    OWNER'S MANUAL CD001030-1040 ©Graham-Field 10/03 IMPOR T ANT : READ THIS MANUAL BEFORE OPERA TING YOUR P A TIENT LIFT A G raham-F ield B rand LF1030 HYDRAULIC P A TIENT LIFT LF1040 ELECTRIC P A TIENT LIFT ...

  • Graham Field LUMEX LF1030 - page 2

    ...

  • Graham Field LUMEX LF1030 - page 3

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 3 CONTENTS INTR ODUCTION ................................................................................................................... 4 SPECIAL NO TES ........................................................................................................ ...

  • Graham Field LUMEX LF1030 - page 4

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 4 INTR ODUCTION Congratulations on your purchase of a Lumex patient lift. The following pages will provide y ou with important safety and operating instructions on the use of patient lifts, slings, and accessories, as well as maintenance and warranty information ...

  • Graham Field LUMEX LF1030 - page 5

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 5 SPECIAL NO TES Warning: Do not operate this equipment without first reading and understanding this manual. If you are unable to understand the warnings, cautions and instructions, contact your dealer or a healthcare professional before attempting to use this e ...

  • Graham Field LUMEX LF1030 - page 6

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 6 SAFETY PRECA UTIONS IMPOR T ANT : Before using patient lift, please r ead and adhere to the following safety pr ecautions and warnings. F ailure to do so could result in serious injury or damage to your patient lift. Always consult your healthcare pr ofessiona ...

  • Graham Field LUMEX LF1030 - page 7

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 7 Warning: Unauthorized modification of your patient lift or the use of non-Lumex replace- ment parts may change the structure of the lift and could create a hazardous condition, which may result in serious injury and will void the warranty. Warning: LF1030 Hydr ...

  • Graham Field LUMEX LF1030 - page 8

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 8 FEA TURES 6-point suspension Hydraulic pump (LF1030 hydraulic patient lift only ) See above left for LF1040 electric patient lift components Battery, battery charger, & control box (LF1040 electric patient lift only ) Actuator (LF1040 electric patient lift ...

  • Graham Field LUMEX LF1030 - page 9

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 9 HANDLING PR OCEDURES SHIPPING AND TRANSPORTATION INSTRUCTIONS If the patient lift is to be reshipped b y common carrier , it should be packed in the same carton/s in which it was receiv ed (the LF1030 is shipped in one carton; the LF1040 motor may be packaged ...

  • Graham Field LUMEX LF1030 - page 10

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 10 ASSEMBL Y INSTR UCTIONS The LF1030 patient lift is packed in one carton. The LF1040 motor may be in a separate carton. Com- ponents are shown in illustration belo w and called out in list below illustration. LF1030 hydraulic patient lift components 1 2 7 3 4 ...

  • Graham Field LUMEX LF1030 - page 11

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 11 Assembly 1. Remo ve the locking pin from the mast sleev e located on the rear of the base. 2. Install the L-inserts in the base. 3. With the boom pointed in the same dir ection as the base legs, insert the mast into the sleeve. 4. Insert the locking pin throu ...

  • Graham Field LUMEX LF1030 - page 12

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 12 6. Screw the push handles into the mast and tighten the nuts. 7. Remo ve the bolt from the top of the spr eader bar , taking care not to drop the pin assembly . Install the spreader bar in the U bracket at the end of the boom. Align holes. Insert bolt and tig ...

  • Graham Field LUMEX LF1030 - page 13

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 13 Note: Steps 10-13 apply to LF1040 Electric Patient Lift only. 10. Attach control box using 5mm cap scr ews as shown at right. 11. Attach battery charger using 5mm cap screws as shown at right. 12. Clip battery onto unit. 13. See follo wing E lectrical connect ...

  • Graham Field LUMEX LF1030 - page 14

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 14 Charging the battery with lift-mounted batter y charger: W ith control box, battery charger , and batter y assembled, insert the batter y charger plug into a standard 110V r eceptacle. A green light signifying power and a yello w light signifying that the bat ...

  • Graham Field LUMEX LF1030 - page 15

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 15 OPERA TING INSTRUCTIONS LIFTING THE PATIENT Lifting the patient from a bed 1. Roll the patient onto their side. P lace the folded sling behind the patient’ s back. The center of the sling should be parallel with the patient’ s spine. Roll the patient onto ...

  • Graham Field LUMEX LF1030 - page 16

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 16 Lowering the patient onto a bed 1. Raise the patient lift until the patient ’ s buttocks are above the surface of the bed. E nsure that there are no obstacles underneath or around the bed. M ove the patient lift under the bed. Be car eful the patient does n ...

  • Graham Field LUMEX LF1030 - page 17

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 17 8. Slo wly lower the patient into the wheelchair . 9. With one (1) attendant behind the wheelchair and the other operating the lift, the attendant behind the chair should pull back on the handle or sides of the sling to place the patient into the back of the ...

  • Graham Field LUMEX LF1030 - page 18

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 18 CARE AND MAINTENANCE Pr oper care and maintenance is essential to keeping y our Lumex patient lift in a safe operating condi- tion. In addition to inspecting the unit befor e each use, periodic maintenance checks should be done. We further recommend that this ...

  • Graham Field LUMEX LF1030 - page 19

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 19 Cleaning R egular cleaning is r ecommended. A soft cloth, dampened with water and a mild NON-ABRASIVE detergent (household cleaner , soapy water , or hospital grade disinfectant) is all that is needed to clean your patient lift. A utomobile wax or furniture p ...

  • Graham Field LUMEX LF1030 - page 20

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 20 MAINTENANCE SCHEDULE Note: Only qualified persons should service and repair your Lumex patient lift. Regular maintenance of your patient lift is necessary to ensure continuing proper and safe operation. Read and observe the following recommended maintenance s ...

  • Graham Field LUMEX LF1030 - page 21

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 21 TR OUBLESHOO TING Warning: If you experience a problem with your lift and are unable to service it yourself, contact your Lumex dealer. LF1030 HYDRAULIC PATIENT LIFT SYMPTOM POSSIBLE CAUSE ACTION Patient lift does not Air in the cylinder Bleed the cylinder by ...

  • Graham Field LUMEX LF1030 - page 22

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 22 SPECIFICA TIONS Note: All dimensions given in inches unless otherwise specified. LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT Base width Open: 39.4" Closed: 25.6" Base height (Clearance): 5-1/2" Base length 45.30" Lifting h ...

  • Graham Field LUMEX LF1030 - page 23

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 23 LIMITED W ARRANTY NO TE: The warranty below has been drafted to comply with the F ederal Law applicable to products manufactured after J uly 4, 1975. This warranty is extended only to the original purchaser/consumer or dealer/non-consumer who does not buy for ...

  • Graham Field LUMEX LF1030 - page 24

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 24 THE FOREGOING W ARRANTY IS EXCL USIVE AND IN LIEU OF ALL OTHER EXPRESSED W ARRANTIES AND IMPLIED W ARRANTIES, INCLUDING BUT NO T LIMITED T O THE IMPLIED W ARRANTIES OF MER CHANT ABILITY AND FITNESS FOR A P AR TICULAR PURPOSE, AND SHALL NO T EXTEND BEY OND THE ...

  • Graham Field LUMEX LF1030 - page 25

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 25 INDEX A Actuator , maintenance (LF1040 only) 20 Assembly instructions 10 B Battery , charging (LF1040 only) 13 C Care and maintenance 18 Care of sling 19 Caution, as used in this manual 5 Checking for damage and wear 19 Cleaning 19 E Electrical connections (L ...

  • Graham Field LUMEX LF1030 - page 26

    LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03 26 NOTES ...

  • Graham Field LUMEX LF1030 - page 27

    ...

  • Graham Field LUMEX LF1030 - page 28

    U.S.A., Corporate Headquar ters: Graham-Field Health Pr oducts 2935 Northeast Parkway Atlanta, Geor gia 30360 telephone: 800-347-5678, 770-447-1609 fax: 800-726-0601, 678-291-3232 http://www .grahamfield.com ...

Manufacturer Graham Field Category Personal Lift

Documents that we receive from a manufacturer of a Graham Field LUMEX LF1030 can be divided into several groups. They are, among others:
- Graham Field technical drawings
- LUMEX LF1030 manuals
- Graham Field product data sheets
- information booklets
- or energy labels Graham Field LUMEX LF1030
All of them are important, but the most important information from the point of view of use of the device are in the user manual Graham Field LUMEX LF1030.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Graham Field LUMEX LF1030, service manual, brief instructions and user manuals Graham Field LUMEX LF1030. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Graham Field LUMEX LF1030.

A complete manual for the device Graham Field LUMEX LF1030, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Graham Field LUMEX LF1030 by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Graham Field LUMEX LF1030.

A complete Graham Field manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Graham Field LUMEX LF1030 - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Graham Field LUMEX LF1030, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Graham Field LUMEX LF1030, that we can find in the current document
3. Tips how to use the basic functions of the device Graham Field LUMEX LF1030 - which should help us in our first steps of using Graham Field LUMEX LF1030
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Graham Field LUMEX LF1030
5. FAQ - Frequently Asked Questions
6. Contact detailsInformation about where to look for contact to the manufacturer/service of Graham Field LUMEX LF1030 in a specific country, if it was not possible to solve the problem on our own.

Do you have a question concerning Graham Field LUMEX LF1030?

Use the form below

If you did not solve your problem by using a manual Graham Field LUMEX LF1030, ask a question using the form below. If a user had a similar problem with Graham Field LUMEX LF1030 it is likely that he will want to share the way to solve it.

Copy the text from the picture

Comments (0)