  • Omega FCLTX-100 - page 1

    PRODUCT INSTRUCTION SHEET Percent of free chlorine in protonated form (HClO) 0 10 20 30 40 50 60 70 80 90 100 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 pH percent SECTION 1.0 THEORY OF OPERA TION 1.0 FREE CHL ORINE DEFINED . Fr ee Chlorine or "freely active chlorine" is defined as the sum of molecular chlorine (Cl 2 ), hypochlorous acid (HOCl) and ...

  • Omega FCLTX-100 - page 2

    SECTION 3.0 SENSOR PREP ARA TION FIG. 2 3.0 FREE CHL ORINE SENSOR ASSEMBL Y . The Free Chlorine Sensor is shipped with the membrane cap pre-installed and cov ered with a cap with water inside to keep the membrane wet. Make sure to keep sensor cap, anode and cathode inside the sensor body , away oily or gr easy materials. Contact with oil or grease ...

  • Omega FCLTX-100 - page 3

    Po wer Supply PL C, DVM,... sensor red wire sensor black wire FIG. 3 PRODUCT INSTRUCTION SHEET 6.0 ELECTRICAL INST ALLA TION. The sensor produces an ap- proximate output of 4 mA in air and 20mA at the t op range of free chlorine output (0-2ppm, 0-5ppm, 0-10ppm). NO TE: The supply voltage to the Sensor must be 12-24 V DC with minimum of 250 mA. M ax ...

  • Omega FCLTX-100 - page 4

    ZERO REA D ClO 2 C L O 2 to Increase output to Decrease output SECTION 8.0 CALIBRA TION FIG. 6 SECTION 9.0 SENSOR ST ORAGE IMPORT ANT NOTE: SENSORS ARE SUPPLIED F ACT ORY CALIBRA TED WITH A 4-20mA SIGNAL OUTPUT CORRESPONDING TO THEIR SPECIFIC RANGE (0-2, 0-5 OR 0-10ppm). SP AN CALIBRA TION IS NECESSARY WHEN RECEIVING A NEW SENSOR SINCE Y OUR CONDIT ...

  • Omega FCLTX-100 - page 5

    PRODUCT INSTRUCTION SHEET 10.1 GENERAL MAINTENANCE. Ser vice the sensor regularly (weekly is suggested) 10.2 MEMBRANE CAP REPLACEMENT . If membrane replace- ment is required , a new cap with preinstalled membrane must be used. Order FCLA-2016 replacement membrane cap and FCLA-2015 refill solution. T o change membrane cap (do the following ov er a s ...

  • Omega FCLTX-100 - page 6

    PRODUCT INSTRUCTION SHEET Operating temperature range: 0-45 degC Maximum operating pressure: 1 bar/14.7 psi/1atm Flow ra te minimum: .20 gpm Flow ra te maximum: 0.6 ppm pH range: 5.5-9.0 Output signal: 4.0+/- 0.2mA in air (zero) 20mA +/- 0.2mA at high range (2, 5 or 10ppm) Po wer Requirement: 12-24 VDC, 250 mA minimum Cross-Sensitivity : ClO2, ozon ...

Manufacturer Omega Category Stud Sensor

Documents that we receive from a manufacturer of a Omega FCLTX-100 can be divided into several groups. They are, among others:
- Omega technical drawings
- FCLTX-100 manuals
- Omega product data sheets
- information booklets
- or energy labels Omega FCLTX-100
All of them are important, but the most important information from the point of view of use of the device are in the user manual Omega FCLTX-100.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Omega FCLTX-100, service manual, brief instructions and user manuals Omega FCLTX-100. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Omega FCLTX-100.

Similar manuals

A complete manual for the device Omega FCLTX-100, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Omega FCLTX-100 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Omega FCLTX-100.

A complete Omega manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Omega FCLTX-100 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Omega FCLTX-100, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Omega FCLTX-100, that we can find in the current document
3. Tips how to use the basic functions of the device Omega FCLTX-100 - which should help us in our first steps of using Omega FCLTX-100
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Omega FCLTX-100
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Omega FCLTX-100 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Omega FCLTX-100?

Use the form below

If you did not solve your problem by using a manual Omega FCLTX-100, ask a question using the form below. If a user had a similar problem with Omega FCLTX-100 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)