Manual Omega Engineering HH502

16 pages 0.59 mb
Download

Go to site of 16

Summary
  • Omega Engineering HH502 - page 1

    omega.com e-mail: info@omega.com For latest product manuals: omegamanual.info U ser ’ s Guide HH502 Digital Thermometer Shop online at ...

  • Omega Engineering HH502 - page 2

    OMEGAnet ® On-Line Service Int ern et e-m ail omega.com info@omega.com It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon ce ...

  • Omega Engineering HH502 - page 3

    The symbol on the instrument indicates that the operator must refer to an explanation in this manual. WARNING To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC. WARNING To avoid damage or burns, do not make temperature measurement in microwave ovens. CAUTION Repeated sharp flexi ...

  • Omega Engineering HH502 - page 4

    SPECIFICATIONS ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable. Measurement Range: J-TYPE -200°C to 1050°C, (-328°F to 1922°F) K-TYPE -200°C to 1370°C, (-328°F to 2498°F) Resolution: 0.1°C or 0.2°F. Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1 year, ...

  • Omega Engineering HH502 - page 5

    3 ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 4½ digit liquid crystal display (LCD) with maximum reading of 19999. Overload: "----.-" is display. Battery: Standard 9V battery. Battery Life: 100 hours typical. Auto ...

  • Omega Engineering HH502 - page 6

    OPERATING INSTRUCTIONS Power Switch The key turns the thermometer on or off. In the data SET mode, can not power off, must leave data SET mode then power off. °C/°F Selecting the Temperature Scale Readings are dual displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set to the temperature ...

  • Omega Engineering HH502 - page 7

    5 4 5 6 T1 /T2/T1-T2 Main display Input Selection The input selection indicates which input is selected for main display; T1 thermocouple, T2 thermocouple or the difference between the two thermo- couples (T1-T2), when the thermometer is turned on, it is set to T1, When main display input selected T1, then T1 input can select alternate of thermocou ...

  • Omega Engineering HH502 - page 8

    6 7 In the this mode, press HOLD key to stop the recording of readings, all values are frozen, press again to restart recording. To prevent accidental loss of MIN, MAX and AVG data, in this mode can only be cancelled by pressing and hold down the MIN MAX key for 2 seconds to exit and erased recorded readings. WARNING MIN, MAX and AVG readings can b ...

  • Omega Engineering HH502 - page 9

    7 8 9 Sec. Minu. Selecting the Time scale Reading the third displayed (Time) in either second or minute. When the thermometer turned on, it is set to second. To change the Time scale, press sec. Minu. key. Maximum Time reading is 100 hours. If 100 hours is exceeded, reset Time to zero and restart. SET mode (Relative value set, Time set and Hi/Lo Li ...

  • Omega Engineering HH502 - page 10

    8 10 11 4. When the thermometer is turned on. The Relative set value and Hi/Lo Limits set value that was in use when thermometer was last turned off set values. T1/T2/T1-T2 second display Input Selection The input selection indicates which input is selected for second display; T1 thermocouple, T2 thermocouple or the difference between the two therm ...

  • Omega Engineering HH502 - page 11

    9 OPERATOR MAINTENANCE WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the thermometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two sc ...

  • Omega Engineering HH502 - page 12

    10 CALIBRATION PROCEDURE Note: The following calibration procedures should be performed only by qualified technicians who have access to the following: Equipment: The class of calibration should be 10 times greater than the measured meter. Step 1: Short the J3 with jumper. Step 2: Apply 65 mV to the T1 input from Volts Calibrator and wait till the ...

  • Omega Engineering HH502 - page 13

    11 NOTES: ...

  • Omega Engineering HH502 - page 14

    12 NOTES: ...

  • Omega Engineering HH502 - page 15

    FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA: 1. Purchase Order number under which the product was PURCHASED, 2. Model and serial number of the product under w arranty, and 3. Repair instructions and/or specific problems relative to the product. FOR NON-WARRANTY REPAIRS, consult OMEGA for current rep ...

  • Omega Engineering HH502 - page 16

    Where Do I Find Everything I Need for Process Measurement and Contr ol? OM EGA…Of Course! Shop online at omega.com sm M2807/1008 TEMPERATURE 䡺 ⻬ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies 䡺 ⻬ Wire: Thermocouple, RTD & Thermistor 䡺 ⻬ Calibrators & Ice Point References 䡺 ⻬ Recorders, Contr ...

Manufacturer Omega Engineering Category Thermometer

Documents that we receive from a manufacturer of a Omega Engineering HH502 can be divided into several groups. They are, among others:
- Omega Engineering technical drawings
- HH502 manuals
- Omega Engineering product data sheets
- information booklets
- or energy labels Omega Engineering HH502
All of them are important, but the most important information from the point of view of use of the device are in the user manual Omega Engineering HH502.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Omega Engineering HH502, service manual, brief instructions and user manuals Omega Engineering HH502. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Omega Engineering HH502.

A complete manual for the device Omega Engineering HH502, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Omega Engineering HH502 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Omega Engineering HH502.

A complete Omega Engineering manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Omega Engineering HH502 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Omega Engineering HH502, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Omega Engineering HH502, that we can find in the current document
3. Tips how to use the basic functions of the device Omega Engineering HH502 - which should help us in our first steps of using Omega Engineering HH502
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Omega Engineering HH502
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Omega Engineering HH502 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Omega Engineering HH502?

Use the form below

If you did not solve your problem by using a manual Omega Engineering HH502, ask a question using the form below. If a user had a similar problem with Omega Engineering HH502 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)