CHLORINE ANAL YZER T ABLE OF CONTENTS Section Title Page 1.0 DESCRIPTION AND SPECIFICA TIONS ................................................................. 1 1.1 Features and Applications ...................................................................................... 2 1.2 S pecifications — ...

  • Emerson TCL - page 6

    ii MODEL TCL T ABLE OF CONTENTS T ABLE OF CONTENTS CONT'D. 8.0 DIGIT AL COMMUNICA TIONS ................................................................................ 41 9.0 CALIBRA TION ....................................................................................................... 42 9.1 Introduction ................................ ...

  • Emerson TCL - page 7

    MODEL TCL T ABLE OF CONTENTS LIST OF FIGURES Number Title Page 2-1 Schematic of Sample Conditioning System and Analyzer .................................... 7 3-1 Panel Mount Dimensions ....................................................................................... 9 3-2 Pipe and W all Mounting Dimensions ................................... ...

  • Emerson TCL - page 8

    ...

  • Emerson TCL - page 9

    MODEL TCL SECTION 1 DESCRIPTION AND SPECIFICA TIONS 1 SECTION 1. DESCRIPTION AND SPECIFICA TIONS • MEMBRANE-COVERED AMPEROMETRIC SENSOR . • NO TOOLS REQUIRED to change membrane. • MAINTENANCE T AKES ONL Y A FEW MINUTES a month. • V ARIOPOL CONNECTOR OPTION allows the sensor to be replaced without removing and rewiring cable. Model 1056 Chlo ...

  • Emerson TCL - page 10

    2 MODEL TCL SECTION 1 DESCRIPTION AND SPECIFICA TIONS used to signal a fault condition. A fault alarm activates when an analyzer or sensor fault occurs. When used for the determination of total chlorine, the Model 1056 analyzer is a single input instrument. The ana- lyzer is also available in a dual input version, where the second input can be pH, ...

  • Emerson TCL - page 11

    3 Case: Polycarbonate NEMA 4X/CSA 4 (IP65). Dimensions: 6.10 x 6.10 x 5.15 in. (155 x 155 x 131 mm) Conduit openings: Accepts PG13.5 or 1/2 in. conduit fittings Display: Monochromatic back-lit LCD. Main character height 0.6 in (15mm). Display is user-programable Languages: English, German, Italian, S panish, French, Portuguese Ambient temperature a ...

  • Emerson TCL - page 12

    4 MODEL TCL SECTION 1 DESCRIPTION AND SPECIFICA TIONS 1.4 SPECIFICA TIONS — MODEL 499ACL-02 SENSOR 1.5 PERFORMANCE SPECIFICA TIONS — COMPLETE SYSTEM Wetted Part s: Gold, Noryl ® 1 (PPO), V iton ® 2 , EPDM, Silicone Dimensions: 1.0 x 5.6 in. (25.4 x 143 mm) Cable: 25 ft. (7.6m) st andard Pressure Rating: 0 to 65 psig (101 to 549 kPa) T emperat ...

  • Emerson TCL - page 13

    MODEL TCL SECTION 1 DESCRIPTION AND SPECIFICA TIONS 1.6 ORDERING INFORMA TION AND ACCESSORIES Model TCL Reagent-Based Chlorine System. The TCL is used for the continuous determination of total chlorine in water . The TCL consists of a sample conditioning system, a reagent carboy , a sensor , and an analyzer . Reagent s must be ordered sep arately . ...

  • Emerson TCL - page 14

    6 ACCESSORIES — Sensor PN DESCRIPTION WEIGHT* SHIP WEIGHT* 23501-02 T otal Chlorine Membrane, includes one membrane assembly and 1 lb (0.5 kg) 1 lb (0.5 kg) one O-ring 23502-02 T otal Chlorine Membrane Kit, includes 3 membrane assemblies and 1 lb (0.5 kg) 1 lb (0.5 kg) three O-rings 9210438 T otal Chlorine Sensor Fill Solution, 4 oz (120 mL) 1 lb ...

  • Emerson TCL - page 15

    T otal chlorine by definition is the iodine pro- duced in a sample when it is treated with potassium iodide at a pH between 3.5 and 4.5. T ypically , acetic acid (or vinegar) is used to adjust the pH. The total chlorine analyzer consist s of a sam- ple conditioning system, which injects the reagent into the sample, and a sensor and analyzer , which ...

  • Emerson TCL - page 16

    3.1 UNP ACKING AND INSPECTION Inspect the shipping containers. If there is damage, cont act the shipper immediately for instructions. Save the boxes. If there is no apparent damage, unp ack the containers. Be sure all items shown on the p acking list are pres- ent. If items are missing, notify Rosemount Analytical immediately . 3.2 INST ALLA TION. ...

  • Emerson TCL - page 17

    MODEL TCL SECTION 3 INST ALLA TION Bottom View Front View Side View FIGURE 3-1 P ANEL MOUNTING DIMENSIONS INCH MILLIMETER 154.9 6.1 154.9 6.1 126.4 5.0 17.13 1.1 126.4 5.0 ) ( 76.2 3.0 41.4 1.6 152.73 6.0 9 ...

  • Emerson TCL - page 18

    Bottom View Front View Side View Side View W all / Surface Mount Pipe Mount 154.9 6.1 102 4.0 187 7.4 154.9 6.1 232 9.1 33.5 1.3 130 5.1 165 6.5 232 9.1 130 5.1 33.5 1.3 165 6.5 108.9 4.3 45.21 1.8 80.01 3.2 71.37 2.8 10 INCH MILLIMETER FIGURE 3-2 PIPE AND W ALL MOUNTING DIMENSIONS (Mounting bracket PN:23820-00) ...

  • Emerson TCL - page 19

    MODEL TCL SECTION 3 INST ALLA TION 3.2.3 Inst all the Sample Conditioning Enclosure 1. Refer to Figures 3-3, 3-4, and 3-5 for installation det ails. 2. Connect the sample line to the sample conditioning system. Use ¼ -inch OD hard plastic or stainless steel tubing. If dechlorinated water is being measured, provide a way for occasionally substituti ...

  • Emerson TCL - page 20

    MODEL TCL SECTION 3 INST ALLA TION FIGURE 3-4. TCL Case Dimensions INCH MILLIMETER FIGURE 3-5. Reagent tubing assembly 12 ...

  • Emerson TCL - page 21

    13 MODEL TCL SECTION 4 WIRING SECTION 4. WIRING 4.1 PREP ARE ANAL YZER CONDUIT OPENINGS The analyzer enclosure has six conduit openings. Four conduit openings are fitted with conduit plugs. Conduit openings accept 1/2-inch conduit fittings or PG 13.5 cable glands. T o keep the case watertight, block unused openings with NEMA 4X or IP65 conduit plug ...

  • Emerson TCL - page 22

    14 MODEL TCL SECTION 4 WIRING 4.3 MAKE POWER, ALARM, OUTPUT , AND SENSOR CONNECTIONS IN THE ANAL YZER 4.3.1 Power Wire AC mains power to the power supply board, which is mounted vertically on the left hand side of the analyzer enclosure. The power connector is at the top of the board. Unplug the connector from the board and wire the power cable to ...

  • Emerson TCL - page 23

    MODEL TCL SECTION 4 WIRING 15 4.4 SENSOR WIRING 1. Shut off power to the analyzer . 2. Locate the chlorine signal board. 3. Insert the sensor cable through the conduit opening nearest the chlorine board. 4. Slide the board forward to gain access to the wires and terminal screws. 5. Connect the sensor cable to the chlorine board. Refer to Figure 4-4 ...

  • Emerson TCL - page 24

    16 MODEL TCL SECTION 4 WIRING 4.5 APPL Y POWER T O THE ANAL YZER AND COMPLETE QUICK ST ART 1. Once all wiring connections are secured and verified, apply power to the analyzer . 2. When the analyzer is powered up for the first time, Quick St art screens appear . Using Quick St art is easy . a. A backlit field shows the position of the cursor . b. T ...

  • Emerson TCL - page 25

    17 MODEL TCL SECTION 4 WIRING 6. Choose the desired temperature units. 7. The main display appears. The outputs and alarms (if an alarm board is present) are assigned to default values. 8. T o change outputs, alarms, and other settings go to the main menu and choose Program . Follow the prompts. A menu tree is on the following page. T o calibrate t ...

  • Emerson TCL - page 26

    MENU TREE Calibrate Sensor 1 (T ot al chlorine) Chlorine Zero In process T emperature Output 1 Output 2 Hold Sensor 1 Sensor 2 Display Main format configuration Language selection W arning (enable or disable) Screen contrast Program Output s Range (assign values to 4 and 20 mA) Configure Output 1 or 2 Assign sensor and measurement Range Scale Dampe ...

  • Emerson TCL - page 27

    MODEL TCL SECTION 5 ST ART -UP SECTION 5. ST ART -UP 5.1 PREP ARE THE REAGENT 1. DO NOT PREP ARE THE SOLUTION UNTIL READY T O USE. 2. Position the blue plastic carboy under the sample conditioning cabinet. Unscrew the cap and reagent tube assembly . 3. Add the potassium iodide reagent to the carboy . See the table. 4. Add five gallons (19 L) of dis ...

  • Emerson TCL - page 28

    20 MODEL TCL SECTION 6 ST ART -UP SECTION 6. DISPLA Y AND OPERA TION When the analyzer is being programmed or calibrated, the display changes to a screen similar to the one shown in Figure 6-2. The live readings appear in small font at the top of the screen. The rest of the dis- play shows programming and calibration information. Programming items ...

  • Emerson TCL - page 29

    MODEL TCL SECTION 6 ST ART -UP 21 6.2 KEYP AD Local communication with the analyzer is through the membrane keypad. Figures 6-4 and 6-5 explain the operation of the keys. FIGURE 6-4. Analyzer keyp ad. Four navigation keys move the cursor around the screen. The position of the cursor is shown in reverse video. The navigation keys are also used to in ...

  • Emerson TCL - page 30

    22 MODEL TCL SECTION 6 ST ART -UP 6.3 PROGRAMMING THE ANAL YZER—TUTORIAL Setting up and calibrating the analyzer is easy . The following tutorial describes how to move around in the programming menus. For practice, the tutorial also describes how to assign ppm chlorine values to the 4 and 20 mA analog outputs . 1. Press MENU. The main menu screen ...

  • Emerson TCL - page 31

    23 MODEL TCL SECTION 6 ST ART -UP 6.4 SECURITY 6.4.1 How the Security Code Works Security codes prevent accidental or unwanted changes to program settings or calibrations. There are three levels of security . a. A user can view the default display and diagnostic screens only . b. A user has access to the calibration and hold menus only . c. A user ...

  • Emerson TCL - page 32

    24 MODEL TCL SECTION 6 ST ART -UP 6.5 USING HOLD 6.5.1 Purpose T o prevent unwanted alarms and improper operation of control systems or dosing pumps, place the alarms and outputs assigned to the sensor in hold before removing it for maintenance. During hold, output s assigned to the sensor remain at the last value, and alarms assigned to the sensor ...

  • Emerson TCL - page 33

    25 MODEL TCL SECTION 6 ST ART -UP 1. Press MENU. The main menu screen appears. Move the cursor to Display and press ENTER. 2. The screen shows the present configuration. There are four items: Main Format, Language, W arning, and Contrast. T o make a change, move the cursor to the desired line and press ENTER. A screen appears in which the present s ...

  • Emerson TCL - page 34

    26 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER SECTION 7. PROGRAMMING THE ANAL YZER 7.1 GENERAL This section describes how to make the following program settings using the local keypad. a. Configure and assign values to the analog current outputs. b. Configure and assign values to the alarm relays (if the alarm board is installed). c. Choose the ...

  • Emerson TCL - page 35

    27 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER T ABLE 7-1. DEF AUL T SETTINGS ITEM CHOICES DEF AUL T Output s 1. Assignments a. output 1 chlorine, temp chlorine b. output 2 chlorine, temp temp 2. Range 0-20 or 4-20 mA 4 – 20 mA 3. 0 or 4 mA setting a. chlorine -9999 to +9999 0 b. temperature -999.9 to +999.9 0 4. 20 mA setting a. chlorine -9999 ...

  • Emerson TCL - page 36

    28 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER T ABLE 7-1. DEF AUL T SETTINGS (continued) ITEM CHOICES DEF AUL T T emperature related settings 1. Units ºC or ºF ºC 2. T emperature compensation automatic or manual automatic Security Code 1. Calibrate/Hold 000 to 999 000 2. Program/Display 000 to 999 000 Calibration–Analog Output s 1. 4 mA 0.0 ...

  • Emerson TCL - page 37

    29 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.3 CONFIGURING , RANGING , AND SIMULA TING OUTPUTS. 7.3.1 Purpose This section describes how to configure, range, and simulate the two analog current outputs. CONFIGURE THE OUTPUTS FIRST . 1. Configuring an output means… a. Assigning a sensor and measurement (chlorine or temperature) to an output. ...

  • Emerson TCL - page 38

    30 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.3.3. Procedure – Configure Output s. 1. Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER. 2. The cursor will be on Output s. Press ENTER. 3. Choose Configure. 4. Choose Output 1 or Output 2 . 5. The screen shows the present configuration. There are six items: A ...

  • Emerson TCL - page 39

    31 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.3.3. Procedure – Ranging Output s. 1. Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER. 2. The cursor will be on Output s . Press ENTER. 3. Choose Range . 4. Choose Output 1 or Output 2 . 5. The screen shows the present settings for the outputs. O1 is output 1, ...

  • Emerson TCL - page 40

    MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 32 7.3.4 Procedure – Simulating Output s 1. Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER. 2. The cursor will be on Output s . Press ENTER. 3. Choose Simulate . 4. Choose Output 1 or Output 2 . 5. Enter the desired simulated output current. T o end the simulated ...

  • Emerson TCL - page 41

    33 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.4.2 Definitions 1. ASSIGNING ALARMS. There are four alarms relays. The relays are freely assignable to any sensor and to either the measurement (chlorine) or temperature. Alarm relays can also be assigned to operate as interval timers or as fault alarms. A fault alarm activates when the analyzer de ...

  • Emerson TCL - page 42

    34 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.4.3 Procedure – Configuring Alarms and Assigning Setpoint s 1. Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER. 2. Choose Alarms. 3. Choose Configure/Setpoint. 4. Choose Alarm 1, Alarm 2, Alarm 3, or Alarm 4. 5. The screen summarizes the present configuration ...

  • Emerson TCL - page 43

    35 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.4.4 Procedure – Simulating Alarms 1. Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER. 2. Choose Alarms. 3. Choose Simulate. 4. Choose Alarm 1, Alarm 2, Alarm 3, or Alarm 4. 5. Choose Don’t simulate, De-energize, or Energize. Press MENU or EXIT to end simulat ...

  • Emerson TCL - page 44

    36 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.4.5 Procedure – Synchronizing Timers 1. Synch Timers is available only if two or more alarm relays have been configured as interval timers. 2. Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER. 3. Choose Alarms. 4. The summary display shows the current Synch Tim ...

  • Emerson TCL - page 45

    37 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.5.3 Procedure – Configuring the Measurement 1. Press MENU. The main menu screen appears. Move the cursor to Program and press ENTER. 2. Choose Measurement. 3. The screen at left appears only if you have a dual input 1056. Choose the Chlorine sensor . 4. The screen summarizes the present configura ...

  • Emerson TCL - page 46

    38 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.6 CONFIGURING TEMPERA TURE RELA TED SETTINGS 7.6.1 Purpose This section describes how to do the following: 1. Choose temperature units. 2. Choose automatic or manual temperature correction for membrane permeability (chlorine sensor). 3. Enter a temperature for manual temperature compensation. 7.6.2 ...

  • Emerson TCL - page 47

    39 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.7 CONFIGURING SECURITY SETTINGS 7.7.1 Purpose This section describes how to set security codes. There are three levels of security . a. A user can view the default display and diagnostic screens only . b. A user has access to the calibration and hold menus only . c. A user has access to all menus. ...

  • Emerson TCL - page 48

    40 MODEL TCL SECTION 7 PROGRAMMING THE ANAL YZER 7.8 RESETTING THE ANAL YZER 7.8.1 Purpose This section describes how to clear user-entered values and restore default settings. There are three reset s: 1. Resetting to factory default values clears ALL user entered settings, including sensor and analog output calibration, and returns ALL settings an ...

  • Emerson TCL - page 49

    41 MODEL TCL SECTION 8 DIGIT AL COMMUNICA TIONS SECTION 8. DIGIT AL COMMUNICA TIONS For information about digit al communications, refer to the following manuals: HART 51-1056HT Model 56 HAR T Addendum Profibus DP 51-1056DP Model 56 Profibus DP Addendum ...

  • Emerson TCL - page 50

    42 MODEL TCL SECTION 9 CALIBRA TION SECTION 9. CALIBRA TION 9.1 INTRODUCTION The calibrate menu allows the user to do the following: 1. Calibrate the temperature sensing element in the total chlorine sensor . 2. Calibrate the chlorine sensor . 3. Calibrate the analog outputs. 9.2 CALIBRA TING TEMPERA TURE 9.2.1 Purpose The total chlorine sensor is ...

  • Emerson TCL - page 51

    43 MODEL TCL SECTION 9 CALIBRA TION 3. Press MENU. The main menu screen appears. The cursor will be on Calibrate. Press ENTER. 4. Choose the sensor you wish to calibrate. Sensor 2 appears only if you are using a dual input analyzer . 5. Choose T emperature . 6. Change the display to match the temperature read from the calibrated thermometer . Press ...

  • Emerson TCL - page 52

    44 MODEL TCL SECTION 9 CALIBRA TION 9.3 CALIBRA TING TOT AL CHLORINE 9.3.1 Purpose The continuous determination of total chlorine requires two step s. See Figure 9-1. First, the sample flows into a conditioning system (the Model TCL) where it is treated with acetic acid (vinegar) and pot assium iodide. The acid lowers the pH, which allows total chl ...

  • Emerson TCL - page 53

    45 MODEL TCL SECTION 9 CALIBRA TION 9.3.2 Procedure — Zeroing the Sensor 1. Place the sensor in a beaker of deionized water . Be sure no air bubbles are trapped against the membrane. 2. Observe the sensor current. The current will drop rapidly at first and then gradually reach a stable zero value. T o monitor the sensor current, press the DIAG ke ...

  • Emerson TCL - page 54

    46 MODEL TCL SECTION 9 CALIBRA TION If the zero current is much larger than expected, the zero calibration failure screen appears. The analyzer will not update the zero current. For troubleshooting assistance, see Section 1 1.5. 8. T o return to the main display , press MENU then EXIT . S1 Zero Cal Sensor zero failed Press Exit 9.3.3 Procedure — ...

  • Emerson TCL - page 55

    47 MODEL TCL SECTION 9 CALIBRA TION 7. Follow the screen prompts: Once the reading is stable, press ENTER. T ake the sample and press ENTER. At this point, the analyzer will store the present sensor current and temperature and use those values in the calibration. Determine the total chlorine concentration in the sample and enter the value in the sc ...

  • Emerson TCL - page 56

    48 MODEL TCL SECTION 9 CALIBRA TION 6. If the calibration was successful, the screen at left will appear . 7. If the user entered value is more that ±1 mA different from the nominal value, a possible error screen will appear . T o force the analyzer to accept the calibration, choose Ye s . 8. T o return to the main display , press MENU then EXIT . ...

  • Emerson TCL - page 57

    49 MODEL TCL SECTION 10 MAINTENANCE SECTION 10. MAINTENANCE 10.1 ANAL YZER The Model 1056 analyzer used with the TCL needs little routine maintenance. Clean the analyzer case and front panel by wiping with a clean sof t cloth dampened with water ONL Y . Do not use solvents, like alcohol, that might cause a buildup of st atic charge. The chlorine se ...

  • Emerson TCL - page 58

    10.2.4 Replacing the membrane and electrolyte solution. 1. Unscrew the membrane retainer and remove the membrane assembly and O-ring. See Figure 10-2. 2. Hold the sensor over a container with the cathode pointing down. 3. Remove the fill plug and allow the electrolyte solution to drain out. 4. Wrap the plug with two turns of pipe tape and set aside ...

  • Emerson TCL - page 59

    51 NOTE Do not use a sharp probe. It will puncture the bladder and destroy the sensor . Gently press the probe against the bladder several times to force liquid through the holes at the base of the cathode stem. Keep pressing the bladder until no air bubbles can be seen leaving the holes. Be sure the holes remain covered with electrolyte solution. ...

  • Emerson TCL - page 60

    52 MODEL TCL SECTION 10 MAINTENANCE FIGURE 10-3. Replacing Reagent T ubing 10.3 SAMPLE CONDITIONING SYSTEM 10.3.1 Reagent The sample conditioning reagent lasts about 2 months. Before putting fresh reagent in the carboy , discard any small amount of remaining reagent. T o prepare the reagent refer to the procedure in Section 5.1. See T able 10-2 for ...

  • Emerson TCL - page 61

    53 MODEL TCL SECTION 10 MAINTENANCE FIGURE 10-4. Replacing Sample T ubing T o replace sample tubing: 1. Sample tubing is shown in Figure 10-4. 2. Turn of f the sample and reagent pumps. 3. Luer fittings connect the sample tubing to the pump. Disconnect the tubing by turning the fitting in the direction of the arrows shown in Figure 10-5. 4. Disconn ...

  • Emerson TCL - page 62

    54 MODEL TCL SECTION 10 MAINTENANCE FIGURE 10-8. FIGURE 10-7. FIGURE 10-6. FIGURE 10-5. 10.3.3 Perist altic pump tubing. The expected life of the peristaltic pump tubing is one year . T o replace pump tubing: 1. Turn of f the sample and reagent pumps. 2. The reagent and sample tubing is connected to the pump tubing with luer fittings. See Figure 10 ...

  • Emerson TCL - page 63

    55 MODEL TCL SECTION 10 MAINTENANCE FIGURE 10-1 1. FIGURE 10-9. FIGURE 10-10. 7. Replace the pump cover . a. Place the cover on the pump casing. See Figure 10-9. b. Be sure the pins at the bottom of the cover (Figure 10-10) ride on the tracks in the pump casing. c. The position of the track is outlined in Figure 10-1 1. The pins on the pump cover m ...

  • Emerson TCL - page 64

    56 MODEL TCL SECTION 10 MAINTENANCE 10.3.4 Replacing the air pump 1. Disconnect power to the analyzer . 2. Refer to Figure 10-12. Disconnect the reagent and air injection tubes. Disconnect the suction and discharge tubing by turning the Luer fitting in the direction shown in the figure. Disconnect the air pump inlet tubing from the barbed fitting i ...

  • Emerson TCL - page 65

    57 MODEL TCL SECTION 10 MAINTENANCE FIGURE 10-16. FIGURE 10-17. 10.3.5 Replacing the air pump diaphragm and check valves. 1. Disconnect power to the analyzer . 2. Refer to Figure 10-12. Disconnect the reagent and air injection tubes. Disconnect the suction and discharge tubing by turning the Luer fitting in the direction shown in the figure. Discon ...

  • Emerson TCL - page 66

    58 MODEL TCL SECTION 10 MAINTENANCE PN Description 24134-00 Air pump, 1 15 V ac, 60 Hz 24134-01 Air pump, 230 V ac, 50 Hz 9160578 Air pump repair kit 9322052 Check valve for air injection line 24153-00 Carboy for reagent, 5 gal/19 L, includes cap 9100204 Fuse, 0.25 A, 250 V , 3AG , slow blow for option -1 1 (1 15 V ac) 9100132 Fuse, 0.125 A, 250 V ...

  • Emerson TCL - page 67

    59 MODEL TCL SECTION 1 1 TROUBLESHOOTING SECTION 1 1. TROUBLESHOOTING 1 1.1 OVERVIEW The analyzer continuously monitors itself and the sensor(s) for problems. When the analyzer identifies a problem, the word warning or fault appears intermittently in the lower line of the main display . When the fault or warning message appears, press the DIAG (dia ...

  • Emerson TCL - page 68

    60 MODEL TCL SECTION 1 1 TROUBLESHOOTING Fault message Explanation Section Main Board CPU Error Main board software is corrupted 1 1.3.1 Main Board Factory Data Main board factory eeprom data is corrupted 1 1.3.1 Main Board User Data Main board user eeprom data is corrupted 1 1.3.1 Sensor Hardware Error Missing or bad hardware component 1 1.3.2 Sen ...

  • Emerson TCL - page 69

    61 MODEL TCL SECTION 1 1 TROUBLESHOOTING 1 1.3.4 Sensor Incompatible This error message means that the sensor board software is not supported by the main board sof tware. Either the sensor board or the main board software is too old. Replace the main board with one compatible with the sensor board. Call the factory for assist ance. Y ou will be ask ...

  • Emerson TCL - page 70

    62 MODEL TCL SECTION 1 1 TROUBLESHOOTING W arning message Explanation Section Sensor Need Factory Cal The sensor board was not calibrated at the factory . 1 1.4.1 Sensor Negative Reading The chlorine reading is less than -0.5 ppm. 1 1.4.2 Sensor RTD Sense Open RTD sensor line is broken or not connected 1 1.4.3 Sensor T emperature High T emperature ...

  • Emerson TCL - page 71

    63 MODEL TCL SECTION 1 1 TROUBLESHOOTING 1 1.5 TROUBLESHOOTING WHEN NO ERROR MESSAGE IS SHOWING 15.5.1 Zero current is too high 1. Is the sensor properly wired to the analyzer? See Section 4.3. 2. Is the zero solution chlorine free? T ake a sample of the zero solution and test it for total chlorine. The con- centration should be less than 0.05 ppm. ...

  • Emerson TCL - page 72

    64 MODEL TCL SECTION 1 1 TROUBLESHOOTING 1 1.5.3 Sensitivity is low or readings are low 1. Does the reagent carboy contain reagent? Is the reagent uptake tubing below the level of the reagent? Has potassium iodide been added to the acetic acid (vinegar) reagent? 2. Is there adequate flow to the overflow sampler? Excess sample should be flowing down ...

  • Emerson TCL - page 73

    MODEL TCL SECTION 1 1 TROUBLESHOOTING 1 1.5.4 Process readings are erratic or wander 1. Is the sensor properly wired to the analyzer? See Section 4.3. V erify that all connections are tight. 2. Readings can be erratic when a new sensor is first placed in service. Readings usually stabilize after about an hour . 3. Is the air pump working? There sho ...

  • Emerson TCL - page 74

    MODEL TCL SECTION 1 1 TROUBLESHOOTING 1 1.5.6 Readings are too high 1. Is the sample conditioning reagent clear and colorless? If the reagent is pale yellow , result s will be high. The pale yellow color is caused by iodine, which comes from the reaction between atmospheric oxygen and pot as- sium iodide. The reaction is cat alyzed by sunlight. The ...

  • Emerson TCL - page 75

    67 MODEL TCL SECTION 1 1 TROUBLESHOOTING T emp. (°C) Pt 100 ( Ω ) 0 100.0 10 103.9 20 107.8 25 109.7 30 1 1 1.7 40 1 15.5 50 1 19.4 60 123.2 70 127.1 80 130.9 85 132.8 90 134.7 100 138.5 1 1.6 SIMULA TING TEMPERA TURE 1 1.6.1 General. The Model 1056 accepts a Pt100 R TD. The Pt100 R TD is in a three-wire configuration. See Figure 1 1-3. 1 1.6.2 S ...

  • Emerson TCL - page 76

    68 MODEL SOLU COMP II SECTION 12 RETURN OF MA TERIAL SECTION 12. RETURN OF MA TERIAL 12.1 GENERAL. T o expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for rep air , call 1-949-757-8500 for a Return Materials Authorization (RMA) number . 12.2 W ARRANTY ...

  • Emerson TCL - page 77

    W ARRANTY Seller warrants that the firmware will execute the programming instructions provided by Seller , and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira- tion of the applicable warranty period. Goods are warranted for twelve (12) months ...

  • Emerson TCL - page 78

    Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. 8 Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .raihome.com © ...

Manufacturer Emerson Category TV Converter Box

Documents that we receive from a manufacturer of a Emerson TCL can be divided into several groups. They are, among others:
- Emerson technical drawings
- TCL manuals
- Emerson product data sheets
- information booklets
- or energy labels Emerson TCL
All of them are important, but the most important information from the point of view of use of the device are in the user manual Emerson TCL.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Emerson TCL, service manual, brief instructions and user manuals Emerson TCL. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Emerson TCL.

Similar manuals

A complete manual for the device Emerson TCL, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Emerson TCL by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Emerson TCL.

A complete Emerson manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Emerson TCL - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Emerson TCL, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Emerson TCL, that we can find in the current document
3. Tips how to use the basic functions of the device Emerson TCL - which should help us in our first steps of using Emerson TCL
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Emerson TCL