  • Agilent Technologies 85036B - page 1

    Us er ’s a nd S erv ice G uid e Agi lent T echno log ies 85 036B/ E 75 Ω Type- N Calibr at ion Kit s This manual a pplies directly to Agi lent 85036B calibrati on kits with seria l number prefi x 3514A, and Ag ilent 85036E calib ration kit s with ser ial number prefix 3 142A. The c alibrati on devices in thi s kit are individ ually ser ialized ...

  • Agilent Technologies 85036B - page 2

    ii 85036B/ E Docu me ntat ion W arranty THE MATERIAL CO NTAINED IN THIS DOCUMENT IS PROVIDED "AS IS , " AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS . FURTHER, TO THE MAXIMUM EXTENT PERM ITTED BY APPLIC ABLE LA W , AGILENT DISCLAIMS AL L W ARRANTIES , EITHER EXPRESS OR IMP LIED WITH REGARD TO THIS MANUAL AND ANY INF ...

  • Agilent Technologies 85036B - page 3

    Cont ents 85036B/E iii 1. Gen eral In form atio n Cali bration Kit Over v iew. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Kit Con tent s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies 85036B - page 4

    Contents 85036B/E iv 6. Rep la cea bl e P ar ts Int r oduc tion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2 A. Standard Definitions Sta ndar d Clas s Assig n ments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...

  • Agilent Technologies 85036B - page 5

    1- 1 1 Gen e ral In forma tio n ...

  • Agilent Technologies 85036B - page 6

    1- 2 850 36B/E General Inf or mation Calibrati on Kit Overview Calibr ati on Kit Overview The Agil ent 85036B and 85036E type- N calibr ation ki ts are used to cal ibrate Agi lent network analyzers up to 3 G Hz for measureme nts of c omponents wit h 75 Ω type-N connector s. Kit Contents The 85036B calib ration k it co ntains the fo llo wing : • ...

  • Agilent Technologies 85036B - page 7

    85036B/ E 1-3 General Information Calibrat ion Kit Overvie w Adap ters Like the other devic es in the kit, t he adapters are b uilt t o very ti g ht tolerance s to pr ovide good broadb and performanc e. T he adapter s utili ze a dual -beaded connector struc ture to ensure st able , r ep ea tabl e conn ections . The beads are d esi g ned to mini mi ...

  • Agilent Technologies 85036B - page 8

    1- 4 850 36B/E General Inf or mation Incoming Inspection Inc o mi ng Ins p e c tio n V er ify that the shi pment is c omplete by ref err ing to Figure 1- 1 o n pag e 1-5 . Check for dama ge. T he foa m-lined stor age case pr ovides protect ion during shi p ping . If the case or any devi ce appea rs damaged, or i f the shipment is inco mplete, conta ...

  • Agilent Technologies 85036B - page 9

    85036B/ E 1-5 General Information Incoming Inspection Figur e 1-1 C alib ration K it Co ntent s ...

  • Agilent Technologies 85036B - page 10

    1- 6 850 36B/E General Inf or mation Recordi ng the Device Serial Numbers Record ing the Device Serial Nu mb ers In addit i on to the kit seri al number , the d evi ces in t his kit a re indi vidually s erializ ed (ser ial numbers a re l abeled onto the b ody of ea ch devi ce). Rec ord these serial numbers in T ab le 1-1 for t h e 8 50 36B a nd T a ...

  • Agilent Technologies 85036B - page 11

    85036B/ E 1-7 General Information C ali br ati on K it s D o cum en te d i n th is M a nu al Cali bra tion Ki ts Do cumented i n t his Manu a l This manual a pplies to any 8503 6B or 85036E ca libration ki t whose seria l prefi x is l isted on the ti tl e page. If your cal i bration kit ha s a dif f erent serial num ber prefix than the one listed o ...

  • Agilent Technologies 85036B - page 12

    1- 8 850 36B/E General Inf or mation Clarifyin g the Sex of a Connector Clarify ing the Sex of a C onn e ctor In thi s manual, the sex of cali bra tion de vi ces and ada pters a r e r ef erred to in terms of thei r connector interface . F or example , a mal e open has a m ale connector . However , duri ng a measureme nt calibr ation, the net work a ...

  • Agilent Technologies 85036B - page 13

    2-1 2 Spec ifica tion s ...

  • Agilent Technologies 85036B - page 14

    2-2 85036B/ E Specificati ons En vironm ental R equi rements Environment al Requirement s T emperature —What to W atch Out F or Changes i n temperatur e can aff ect el ectric al charac teristics. T herefore , the operati ng temperatur e is a c ritic al fa ctor in performance . During a meas urement cali bration, the temperatur e of t he calibrati ...

  • Agilent Technologies 85036B - page 15

    85036B/ E 2-3 Specifications Mechanical Specifi cations Mechani cal Sp ecific atio ns The mechanic al spec ificati ons in T a ble 2-2 a pply to the devices in the 8 5036B and the 8503 6E 75 Ω type-N ca libration kits. Supplemental Mechanical Characterist ics Supplemental c har acteristi c s are val ues which a re typi cally met by a major ity of ...

  • Agilent Technologies 85036B - page 16

    2-4 85036B/ E Specificati ons Electrical Specifi cat ions Electr ic al Specific ation s The el ectrical specif ications i n T able 2-4 apply to t he devices in your c alibrati on kit when connected with an Agilent p recisi on inter face . Certification Agilent T ec hnolog ie s certi f ies that this p roduc t m et i ts pub li s hed s peci fic ati on ...

  • Agilent Technologies 85036B - page 17

    3-1 3 Us e, Main ten anc e, an d Ca re of the De vic es ...

  • Agilent Technologies 85036B - page 18

    3-2 85036B/ E Use, Maintenance, and Care of the D evi ces Electr ostatic Disc harge Elect rostati c Discha rg e Protect ion agai nst ESD ( electrosta tic di scharge) i s essential whi le c onnecting , inspecting , or cl eaning connectors a ttached to a static-sens itive c ircuit (such as thos e found in test sets) . Stati c el ectricity can bui ld ...

  • Agilent Technologies 85036B - page 19

    85036B/ E 3-3 Use, Maintenance, and Care of t he Devices Visual Inspection Visual Ins pect ion Visual inspe ction and, i f necessary , cleaning should be d one ever y time a connection i s made. Metal parti cl es from the c onnector t hr eads may fal l into the c onnector when i t is disconnect ed. One c onnection made with a d irty or damaged conn ...

  • Agilent Technologies 85036B - page 20

    3-4 85036B/ E Use, Maintenance, and Care of the D evi ces Cleani ng Connect or s Inspect the Pr ecision S lotless C onnectors (female) Preci sion slotle ss female connectors are used to improv e accuracy . T he slotl ess contact s are not aff ected by the s light va riations i n male c ontact p in diameter . Howev er , it is still advi sable to i n ...

  • Agilent Technologies 85036B - page 21

    85036B/ E 3-5 Use, Maintenance, and Care of t he Devices Cleani ng Connect ors Use a l i nt-free sw ab or c lea ning cl oth moistened w ith isop ropyl al cohol to remove any dirt or stubbor n conta minants on a connec tor t ha t cannot be r emoved with compressed air or nitrogen. Refer to T a ble 6-3 on page 6-5 for part numbers for isopropy l alco ...

  • Agilent Technologies 85036B - page 22

    3-6 85036B/ E Use, Maintenance, and Care of the D evi ces Connect ions Co nne c ti on s Good connecti ons requi re a ski lled op erator . The mo st com m o n caus e of me asu re me n t erro r is bad c onnections . The followi ng proced ures il lustrate how to make good connecti ons. CA UTION Never mate a 50 ohm connect or with a 7 5 ohm connec tor ...

  • Agilent Technologies 85036B - page 23

    85036B/ E 3-7 Use, Maintenance, and Care of t he Devices Connections Using a torque wrenc h guara ntees that the connection is not too tight, p reventing possib le connect or damage . It a lso guarantees tha t all co nnections a re equal ly tight each ti m e. 2. Prevent the rotation of anything o ther tha n the conne ctor nut that you a re tighteni ...

  • Agilent Technologies 85036B - page 24

    3-8 85036B/ E Use, Maintenance, and Care of the D evi ces Connect ions 4. Hold th e torque wrench l i ghtly , at the end of th e handle onl y (beyo nd the groov e). Se e Figu r e 3 -3 . Figure 3 -3 Usi ng the T o r que W ren ch 5. Appl y do wnwa rd f orce p erpend icul a r t o the wrench han dle . See Fi gu re 3- 3 . Th is ap plies torque to the co ...

  • Agilent Technologies 85036B - page 25

    85036B/ E 3-9 Use, Maintenance, and Care of t he Devices Connections Co nnect ing a n d Di sc o nnect ing th e Tw o-Pi ece Female O pen ( 85036B ) The female open st andard in the 85036B calibr ation ki t i s c omposed of two parts : the open body (outer cond uctor) a nd the cente r conductor ext ender . Refer to Figure 3 -4 . Figur e 3-4 C onn ect ...

  • Agilent Technologies 85036B - page 26

    3-10 85036B/ E Use, Maintenance, and Care of the D evi ces Handli ng and Storage Ha ndling and S torage • Install the p rotective end c aps and store the c alibrat ion devices in the foam-line d storage c ase when not i n use. • Never stor e connector s lo ose in a box, desk, or bench dra wer . This i s the mos t common cause of c onnector d am ...

  • Agilent Technologies 85036B - page 27

    4-1 4 P erf orm anc e V erific at ion ...

  • Agilent Technologies 85036B - page 28

    4-2 85036B/ E P erformance V er ification Int roduct ion In troduct i on The perfor mance of your cali bra tio n kit c an on ly be verif ied b y retur ning the kit to Ag ilen t T echnol ogies f or rec ertifi cation. T he equipment r equired to ver ify the sp ecifica tions of the devic es in the ki t has been speci ally manufa ctured and is not c om ...

  • Agilent Technologies 85036B - page 29

    85036B/ E 4-3 P erformance V er ification Recer ti fication Rec ert ification The foll owing will be prov ided with a recer tified kit: • a new calibrati on sticker aff ixed to t he case • a cert i fic at e of ca l ib rati on • a calibration r eport for ea ch devi ce in the kit l isting measured values , specif icati ons , and unc ertaintie s ...

  • Agilent Technologies 85036B - page 30

    4-4 85036B/ E P erformance V er ification Recertification ...

  • Agilent Technologies 85036B - page 31

    5-1 5 T roub leshoot ing ...

  • Agilent Technologies 85036B - page 32

    5-2 85036B/ E T roub l eshooting T roubl eshooting Process Tr oub l esh ooting Pr ocess If you suspec t a bad c alibration, or if y our network ana lyzer doe s not pass p erformance verif ication, follow the steps in Figure 5-1 . Fi gure 5-1 T roub leshooting Flowchart ...

  • Agilent Technologies 85036B - page 33

    85036B/ E 5-3 T roubleshooting Returning a Kit o r Device to Agilent Ret u rning a K it or Device t o Agilent If your ki t or device requi res service , conta ct the Agile nt T ec hnologies office nearest yo u for informati on on wher e to send i t. See T ab le 5-1 . Incl ude a service t a g (located near the end of thi s manual) on w hich you prov ...

  • Agilent Technologies 85036B - page 34

    5-4 85036B/ E T roub l eshooting Returni ng a Kit or Devi ce to A gilent ...

  • Agilent Technologies 85036B - page 35

    6-1 6 R eplacea ble P arts ...

  • Agilent Technologies 85036B - page 36

    6-2 85036B/ E Replaceab le P ar ts Int roduct ion In troduct i on T ab le 6-1 list s the rep lacement part numbers f or items included in the 8 5036B cali bration kit and Figu r e 6 -1 i llustrates each of t hese items . T ab le 6-2 list s the rep lacement part numbers f or items included in the 8 5036E cali bration kit and Figu r e 6 -2 i llustrat ...

  • Agilent Technologies 85036B - page 37

    85036B/ E 6-3 Replaceable Parts Intr oducti on Figur e 6-1 R epl acea b le P arts f or the 85036B C a libration Kit ...

  • Agilent Technologies 85036B - page 38

    6-4 85036B/ E Replaceab le P ar ts Int roduct ion Figure 6 -2 Repla ce able P ar ts f or t he 85036 E Ca libr a tion Kit T able 6 -2 Rep lace able P arts fo r the 8 5036 E Ca libra tion Kit Des cri ptio n Qt y P er Kit Agi le nt Part Number Calibration Devices (7 5 Ω Type-N) Male broad band load 1 00909-60 019 Male com bination open/sho rt 1 8503 ...

  • Agilent Technologies 85036B - page 39

    85036B/ E 6-5 Replaceable Parts Intr oducti on Ta b l e 6 - 3 R eplac ea ble Parts— Ite ms Not I nclud ed in the 8 5036 B an d 8 503 6E Calib ration K its De sc rip ti on Qt y Ag i lent P art N um ber Connector Gages (75 Ω Type-N) Connecto r gage set 1 85036B K01 Wr enc hes 3/4 in, 135 N-cm (12 in- lb) torqu e wrenc h 1 8710-1766 1/2 in and 9/1 ...

  • Agilent Technologies 85036B - page 40

    6-6 85036B/ E Replaceab le P ar ts Int roduct ion ...

  • Agilent Technologies 85036B - page 41

    A-1 A S tand ard D efini tio ns ...

  • Agilent Technologies 85036B - page 42

    A-2 85036B/ E Standard De finitions Standar d Class Assignments Stand ard C lass A ssignm ents Class a ssignment orga nizes cal i bration sta ndards into a format c ompatibl e with the er ror models us ed i n the measurement cal i brat ion. A c l a ss or g rou p of c las ses corr es ponds t o the syste matic err ors to b e removed from the measured ...

  • Agilent Technologies 85036B - page 43

    85036B/ E A-3 Standard Definitions Standar d Class Assignments Blank F orm The standard c lass a s signments listed i n T able A-1 may be c hanged to meet y our specific requirements . T a ble A-2 i s provided to record the m odified standa rd class assignments . T able A-2 Standard Class Assignments Blank F orm Calibr ation Ki t Labe l: _ _______ ...

  • Agilent Technologies 85036B - page 44

    A-4 85036B/ E Standard De finitions Nominal Standar d Defi nitio ns Nomina l Stand ard Defi nit ions Standard definiti ons provide the c onstants needed to m athematicall y model t he elect rical charac teristi cs (del ay , attenuation, a nd impedance) of each calibr ation sta ndard. The nominal values o f these c onstants are theor eticall y deriv ...

  • Agilent Technologies 85036B - page 45

    85036B/ E A-5 Standard Definitions Nominal Standar d Definiti ons T able A-3 Standard Definition s Syst em Z 0 a = 75 Ω Ca li br ati on Ki t L a be l : N 75 Ω Standa rd b C0 × 10 − 15 F C1 × 10 − 27 F/Hz C2 × 10 − 36 F/Hz 2 C3 × 10 − 45 F/Hz 3 Fixed or Sl iding c Offset Fr eq (GHz) Coax or W a vegui de Standar d Label d Number Ty p ...

  • Agilent Technologies 85036B - page 46

    A-6 85036B/ E Standard De finitions Nominal Standar d Defi nitio ns Blank F orm The sta ndard defi nitions l isted in T able A- 3 m ay be c hanged to m eet your spec ific requir ements . T able A-4 is p rovided to rec ord the modi fied sta ndard defini ti ons. T ab le A-4 Standard Defini tions Blank F orm Syst em Z 0 a = _ ___ _______ ___ a. En sur ...

  • Agilent Technologies 85036B - page 47

    85036B/E Inde x- 1 A adapters , 1-3 part numbers , 6-5 alcoho l isopr opyl as clea nin g solven t , 3-4 altitud e , 2- 2 B blank form sta nda r d cla ss assig nme n ts , A-3 standard definitions , A-6 broa dband lo ads , 1- 2 C cal k it con te nts , 1-2 over view , 1-2 retu rn t o Agi l ent , 5-3 ser ial numb er , 1- 6 calibratio n bad , 5-2 certif ...

  • Agilent Technologies 85036B - page 48

    Index 85036B/E Inde x-2 when mak i ng c onnection s , 3- 6 envir onmental requir ement s , 2- 2 equi pment req ui red , 1- 3 but n ot supp lied , 1-3 , 6-2 , 6- 5 sup plie d , 1-2 , 6- 2 , 6-4 ESD , 3-2 precau tions , 3-2 , 3- 4 sup plie s part numbers , 6-5 when mak i ng c onnection s , 3- 6 F female open , 3-9 con nect ing , 3-9 disc onnec tin g ...

  • Agilent Technologies 85036B - page 49

    Index 85036B/E Inde x- 3 requir ements envir onmental , 2- 2 retu rn lo ss sp ecif ic at ions , 2-4 retu rn , ki t or d evi ce , 5- 3 S sales and service offi ces , 5- 3 sales an d servic e off ice s , 5-3 seria l n umbe rs , 1-6 devic es , 1-6 rec ord in g , 1- 6 serv ice , 5-2 and sales offices , 5-3 serv ice tag , 1-4 , 4-3 , 5-3 sex con ne ct o ...

Manufacturer Agilent Technologies Category Work Light

Documents that we receive from a manufacturer of a Agilent Technologies 85036B can be divided into several groups. They are, among others:
- Agilent Technologies technical drawings
- 85036B manuals
- Agilent Technologies product data sheets
- information booklets
- or energy labels Agilent Technologies 85036B
All of them are important, but the most important information from the point of view of use of the device are in the user manual Agilent Technologies 85036B.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Agilent Technologies 85036B, service manual, brief instructions and user manuals Agilent Technologies 85036B. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Agilent Technologies 85036B.

A complete manual for the device Agilent Technologies 85036B, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Agilent Technologies 85036B by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Agilent Technologies 85036B.

A complete Agilent Technologies manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Agilent Technologies 85036B - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Agilent Technologies 85036B, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Agilent Technologies 85036B, that we can find in the current document
3. Tips how to use the basic functions of the device Agilent Technologies 85036B - which should help us in our first steps of using Agilent Technologies 85036B
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Agilent Technologies 85036B
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Agilent Technologies 85036B in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Agilent Technologies 85036B?

Use the form below

If you did not solve your problem by using a manual Agilent Technologies 85036B, ask a question using the form below. If a user had a similar problem with Agilent Technologies 85036B it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)