  • Omega Engineering HH507R - page 1

    OMEGA HH507R Digital Thermometer ...

  • Omega Engineering HH507R - page 2

    OMEGAnet SM On-Line Service http://www.omega.com Internet e-mail info@omega.com Servicing Nor th America: USA: ISO 9001 Certified Canada: One Omega Drive, Box 4047 976 Bergar Stamford, CT 06907-0047 Laval (Quebec) H7L5A1 Tel: (203) 359-1660 Tel: (514) 856-6928 FAX: (203)359-7700 FAX: (514) 856-6886 e-mail: info@omega.com e-mail: info@omega.com omeg ...

  • Omega Engineering HH507R - page 3

    1 The symbol on the instrument indicates that the operator must refer to an explanation in this manual. WARNING To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC. WARNING To avoid damage or burns, do not make temperature measurement in microwave ovens. CAUTION Repeated sharp fle ...

  • Omega Engineering HH507R - page 4

    2 SPECIFICATIONS ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable Measurement Range: Thermocouple Range K-TYPE -200°C to 1372°C, -328°F to 2501°F J-TYPE -210°C to 1050°C, -346°F to 1922°F T-TYPE -200°C to 400°C, -328°F to 752°F E-TYPE -210°C to 790°C, -346°F to 1454°F R-TYPE 0°C to 1767°C, 32°F to 3212°F S-TYP ...

  • Omega Engineering HH507R - page 5

    3 Accuracy: Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error. K/J/T/E-TYPE ±(0.05% rdg + 0.3°C) -50°C to 1370°C ±(0.05% rdg + 0.7°C) -50°C to -210°C ±(0.05% rdg + 0.6°F) -58°F to 2498°F ±(0.05% rdg + 1.4°F) -58°F to -346°F N-TYPE ±(0.05% ...

  • Omega Engineering HH507R - page 6

    4 ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 4½ digit liquid crystal display (LCD) with maximum reading of 9999.9. Overload: "----.-" is display. Battery: Standard 9V battery. Battery Life: 100 hours typical with ...

  • Omega Engineering HH507R - page 7

    5 OPERATING INSTRUCTIONS Power Switch The key turns the thermometer on or off. In the MIN MAX record mode can not power off, must leave MIN MAX record mode then power off. °C/°F Selecting the Temperature Scale Readings are dual displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set to the ...

  • Omega Engineering HH507R - page 8

    6 TIMER STOP-WATCH Mode Press TIMER STOP-W key to enter stop-watch mode stop-watch is displayed in second display. Press start/stop key to toggle stop-watch starts and stops (Time goes up to 99 minutes & 59.9 seconds). In the this mode the automatic power-off feature is disabled. Pess Clear key to zero the stop-watch. Press TIMER STOP-W key aga ...

  • Omega Engineering HH507R - page 9

    7 Push MIN MAX key to cycle through the Max/Min/Avg/Present readings. If an overload is recorded, the averaging function is stopped and average value display"----.-". The true average of all the reading taken over at least 22 hours preiod can be displayed. If 22 hours is exceeded, new averages are no longer calculated. The last calcalated ...

  • Omega Engineering HH507R - page 10

    8 Hi/Lo LIMITS mode (only Main display) Press Hi/Lo LIMITS key to enter the Hi/Lo Limits comparative mode, "LIMIT" is displayed. In the this mode the automatic power-off feature is disabled and REL key, HOLD key, MAX/MIN key are disabled, when input temperature value exceed Hi. The beeper emits a continuity pulse tone and "Hi" i ...

  • Omega Engineering HH507R - page 11

    9 3. Time set mode, (Press ENTER key can escape Time set mode) = = : = = = = is displayed in third display. Time (hours, minutes, seconds) value is entered via overlay numbers, then press overlay ENTER key. Time start from set time value and exit set mode. Relative value Recall display mode Press REL RECALL key to display the Relative set value on ...

  • Omega Engineering HH507R - page 12

    10 OPERATOR MAINTENANCE WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the thermometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. The " " appears on the LCD display when replacement is needed. To replace the battery, remove the two s ...

  • Omega Engineering HH507R - page 13

    RS-232 Operation Using Quick BASIC The following example shows how to send command instruction and receive data responses over the RS-232 interface using Quick BASIC. The program example is set up on "COM1". CLS E$ = A ST1: CLOSE #1 OPEN COM1:1200,E,7,1,DS,RS FOR OUTPUT AS #1 PRINT #1, E$; CLOSE #1 OPEN COM1:1200,E,7,1,DS,R ...

  • Omega Engineering HH507R - page 14

    12 IF C$ = - THEN DEC = DEC * -1 END IF CLS LOCATE 13, 34 PRINT USING ######.#; DEC E$ = INKEY$ IF E$ =  THEN GOTO ST2 END IF GOTO ST1 ...

  • Omega Engineering HH507R - page 15

    13 RS-232 Transmit and Receive Comammand Summary 1. Control letter should use Capital Letter. 2. If you write you own program, the host computer must send the "A" letter to activate data transmission. 3. If you should stop data transmission the host computer must send the "B" letter to extend battery life. 4. Communication Param ...

  • Omega Engineering HH507R - page 16

    08 — T1 Thermocouple Type "K" ® K type "J" ® J type "T" ® T type "E" ® E type "R" ® R type "S" ® S type "N" ® N type OUTPUT DATA FORMATS The data format consists of 32 bytes. byte S: Negative sign or positive sign H: Hexadecimal digits D: Numeric digits A: Capital Letter ...

  • Omega Engineering HH507R - page 17

    17 — MSD 18 — LSD 19 — MSD 20 — LSD 21 — MSD 22 — LSD Hours Minutes 23 — Record mode "R" ® Record mode "M" ® MAX mode " I " ® MIN mode "A" ® AVG mode " - " ® normal mode 24 — REL mode "R" ® Relative mode " - " ® normal mode 25 — HOLD mode "H" ® ...

  • Omega Engineering HH507R - page 18

    COMMUNICATION 507R RS-232C "COM1" PORT 4612-IR232 RS232C CABLE 1. Install the RSR232C cable between computer "COM1" and Thermometer RS232C port. 2. Turn on Thermometer. 3. Utilize the Quick BASIC for inputting programs. 4. Use the capital control letter and input "A" letter from the keyboard to start sending messages t ...

  • Omega Engineering HH507R - page 19

    1 HH 506(R)/HH 507(R) CA LIBRA TION PROCEDURE Note: The following calibration procedure s hould perform only by qualified technicians who have access to the items as fol lowing. Equipment: The class of calibrator ha d better 10 times g reater than the measur ed meter. 1. Set the jumper position on J1 position. 2. Input DCV 17m V to the T 1 after th ...

  • Omega Engineering HH507R - page 20

    WARRANTY / DISCLAIMER OMEGA ENGINEERING, INC . warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA W arranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA's customers ...

  • Omega Engineering HH507R - page 21

    M-2814/0799 Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE þ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies þ Wire: Thermocouple, RTD & Thermistor þ Calibrators & Ice Point References þ Recorders, Controllers & Process Monitors þ Infrared Pyromet ...

Manufacturer Omega Engineering Category Thermometer

Documents that we receive from a manufacturer of a Omega Engineering HH507R can be divided into several groups. They are, among others:
- Omega Engineering technical drawings
- HH507R manuals
- Omega Engineering product data sheets
- information booklets
- or energy labels Omega Engineering HH507R
All of them are important, but the most important information from the point of view of use of the device are in the user manual Omega Engineering HH507R.

A group of documents referred to as user manuals is also divided into more specific types, such as: Installation manuals Omega Engineering HH507R, service manual, brief instructions and user manuals Omega Engineering HH507R. Depending on your needs, you should look for the document you need. In our website you can view the most popular manual of the product Omega Engineering HH507R.

A complete manual for the device Omega Engineering HH507R, how should it look like?
A manual, also referred to as a user manual, or simply "instructions" is a technical document designed to assist in the use Omega Engineering HH507R by users. Manuals are usually written by a technical writer, but in a language understandable to all users of Omega Engineering HH507R.

A complete Omega Engineering manual, should contain several basic components. Some of them are less important, such as: cover / title page or copyright page. However, the remaining part should provide us with information that is important from the point of view of the user.

1. Preface and tips on how to use the manual Omega Engineering HH507R - At the beginning of each manual we should find clues about how to use the guidelines. It should include information about the location of the Contents of the Omega Engineering HH507R, FAQ or common problems, i.e. places that are most often searched by users in each manual
2. Contents - index of all tips concerning the Omega Engineering HH507R, that we can find in the current document
3. Tips how to use the basic functions of the device Omega Engineering HH507R - which should help us in our first steps of using Omega Engineering HH507R
4. Troubleshooting - systematic sequence of activities that will help us diagnose and subsequently solve the most important problems with Omega Engineering HH507R
5. FAQ - Frequently Asked Questions