Precision Medical PM5200 - page 1
USER MANUAL AIR-O XYGEN BLENDER (DIS S and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS CA UTION F ederal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive T el: (+001) 610-262-6090 Northampton, P A 18067 USA F ax: (+001) 610-262-6080 www .precisionmedical.com ...
-
Precision Medical PM5200 - page 2
1 Blender Air-Oxygen CONTENTS RECEIVING / INSPECTION ......................................................... 2 INTENDED USE .......................................................................... 2 READ ALL INSTRUCTIONS BEFORE USING ............................ 2 EXPLANA TION OF ABBREVIA TIONS ......................................... 2 SAFETY ...
-
Precision Medical PM5200 - page 3
2 Blender Air-Oxygen RECEIVING / INSPECTION Remove the Precision Medical, Inc. Air -Oxygen Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider . INTENDED USE Precision Medical, Inc. Air -Oxygen Blender dispenses a continuous and precise blend of medical air and USP oxygen via outlet ports ...
-
Precision Medical PM5200 - page 4
3 Blender Air-Oxygen SAFETY INFORMA TION - W ARNINGS AND CAUTIONS DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury . W ARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury . CA UTION Indicates a potentially hazardous situatio ...
-
Precision Medical PM5200 - page 5
4 Blender Air-Oxygen W ARNING An Oxygen Analyzer/Monitor must be used to verify oxygen concentration. Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the High Flow Blender , and 3 lpm for the Low Flow Blender . DO NOT obstruct the alarm. DO NO T use Blender when alarm is sounding. DO NO ...
-
Precision Medical PM5200 - page 6
5 Blender Air-Oxygen SPECIFICA TIONS Model PM5200 High Flow PM5300 Low Flow Primary Outlet Flow Range 15 - 120 lpm 3 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 . 4 b a r ) w i t h N O B L E E D Auxiliary Outlet Flow Range 2 - 100 lpm 0 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 . 4 b a r ) ...
-
Precision Medical PM5200 - page 7
6 Blender Air-Oxygen SPECIFICA TIONS continued Dimensions: (without fittings) Depth: 4.9 in (12.5 cm) Width: 2.3 in (5.7 cm) Height: 4.1 in (10.4 cm) W eight: 2.9 lbs (1.3 kg) Shipping W eight: 3.5 lbs (1.6 kg) Operating T emperature Range: 59°F to 104°F (15°C to 40°C) T ransport / Storage Requirements T emperature Range: -10°F to 140°F (-23 ...
-
Precision Medical PM5200 - page 8
7 Blender Air-Oxygen DIAGRAMS CA UTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. Depending on model, your fittings may differ from these diagrams. PM5300 Model B F A E PM5200 Model B G D C I H ...
-
Precision Medical PM5200 - page 9
8 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION A Oxygen Concentration Dial A d ia l u se d f or se le ct ing ox yg en c on ce ntr at ion s b et we e n 21%-100%. The FiO 2 scale is used for reference only . This Dial does not rotate 360°. The dial starts at 21% and ends at 100%. B Primary Outlet Port A ma le DISS oxy ge n fit tin g wi ...
-
Precision Medical PM5200 - page 10
9 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION F Air Inlet Fitting A male DISS or NIST air fitting with one way valve that is used to connect an air supply hose . G Alarm An audible alarm that sounds due to an excessive pressure drop or deletion of either gas supply . H Manifold Outlet (Optional) Manifold with 3 primary outlets. I Rea ...
-
Precision Medical PM5200 - page 11
10 Blender Air-Oxygen ALARM TEST 1. The Air -Oxygen Blender is installed and the flowmeter is turned on. Disconnect or turn off the air supply line to the Air -Oxygen Blender . 2. The Blender should alarm with a loud whistle noise. The whistle indicates the alarm is operating correctly . 3. Reconnect and activate the air supply line to the Blender ...
-
Precision Medical PM5200 - page 12
11 Blender Air-Oxygen 3. T o activate the bleed, turn the knurled collar until the bleed pin is engaged. Then slide the collar back until it contacts the cover . 4. Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor . 5. Confirm the flow of air and/or oxygen mixture to the patient. 6. T urn “OFF” the Air and Oxygen suppl ...
-
Precision Medical PM5200 - page 13
12 Blender Air-Oxygen 1. Disconnect all gas connections and equipment before cleaning. 2. Clean exterior surfaces with a cloth dampened with mild detergent and water . 3. Wipe dry with a clean cloth. MAINTENANCE The following maintenance on the Air -Oxygen Blender must be performed by a trained service technician: Every month verify proper function ...
-
Precision Medical PM5200 - page 14
13 Blender Air-Oxygen TROUBLESHOOTING If the Air -Oxygen Blender fails to function, consult the T roubleshooting Guide below . If problem cannot be solved by using T roubleshooting Guide refer to the Air -Oxygen Blender Service Manual (P/N 504827) available on the Internet; www .precisionmedical.com or consult your Provider . Problem Probable Cause ...
-
Precision Medical PM5200 - page 15
14 Blender Air-Oxygen LIMITED W ARRANTY AND LIMIT A TION OF LIABILITY Precision Medical, Inc. warrants that the Blender , (the Product), will be free of defects in workmanship and/or material for the following period: T wo (2) years from shipment Should any failure to conform to this warranty appear within the applicable period, Precision Medical, ...
-
Precision Medical PM5200 - page 16
DECLARA TION OF CONFORMITY Manufacturer: Precision Medical, Inc. 300 Held Drive, Northampton, P A 18067, USA CONT ACT : Quality Manager Phone: 610-262-6090 Authorized European Representative: Emergo Europe Molenstraat 15 2513 BH, The Hague The Netherlands Product: Gas Mixers for Medical Use Model(s): PM5200EN, PM5200MEN, PM5300EN, PM5300MEN MDD Cla ...
Do you have a question concerning Precision Medical PM5200?
Use the form below
If you did not solve your problem by using a manual Precision Medical PM5200, ask a question using the form below. If a user had a similar problem with Precision Medical PM5200 it is likely that he will want to share the way to solve it.