    IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-320 MODEL: 61-322 MODEL: 61-324 The Service Information provid es the following in formation: • Precautions and safety information • Specificatio ns • Performance test procedure • Calibration and cali bration adjustm ent procedure • Basic maintenance (cleaning, replacing the battery and fu ...

  • IDEAL INDUSTRIES 61-324 - page 2

    TABLE OF CONTENTS Page # Introduction 1 Precautions an d Safety Inf ormation 1 Symbols 1 Safety 2 Specifications 3 General Specification 3 Voltage Specifications 4 Current Specifications 5 Resistance, Diode, Continuity Specifications 5 Capacitance , Frequency S pecifications 6 Physical and environm ent characteristics 7 Certification and compliance ...

  • IDEAL INDUSTRIES 61-324 - page 3

    Page 1 Introduction Warning To avoid shoc k or injury , do not per form the verification tests or cal ibration pr ocedures described in this manual unless you are qualified to do so. The information provided in th is document is for the use of qualified personnel only. Caution The 61-320, 61-322, and 61-3 24 contain part s that can be damaged by st ...

  • IDEAL INDUSTRIES 61-324 - page 4

    Page 2 SAFETY Review the fol lowing safety preca utions to avoid inj ury and pre vent dam age to this product or any product s connected to it. To avoid potential h azards, use the product only as specified. CAUTION. These statements iden tify conditions or practices that could result in d amage to the e quipm ent or other pr operty. WARNING. These ...

  • IDEAL INDUSTRIES 61-324 - page 5

    Page 3 SPECIFICATIONS All specificatio ns are warrante d unless n oted typi cal and appl y to the 61-32 0, 61-322 an d 61-324. Stated accuracies are at 23 ºC ± 5 ºC at less than 80% relative humidity and without the ba ttery indicator di splayed. General specifications Characteristic s Description Display c ount 6,000 Numeric updat e rate 1.5 ti ...

  • IDEAL INDUSTRIES 61-324 - page 6

    Page 4 Measurement Characteristics Accuracy is ± (% reading + number of digits) at 23 ° C ± 5 ° C , less than 80% R.H. DC Volts Range Resolution Accuracy Over voltage protection 600.0m V 100µV 6.000V 1mV 60.00V 10mV 600.0V 100mV 1000V 1V ±(0.5% + 2dgt) 1000V rms AC Volts Range Resolution Accuracy Over voltage protection 600.0m V 100µV Unspec ...

  • IDEAL INDUSTRIES 61-324 - page 7

    Page 5 DC Current Range Resolution Accuracy Over voltage protection 600.0µA 1m V 6000µA 1mV 60.00m A 10mV 600.0m A 100mV ±(1.0%+2 dgt ) 600V rms1A (600) Fast blow fuse AC Current Range Resolution Accuracy Over voltage protection 600.0µA 1m V 6000µA 1mV 60.00m A 10mV 600.0m A 100mV ±(1.5%+2 dgt ) 50Hz ~ 50 0Hz * 600V rms 1A (600) Fa st blow fu ...

  • IDEAL INDUSTRIES 61-324 - page 8

    Page 6 Frequency Range Sensitivity Accuracy Overload protection 6000Hz 60.00KHz 600.0KHz 100mV rm s * 6.000MHz 25 0mV rms 60.00MHz 1V rms Frequency : 0.01%±1di git 600V rms * Less t han 20Hz, the sensitivity is 1.5V rms. Capacitance Range Accura cy Overload protection 6.000nF 60.00nF 600.0nF 6.000µF 60.00µF 600.0µF 6.000m F * ±(1.9% + 8 dgt) 6 ...

  • IDEAL INDUSTRIES 61-324 - page 9

    Page 7 Physical and E nvironmental C haracteristi cs Characteristic s Description Dime nsions (H x W x D) 158mm x 76m m x 38mm 164mm x 82mm x 44mm (with holst er) Weight (with battery) 0.3Kg With holster 0.5Kg Environmental Characteristics Descripti on Temperature operating 0 to + 5 0ºC Non-Operating -20 to + 60 º C Humidity (operating) < 80% ...

  • IDEAL INDUSTRIES 61-324 - page 10

    Page 8 Required Equipment Required equipment is listed in Table B. I f the recom mended m odels are not avail able, equi pment wi th equivale nt specifications may be used. Repai rs or ser vicing sh ould be performed o nly by q ualified per sonnel. Table B. Required Equipment Equipment Required Characteristics Recommended Model Calibrator AC Voltag ...

  • IDEAL INDUSTRIES 61-324 - page 11

    Page 9 Basic Maintenance Warning To avoid shoc k, rem ove the test leads and an y input signals before opening the case or re placing the battery or fuses. Opening the Meter Case Caution To avoid unintentional short circ uit, always place the uncovered m eter assembly on a protective surface . When the case of the m eter is open, circuit connection ...

  • IDEAL INDUSTRIES 61-324 - page 12

    Page 10 Testing Fuses To test the internal fuses of the meter: 1. Turn the rotary selector switch to the Ω positio n. 2. To test FS1, plug a test lead into V Ω Hz input terminal and touch the probe to th e mA input terminal. The display sh ould indicate bet ween 0.0 to 0.2 Ω . FS1 (1 A 600V - Bussmann BBS-1 recomm ended). If display reads hig ...

  • IDEAL INDUSTRIES 61-324 - page 13

    Page 11 Performance Tests The following performance tests ve rify the complete ope ra bility of the meter and c heck the accuracy of each meter function against the meter’s specifications. Accuracy specifications are valid for a period of one y ear after calibration, when measured at an operating temp er atu r e of 1 8ºC to 28ºC and at a maximu ...

  • IDEAL INDUSTRIES 61-324 - page 14

    Page 12 Table 1 AC Voltage Test: Step Input Frequency Reading 1 5.800V 50Hz 5.743 to 5.857 2 5.800V 500Hz 5.743 to 5.857 3 58.00V 50Hz 57.43 to 58.57 4 58.00V 500Hz 57.43 to 58.57 5 580.0V 50Hz 574.3 to 585.7 6 580.0V 500Hz 574.3 to 585.7 7 750V 50Hz 738 to 762 8 750V 500Hz 738 to 762 6. Turn the rotary switch to “V ” position. 7. Set the calib ...

  • IDEAL INDUSTRIES 61-324 - page 15

    Page 13 Testing the Capacitance Function The meter measures capacitance by c h arging the capac itor with a known direct current, measuring the resultant voltage, and calcu lating the capacitance. If the same capacitan ce is m easured on an impedance bridge, a different rea ding may result. This variance is likely to be greater at higher frequencie ...

  • IDEAL INDUSTRIES 61-324 - page 16

    Page 14 Table 5 AC mA Test: Step Source Frequency Reading 1 58.00mA 50Hz 56.93 to 59.07 2 58.00mA 500Hz 56.93 to 59.07 3 580.0m A 50Hz 569.3 to 590.7 4 580.0m A 500Hz 569.3 to 590. 7 6. Turn the rotary switch to mA . 7. Apply the inputs for steps 1-2 in Table 6. 8. For each input, compare the reading on the meter display to the read ing in Table 6. ...

  • IDEAL INDUSTRIES 61-324 - page 17

    Page 15 Table 8 Frequency Test: Step Source Level Reading 1 5.600KHz 100m V rms 5.598 to 5.602 2 56.00KHz 100m V rms 55.98 to 56 .02 3 560.0KHz 100m V rms 559.8 to 56.02 4 5.600M Hz 250mV rm s 5.598 to 5.602 5 56.00M Hz 1V rms 55.98 to 56.02 Calibration Calibrate the meter once a year to ensure that it performs according to specifications. Per for ...

  • IDEAL INDUSTRIES 61-324 - page 18

    Page 16 Figure 3 - Calibrati on Adjustment Point s Form number TM61320-2-4 Rev 2 September 2002 ...

  • IDEAL INDUSTRIES 61-324 - page 19

    Form number TM61320-2-4 Rev 2 September 2002 Page 17 ...

Manufacturer IDEAL INDUSTRIES Category Automobile Accessories

Documents that we receive from a manufacturer of a IDEAL INDUSTRIES 61-324 can be divided into several groups. They are, among others:
- IDEAL INDUSTRIES technical drawings
- 61-324 manuals
- IDEAL INDUSTRIES product data sheets
- information booklets
- or energy labels IDEAL INDUSTRIES 61-324
All of them are important, but the most important information from the point of view of use of the device are in the user manual IDEAL INDUSTRIES 61-324.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals IDEAL INDUSTRIES 61-324, service manual, brief instructions and user manuals IDEAL INDUSTRIES 61-324. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product IDEAL INDUSTRIES 61-324.

A complete manual for the device IDEAL INDUSTRIES 61-324, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use IDEAL INDUSTRIES 61-324 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of IDEAL INDUSTRIES 61-324.

A complete IDEAL INDUSTRIES manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual IDEAL INDUSTRIES 61-324 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the IDEAL INDUSTRIES 61-324, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the IDEAL INDUSTRIES 61-324, that we can find in the current document
3. Tips how to use the basic functions of the device IDEAL INDUSTRIES 61-324 - which should help us in our first steps of using IDEAL INDUSTRIES 61-324
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with IDEAL INDUSTRIES 61-324
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of IDEAL INDUSTRIES 61-324 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning IDEAL INDUSTRIES 61-324?

Use the form below

If you did not solve your problem by using a manual IDEAL INDUSTRIES 61-324, ask a question using the form below. If a user had a similar problem with IDEAL INDUSTRIES 61-324 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (1)

Emanuele Zanetti

Ho una lavastoviglie tekadw7 45s la parte sotto non lavava bene cosa devo guardare?