Manual Sanyo MDF-136

27 pages 0.45 mb
Download

Go to site of 27

Summary
  • Sanyo MDF-136 - page 1

    MDF-136 MDF-236 MDF-436 BIOMEDICAL FREEZER INSTRUCTION MANUAL MDF-236 ...

  • Sanyo MDF-136 - page 2

    Note: 1. No part of this manual may be reproduced in any form w ithout the expressed w ritten permission of SANYO. 2. The contents of this manual are subject to change w ithout notice. 3. Please contact SANYO if any point in this manual is unclear or if there are any inaccuracies. SANYO Electric Biomedical Co., Ltd. All rights reserved. Printed in ...

  • Sanyo MDF-136 - page 3

    1 CONTENTS PRECAUTIONS FOR SAFE OPERATION P. 2 CAUTIONS FOR USAGE P. 6 ENVIRONMENTAL CONDITIONS P. 7 FREEZER COMPONENTS P. 8 INSTALLATION P. 11 OPTIONAL COMPONENTS P. 13 START-UP OF UNIT P. 15 REMOTE ALARM TERMINAL P. 15 TEMPERATURE SETTING P. 16 ALARMS & SAFETY FUNCTIONS P. 18 SETTING OF ALARM RESUME TIME P. 19 ROUTINE MAINTENANCE P. 20 TROUBL ...

  • Sanyo MDF-136 - page 4

    2 PRECAUTIONS FOR SAFE OPERATION It is imperative that the user complies with this manual as it contains important safety advice. Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person. Precautions are illustra ...

  • Sanyo MDF-136 - page 5

    3 PRECAUTIONS FOR SAFE OPERATION Do not use the unit outdoors. Current leakag e or electric shock may result if the unit is exposed to rain water. Only qualified engineers or service personnel should install the unit. The installation by unqualified personnel may cause electric shock or fire. Install the unit on a sturdy floor. If the floor is not ...

  • Sanyo MDF-136 - page 6

    4 PRECAUTIONS FOR SAFE OPERATION Never splash water directly onto the unit as this may cause electric shock or short circuit. Never disassemble, repair, or modify the unit yourself. Any such work carried out by an unauthorized person may result in fire or injury due to a malfunction. Disconnect the pow er supply plug if there is something w rong w ...

  • Sanyo MDF-136 - page 7

    5 PRECAUTIONS FOR SAFE OPERATION Do not climb onto the unit or do not put articles on the unit. This may cause injury by tipping or damage to the unit. Put on dry glov es w hen y ou take out refrigerated articles from the freezer. Handing frozen contents or the inside walls with naked hands may cause frostbite. Do not defrost inside w alls using a ...

  • Sanyo MDF-136 - page 8

    6 CAUTIONS FOR USAGE 1. Sometimes the alarm may not operate at the time of first start-up. This does not mean malfunction. It is due to the complete discharge of incorporated ba ttery. 2-day continuous oper ation of the freezer is necessary to charge the battery fully. 2. The digital thermometer of medical freezer is desi gned to display the temper ...

  • Sanyo MDF-136 - page 9

    7 ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the following conditions (based on the IEC 1010-1): 1. Indoor use; 2. Altitude up to 2000 m; 3. Ambient temperature 5 o C to 40 o C 4. Maximum relative humidity 80% for temperature up to 31 o C decreasing linearly to 50% relative humidity at 40 o C; 5. Mains supply volt ...

  • Sanyo MDF-136 - page 10

    8 FREEZER COMPONENTS 1 6 3 11 12 MDF-236 4 5 2 7 8 9 10 13 14 ...

  • Sanyo MDF-136 - page 11

    9 FREEZER COMPONENTS 1. Door: To open the door, grip the handle. 2. Handle: Always grip this handle to open and close the outer door. 3. Door gasket: This provides a tight door seal and prevents cold air leak. Keep clean. 4. Access port: This is used for leading the meas uring cable from the freezing chamber to the outside. 5. Lock: Turn counterclo ...

  • Sanyo MDF-136 - page 12

    10 FREEZER COMPONENTS Control panel and keypad 1. Alarm test key (ALARM TEST): Test key for alarm device. By pressing this key, the alarm lamp is flashed, remote alarm is activated and buzzer sound. This means all alarm function operate correctly. 2. Buzzer stop key (BUZZER): To silence the audible alarm, press this key. The remote alarm is als o s ...

  • Sanyo MDF-136 - page 13

    11 INSTALLATION Installation site T o operate this unit properly and to obtain maximum performance, install the unit in a location with the following conditions: 1. A location not subjected to direct sunlight Installation in a location subjected to direct sunlight may lead to inadequate cooling. 2. A location w ith adequate ventilation Leave at lea ...

  • Sanyo MDF-136 - page 14

    12 INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and t apes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a neutral detergent and wipe it up with a wet cloth. 2. Adjust the leveling feet Extend the leveling feet by rotating them counterc ...

  • Sanyo MDF-136 - page 15

    13 OPTIONAL COMPONENTS Temperature recorder WARNING Alw ays disconnect the pow er supply to the unit prior to attachment of a temperature recorder in order to prevent electric shock or injury. An automatic temperature recorders is available for this freezer as the optional component. The type of the recorder is MTR-G85. For the attachment, the moun ...

  • Sanyo MDF-136 - page 16

    14 OPTIONAL COMPONENTS 6. As shown in the Fig. 4, fix the recorder sensor on the chamber wall by using 2 clips with a 10mm spacer located between the wall and the clip. The clips and spacers are enclosed with the mounting kit. 7. Cover the gap of the access port on the chamber side completely by a silicon sealant after pushing the insulation into t ...

  • Sanyo MDF-136 - page 17

    15 START-UP OF UNIT Follow the procedures for the initial and consequent operations of the unit. 1. Turn the power switch ON with the chamber empty. 2. Turn ON the battery switch. 3. The alarm sometimes may operate. In this case, silence the alarm by pressing BUZZER key. 4. Set the chamber temperature to a desired value. 5. Allow the unit to achiev ...

  • Sanyo MDF-136 - page 18

    16 TEMPERATURE SETTING Chamber temperature Table 1 shows t he basic procedure for se tting the chamber temperature. Perform key operations in the sequence indicated in the table. The example in the table is based on the assumption that the desired temperature is -35 o C. Note: The unit is set at the factory that the chamber temperature -30 o C. Tab ...

  • Sanyo MDF-136 - page 19

    17 TEMPERATURE SETTING Alarm temperature setting This unit is provided with both high and low temperatur e alarms. The temperature at which the alarm is activated may be changed. The available set range for high temperature alarm is between +5 o C and +15 o C, and -5 o C and -15 o C for low temperature alarm against the chamber temperature. Note: T ...

  • Sanyo MDF-136 - page 20

    18 ALARMS & SAFETY FUNCTIONS This unit has the alarms and safety functions shown in Table 5, and also self diagnostic functions. Table 5 Alarms and safety functions Alarm & Safety Situation Indication Buzzer Safety operation High temperature alarm If the chamber temperature is higher than the temperature at which the high temperature alarm ...

  • Sanyo MDF-136 - page 21

    19 SETTING OF ALARM RESUME TIME The alarm buzzer and remote alarm are silenced by pressing BUZZER key on the control panel during alarm condition. The buzzer and remote alarm will be activated again after certain suspension if the alarm condition is continued. The suspension time can be set by following the proce dure shown in the Table 6 below. Th ...

  • Sanyo MDF-136 - page 22

    20 ROUTINE MAINTENANCE WARNING Alw ays disconnect the pow er supply to the unit prior to any repair or maintenance of the unit in order to prevent electric shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance. These may be harmful to your health. CAUTION Alw ay s put on the dry ...

  • Sanyo MDF-136 - page 23

    21 TROUBLE SHOOTING If the unit malfunctions, check out the following before calling for service. The chamber is not cooled at all 1. The circuit breaker of power source is active. 2. The voltage is too low. (In this case, call an electrician.) 3. The power switch is not ON. 4. The large amount of articles (load) is stored in the chamber at one tim ...

  • Sanyo MDF-136 - page 24

    22 SPECIFICATIONS Name Biomedical Freezer Model MDF-136 MDF-236 MDF-436 External dimensions W640 x D687 x H881 (mm) W905 x D687 x H881 (mm) W1265 x D807 x H905 (mm) Internal dimensions W525 x D440 x H715 (mm) W790 x D440 x H715 (mm) W1140 x D550 x H735 (mm) Effective capacity 138 L 221 L 426 L Exterior Polyester finish baked on zinc galvanized stee ...

  • Sanyo MDF-136 - page 25

    23 PERFORMANCE MDF-136 Cooling performance -35 o C (ambient temperature; 35 o C, no load) Temperature control range -20 o C to -35 o C Rated voltage AC 110 V AC 115 V AC 220 V AC 220 to 240 V Rated frequency 60 Hz 60 Hz 60 Hz 50 Hz Rated power consumption 140 W 140 W 150 W 145 W Noise level 37 dB [A] (background noise; 20 dB) Maximum pressure 2570 ...

  • Sanyo MDF-136 - page 26

    24 CAUTION Please fill in this form before servicing. Hand over this form to the service engi neer to keep for his an d your safety . Safety check sheet 1. Refrigerator contents : Risk of infection: Risk of toxicity: Risk from radioactive sources: (List all potentially hazardous materials that have been stored in this unit.) Notes : 2. Contaminatio ...

  • Sanyo MDF-136 - page 27

    7FB6P101279004 Recycled paper CEM SANYO Electric Biomedical Co., Ltd. 5-5, Keihan-Hondori 2-Chome Moriguchi City , Osaka 570-8677 Japan Printed in Japan ...

Manufacturer Sanyo Category Freezer

Documents that we receive from a manufacturer of a Sanyo MDF-136 can be divided into several groups. They are, among others:
- Sanyo technical drawings
- MDF-136 manuals
- Sanyo product data sheets
- information booklets
- or energy labels Sanyo MDF-136
All of them are important, but the most important information from the point of view of use of the device are in the user manual Sanyo MDF-136.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Sanyo MDF-136, service manual, brief instructions and user manuals Sanyo MDF-136. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Sanyo MDF-136.

Similar manuals

A complete manual for the device Sanyo MDF-136, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Sanyo MDF-136 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Sanyo MDF-136.

A complete Sanyo manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Sanyo MDF-136 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Sanyo MDF-136, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Sanyo MDF-136, that we can find in the current document
3. Tips how to use the basic functions of the device Sanyo MDF-136 - which should help us in our first steps of using Sanyo MDF-136
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Sanyo MDF-136
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Sanyo MDF-136 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Sanyo MDF-136?

Use the form below

If you did not solve your problem by using a manual Sanyo MDF-136, ask a question using the form below. If a user had a similar problem with Sanyo MDF-136 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)