Fluke 1551A EX - page 1
November 2010 © 2010 Fluke Corporation. All rights reserved. Spe c ifications are subject to change without no tice. All product names are trademarks of their respectiv e companies. 1551A Ex/1552A Ex Stik Thermometer Users Manual ...
-
Fluke 1551A EX - page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defec ts in material a nd workmanship under norm al use and service. The warranty period is one year and begi ns on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the origi nal buyer or en ...
-
Fluke 1551A EX - page 3
i Table of Contents Title Page Introduction ........................................................................................................ 1 Features .............................................................................................................. 1 How to Contact Fluke ........................................................ ...
-
Fluke 1551A EX - page 4
1551A Ex/1552A Ex Users Manual ii ...
-
Fluke 1551A EX - page 5
iii List of Tables Table Title Page 1. Symbols .................................................................................................................. 4 2. Approved Batteries................................................................................................. 12 ...
-
Fluke 1551A EX - page 6
1551A Ex/1552A Ex Users Manual iv ...
-
Fluke 1551A EX - page 7
v List of Figures Figure Title Page 1. Probe Measurements .............................................................................................. 5 2. Immersion Depth .................................................................................................... 6 3. Display and Controls ................................................. ...
-
Fluke 1551A EX - page 8
1551A Ex/1552A Ex Users Manual vi ...
-
Fluke 1551A EX - page 9
1 Introduction The Stik Thermometer (the Product) is a precision digital therm ometer that can be used as an alternative to mercury liquid-in-glass (L IG) thermometers. It can also be used as a reference standard for other types of digi tal or analog ther mometers. Because of its intrinsically safe certification, it can be used in many of the same ...
-
Fluke 1551A EX - page 10
1551A/1552A Ex Users Manual 2 • 300 hour battery life • Percent battery life and low-battery indicator • NVLAP accredited calibration (NIST traceable) How to Contact Fluke To contact Fluke, call one of the following telephone num bers: • Technical Support USA: 1-800-44-FLUKE (1-800-443-5853) • Calibration/Repair USA: 1-888-99-FLUKE (1-888 ...
-
Fluke 1551A EX - page 11
Stik Thermometer Safety Information 3 • Check for proper operation of the Product before taking it into a hazardous area. Should the Product not operate as expected do not take the Product into a hazardous area and return it to the manufacturer for inspection. • The Product is intended for installation only in locations providing adequate prote ...
-
Fluke 1551A EX - page 12
1551A/1552A Ex Users Manual 4 Symbols International symbols used on the Product and in this manual are explained in T able 1. Table 1. Symbols Symbol Meaning Symbol Meaning J Earth ground M Battery W Risk of Danger. Important information. See Manual. X Hazardous Voltage ~ Do not dispose of this product as unsorted municipal waste. Go to Fluke’s w ...
-
Fluke 1551A EX - page 13
Stik Thermometer Operation 5 Operation Immersion Depth and Usage W Caution To prevent possible damage to the Product or to equipment under test, do not immerse the Probe beyond the maximum immersion depth. This may damage its internal electronics. Allow enough time for the Probe to stabilize before recording me asurements. Use the Stability Display ...
-
Fluke 1551A EX - page 14
1551A/1552A Ex Users Manual 6 The minimum immersion depth for t he Probe is 3.5 inches in a stirred fluid heat source. It is recommended that you make sure of the minimum immersion depth for the specific application. See Figure 2. • The maximum immersion depth for the 1551A Ex is 2 inches less than the total length. Make sure the last 2 inches of ...
-
Fluke 1551A EX - page 15
Stik Thermometer Operation 7 Display and Controls The Display and Controls are explained below and i n the Setup and Configurati on section. See Figure 3. 5 6 7 10 9 4 3 8 2 1 gjo001.eps Number Item Description A D Push to power the Product on and off. Hold the button down at power-up to check the display to see that all display segments are sho wn ...
-
Fluke 1551A EX - page 16
1551A/1552A Ex Users Manual 8 Primary Display Push A to toggle the numeric display between the current reading, minimum, maximum, and 1-minute trend values in ° C or ° F as calculated since power up or since the last reset of values. Push and hold the A button for 2 seconds until " CLR " shows to reset the MIN/MAX/TREND values. Stabilit ...
-
Fluke 1551A EX - page 17
Stik Thermometer Operation 9 Auto-Off Selection The Auto-off parameters control when the Pr oduct automatically turns off after inactivity on the keypad. K Increase the auto-off setting and the primary data field from "OFF" to 1 minute and stopping at 20 minutes. L Decrease the auto-off setting and the primary data field from 1 minute to ...
-
Fluke 1551A EX - page 18
1551A/1552A Ex Users Manual 10 Display Resolution Display resolution is the number of digits to the right of the decimal point . The settings are 0.1, 0.01, or 0.001. K Change the resolution setting to the subsequent item in the list, cycling from the last to the first. L Change the resolution setting to the previous item in the list, cycling from ...
-
Fluke 1551A EX - page 19
Stik Thermometer Maintenance 11 Logging Interval Push C to enter the data edit mode. ENTER is ignored if presently logging data. K Change the logging interval setting to th e subsequent item in the list, cycling from the last to the first. L Change the logging interval setting to th e previous item in the list, cycling from the first to the last. C ...
-
Fluke 1551A EX - page 20
1551A/1552A Ex Users Manual 12 To change the batteries, see Figure 4: 1. Use a flat-blade screwdriver to remove the battery door screw. 2. Remove the battery door to access the three AAA batteries. 3. Replace only with approved AAA batteries listed in this document. See Table 2. Any substitution voids the Product safety rating. 4. Make sure the bat ...
-
Fluke 1551A EX - page 21
Stik Thermometer SCPI Commands 13 Battery Life Battery life is approximately 300 hours (12. 5 da ys) of continuous operation with the backlight off. A low-battery icon ( E ) shows in the lower right of the display when battery level is low. Replace batteries per recommendations found in the specifications section of this manual. Calibration For cal ...
-
Fluke 1551A EX - page 22
1551A/1552A Ex Users Manual 14 Battery Life ................................................................................ Approximately 300 hours without back light Battery Sa ve (auto-off) Range ................................................. Selectable from 1 to 30 minutes or can be disabled Size (readout onl y) ............................. ...